The MBC450 Series of open-frame medical power supplies, with its
wide universal 90-264 VAC input range, is available at 300 W of output
power and a variety of single output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These medical power supplies are ideal for monitoring, home health
equipment as well as surgical devices.
4 x 6.5 x 1.61 inches
Approved to EN/IEC 60601-1
Dual Fusing
Current Sharing Option
Cover and Fan Options
Peak Power Capability
Class B EMI & Medical (BF) Safety Approvals
Side Fan or Top Fan Mounting Option Product
(–S or –T to be added to model number)
Current Sharing Option Product (–I to be added to model number)
ITE Safety Agency Approvals
RoHS Compliant
CE marked
Diagnostic
Drug Pump
Dialysis
Hospital Beds
Home Health Care
Monitoring
Imaging
Therapy Devices
2
MBC450 Series
MODEL
MBC450-1T05G
MBC450-1T12G
MBC450-1T15G
MBC450-1T24G
MBC450-1T30G
MBC450-1T48G
NOTES:
1
2
OUTPUT VOLTAGE
5 VDC
12 VDC
15 VDC
24 VDC
30 VDC
48 VDC
MAX LOAD
(CONVECTION)
1,2,3
31.0 A
20.83 A
16.66 A
12.30 A
10.0 A
6.25 A
MAX LOAD
(420 LFM)
1,2,3
55.0 A
37.5 A
30.0 A
18.75 A
15.0 A
9.37 A
MINIMUM LOAD
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
RIPPLE & NOISE
4
2%
2%
2%
2%
2%
2%
3
4
Combined output power of main output, fan supply and standby supply shall not exceed max. power rating.
Standby output voltage 5 V/ 1.5A(convection) / 2A(420LFM) with tolerance including set point accuracy, line and load regulation
is +/-10%. Ripple and noise is less than 5%.
Fan supply output voltage 12V/ 500mA with tolerance including set point accuracy, line and load regulation is +/-30% and needs
min. 1% load on main output to be within regulation band. Ripple and noise is less than 10%.
Ripple is peak to peak with 20 MHz bandwidth and 10
μF
(Tantalum capacitor) in parallel with a 0.1
μF
capacitor at rated line
voltage and load ranges. Please contact factory/ sales representative for minimum load required for ripple to be within
specification.
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
Input Current
No Load Power
Inrush Current
Leakage Current
Input Protection
Power Factor
Switching Frequency
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
Earth Leakage Current
Touch Leakage Current
Dual fusing, in AC Line and AC Neutral
120 VAC
230 VAC
PFC converter: Variable
Resonant converter: Variable
DESCRIPTION / CONDITION
Universal
SPECIFICATIONS
90-264 VAC / 120-390 VDC
47 to 63 Hz
4.5 A max.
2.3 A max.
0.4 W
0.8 W
40 A max.
75 A max.
270
μA,
45
μA
@120 VAC / 63 Hz
T8A / 250 V
0.98
0.95
45-160 kHz typical
35-250 kHz, 90 kHz typical
tech.support@psbel.com
MBC450 Series
PARAMETER
Output Power
5
3
DESCRIPTION / CONDITION
475 W for 24 V, 30 V & 500 W for 48 V model
only for 5 seconds max.
24 V, 48 V, 30 V
120 VAC
12 V, 15 V
5V
230 VAC
24 V, 48 V, 30 V
SPECIFICATIONS
155 to 450 W
88%
86%
83% typical
90%
10 ms
+/-0.5%
+/-3%
Efficiency (Full Load)
Hold Up Time
Line Regulation
Load Regulation
Transient Response
Rise Time
Set Point Tolerance
Voltage Adjustment
Over Voltage Protection
Over Current Protection
Short Circuit Protection
Over Temperature Protection
Current Share
5
120 / 230 VAC
<10%, 50% to 100% load change, 50 Hz, 50% duty
cycle, 0.1 A/μs
Recovery time < 5 ms
< 100 ms
+/-1%
V1
Latch Type
Hic-Up type
Short term, auto recovery
Automatic recovery
Up to 2 supplies connected in parallel (optional)
Derate output power linearly to 80% from 90 VAC to 80 VAC input.
±3%
>114%
120 to 150%
130°C primary heat sink
PARAMETER
Operating Temperature
Storage Temperature
Humidity
Altitude
DESCRIPTION / CONDITION
Refer to derating curve
(Figure 1)
SPECIFICATIONS
0 to +70°C
-40 to 85° C
Non Condensing
Operating:
Non-Operating:
5 V model
95% HR
10,000 ft.
40,000 ft.
Convection: 155 W
420 LFM: 275 W
Convection: 250 W
420 LFM: 450 W
Convection: 300 W
420 LFM: 450 W
1.28 million hours
Cooling
12 V & 15 V models
24 V, 30 V & 48 V models
Reliability
MTBF according to Telcordia -SR332-Issue 3
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
BCD.00335_AE
© 2017 Bel Power Solutions & Protection
4
MBC450 Series
Figure 1. Output Power Vs. Temperature Derating Curves
PARAMETER
Conducted Emissions
Radiated Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Harmonic Current
AC Flicker
DESCRIPTION / CONDITION
EN55022-B, CISPR22-B, FCC PART15-B
EN55022-B, CISPR22-B, FCC PART15-B
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-3-2
EN61000-3-3
SPECIFICATION
To be controlled in end system
Level-3
Level-3
Level-3
Level 3
Class D
Pass
PARAMETER
Isolation Voltage
Safety Standards
Agency Approvals
CE mark
DESCRIPTION / CONDITION
Input to Output, 2MOPP
Input to Earth,1MOPP
Output to GDN for type BF
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, Nemko-CCL
Complies with LVD Directive
SPECIFICATION
5940 VDC
2121 VDC
1500 VAC
tech.support@psbel.com
MBC450 Series
PARAMETER
Power Good Signal
Remote Sense
Remote on/off
DESCRIPTION / CONDITION
TTL signal goes high after main output is within regulation band, delay is 0.1 to 0.3 s
Compensates for 200 mV drop
To turn on PSU short remote pin to ground
5
CONNECTOR
AC Input Connector
PIN
J1
DESCRIPTION / CONDITION
Pin 1
Pin 3
Pin 5
Lug 1
Lug 2
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
Pin 9
Pin 10
Pin 1
Pin 2
AC line
AC neutral
Earth
+V1
RTN
NC
Power Fail
Power Good
DC Return
+5Vstby
+VE Remote Sense
-VE Remote Sense
CS
DC Return
Remote On/Off
+VE
-VE
MANUFACTURER / PN
Tyco: 1-1123724-3
Mating: 1-1123722-5
6-32 inches Screw Pan HD
Mating: 16 AWG wire crimped to Ring Tongue Terminal AMP: 8-
31886-1
DC Output Connector
J2
Signals
J3
6
Molex: 22-23-2081
Mating: 22-01-2087; Pins: 08-50-0113
Fan
Earth
6
7
J4
J5
7
Mating Connector: Molex 22-01-2025
Pins = 08-50-0113
Molex: 19705-4301
Mating: 190030001
PSU is supplied with J3, pin-9 and pin-10 shorted to enable main output without remote on/off feature
The J5 (Earth) spade connector can be used for U-Channel option products only. When fan options are required the
earth connection provided in the input AC connector should be used (Pin 5 – J1)
PARAMETER
Weight
Dimensions
DESCRIPTION / CONDITION
900 g (1.98 lbs)
101.6 x 165.0 x 41.0 mm (4.0 x 6.5 x 1.6 inch)
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
BCD.00335_AE
© 2017 Bel Power Solutions & Protection