For user manuals and dimensional drawings, visit the product page resources tab on ni.com.
Last Revised: 2012-10-29 14:54:43.0
NI SCXI-1100
32-Channel, ±10 V Analog Input Module
166 kS/s maximum sampling rate at full bandwidth with gain up to 100
6.6 kS/s maximum sampling rate with 10 kHz filter
3 S/s maximum sampling rate with 4 Hz filter
NI-DAQmx measurement services software to simplify configuration and
measurements
Overview
The NI SCXI-1100 analog input module is an economical solution for millivolt, volt, and current inputs. All 32 channels are multiplexed into a single software-programmable gain
instrumentation amplifier (PGIA) and jumper-selectable lowpass filter. Because each module multiplexes the 32 channels into a single channel of the DAQ device, you can add
modules to increase channel count. For thermocouple measurements, the NI SCXI-1102 offers gain and filter settings on a per channel basis and provides better performance and
higher sampling rates.
An SCXI chassis is required to house the SCXI modules. The NI SCXI-1000 chassis can hold up to four modules while the NI SCXI-1001 can hold up to 12. You can choose NI X
Series, M Series, E Series, or S Series DAQ modules to control the chassis.
Back to Top
Support and Services
System Assurance Programs
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI,
CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely
assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application
development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled.
When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended
maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty
coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
Calibration
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your
measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more
about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Technical Support
Get answers to your technical questions using the following National Instruments resources.
Support
- Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales
offices around the world and speak the local language.
Discussion Forums
- Visit forums.ni.com for a diverse set of discussion boards on topics you care about.
Online Community
- Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.
Repair
1/7
www.ni.com
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained
technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
Training and Certifications
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills
to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
Classroom training in cities worldwide
- the most comprehensive hands-on training taught by engineers.
On-site training at your facility
- an excellent option to train multiple employees at the same time.
Online instructor-led training
- lower-cost, remote training if classroom or on-site courses are not possible.
Course kits
- lowest-cost, self-paced training that you can use as reference guides.
Training memberships
and training credits - to buy now and schedule training later.
Visit ni.com/training for more information.
Extended Warranty
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may
change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
OEM
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM
customers, visit ni.com/oem.
Alliance
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than
700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
Back to Top
Detailed Specifications
This topic lists the specifications for the SCXI-1102/B/C modules. These specifications are typical at 25
°
C unless otherwise noted.
Analog Input
Input Characteristics
Number of channels
Input signal ranges
Input damage level
Powered on
Powered off
Inputs protected
Transfer Characteristics
Nonlinearity
Offset error
Gain = 1
After calibration
Before calibration
Gain = 100
After calibration
Before calibration
Gain error (relative to calibration reference)
Gain = 1
After calibration
Before calibration
Gain = 100
After calibration
Before calibration
0.020% of reading max
0.1% of reading
0.015% of reading max
0.04% of reading
15 μV max
100 μV
300 μV max
600 μV
0.005% FSR
±42 VDC
±27 VDC
CH<0..31>, CJ SENSOR
32 differential
±100 mV (gain = 100) or ±10 V (gain = 1)
2/7
www.ni.com
Amplifier Characteristics
Input impedance
Normal powered on
Powered off
Overload
Input bias current
Input offset current
>1 GΩ
10 kΩ
10 kΩ
±0.5 nA
±1.0 nA
CMRR
Characteristics
1102
1102B
90 dB
75 dB min
1102C
90 dB
75 dB min
50 to 60 Hz, either gain 110 dB
DC, gain 1
DC, gain 100
75 dB min
100 dB min 100 dB min 100 dB min
Output range
Output impedance
Dynamic Characteristics
Bandwidth
Minimum scan interval (per channel, any gain)
±0.012% accuracy
±0.0061% accuracy
±10 V
91 Ω
2 Hz (1102), 200 Hz (1102B), 10 kHz (1102C)
3 μs
10 μs
System noise (RTI)
Characteristics
Gain = 1
Gain = 100
1102
1102B
1102C
50 μV
rms
50 μV
rms
70 μV
rms
5 μV
rms
5 μV
rms
10 μV
rms
Filters
Cutoff frequency (–3 dB)
NMR (60 Hz)
2 Hz (1102), 200 Hz (1102B), 10 KHz (1102C)
40 dB (1102)
Step response (either gain)
Characteristics 1102
To 0.1%
To 0.01%
1s
10 s
1102B
10 ms
1102C
200 μs
100 ms 1 μs
Stability
Recommended warm-up time
Offset temperature coefficient
Gain = 1
Gain = 100
Gain temperature coefficient
20 μV/°C
1 μV/°C
10 ppm/°C
20 min
Power Requirements
5 V supply
±15 V supply (regulated from ±24 V supply)
15 mA max
150 mA max
3/7
www.ni.com
Physical
SCXI-1102/B/C Dimensions
Weight
611 gm (24.6 oz)
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the common-mode voltage.
Signal + common mode
Each input should remain within ±10 V of CH GND
Environmental
Operating temperature
Storage temperature
Humidity
Maximum altitude
Pollution Degree (indoor use only)
0 to 50 °C
–20 to 70 °C
10 to 90% RH, noncondensing
2,000 m
2
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
Online Product Certification
section.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:
EN 61326 EMC requirements; Minimum Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note
For EMC compliance, operate this device with shielded cables.
CE Compliance
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
73/23/EEC
Low-Voltage Directive (safety)
Electromagnetic Compatibility Directive (EMC)
89/336/EEC
Note
Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, visit
ni.com/certification
, search by model number or product line, and click the appropriate link in the Certification column.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers