A high performance AC to DC external power supply family designed for medical applications. The ME10 Medical Series
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition, with isolation levels which satisfy the 2
MOPP requirements; and designed to UL/EN/IEC60601-1-2, 4th edition for EMC
*
. The ME10 Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +70°C, delivering full rated output power up
to +40°C and applicable output power derating up to 70°C. These models are available in desktop and wall-plug versions,
include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors.
*
Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME10A0503F01
ME10A0603F01
ME10A0703F01
ME10A0903F01
ME10A1203F01
ME10A1503F01
ME10A2403F01
ME10A0503N01
ME10A0603N01
ME10A0703N01
ME10A0903N01
ME10A1203N01
ME10A1503N01
ME10A2403N01
ME10A0503Q01
ME10A0603Q01
ME10A0703Q01
ME10A0903Q01
ME10A1203Q01
ME10A1503Q01
ME10A2403Q01
Volts
5.0V
5.9V
7.5V
9.0V
12.0V
15.0V
24.0V
5.0V
5.9V
7.5V
9.0V
12.0V
15.0V
24.0V
5.0V
5.9V
7.5V
9.0V
12.0V
15.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.1A
1.0A
0.8A
0.5A
2.0A
1.6A
1.3A
1.1A
1.0A
0.8A
0.5A
2.0A
1.6A
1.3A
1.1A
1.0A
0.8A
0.5A
Output
Power
10W
10W
10W
10W
12W
12W
12W
10W
10W
10W
10W
12W
12W
12W
10W
10W
10W
10W
12W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Output
Connector
Input
Configuration
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
www.slpower.com
ME10 Rev 2.8
09-NOV-17
1
ME10
10W-12W Single Output External Power
Medical Series
Model Selection
(continued)
Model
Number
ME10A0503B01
ME10A0603B01
ME10A0703B01
ME10A0903B01
ME10A1203B01
ME10A1503B01
ME10A2403B01
ME10A0503C01
ME10A0603C01
ME10A0703C01
ME10A0903C01
ME10A1203C01
ME10A1503C01
ME10A2403C01
Notes:
Volts
5.0V
5.9V
7.5V
9.0V
12.0V
15.0V
24.0V
5.0V
5.9V
7.5V
9.0V
12.0V
15.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.1A
1.0A
0.8A
0.5A
2.0A
1.6A
1.3A
1.1A
1.0A
0.8A
0.5A
Output
Power
10W
10W
10W
10W
12W
12W
12W
10W
10W
10W
10W
12W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Output
Connector
Input
Configuration
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades
(North American
Blade included)²
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North American
Blades³
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
Parameter
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
(Input to Ground)
Patient Leakage
Current
(Output to Earth)
Output Power
Efficiency
No Load Input
Power
Ripple and Noise
Output Voltage
Transient
Response
Regulation
Specification
100-240Vac, ±10%, 47-63Hz, 1∅
115Vac: 0.45A, 230Vac: 0.22A
264Vac, cold start: will not exceed 40A
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
<500μA@264Vac, 60Hz, NC
<1mA@264Vac, 60Hz, SFC
<100μA@264Vac, 60Hz, NC
<500µA@264Vac, 60Hz, SFC
10 to 12W continuous – See models chart for
specific voltage model ratings.
>87%, typical
<0.1W per DoE Efficiency Level VI
Requirements
See models chart
See models chart
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load,
∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
See models chart
Operating: 0.003g/Hz, 1.5grms overall, 3 axes,
10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 min. per axis,
3 axes and Sine waveform, Vib. Freq.
/acceleration: 10-500Hz/1g, sweep rate of 1 oct
/min., Vib. time of 10 sweeps / axes, 3 axes
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.
ME10 Rev 2.8
09-NOV-17
Parameter
Turn On Time
Hold-up Time
Overtemperature
Protection
Overload Protection
Short Circuit
Protection
Overvoltage
Protection
Isolation
Safety Standards
Operating
Temperature
Storage
Temperature
Altitude
Relative Humidity
Drop Test
Specification
Less than 700mS @115Vac, full load
20mS min., at full Load, 100Vac input
Will shutdown upon an overtemperature
condition, auto-recovery.
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
130 to 150% of output voltage, hiccup mode
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
EN/IEC/UL60601-1, 3rd edition
-20°C to +70°C. See curve for derating
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
1.4m from table top to wooden platform, 6
faces.
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
>10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
2
Vibration
Shock
MTBF
www.slpower.com
E-Cap Life
ME10
10W-12W Single Output External Power
Medical Series
EMI/EMC Compliance
Parameter
Conducted Emissions:
Radiated Emissions:
Common Mode Noise:
Electro-Static Discharge (ESD)
Immunity on Power ports:
Radiated RF EM Fields Susceptibility
Electrical Fast Transients (EFT) /Bursts:
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Specification
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4
th
Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4
th
Edition, Table 5
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
EN55011/EN61000-3-2, Class A
EN61000-3-3
Voltage Interruptions, Dips, Sags &
Surges
Harmonic Current Emissions
Flicker Test
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Note: Performance criteria are based are defined as following:
A – Normal performance during and after the test
B – Temporary degradation, self-recoverable
C – Temporary degradation, operator intervention required to recover the operation
D – Permanent damage
Mechanical Drawing
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Notes:
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
1. All dimensions in mm.
2. Weight = 110g
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
3. The unit should not be covered or enclosed to protect against excessive case temperature rise.
www.slpower.com
ME10 Rev 2.8
09-NOV-17
3
ME10
10W-12W Single Output External Power
Medical Series
Derating Chart:
14
12
10
Watts
8
6
4
2
0
-20
-10
0
10
20
30
40
50
60
70
5V
7V
12V
24V
Ambient Temperature ͦ
C
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (No. 3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
These are the most common standard connectors. SL Power has the capability to incorporate any non-standard output connector. All output connectors are limited by wattage
range and application type. The SL Power applications team is available to provide professional support and can be contacted here: info@slpower.com.