Data Sheet
50 MHz Programmable Pulse Generators
Models 4033 & 4034
Description
The 4033 and 4034 are high performance pro-
grammable pulse generators for testing digital
systems and circuits based on TTL, CMOS, or
ECL technologies. Both instruments generate
clean and accurate pulses at up to 6 digits reso-
lution with a repetition rate up to 50 MHz, vari-
able pulse widths from 10 ns to 10 s, and pulse
delays from 0 ns to 10 s. Output levels are
adjustable from -10 V to +10 V, with pulse
amplitudes settable from 0.1 Vpp to 10 Vpp into
a 50 ohm load. All parameters, modes, and
functions are programmable via the front panel
or remote control commands. Additionally, the
pulse generators provide selectable complemen-
tary pulse and double pulse generation in contin-
uous, triggered, gated, and counted burst
modes.
Features & Benefits
I
I
Dual-Channel Model 4034
I
Repetition rate of 0.1 Hz to 50 MHz
Flexible trigger modes: Continuous, Triggered
(internal, external, manual), Gated Burst and
External Width
Both channels offer full functionality and all
parameters such as pulse width and transition
time can be set independently
I
Synchronize both channels with the push of a
button
Saves cost and bench space
I
I
Pulse width programmable from 10 ns to 10 s
Transition times (rise and fall times) variable
from 6 ns to 100 ms
Programmable delay and double pulse
Predefined amplitude levels for ECL, TTL, and
CMOS signals
Store up to 99 different test setups with auto
retention of last power down setup
10 Vpp into 50 ohm output
Closed case calibration
Programmable via GPIB and RS-232
SCPI compatible
Three year warranty
I
I
I
Applications
I
I
I
I
Automatic Test Equipment (ATE)
Avionics and radar testing
Switching power supply testing
Characterization of active components
I
I
I
I
I
I
Technical data subject to change
© B&K Precision Corp. 2010
w w w.bk prec ision.c om
50 MHz Programmable Pulse Generators
Models 4033 & 4034
Industry Leading Performance
External Width
In the external width pulse mode, pulse period
and width are determined by an externally
applied signal. The pulse generator then applies
transition and level parameters to this signal in
order to generate the pulse. A power supply
designer for example could use this feature to
shape the width and period of the output signal
to drive the power supply’s FET transistors while
remaining synchronized with its control circuits.
Variable Transitions
For added flexibility, variable rise and fall times
can be programmed from 6 ns to 100 ms.
Various shapes of pulses can be obtained for
applications where parameters such as linearity,
switching times, or reflection times must be ana-
lyzed. Programmable rise and fall times could be
used to measure operational amplifier slew rates
and easily test thresholds of devices and circuits.
Straightforward User Interface
These generators use a menu-driven front panel
keypad and control knob to adjust all parameters.
The bright, easy-to-read display shows all rele-
vant parameters along with a graphical represen-
tation of the output pulse. If an entered parame-
ter is not compatible with the existing setup sta-
tus, the operator is informed by an error mes-
sage.
Users can conveniently select predefined ampli-
tude levels for TTL, ECL, and CMOS signals, or
choose a custom amplitude level for specialized
applications.
Front Panel Interface
Menu Keys – Selects menu
options for waveform parameters
Output On Button – Turns the
main output signal on or off
F1-F5 Key – Select menu options
that appear on the bottom section
of the LCD display
Manual Trigger Key – Sends
manual trigger pulse when
pushed (requires instrument
to be in manual trigger
mode)
4033 Output Section –
Main output and Sync
output
2
w w w .b k p r ec is i o n .c o m
50 MHz Programmable Pulse Generators
Models 4033 & 4034
Specifications
The specifications describe the instrument performance after 30 minutes warm-up period into a 50
Ω
load. All timing characteristics are measured at 50% of
amplitude with fastest edges.
Mode ls
CHANNELS
FREQUENCY
T I M I NG C H A R A C T E R I S T I C S
Range (single pulse)
Range (double
pulse)
PERIOD
Resolution
Accuracy
Jitter
Range
WIDTH
Resolution
Accuracy
Double Pulse
Range
DELAY
Resolution
Accuracy
Range
DUTY CYCLE
Resolution
Accuracy
O U TPU T C HA RAC T ERIS T IC S
High Level Range
Low Level Range
Amplitude Range
AMPLITUDE
Resolution
Accuracy
Aberrations
Output Resistance
Offset Accuracy
O PERA TI N G M O D ES
Continuous
Triggered
Output continuous at programmed period rate
Output quiescent until triggered by an internal, external, GPIB
or manual trigger, then generates one cycle at programmed
period rate
Same as triggered mode except pulses are output for the dura-
tion of the gated signal. The last cycle started is completed
Same as triggered mode for programmed number of cycles from
2 to 999,999 as set by the N-BURST function
Trigger duration and rate sets pulse width and repetition
-9.90 V to +10 V into 50
Ω
load
(-19.80 V to +20 V into open circuit)
-10 V to +9.90 V into 50
Ω
load
(-20 V to +19.80 V into open circuit)
0.1 V to 10 V p-p into 50
Ω
load
(20 V p-p max into open circuit)
3 digits limited to 10 mV
±1% of setting ±10 mV into 50
Ω
<5% + 20 mV into 50
Ω
load, for pulse levels between ±5V
50
Ω
±1% ±25 mV
Power Requirements
Dimensions WxHxD
Net Weight
EMC
Electrical Discharge Immunity
Safety Specifications
Operating Temperature
Storage Temperature
Humidity
One pulse at each selected period up to 50 MHz repetition rate
One pair of pulses at each period up to 25 MHz repetition rate.
Both pulses have the same selected width; the position of
the second pulse set by the delay control.
20 ns to 10 s (50 MHz to 0.1 Hz repetition rate)
40 ns to 10 s (25 MHz to 0.1 Hz repetition rate)
Up to 6 digits, limited to 100 ps
±0.01 %
< 0.01 % of setting +20 ps on Period, Width and Delay
10 ns to (Period – 10 ns)
Up to 6 digits, limited to 100 ps
±(0.5% of setting +500 ps)
±(0.5% of setting +3 ns) for the second pulse
0 ns to (Period – Width – 10 ns)
Up to 6 digits, limited to 100 ps
±(0.5% of setting +500 ps)
1 to 99%
3 digits (0.1%)
Limited by width and pulse accuracy
TRIGGER
INPUT
Linearity
I NT E R N A L TR I G G E R
Range
Resolution
Accuracy
I NP UT A N D O UT P U T
Sensitivity
Minimum Width
Maximum Rate
Input Impedance
Input Protection
Range
Resolution
Slope Selection
200 mVpp minimum
10 ns
50 MHz
10 kΩ
±15V DC plus peak AC
Selectable from -10 V to +10 V
3 digits limited to 10 mV
Positive or Negative
A TTL level pulse at the programmed period. Output impedance
is 50
Ω,
protected against short circuit and up to ±15 V
accidental input. The high level is >2 V into 50
Ω
and with 3.5
ns typical transition times.
100 ns to 100 s
4 digits limited to 100 ns
±0.01%
Resolution
Accuracy
4033
1
0.1 Hz to 50 MHz
4 03 4
2
T RANS IT IO N T IME S
<6 ns to 100 ms variable. Leading and trailing edges settable
separately and limited to 20:1 ratio between settings into one of
the following ranges: 5 ns-100 ns; 50 ns-1.0 us; 500 ns-10 us;
5.0 us-100 us; 50 us-1.0 ms; 500 us-10 ms, 5 ms – 100 ms
3 digits limited to 10 ps
±(5% of setting +2 ns)
<5% deviation from a straight line between 10% and 90%
points, for transitions > 50 ns
Range
SYNC OUTPUT
R E M OT E P R O G R AM M I N G
Interface
GPIB Function Codes
G E NE R A L
Memory
Non-volatile, stores up to 99 complete panel settings. Last user
setup also retained at power down.
100-240 V, ±10%, 48-66 Hz, 50 VA maximum
8.4 x 11.8 x 3.5 inches (213 x 300 x 88 mm)
6.61 lbs (3 kg)
Conforms to EN55011 class B for radiated and conducted
emissions
Conforms to EN55082
Conforms to EN61010, CE Approved
32 °F to 122 ° F (0 °C to 50 °C)
-4 ° F to 140 °F (-20 °C to 60 °C)
90% RH at 32 °F to 86 °F (0 °C to 30 °C)
GPIB and RS-232, IEEE-488.2 and SCPI compatible
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1,DT1, C0, E2
Gated
Burst
External Width
P U L S E F U N C TI ON S
Single
Double
Three Year Warranty
Included accessories : AC Power cord, CD containing instruction manual, RS-232 cable,
test report and certificate of calibration