Data Sheet
Multimeters
Digital Clamp Meters
369B
367A
350B
330B
B&K Precision offers a variety of current clamps
from small to large, for safe non-invasive current
measurements. These meters also measure
voltage and resistance. Some models even
support frequency and capacitance
measurements, and include special features
such as recording capabilities of minimum and
maximum values, peak values, and diode testing.
Model
AC Current
DC Current
AC Voltage
DC Voltage
Resistance
Frequency
Capacitance
True RMS
Continuity
Diode
Count Resolution
Bargraph
Data Hold
Peak Hold
Max Hold
Range Hold
Max/Min Hold
Relative
Auto Off
Supplied Accessories
Optional Accessories
369B
1000 A
1000 A
750 V
1000 V
40 MΩ
√
√
√
√
√
4000
√
√
√
--
√
√
√
30 min.
367A
2000 A
2000 A
750 V
1000 V
40 MΩ
√
√
√
√
√
4000
√
√
√
--
√
√
√
30 min.
350B
1000 A
--
750 V
1000 V
30 MΩ
√
--
--
√
√
3200
√
√
√
--
√
--
--
10 min.
330B
1000 A
--
750 V
1000 V
2000 MΩ
--
--
--
√
√
2000
--
√
--
√
--
--
--
--
Common Features & Benefits
n
AC
voltage and current measurements
measurements
continuity
n
Resistance
n
Audible
n
Data
n
Low
hold
protection
battery indicator
n
Overload
n
CE
Approval
One-Year Warranty
Battery, Test leads, Instruction Manual, Carrying Case
TL 1 replacement test leads , TL 2A deluxe test leads, TL 3 accessory tIp kit (for TL 2A),
TL 130A general purpose DMM kit, PR 28A high voltage probe, LC 33 carrying case
Technical data subject to change
© B&K Precision Corp. 2016
www.bkprecision.com
Multimeters
Digital Clamp Meters
Specifications
Model
AC Current (using clamp)
Ranges
Resolution
Basic Accuracy (50-60 Hz)
Overload Protection (for 60 sec max.)
DC Current (using clamp)
Range
Resolution
Basic Accuracy
Overload Protection (for 60 sec max.)
AC Voltage (using test leads)
Range
Resolution
Basic Accuracy
Overload Protection
Input Impedance
DC Voltage (using test leads)
Range
Resolution
Basic Accuracy
Overload Protection
Input Impedance
Resistance
Range
Resolution
Basic Accuracy
Overload Protection
Frequency
Range
Resolution
Basic Accuracy
Overload Protection
Capacitance
Range
Resolution
Basic Accuracy
Overload Protection
General
Jaw Opening
Display
Measure Rate
Operating Temperature
Storage Temperature
Power Requirements
Battery Life
Dimensions
Weight (includes battery)
3 3/4 digit LCD,
4000 count (9999 in frequency)
with 42 segment bargraph
Nominal 2/sec
32° to 122°F, 70% RH, (0° to 50°C)
-4° to 140°F, 80% RH, (-20° to 60°C)
One 9 V battery, alkaline recommended type NEDA 1604A
500 Hours
10.9 x 4 x 1.9” (277 x 102 x 49 mm)
18.9 oz (540 g)
2.24” (57 mm)
3 1/2 digit LCD
3200 count
with 34 segment bargraph
3 3/4 digit LCD
2000 count
Nominal 2.5/sec
4n F, 40 nF, 400 nF, 4 µF, 40 µF
0.001 nF. 0.01 nF, 0.1 nF, 0.001 µF, 0.01 µF
±(1.5% reading + 4 digits)
500 V DC/AC RMS
4 nF, 40 nF, 400 nF, 4 µF, 40 µF
0.001 nF. 0.01 nF, 0.1 nF, 0.001 µF, 0.01 µF
±(1.5% reading + 4 digits)
500 V DC/AC RMS
--
--
--
--
--
--
--
--
100 Hz, 1 kHz, 10 kHz, 100 kHz, 400 kHz
0.01 Hz, 0.1 Hz, 1 Hz, 10 Hz, 100 Hz
±(0.1% reading + 3 digits)
500 V DC/AC RMS
100 Hz, 1 kHz, 10 kHz, 100 kHz, 400 kHz
0.01 Hz, 0.1 Hz, 1 Hz, 10 Hz, 100 Hz
±(0.1% reading + 3 digits)
500 V DC/AC RMS
320 Hz, 3.2 kHz, 32 kHz
0.1 Hz, 1 Hz, 10 Hz
±(1% reading + 4 digits)
500 V DC/AC RMS
--
--
--
--
400
Ω,
4 kΩ, 40 kΩ, 400 k, 4 M, 40 MΩ
0.1
Ω,
1
Ω,
10
Ω,
100
Ω,1
kΩ, 10 kΩ
±(1% reading + 2 digits)
500 V DC/ AC RMS
400
Ω,
4k
Ω,
40 kΩ, 400 k, 4 M 40 MΩ
0.1
Ω,
1
Ω,
10
Ω,
100
Ω,1
kΩ, 10 kΩ
±(1% reading + 2 digits)
500 V DC/ AC RMS
320
Ω,
3.2 kΩ, 32 kΩ, 320 kΩ, 3.2 MΩ, 30 MΩ
0.1
Ω,
1
Ω,
10
Ω,
100
Ω,1
kΩ, 10 kΩ
±(1% reading + 3 digits)
500 V DC/AC RMS
200
Ω,
20 kΩ, 200 kΩ, 20 MΩ, 2000 MΩ
0.1
Ω,
1
Ω,
10
Ω,
10 kΩ, 100 kΩ
±(1% reading + 3 digits)
1000 V DC/750 AC RMS
400 mV, 4 V, 40 V, 400 V, 1000 V
100 µV, 1 mV, 10 mV, 100 mV, 1 V
±(0.5% reading + 1 digits)
1000 V DC/750 V RMS
10 MΩ
400 mV, 4 V, 40 V, 400 V, 1000 V
100 µV, 1 mV, 10 mV, 100 mV, 1 V
±(0.5% reading +1 digits)
1000 V DC/750 V RMS
10 MΩ
320 mV, 3.2 V, 32 V, 320 V, 1000 V
100 µV, 1 mV, 10 mV, 100 mV, 1 V
±(0.5% reading + 1 digits)
1000 V DC/750 V RMS
10 MΩ
1000 V
1V
±(0.5% reading + 1 digits)
1000 V DC/750 V RMS
10 MΩ
400 mV, 4 V, 40 V, 400 V, 750 V
100 µV, 1 mV, 10 mV, 100 mV, 1 V
±(1.5% reading + 4 digits)
1000 V DC/750 V RMS
10 MΩ
400 mV, 4 V, 40 V, 400 V, 750 V
100 µV, 1 mV, 10 mV, 100 mV, 1 V
±(1.5% reading + 4 digits)
1000 V DC /750 V RMS
10 MΩ
3.2 V, 32 V, 320 V, 750 V
1 mV, 10 mV, 100 mV, 1 V
±(1.5% reading + 4 digits)
1000 V DC /750 V RMS
10 MΩ
750 V
1V
±(1.5% reading + 5 digits)
1000 V DC /750 V RMS
4.5 MΩ
400 A, 1000 A
100 mA, 1 A
±(1.5% reading + 5 digits)
1200 A
400 A, 2000 A
100 mA, 1 A
±(1.5% reading + 5 digits)
2000 A
--
--
--
--
--
--
--
--
400 A, 1000 A
100 mA, 1 A
±(1.5% reading + 5 digits)
1200 A
400 A, 2000 A
100 mA, 1 A
±(1.5% reading + 5 digits)
1200 A
32 A, 320 A, 1000 A
10 mA, 100 mA, 1 A
±(1.5% reading + 5 digits)
1200 A
200 A, 1000 A
100 mA, 1 A
±(1.5% reading + 5 digits)
1200 A
369B
367A
350B
330B
2
v012516