MINT1150
150W Single Output
Medical Series
Features
2” x 4” x 1.3” Package
For 1U Applications
150W w/air, 100W convection cooled
Universal Input 90-264 VAC
Power Fail/Output Good Signal
Approved to CSA/EN/IEC/UL60601-1,
3
rd
Edition
2 x MOPP Input to Output Isolation
3 Year Warranty
RoHS Compliant
Description
The MINT1150 family is a high power density for a power supply in a 2” x 4” size. Approved to EN/IEC/UL60601-1, 3rd
edition, with isolation levels which satisfy the 2 MOPP requirements. The MINT1150 family is ideal for Portable Medical
Devices, and many other applications where medical certifications, power density and cost are critical. The MINT1150
operates at universal input range of 90 to 264Vac and wide temperature range -10°C to +70°C, delivering full rated output
power up to +50°C. In addition, these models feature Power Fail and DC OK signals.
Model Selection
Model
Number
MINT1150A1206K01
MINT1150A1506K01
MINT1150A2406K01
MINT1150A4806K01
MINT1150A5606K01
Notes:
*
Volts
12V
15V
24V
48V
56V
Output Current
w/200LFM air
Convection*
12.5A
10.0A
6.25A
3.13A
2.68A
8.33A
6.67A
4.17A
2.08A
1.79A
Ripple &
Noise**
0.5%RMS, 1.2% pk-pk
0.5%RMS, 1.0% pk-pk
0.5%RMS, 1.0% pk-pk
0.5%RMS, 1.0% pk-pk
0.5%RMS, 1.0% pk-pk
Total
Regulation
5%
5%
5%
5%
5%
OVP
Threshold**
14.0
1.1V
18.0 ± 1.5V
28.0 ± 2.5V
55.0 ± 4.0V
<59.9V
Maximum output power is 95 Watts for input voltage of 90-105VAC at 50°C convection. For input voltage of 105Vac or more, the total power is
100 Watts at 50°C convection.
** Measured with noise probe directly across output terminals, and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
General Specifications
AC Input
100-240Vac, ±10%, 47-63Hz, 1
120–370Vdc
Turn On Time
Less than 2 sec. @115Vac (inversely
proportional to input voltage and thermistor
temperature)
>12mS at full load, 120Vac
Input Current
115Vac: 2A, 230Vac: 1A
Hold-up Time
Inrush Current
264Vac, cold start: will not exceed 50A
Signals
AC Power Fail, DC OK
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
1
MINT1150
150W Single Output
Medical Series
General Specifications
(continued)
Input Fuses
Earth Leakage
Current
Efficiency
F1, F2: 4A, 250Vac fuses provided on all
models
<300μA@264Vac, 60Hz, NC
Overload
Protection
Short Circuit
Protection
Overvoltage
Protection
Switching
Frequency
Isolation
Operating
Temperature
Temperature
Derating
Storage
Temperature
Altitude
Hiccup Mode
Provided - no damage will occur if the output is
shorted. Hiccup Mode
OVP firing reduces output voltage to <50% of
nominal in <50mS. See chart for trip range.
PFC: Variable 30-400kHz.
Main Converter: Variable 35-180kHz, 65-70kHz
at full load.
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 500Vac
-10°C to +70°C
89% typical at 115Vac
150W continuous with 200 lfm airflow, 100W
convection cooled – See chart for specific
voltage model ratings.
50% load step.
i/t:
<0.2A/μS
Max Volt Deviation = 3%
Output Power
Transient
Response
Ripple and Noise
See chart
Output Voltage
Voltage
Adjustability
Minimum Load
See chart
Derate output power linearly above 50°C to 50%
at 70°C
+/-5% from nominal
-40°C to +85°C
Not required
Operating: -500 to 10,000 ft
Non-operating: -500 to 40,000 ft.
Total Regulation
+/- 5%. See chart
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g2/Hz, 5.0grms overall,
3 axes, 1 hr/axis
W: 2.0” x L: 4.0” x H: 1.3”
Relative Humidity
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
EN/CSA/UL/IEC60601-1, 3
rd
Edition
Vibration
Shock
Dimensions
Safety Standards
Weight
183g
MTBF
640,000 hours at 100W convection, 1,500,000
hours at 150W with 200LFM air
Auxiliary Signals
AC Power Fail:
During normal operations, stays HIGH.
Power Fail:
Goes LOW with 5 mS warning before loss
of output power due to AC failure
Open collector logic signal goes and stays HIGH
100mS to 500mS after main output reaches
regulation.
DC OK:
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
2
MINT1150
150W Single Output
Medical Series
EMI/EMC Compliance
Conducted Emissions
Radiated Emissions
Static Discharge Immunity
Radiated RF Immunity
EFT/Burst Immunity
Line Surge Immunity
Conducted RF Immunity
Power Frequency Magnetic Field Immunity
Voltage Dip Immunity
Line Harmonic Emissions
Flicker Test
EN55011/22 Class B, FCC Part 15, Subpart B, Class B
EN55011/22 Class A, FCC Part 15, Subpart B, Class A w/6db margin
EN61000-4-2, Criteria A, 6kV Contact Discharge, 8kV air discharge
EN61000-4-3, 3V/m. Criteria A
EN61000-4-4, 2kV/5kHz, Criteria A
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
EN61000-4-6, 3Vrms, Criteria A
EN61000-4-8, 3A/m, Criteria A
EN61000-4-11, 0% Vin, 0.5cycle; 40% Vin, 5 cycles; 70% Vin, 25 cycles;
Criteria A
EN61000-3-2, Class A, B, C, & D
EN61000-3-3, Complies (dmax<6%)
Mechanical Drawing
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
3
MINT1150
150W Single Output
Medical Series
Connector Information
Input Connector
J100
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
Mating Connector:
Molex 09-50-3031
Pins= 08-52-0072
Ground
J101
0.187” FASTON TAB
DC Output Connector
J200
PIN 1) +Vout PIN 4) -Vout
PIN 2) +Vout PIN 5) -Vout
PIN 3) +Vout PIN 6) -Vout
Mating Connector:
AMP 640250-6
Pins = 640252-1
Signal Connector
J300
PIN 1) PF/DC OK
PIN 2) Common
Mating Connector:
Molex 1375820-2
Pins = 1375819
Mating Connector:
Molex 01-90020005
Power Fail/DC OK Signals – J300
J300–Pin 1
J300 – Pin 2
AC P
OWER
F
AIL
S
IGNAL
DC OK S
IGNAL
Isolation Specifications
Parameter
Insulation Safety Rating
Conditions/Description
Input/Ground
Input/Output
Output/Ground
Input/Ground
Input/Output
Output/Ground
Min
Nom
2 MOPP
1 MOPP
N/A
Max
Units
Electric Strength Test Voltage
1800
4000
500
-
-
Vac
Vac
Vac
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
4
MINT1150
150W Single Output
Medical Series
Input Specifications
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Input Voltage
Turn-On Input Voltage
Turn-Off Input Voltage
Input Frequency
Inrush Current Limitation
Power Factor
Conditions/Description
Ramping up
Ramping down
Min
90
Nom
115/230
82.7
67.0
Max
264
Units
Vac
Vac
Vac
47
264Vac, cold start
Vi
nom,
Io
nom
Vi
nom,
Io
nom
MINT1150A1206K01
MINT1150A2406K01
MINT1150A4806K01
MINT1150A5606K01
-
0.9
50/60
-
-
63
50
-
Hz
A
Efficiency
-
89%
-
%
Protection
Parameter
Input Fuse
Input Transient Protection
Conditions/Description
Not user accessible
2KV(CM) and 1KV(DM) surge
No-load
Min
Nom
Max
2
Normal
Hiccup Mode
Hiccup Mode
Units
KV (CM)
Output
Short Circuit
Overload
Latch style
Automatic power shutdown at TC =155°C
Overvoltage Protection
Over temperature Protection
See Models chart for trip ranges
Characteristic Curves
Output vs. Temperature
100W convection cooled and 150W continuous with 200 LFM airflow, Derating output power to 50% at 70C.
www.slpower.com
MINT1150 REV 1.4, 13-NOV-12
5