The
MBC40 Series
of open-frame medical
power supplies, with its wide universal 90-264
VAC input range, is available at 40 W of output
power and a variety of single and multiple output
voltages.
The MBC series was designed to 3rd edition
medical approvals and provides 2x MOPP
(Means of Patient Protection) isolation for class
1 and class 2 installations.
These medical power supplies are ideal for
monitoring, home health equipment as well as
surgical devices
.
40 W Convection Cooled
90-264 VAC Input
20 to 50°C Full Load Operation
2x MOPP
4 x 2 x 1.2” (101.6 mm x 50.8 mm x 30.5 mm)
No Minimum Load Required - Single Output Models
No Load Power < 0.3 W
IEC Protection Class Options:
o
o
Class I: Earthing Tab J4 (no suffix)
Class II: No Earthing Tab (-2 suffix)
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
RoHS Compliant
CE Marked
Dialysis
Monitoring
Pumps
Surgical Devices
Home Health
Portable Device
North America
+1-866.513.2839
Asia-Pacific
+86.755.29885888
Europe, Middle East
+353 61 225 977
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00286_AC
MBC40 Series
OUTPUT VOLTAGE OUTPUT CURRENT
(VDC)
1
MAX (A)
5.1
8.0
12
15
24
48
5.2
MBC40-3000G
12.5
-12.8
5.2
MBC40-3001G
23.8
-12.8
5.2
MBC40-3002G
14.6
-14.8
3.3
MBC40-3003G
Warranty 2 years.
NOTES:
1
2
MODEL
MBC40-1005G
MBC40-1012G
MBC40-1015G
MBC40-1024G
MBC40-1048G
MINIMUM
LOAD (A)
3
0.0
0.0
0.0
0.0
0.0
0.5
0.1
0.0
0.5
0.1
0.0
0.5
0.1
0.0
1.0
0.1
0.0
RIPPLE & NOISE
2
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1.5%
1%
1%
TOTAL
REGULATION
± 0.8%
± 0.8%
± 0.8%
± 0.8%
± 0.8%
± 0.8%
± 5.3%
± 5.3%
± 0.8%
± 5.3%
± 5.3%
± 0.8%
± 5.3%
± 5.3%
± 0.8%
± 5.3%
± 5.3%
OVP THRESHOLD
130% Typical
130% Typical
130% Typical
130% Typical
130% Typical
130% Typical
3.5
2.7
1.7
0.83
6.0
2.0
0.5
6.0
1.0
0.5
6.0
1.5
0.5
6.0
3.0
0.5
130% Typical
130% Typical
130% Typical
5.2
-12.8
3
Maximum outputs for each output. Max power rating should not be exceeded.
Ripple and Noise may be 3% at initial start-up at -20 C. Output noise measurement is made with a 20 MHz bandwidth using a 6”
twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor.
Minimum load specified to meet cross regulation.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00286_AC
MBC40 Series
Specifications are for nominal input voltage, 25°C and max load unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
4
Input Current
No Load Power
Inrush Current
Leakage Current
Switching Frequency
1
DESCRIPTION / CONDITION
Universal Input
SPECIFICATIONS
90 - 264 VAC
47 to 400 Hz
120 VAC @ 40 W
230 VAC @ 40 W
Single output models
Multiple output models
120 VAC:
230 VAC:
120 VAC:
230 VAC:
Typical
0.85 A
0.45 A
< 0.3 W
< 0.5 W
30 A max.
60 A max.
< 140
μA
< 250
μA
67 kHz
Safety Approved: 47 to 63 Hz.
PARAMETER
Efficiency
Hold Up Time
Output Power
Line Regulation
Load Regulation
Transient Response
Rise Time
Set Point Tolerance
Voltage Adjustment
Remote Sense
Over Current Protection
Over Voltage Protection
Short Circuit Protection
DESCRIPTION / CONDITION
Typical
120/230 VAC input
SPECIFICATIONS
85%
6 ms
40 W
V1:
V2 & V3:
Main output 50 to 100% load change,
50/60 Hz, 50% duty cycle, 0.1A / uSec
V1:
V2 & V3:
V1
V1
+/-0.3%
+/-0.5%
+/-5%
< 10%,
recovery time < 5 ms
< 100 ms
± 3%
± 5%
± 10%
0.5 VDC compensation
130% typical above rating
Typical (V1 only)
Short term, auto recovery
130%
PARAMETER
Operating Temperature
Storage Temperature
Humidity
Altitude
Reliability
DESCRIPTION / CONDITION
Full Load
Derate linearly to 75% at 70° C
Non Condensing
Operating:
Non-Operating:
MTBF according to Telcordia -SR332-Issue 3
SPECIFICATIONS
-20 to 50°C
50 to 70°C
-40 to 70° C
95%
10,000 ft.
40,000 ft.
1.87 million hours
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00286_AC
MBC40 Series
PARAMETER
Conducted Emissions
Radiated Emissions
Harmonic Current
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
DESCRIPTION / CONDITION
EN55022-B, CISPR22-B, FCC PART15-B
EN55022-B, CISPR22-B, FCC PART15-B
EN61000-3-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
SPECIFICATIONS
Level B
To be controlled in end system
Class A
Level 3
Level 3
Level 3
Level 3
PARAMETER
Isolation Voltage
Safety Standards
Agency Approvals
CE mark
DESCRIPTION / CONDITION
Input to Output, 2M0PP
Input to Ground, 1M0PP
Output to Ground
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, UL, C-UL
Complies with LVD Directive
SPECIFICATION
Min. 5700 VDC
1500 VAC
500 VAC
Figure 1 - Output Power vs. Temperature
CONNECTOR
AC Input Connector
PIN
J1
DC Output Connector
J2
Remote Sense
Earthing Tab
J3
J4
DESCRIPTION / CONDITION
Pin 1
AC Neutral
Pin 2
AC Line
Pin 1,2
V1
Pin 3,4
RTN
Pin 5
V3
Pin 6
V2
Pin 1
+V1 Sense
Pin 2
-V1 Sense
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Tyco: 640445-6 or equivalent
Mating: 647402-6; Pins: 3-647409-1
Molex: 22-23-2021 or equivalent
Mating: 22-01-2021
Molex: 19705-4301
Mating: 190030001
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00286_AC
MBC40 Series
PARAMETER
Weight
Dimensions
DESCRIPTION / CONDITION
150 g (0.33 lbs.)
101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch)
Figure 2 - Mechanical Drawing MBC40-1xxxG
Figure 3 - Mechanical Drawing MBC40-3xxxG
NUCLEAR AND MEDICAL APPLICATIONS
- Products are not designed or intended for use as critical components in life support systems, equipment used in
hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS
- The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00286_AC