The
MBC150 Series
of open-frame medical
power supplies, with its wide universal 90-264
VAC input range, is available at 150 W of output
power and a variety of single output voltages. The
MBC series was designed to 3rd edition medical
approvals and provides 2x MOPP (Means of
Patient Protection) isolation for class 1 and class
2 installations.
150 W with Forced-Air Cooling
90-264 VAC Input
-20 to 50°C Full Load Operation
2x MOPP Isolation
4.0 x 2.0 x 1.3 inch (101.6 x 50.8 x 33.6 mm)
No Minimum Load Required
Fan output, 12 VDC @ 0.5 A standard
IEC Protection Class Options:
o
o
Class I: Earthing Tab J4 (no suffix)
Class II: No Earthing Tab (-2 suffix)
Dialysis
Monitoring
Pumps
Surgical Devices
Home Health
Portable Devices
These medical power supplies are ideal for
monitoring, home health equipment as well as
surgical devices.
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
RoHS Compliant
CE Marked
North America
+1-866.513.2839
Asia-Pacific
+86.755.29885888
Europe, Middle East
+353 61 225 977
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00288_AC
MBC150 Series
MODEL
MBC150-1005G
MBC150-1005G-2
MBC150-1T05G
MBC150-1T05G-2
MBC150-1012G
MBC150-1012G-2
MBC150-1T12G
MBC150-1T12G-2
MBC150-1015G
MBC150-1015G-2
MBC150-1T15G
MBC150-1T15G-2
MBC150-1024G
MBC150-1024G-2
MBC150-1T24G
MBC150-1T24G-2
MBC150-1048G
MBC150-1048G-2
MBC150-1T48G
MBC150-1T48G-2
Vfan (all models)
Warranty 2 years.
NOTES:
1
2
3
4
OUTPUT VOLTAGE
(VDC)
5.0
5.0
5.0
5.0
12
12
12
12
15
15
15
15
24
24
24
24
48
48
48
48
12
MAX LOAD
CONVECTION
1
16.0 A
16.0 A
16.0 A
16.0 A
8.33 A
8.33 A
8.33 A
8.33 A
6.67 A
6.67 A
6.67 A
6.67 A
4.17 A
4.17 A
4.17 A
4.17 A
2.08 A
2.08 A
2.08 A
2.08 A
0.5 A
3
MAX LOAD
300 LFM
1,2
16.0 A
16.0 A
20.0 A
20.0 A
12.5 A
12.5 A
12.5 A
12.5 A
10.0 A
10.0 A
10.0 A
10.0 A
6.25 A
6.25 A
6.25 A
6.25 A
3.13 A
3.13 A
3.13 A
3.13 A
0.5 A
3
MINIMUM
LOAD (A)
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
RIPPLE &
NOISE
4
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
CONNECTOR
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
TOTAL
REGULATION
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
20%
Combined power from main output and Vfan should not exceed total power rating.
Fan output tolerance is ± 20%.
Peak current for fan output is 1A.
Ripple is 2% up to 20% load and less than 1% above 20% load. Output noise measurement is made with a 20 MHz bandwidth
using a 6” twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00288_AC
MBC150 Series
Specifications are for nominal input voltage, 25°C and max load unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
Input Current
Inrush Current
Leakage Current
Power Factor
Switching Frequency
DESCRIPTION / CONDITION
Universal
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
PFC converter (variable)
Resonant converter (variable)
SPECIFICATION
90-264 VAC / 120-390 VDC
47 to 63 Hz
1.7 A max.
0.85 A max.
35 A max.
65 A max.
< 140
μA
< 250
μA
0.99
0.95
35 - 250 kHz, 90 kHz typical
35 - 250 kHz, 90 kHz typical
PARAMETER
Output Power
Efficiency
Hold Up Time
Line Regulation
Load Regulation
Transient Response
Rise Time
Set Point Accuracy
Voltage Adjustment
Over Current Protection
Over Voltage Protection
Short Circuit Protection
DESCRIPTION / CONDITION
Peak Power 170 W for 0.2 s
120 VAC:
230 VAC:
120 VAC:
230 VAC:
SPECIFICATION
150 W
84% typical
86% typical
6 ms
10 ms
+/-0.5%
+/-2.0%
< 10%,
recovery time < 5 ms
< 100 ms
± 1%
±3%
110% typical above rating
Main output 50 to 100% load change,
50 Hz, 50% duty cycle, 0.1A / uSec
Main output
V1
V1
Short term, Automatic recovery
110 to 150%
PARAMETER
Operating Temperature
Storage Temperature
DESCRIPTION / CONDITION
Refer to derating curve
Start-up is guaranteed
Convection:
300 LFM:
Convection:
300 LFM:
5 V model
Cooling
Other models
Humidity
Altitude
Reliability
Non Condensing
Operating:
Non-Operating:
MTBF according to Telcordia -SR332-Issue 3
SPECIFICATION
-20 to 70°C
-20 to 0°C
-40 to 70° C
80 W
100 W
100 W
150 W
95%
10,000 ft.
40,000 ft.
2.4 million hours
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00288_AC
MBC150 Series
PARAMETER
Conducted Emissions
Radiated Emissions
Harmonic Current
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
DESCRIPTION / CONDITION
EN55022-B, CISPR22-B, FCC PART15,
EN50082-1
EN55022-B, CISPR22-B, FCC PART15-B
EN61000-3-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
SPECIFICATION
Class B
To be controlled in end system
Class D
Level 3
Level-3
Level-3
Level-3
PARAMETER
Isolation Voltage
Agency Approvals
CE mark
DESCRIPTION/CONDITION
Input to Output:
Input to Ground:
Approved to the latest edition of the following
standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1
Complies with LVD Directive
SPECIFICATION
Min. 5900 VDC
2120 VDC
Figure 1 - Output Power Vs. Temperature
CONNECTOR
AC Input Connector
PIN
J1
DESCRIPTION / CONDITION
Pin 1
AC Line
Pin 2
AC Neutral
Pin 1,2
Pin 3,4
Pin 1
Pin 2
V1
RTN
VFAN (12 V / 0.5 A)
RTN
DC Output Connector
J2
Fan
Earthing Tab
J3
J4
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Option-1: Tyco: 1776112-4 or equivalent
Mating: 13 AWG wire
Option-2: JST: B4P-VH-B (LF) (SN) or B4P-VH (LF) (SN) or equivalent
Mating: VHR-4M; Pins: SVH-41T-P1.1
Tyco: 640456-2 or equivalent
Mating: 640440-2
Molex: 19705-4301 or equivalent
Mating: 190030001
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00288_AC
MBC150 Series
PARAMETER
Weight
Dimensions
DESCRIPTION / CONDITION
150 g (0.33 lbs.)
101.6 x 50.8 x 33.6 mm (4.0 x 2.0 x 1.3 inch)
Figure 2 - Mechanical Drawing (Top and Side View)
NOTE:
Air flow over length of supply recommended (either direction) for forced air rating.
NUCLEAR AND MEDICAL APPLICATIONS
- Products are not designed or intended for use as critical components in life support systems, equipment used in
hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS
- The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00288_AC