The
MBC201 Series
of open-frame medical
power supplies, with its wide universal 90-264
VAC input range, is available at 200 W of output
power and a variety of single output voltages.
The MBC series was designed to 3rd edition
medical approvals and provides 2x MOPP
(Means of Patient Protection) isolation for class
1 and class 2 installations.
These medical power supplies are ideal for
monitoring, home health equipment as well as
surgical devices.
160 W Convection Cooled
-20 to 50°C Full Load Operation
2x MOPP
5.0 x 3.0 x 1.5 inch (127.0 x 76.2 x 38.1 mm)
12 V Fan & 5 V Standby Outputs
Inhibit and Power Good Signals
No Minimum Load Required
IEC Protection Class Options:
o
o
Class I: Earthing Tab J4 (no suffix)
Class II: No Earthing Tab (-2 suffix)
Conducted EMI EN 55022-B, FCC Part 15 Level B
Medical Safety Agency Approvals
RoHS Compliant
CE Marked
Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
Portable Equipment
North America
+1-866.513.2839
Asia-Pacific
+86.755.29885888
Europe, Middle East
+353 61 225 977
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00289_AD
MBC201 Series
OUTPUT VOLTAGE
(VDC)
5.0
5.0
5.0
5.0
12
12
12
12
15
15
15
15
24
24
24
24
30
30
30
30
48
48
48
48
12
5
MAX LOAD
CONVECTION
1
26.0 A
26.0 A
26.0 A
26.0 A
13.33 A
13.33 A
13.33 A
13.33 A
10.66 A
10.66 A
10.66 A
10.66 A
6.66 A
6.66 A
6.66 A
6.66 A
5.33 A
5.33 A
5.33 A
5.33 A
3.33 A
3.33 A
3.33 A
3.33 A
0.5 A
1.0 A
MAX LOAD
300 LFM
1,2
26.0 A
26.0 A
35.0 A
35.0 A
16.67 A
16.67 A
16.67 A
16.67 A
13.33 A
13.33 A
13.33 A
13.33 A
8.33 A
8.33 A
8.33 A
8.33 A
6.66 A
6.66 A
6.66 A
6.66 A
4.17 A
4.17 A
4.17 A
4.17 A
0.5 A
1.0 A
MINIMUM
LOAD (A)
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
RIPPLE &
NOISE
3
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
TOTAL
REGULATION
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 2.5%
± 20%
± 5%
MODEL
MBC201-1005G
MBC201-1005G-2
MBC201-1T05G
MBC201-1T05G-2
MBC201-1012G
MBC201-1012G-2
MBC201-1T12G
MBC201-1T12G-2
MBC201-1015G
MBC201-1015G-2
MBC201-1T15G
MBC201-1T15G-2
MBC201-1024G
MBC201-1024G-2
MBC201-1T24G
MBC201-1T24G-2
MBC201-1030G
MBC201-1030G-2
MBC201-1T30G
MBC201-1T30G-2
MBC201-1048G
MBC201-1048G-2
MBC201-1T48G
MBC201-1T48G-2
Vfan (all models)
V s/b (all models)
Warranty 2 years.
NOTES:
1
2
CONNECTOR
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
JST
JST
Screw Terminal
Screw Terminal
3
Combined power from main output, Vfan and Vs/b should not exceed total power rating.
Fan output tolerance is ± 20%. When V1 full load, Vfan needs 20 mA load to be within regulation specification. Peak current for
fan output is 1 A.
Ripple is 2% up to 20% load and less than 1% above 20% load. Output noise measurement is made with a 20 MHz bandwidth
using a 6” twisted pair, terminated with a 10 uF tantalum capacitor in parallel with a 0.1 uF ceramic capacitor.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00289_AD
MBC201 Series
Specifications are for nominal input voltage, 25°C and max load unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
Input Current
Inrush Current
Leakage Current
Power Factor
Switching Frequency
DESCRIPTION / CONDITION
Universal
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
PFC converter (variable)
Resonant converter (variable)
SPECIFICATION
90-264 VAC / 120-390 VDC
47 to 63 Hz
2.4 A max.
1.2 A max.
35 A max.
65 A max.
< 150
μA
< 300
μA
0.99
0.95
35 to 250 kHz, 90 kHz typical
35 to 250 kHz, 90 kHz typical
PARAMETER
Efficiency
Hold Up Time
Output Power
Line Regulation
Load Regulation
Transient Response
Rise Time
Set Point Accuracy
Voltage Adjustment
Over Voltage Protection
Over Current Protection
Short Circuit Protection
DESCRIPTION / CONDITION
120 VAC:
230 VAC:
120 / 230 VAC
Peak Power 250 W for 0.2 s
SPECIFICATION
84% typical
86% typical
10 ms
160 W to 200 W
+/-0.5%
Main output 50 to 100% load change,
50 Hz, 50% duty cycle, 0.1A / uSec
+/-2.0%
< 10%,
recovery time < 5 ms
< 100 ms
± 1%
V1
V1 only
±3%
110 to 150 %
110% typical above rating
Short term, auto recovery
PARAMETER
Operating Temperature
Storage Temperature
DESCRIPTION / CONDITION
Refer to derating curve (
Figure 1
)
Start-up is guaranteed
Convection:
300 LFM:
Convection:
300 LFM:
5 V model
Cooling
Other models
Humidity
Altitude
Reliability
Non Condensing
Operating:
Non-Operating:
MTBF according to Telcordia -SR332-Issue 3
SPECIFICATION
-20 to 70°C
-20 to 0°C
-40 to 70° C
130 W
175 W
160 W
200 W
95%
10,000 ft.
40,000 ft.
1.6 million hours
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00289_AD
MBC201 Series
PARAMETER
Conducted Emissions
Radiated Emissions
Harmonic Current
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
DESCRIPTION / CONDITION
EN55022-B, CISPR22-B, FCC PART15, EN60601-1-2
EN55022-B, CISPR22-B, FCC PART15-B
EN61000-3-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
SPECIFICATION
Class B
To be controlled in end system
Class D
Level 3
Level-3
Level 3
Level-3
PARAMETER
Isolation Voltage
Safety Standards
Agency Approvals
CE mark
DESCRIPTION / CONDITION
Input to Output:
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, UL, C-UL
Complies with LVD Directive
SPECIFICATION
Min. 5900 VDC
PARAMETER
Power Good Signal
Power Fail Signal
Remote ON / OFF
Remote Sense
DESCRIPTION / CONDITION
TTL signal goes high after main output is in regulation band. Delay is 0.1 to 0.3 sec.
TTL signal goes low 1 msec advance before output goes out of regulation due to mains failure
To turn-on power supply short J3 pin 4 to pin 6 (PSU is shipped in this configuration)
Compensates for 200 mV cable drop
CONNECTOR
AC Input Connector
PIN
J1
DESCRIPTION / CONDITION
Pin 1
AC Neutral
Pin 2
AC Line
Pin 1,2,3
Pin 4,5,6
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
RTN
Vout
+VE Remote Sense
VFAN (+12 V/0.5 A)
-VE Remote Sense
Remote ON/OFF
VSTBY (+5 V/1 A, +/-5%)
RTN (Signal)
Power Fail
Power Good
DC Output Connector
J2
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent;
Mating: 09-50-3031; Pins: 08-50-0106
Option 1: Tyco: 2-1776112-3 or equivalent
Mating: 13 AWG wire
Option 2: JST: B6P-VH-B (LF) (SN) or B6P-VH (LF) (SN) or equivalent
Mating: VHR-6M; Pins: SVH-41T-P1.1
Signals & Aux Power
J3
Molex: 22-23-2081 or equivalent
Mating: 22-01-2087, Pins: 08-50-0113
Earthing Tab
J4
Molex: 19705-4301 or equivalent
Mating: 190030001
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00289_AD
MBC201 Series
Figure 1 - Output Power Vs. Temperature
PARAMETER
Weight
Dimensions
DESCRIPTION / CONDITION
325 g (0.72 lbs.)
127.0 x 76.2 x 38.1 mm (5 x 3 x 1.5 inch)
Figure 2 - Mechanical Drawing
NOTE:
Air flow over length of supply recommended (either direction) for forced air rating.
NUCLEAR AND MEDICAL APPLICATIONS
- Products are not designed or intended for use as critical components in life support systems, equipment used in
hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS
- The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.
+1.866.513.2839
tech.support@psbel.com
© 2015 Bel Power Solutions, Inc.
BCD.00289_AD