The MBC150 Series of open-frame medical power supplies, with its
wide universal input range 90-264 VAC, is available at 150 W of
output power with forced-air cooling and a variety of single output
voltages.
The MBC Series is designed and approved to the latest medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These power supplies are ideal for medical, telecom, datacom,
industrial equipment and other applications.
4 x 2 x 1.3 Inch Form Factor
150 W with Forced-Air Cooling
12 V @ 0.5 A fan voltage auxiliary output
High Efficiency > 86%
Low conducted and radiated noise
Light weight
IEC Protection Class Options:
Class I: Earthing Tab J4 (no suffix)
Class II: No Earthing Tab (-2 suffix)
Medical Safety Agency Approvals
RoHS Compliant
Diagnostic
Drug Pump
Monitoring
Dialysis
Home Health Care
Portable Equipment
2
MBC150 Series
MAX LOAD
CONVECTION
1
16.0 A
16.0 A
8.33 A
8.33 A
6.67 A
6.67 A
4.17 A
4.17 A
2.08 A
2.08 A
MAX LOAD
300 LFM
1,4, 5
16.0 A
20.0 A
12.5 A
12.5 A
10.0 A
10.0 A
6.25 A
6.25 A
3.13 A
3.13 A
RIPPLE &
NOISE
2
1%
1%
1%
1%
1%
1%
1%
1%
1%
1%
MODEL
3
MBC150-1005G
MBC150-1T05G
MBC150-1012G
MBC150-1T12G
MBC150-1015G
MBC150-1T15G
MBC150-1024G
MBC150-1T24G
MBC150-1048G
MBC150-1T48G
COVER-201-XBC
1
2
CONNECTOR
JST
Screw Terminal
JST
Screw Terminal
JST
Screw Terminal
JST
Screw Terminal
JST
Screw Terminal
OUTPUT VOLTAGE
5 VDC
5 VDC
12 VDC
12 VDC
15 VDC
15 VDC
24 VDC
24 VDC
48 VDC
48 VDC
MINIMUM LOAD
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
Metal cover kit accessory
3
4
5
Combined output power from V1, VSTBY and VFAN should not exceed the total output power rating.
Ripple is 2% up to 20% load and < 1% above 20% load. Ripple is peak to peak with 20 MHz bandwidth and 10
μF
(Tantalum
capacitor) in parallel with a 0.1
μF
capacitor at rated line voltage and load ranges.
For Class II (without input Earth pin) add suffix -2 (e.g.: MBC150-1012G-2).
Fan output voltage tolerance is +/-20%.
Peak current for fan output is 1 A.
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
Input Voltage
Input Frequency
Input Current
No Load Power
Inrush Current
Leakage Current
Power Factor
Switching Frequency
120 VAC:
230 VAC:
120 VAC:
230 VAC:
120 VAC:
230 VAC:
PFC converter (variable)
Resonant converter (variable)
120 VAC:
230 VAC:
DESCRIPTION / CONDITION
Universal
SPECIFICATION
90-264 VAC / 120-390 VDC
47 to 63 Hz
1.7 A max.
0.85 A max.
1.2 W
35 A max.
65 A max.
< 140
μA
< 250
μA
0.99
0.95
35 - 250 kHz, 90 kHz typical
35 - 250 kHz, 90 kHz typical
tech.support@psbel.com
MBC150 Series
PARAMETER
Output Power
Efficiency
Hold Up Time
Line Regulation
Load Regulation
Transient Response
Rise Time
Set Point Accuracy
Voltage Adjustment
Over Current Protection
Over Voltage Protection
Short Circuit Protection
V1
Short term, Automatic recovery
Main output
V1
Main output 50 to 100% load change,
50 Hz, 50% duty cycle, 0.1A / uSec
3
DESCRIPTION / CONDITION
Derate output power linearly to 80% from 90 VAC to 80 VAC input
Peak Power 170 W for 0.2 s
120 VAC:
230 VAC:
120 VAC:
230 VAC:
SPECIFICATION
150 W
84% typical
86% typical
6 ms
10 ms
+/-0.5%
+/-2.0%
< 10%,
recovery time < 5 ms
< 100 ms
± 1%
±3%
110% typical above rating
110 to 150%
PARAMETER
Operating Temperature
Storage Temperature
DESCRIPTION / CONDITION
Refer to derating curve
Start-up is guaranteed
Convection:
300 LFM:
Convection:
300 LFM:
SPECIFICATION
-20 to 70°C
-20 to 0°C
-40 to 70° C
80 W
100 W
100 W
150 W
95%
10,000 ft.
40,000 ft.
2.4 million hours
5 V model
Cooling
Other models
Humidity
Altitude
Reliability
Non Condensing
Operating:
Non-Operating:
MTBF according to Telcordia -SR332-Issue 3
PARAMETER
Conducted Emissions
Radiated Emissions
Harmonic Current
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
AC Flicker
DESCRIPTION / CONDITION
EN55022-B, CISPR22-B, FCC PART15, EN50082-1
EN55022-B, CISPR22-B, FCC PART15-B
EN61000-3-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-3-3:
SPECIFICATION
Class B
To be controlled in end system
Class D
Level 3
Level-3
Level 3
Level-3
Pass
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
BCD.00288_AD
© 2017 Bel Power Solutions & Protection
4
MBC150 Series
DESCRIPTION / CONDITION
Input to Output:
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, UL, C-UL
Complies with LVD Directive
PARAMETER
Isolation Voltage
Safety Standards
Agency Approvals
CE mark
SPECIFICATION
Min. 5900 VDC
CONNECTOR
AC Input Connector
PIN
J1
DESCRIPTION / CONDITION
Pin 1
AC Line
Pin 2
AC Neutral
Pin 1,2
Pin 3,4
Pin 1
Pin 2
V1
RTN
VFAN (12 V / 0.5 A)
RTN
DC Output Connector
J2
Fan
Earthing Tab
J3
J4
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Option-1: Tyco: 1776112-4 or equivalent
Mating: 13 AWG wire
Option-2: JST: B4P-VH-B (LF) (SN) or B4P-VH (LF) (SN) or
equivalent
Mating: VHR-4M; Pins: SVH-41T-P1.1
Tyco: 640456-2 or equivalent
Mating: 640440-2
Molex: 19705-4301 or equivalent
Mating: 190030001
PARAMETER
Weight
Dimensions
DESCRIPTION / CONDITION
150 g (0.33 lbs.)
101.6 x 50.8 x 33.6 mm (4.0 x 2.0 x 1.3 inch)
Figure 1 - Output Power Vs. Temperature
tech.support@psbel.com
MBC150 Series
5
Figure 2 - Mechanical Drawing
NOTES:
In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
Stand off, used to mount PCB has OD of 5.4 mm max.
2
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
3
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS
- Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS
- The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
North America
+1 408 785 5200
BCD.00288_AD
© 2017 Bel Power Solutions & Protection