A high performance AC to DC external power supply family designed for medical applications. The ME10 Medical Series
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition, with isolation levels which satisfy the 2
*
MOPP requirements; and designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME10 Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up
to +40°C and applicable output power derating at 50°C. These models are available in desktop and wall-plug versions, include
an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors.
*
Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME10A0503F01
ME10A0603F01
ME10A0703F01
ME10A1203F01
ME10A2403F01
ME10A0503N01
ME10A0603N01
ME10A0703N01
ME10A1203N01
ME10A2403N01
ME10A0503Q01
ME10A0603Q01
ME10A0703Q01
ME10A1203Q01
ME10A2403Q01
ME10A0503B01
ME10A0603B01
ME10A0703B01
ME10A1203B01
ME10A2403B01
Volts
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
Output
Power
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Output
Connector
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Input
Configuration
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades
(North American
Blade included)²
www.slpower.com
ME10 Rev 2.5
31-JAN-16
1
ME10
10W-12W Single Output External Power
Medical Series
Model Selection
(continued)
Model
Number
ME10A0503C01
ME10A0603C01
ME10A0703C01
ME10A1203C01
ME10A2403C01
Notes:
Volts
5.0V
5.9V
7.5V
12.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.0A
0.5A
Output
Power
10W
10W
10W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
Output
Connector
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Input
Configuration
Class II Wall-Plug,
Fixed North American
Blades³
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input
Input Current
Inrush Current
100-240Vac, ±10%, 47-63Hz, 1∅
115Vac: 0.45A, 230Vac: 0.28A
Turn On Time
Hold-up Time
Overtemperature
Protection
Overload Protection
Short Circuit
Protection
Overvoltage
Protection
Isolation
Less than 700mS @115Vac, full load
20mS min., at full Load, 100Vac input
Will shutdown upon an overtemperature
condition, auto-recovery.
264Vac, cold start: will not exceed 40A
Input Fuses
Earth Leakage
Current
Efficiency
Output Power
No Load Input
Power
Ripple and Noise
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
>87%, typical
10 to 12W continuous – See models chart for
specific voltage model ratings.
<0.1W per DoE Efficiency Level VI
Requirements
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
130 to 150% of output voltage, hiccup mode
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
-20°C to +50°C
Start Up at -40°C, full load,
(warmup period before
all parameters are within published specifications).
Safety Standards
Operating
Temperature
Temperature
Derating
Storage Temperature
See models chart on pg 1.
Output Voltage
Transient
Response
Regulation
See models chart on pg 1.
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load,
∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
See models chart on pg 1.
Derate output power above 40°C to TBD at
50°C
-40°C to +85°C
Altitude
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Drop Test
1.4m from table top to wooden platform, 6
faces.
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
>10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
ME10 Rev 2.5
Relative Humidity
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.
2
Vibration
Shock
E-Cap Life
www.slpower.com
MTBF
31-JAN-16
ME10
10W-12W Single Output External Power
Medical Series
EMI/EMC Compliance
Conducted Emissions:
Radiated Emissions:
Common Mode Noise:
Electro-Static Discharge (ESD)
Immunity on Power ports:
Radiated RF EM Fields Susceptibility
Electrical Fast Transients (EFT) /Bursts:
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Voltage Interruptions, Dips, Sags &
Surges
Harmonic Current Emissions
Flicker Test
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac