TRUMPower
TRM60 Series
FEATURES
• 90 to 264VAC univesal input
• Medically approved
• Internal EMI filter
• Single and dual outputs
• Output voltage available from 3 VDC thru 50 VDC
• Short-circuit, OVP and OLP protections
• Desktop style with class I insulation
30-60W Medical Power Adapters
SPECIFICATIONS
Input Voltage……......................................................90-264VAC
Input Frequency.............................................................47-63Hz
Inrush Current…..........20A at 115VAC, 40A at 230VAC, cold start
Efficiency…....................................70% min. for output = 3~11V
>=85% on CEC compliant models
Standby Power.......< 0.5W @ no load on CEC compliant models
Leakage Current..................................................…. 0.1mA Max.
Operating Temperature...0~60ºC; derate linearly from 100% load
at 40ºC to 50% load at 60ºC
Safety Approvals....UL/cUL (UL 60601-1), TUV (EN 60601-1), CE
EMI Standards…………...........................FCC Part 18 class B,
CISPR-11 EN55011 class B emission limits
MODEL LIST
*Product No.
TRM60-S03-X
TRM60-S05-X
TRM60-S07-X
TRM60-S09-X
TRM60-S12-X
TRM60-S15-X
TRM60-S18-X
TRM60-S24-X
TRM60-S48-X
*Output
Voltage
3~5V
5~6V
6~8V
8~11V
11~13V
13~16V
16~21V
21~27V
40~50V
Output
Current
13.33~8A
9~7.5A
8.33~6.25A
6.25~4.54A
5.45~4.61A
4.61~3.75A
3.75~2.85A
2.85~2.5A
1.5~1.2A
EN 60601-1
IEC 60601-1
V
*CEC
Output
Total
Regulation Power
± 5%
± 5%
± 5%
± 5%
± 5%
± 3%
± 3%
± 2%
± 2%
35W
45W
50W
50W
60W
60W
60W
60W
60W
*Specify output voltage when ordering, e.g. TRM60-S19 for 19V output.
*Add suffix “G” for CEC compliant models, e.g. TRM60-S12G.
*AC inlet code: X=4 for IEC320/C14, X=6 for IEC320/C6 (3 round pin).
Output #1
*Product No.
TRM60-D01-X
TRM60-D03-X
TRM60-D07-X
TRM60-D13-X
TRM60-D15-X
TRM60-D17-X
Vo nom.
+3.3V
+3.3V
+3.3V
+5V
+5V
+5V
Io min.
0.2A
0.2A
0.2A
0.2A
0.2A
0.2A
Io max.
6A
6A
6A
6A
6A
6A
Reg. Max.
± 7%
± 7%
± 7%
± 5%
± 5%
± 5%
Vo nom.
+5V
+12V
+24V
+12V
+15V
+24V
Output #2
Io min.
0.2A
0.1A
0.1A
0.1A
0.1A
0.1A
Io max.
6A
4A
2A
4A
3A
2A
Reg. Max.
± 5%
± 5%
± 5%
± 5%
± 5%
± 5%
Maximum
Output
Power
30W
50W
50W
55W
55W
55W
*On multiple output models, each output can load to it’s max. current with the total power limited to the maximum allowable output power.
*AC inlet code: X=4 for IEC320/C14, X=6 for IEC320/C6 (3 round pins).
Mechanical Drawing: (mm)
139.5 ± 0.3
14
40
35.5
62.9 ± 0.3
TIE
TIE
31.5 ± 0.3
PE BAG
1.0
120 ± 20
1200 ± 30
0.5~0.65
4-
φ
2.5
46 ± 0.2
C5.0
74 ± 0.2
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 • Fax: 408-988-6622 • email: sales@trumpower.com • Website: www.TRUMPower.com