DC/DC Converters
THM 6WI Series, 6 Watt
Features
◆
Ultra wide 4:1 input voltage
◆
I/O isolation 5000 VACrms rated for
250 working voltage
ES 60601-1
◆
2 x MOPP Medical safety according to
AAMI/ANSI ES 60601-1:2005(R) and
IEC/EN 60601-1 3rd edition
◆
Low leakage current <2 µA
◆
Very high efficiency up to 88%
◆
Extended operating temperature range
–40°C to 88°C at full load.
◆
Input filter to meet EN55022 class A
◆
3-year product warranty
The THM-6WI series is a range of high performance, regulated 6 Watt DC/DC
converters in a DIP-24 plastic package. The reinforced I/O-isolation system com-
plies with the medical safety requirements for MOPP (Means Of Patient Protec-
tion). Together with a wide 4:1 input voltage range, and an internal EMI filter to
meet EN55022 class A the converters constitute also a reliable solution for many
demanding applications such as transportation systems, industrial control equip-
ments, measurement equipments, and some IGBT driver applications. With A high
efficiency of up to 88% and highest grade components the converters can reliably
operate in an ambient temperature range of –40 up to +88°C at full load.
Models
Order code
THM 6-0510WI
THM 6-0511WI
THM 6-0512WI
THM 6-0513WI
THM 6-0515WI
THM 6-0521WI
THM 6-0522WI
THM 6-0523WI
THM 6-2410WI
THM 6-2411WI
THM 6-2412WI
THM 6-2413WI
THM 6-2415WI
THM 6-2421WI
THM 6-2422WI
THM 6-2423WI
THM 6-4810WI
THM 6-4811WI
THM 6-4812WI
THM 6-4813WI
THM 6-4815WI
THM 6-4821WI
THM 6-4822WI
THM 6-4823WI
http://www.tracopower.com
18 – 75 VDC
(48 VDC nominal)
9 – 36 VDC
(12 VDC nominal)
4.5 – 9 VDC
(5 VDC nominal)
Input voltage range
Output voltage
3.3 VDC
5.0 VDC
12 VDC
15 VDC
24 VDC
±5.0 VDC
±12 VDC
±15 VDC
3.3 VDC
5.0 VDC
12 VDC
15 VDC
24 VDC
±5.0 VDC
±12 VDC
±15 VDC
3.3 VDC
5.0 VDC
12 VDC
15 VDC
24 VDC
±5.0 VDC
±12 VDC
±15 VDC
Output current max.
1800 mA
1200 mA
500 mA
400 mA
250 mA
±600 mA
±250 mA
±200 mA
1800 mA
1200 mA
500 mA
400 mA
250 mA
±600 mA
±250 mA
±200 mA
1800 mA
1200 mA
500 mA
400 mA
250 mA
±600 mA
±250 mA
±200 mA
Efficiency typ.
81.5 %
86.0 %
86.0 %
86.0 %
87.0 %
84.0 %
86.5 %
87.5 %
83.0 %
86.0 %
89.0 %
89.0 %
88.5 %
85.0 %
88.5 %
88.0 %
82.5 %
86.5 %
88.0 %
88.5 %
88.0 %
85.0 %
88.0 %
88.0 %
Page 1 of 3
DC/DC Converters
THM 6WI Series
Input Specifications
Input current at no load
5 Vin models:
20 mA typ.
24 Vin models:
6 mA typ.
48 Vin models:
4 mA typ.
5 Vin models:
4.5 VDC / 4.0 VDC typ.
24 Vin models:
9.0 VDC / 8.0 VDC typ.
48 Vin models:
18 VDC / 16 VDC typ.
5 Vin models:
16 VDC max.
24 Vin models:
50 VDC max.
48 Vin models:
100 VDC max.
EN 55022 class A,
without external components
EN 61000-4-2, air ±8 kV, contact ±6 kV,
perf. criteria A
EN 61000-4-3, 10 V/m, perf. criteria A
EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±2 kV perf. criteria A
5 Vin models: Nippon chemi-con KY 1000 µF/25 V and
reverse diode (Vishay V10P45) in parallel
24 Vin models: Nippon chemi-con KY 470 µF/ 50 V
48 Vin models: Nippon chemi-con KY 330 µF/ 100 V
EN 61000-4-6, 10 V, perf. criteria A
5 Vin models:
5 A
24 Vin models:
2.5 A
48 Vin models:
1.5 A
±1.0 % max.
single output models
dual output models
– Load variation 0 – 100 %:
single output models:
dual output models balanced load:
dual output models unbalanced load:
– Input variation
0.2 % max.
0.5 % max.
0.2 % max..
1.0 % max.
5.0 % max.
not required
30 mS
3.3 & 5.0 VDC models:
30 mVp-p typ.
with cap. 10µF/25V X7R MLCC
12 & 15 VDC models:
40 mVp-p typ.
with cap. 10µF/25V X7R MLCC
24 VDC models:
50 mVp-p typ.
with cap. 4.7µF/50V X7R MLCC
250 µs typ.
150 % Iout nominal typ.
(hiccup mode)
continuous
(automatic recovery)
3.3
5
12
15
24
3.3
5
12
15
24
±5
±12
±15
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
output
output
output
output
output
output
output
output
output
output
output
output
output
models:
models:
models:
models:
models:
models:
models:
models:
models:
models:
models:
models:
models:
3.7 – 5.4 VDC
5.6 – 7.0 VDC
13.5 – 19.6 VDC
18.3 – 22.0 VDC
29.1 – 32.5 VDC
2100 µF max.
1500 µF max.
260 µF max.
210 µF max.
75 µF max.
860 µF max.
(each output)
150 µF max.
(each output)
100 µF max.
(each output)
6 Watt
Start-up voltage / under voltage shut down
Surge voltage
(1 sec. max.)
Conducted noise
ESD
(electrostatic discharge)
Radiated immunity
Fast transient / surge
(with external input capacitor / diode)
– external input capacitor:
Conducted immunity
External input fuse required
(recommended values, slow blow type)
Output Specifications
Voltage set accuracy
Regulation
Minimum load
Start up time
Ripple and noise
(20 MHz Bandwidth)
Transient response
(25% load step change)
Current limitation
Short circuit protection
Over voltage protection
Capacitive load
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
http://www.tracopower.com
Page 2 of 3
DC/DC Converters
THM 6WI Series
General Specifications
Temperature ranges
– Operating
– Casing
– Storage
–40°C to +88°C
(without derating)
+105°C max.
–55°C to +125°C
95 % rel H max.
±0.02 %/K typ.
250 kHz ±25 kHz.
(puls width modulation)
– to meet UL/IEC/EN 60601-1
5000 VACrms, rated for 250 Vrms working
voltage, 2 x MOPP
8 mm min.
2 µA max.
(at 240 VAC, 60 Hz)
– Input/Output
17 pF max.
(at 100 KHz, 1 V)
ANSI/AAMI ES 60601-1:2005/(R)2012,
IEC/EN 60601-1 3rd edition
– UL online certification UL 60601-1
www.ul.com
File E188913, copy:
e188913qqhm2.pdf
>1.1 Mio. h
non conductive plastic
(UL 94V-0-rated)
silicone
(UL 94V-0-rated)
according to MIL-STD-810F
14.0 g
(0.48 oz)
max. 265°C / 10 sec.
– Reach
– RoHS
www.tracopower.com/products/reach-declaration.pdf
according RoHS directive 2011/65/EU
6 Watt
Humidity
(non condensing)
Temperature coefficient
Switching frequency
I/O isolation voltage
(50Hz, 60sec.)
Clearance/creepage
Leakage current
Isolation capacitance
Safety standards
Safety approvals
Casing material
Potting material
Vibration and thermal shock resistance
Weight
Soldering temperature
Environmental compliance
Reliability, calculated MTBF
(MIL-HDBK-217F, at +25°C, ground benign)
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.
Outline Dimensions
2
11
15.24
(0.60)
20.3
(0.80)
Pin-Out
Pin
2
11
14
16
22
Single
–Vin (GND)
No con.
+Vout
–Vout
+Vin (Vcc)
+Vin (Vcc)
Dual
–Vin (GND)
–Vout
+Vout
Common
+Vin (Vcc)
+Vin (Vcc)
Bottom view
23 22
16
14
2.54
31.8
(1.25)
(0.10)
10.2
(0.40)
23
(0.15)
4.5
(0.18)
2.54
(0.10)
15.24
(0.60)
5.08
(0.20)
Dimensions in [mm], () = Inch
Pin diameter ø 0.6 ±0.1 (0.024 ±0.004)
Tolerances ±0.5 (±0.02)
Pin pich tolerances ±0.25 (±0.01)
Specifications can be changed without notice! Make sure you are using the latest documentation, downloadable at www.tracopower.com
3.8
www.tracopower.com
Rev. October 22. 2013
Page 3 of 3