TECHNICAL DATASHEET
MDS Medical AC-DC Adapter
24V 60W / MDS-060AAS24 BA
060AAS
Highlights & Features
Compliant to IEC 60601-1 3rd edition
IT and medical safety approvals
Low earth leakage current (<0.1mA)
Overload, over voltage, over temperature and short circuit
protections
– Risk management report available
– Energy Star Compliance (Level V)
– 2 x MOPP(means of patient protection)
–
–
–
–
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-060AAS24 BA
Unit Weight:
0.36 kg
Dimensions (W x L x H):
62.0x135.0x34.1 mm
2.4x5.3x1.3 in
General Description
The MDS series of external power supply comes with universal AC input at 90Vac to 275Vac. Other features include low
earth leakage, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS
series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency
equipment and EN 55022 for Information Technology Equipment (ITE) radio-frequency equipment. In addition, only
recognized Japanese capacitors are used.
The MDS series come with both medical and ITE safety approvals including UL/CSA/CCC/CE and CB certification and
are fully compliant with RoHS Directive 2011/65/EU for environmental protection.
Model Information
Medical AC-DC Adapter
Model Number
MDS-060AAS24 BA
Input Voltage Range
90-275Vac
Output Voltage
24Vdc
Output Current
2.5A
Model Numbering
MDS
Delta Medical
power Supply
060
Max wattage in
the product
Series.
Maybe lower at
some voltage.
060 → 60W
120 →120W
1K2 → 1,200W
A
Family Code
A½ Z
A
Product Type
A: Adapter
S
Output Code
S: Single Output
2: Dual Output
3: Triple Output
4: Four Output
5: More than
Five Output
24
Output Voltage
Single Output:
- 03 for 3.3V
- 05 for 5V
- 24 for 24V
Multiple Output:
Serial Number
00 to 99
B
Input Connector code
B: Socket C14 / 3 Pin
D: Socket C6 / 3 Pin
A
Output Connector code
A: 4 pin DIN/1200mm
B: 4 pin DIN/1800mm
E = barrel type 5.5*2.5, L=10mm/1200mm
F = barrel type 5.5*2.5, L=10mm/1800mm
J = barrel type 5.5*1.7, L=10mm/1200mm
K = barrel type 5.5*1.7, L=10mm/1800mm
1
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)
TECHNICAL DATASHEET
MDS Medical AC-DC Adapter
24V 60W / MDS-060AAS24 BA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Efficiency (typ.)
Standby Power (max)
Inrush Current (typ.)
Earth Leakage Current (max)
1) NC: normal condition
2) SFC: single fault condition
100-250Vac
90-275Vac
50-60Hz
47-63Hz
1.5A @ 115Vac, 0.75A @ 230Vac
88.3%,
Reference Fig.1
0.3W
50A @ 115Vac, 100A @ 230Vac
0.1mA @ 275Vac NC , 0.3mA @ 275Vac SFC
1)
2)
Fig.1 Efficiency versus output load
Output Ratings / Characteristics
Nominal Output Voltage
Output Voltage Tolerance
Output Current
Output Power
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
Start-up Time (max)
Hold-up Time (min)
Dynamic Response (Overshoot & Undershoot O/P Voltage)
*Periodic and Random Deviation
24Vdc
± 5%
2.5A
60W
±0.5%
±5%
43mV pk-pk @ Full load, Reference Fig. 2
3000ms @ 115Vac
10ms @ 115Vac
± 5% @ 50-100% load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)
TECHNICAL DATASHEET
MDS Medical AC-DC Adapter
24V 60W / MDS-060AAS24 BA
(a) 115V
Fig. 2 Ripple & Noise example, 20MHz BW
(b) 230V
Ripple & Noise measurement circuit
Mechanical
Case Chassis
Case Cover
Dimensions(W x L x H)
Unit Weight
Indicator
Cooling System
Terminal
Input
Output
PC
PC
62.0 x 135.0 x 34.1 mm(2.4x5.3x1.3 in)
0.36 kg
NA
NA
C14 socket
4 Pin Din
Environment
Surrounding Air Temperature
Operating Humidity
Operating Altitude
Shock Test (Non-Operating)
Vibration (Operating)
Operating
Storage
0°C to +40°C
-40°C to +85°C
10-95% RH (Non-Condensing)
3,000 meters
50G, 11ms, 3 shocks for each direction
5-500Hz, 2.09Grms, 20 minute for each three axis
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)
TECHNICAL DATASHEET
MDS Medical AC-DC Adapter
24V 60W / MDS-060AAS24 BA
Protections
Overvoltage (max)
Over load / Over current (max)
Over Temperature
Short Circuit
Degree of Protection
Protection Against Shock
150%, Latch Mode
180% of rated load current, Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Hiccup Mode,
(Non-Latching, Auto-Recovery)
IP40
Class I
Reliability Data
MTBF
(typ.)
1218 kHrs based on Telecordia SR-332
Safety Standards / Directives
Medical Safety
IEC60601-1: (Ed.3,2005), EN0601-1:2006, CAN/CSA-
C22.2 No. 60601-1:08, ANSI/AAMI ES60601-1:
(Ed.3,2005)
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-
2008, GB17625.1-2003
ITE Safety
CE
Material and Parts
Galvanic Isolation
Input to Output
Input to Ground
Output to Ground
MDD Directive 93/42/EEC
RoHS Directive 2011/65/EU Compliant
4000 Vac
1500 Vac
500 Vac
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)
TECHNICAL DATASHEET
MDS Medical AC-DC Adapter
24V 60W / MDS-060AAS24 BA
EMC
EMC / Emissions
Immunity to
Voltage Flicker
Electrostatic Discharge
IEC61000-3-3
IEC61000-4-2
Level 3 Criteria A
Air Discharge: 8kV
Contact Discharge: 6kV
Level 2 Criteria A
80MHz-1GHz, 3V/M with 1kHz tone / 80% modulation
Level 3 Criteria A
2kV
1)
1)
1)
EN55011, EN55022, FCC Title 47: Class B
Radiated Field
Electrical Fast Transient / Burst
Surge
IEC61000-4-3
IEC61000-4-4
IEC61000-4-5
Level 3 Criteria A
2)
Common Mode : 2kV
3)
Differential Mode : 1kV
Level 2 Criteria A
150kHz-80MHz, 3Vrms
Criteria A
Magnetic field strength 3A/Meter
30% 10ms Criteria A ; 60% 100ms and 100% 5000ms
Criteria B
1)
1)
1)
Conducted
Power Frequency Magnetic Fields
Voltage Dips
1) Criteria A: Normal performance within the specification limits
2) Asymmetrical: Common mode (Line to earth)
3) Symmetrical: Differential mode (Line to line)
IEC61000-4-6
IEC61000-4-8
IEC61000-4-11
Block Diagram
5
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)