144S...-PCB Series
Signal conditioned precision pressure transducers
FEATURES
· 0...70 mbar to 0...10 bar,
0...1 to 0...150 psi,
absolute, gage or differential pressure
(custom calibrations available)
· Barometric pressure ranges
· 0...5 V output
· Internal supply regulation
· Precision temperature compensated
and calibrated
SERVICE
Non-corrosive, non-ionic working fluids,
such as dry air and dry gases.
Scale:
1 cm
1 inch
SPECIFICATIONS
Maximum ratings
Supply voltage
Maximum load current
source
sink
Temperature limits
Storage
Operating
Compensated
144SC...BARO
all others
Lead temperature (10 sec soldering)
Humidity limits
pressure inlets only
Proof pressure
1
144SM...
144SB010...
144SC...BARO
144SU01..., 144SU05...
144SU150...
all others
January 2005 / 005
ELECTRICAL CONNECTION
7.5...24 V
20 mA
10 mA
-55 to 100°C
-40 to 85°C
-10 to 60°C
0 to 70 °C
300°C
0 - 100 %RH
1.4 bar
16 bar
2 bar
20 psi
250 psi
2x rated pressure
1/5
+Vs
144S...-
PCB
Vout
-Vs
7.5 V to 24 V
V
www.sensortechnics.com
144S...-PCB Series
Signal conditioned precision pressure transducers
PERFORMANCE CHARACTERISTICS
STANDARD DEVICES
(unless otherwise noted V
S
= 8 V, R
L
> 100 kΩ, t
amb
= 25°C)
Ω
Characteristics
Operating pressure
differential devices
2
144SM070D-PCB
144SM350D-PCB
144SB001D-PCB
144SB002D-PCB
144SB005D-PCB
144SB010D-PCB
144SB001A-PCB
144SB002A-PCB
144SB005A-PCB
144SU01D-PCB
144SU05D-PCB
144SU15D-PCB
144SU30D-PCB
144SU100D-PCB
144SU15A-PCB
144SU30A-PCB
144SU100A-PCB
Min.
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
-0.05
4.9
4.9
Typ.
Max.
70
350
1
2
5
10
1
2
5
1
5
15
30
100
15
30
100
Unit
mbar
bar
absolute devices
3
differential devices
2
p si
absolute devices
3
Zero pressure offset
Full scale span
4
Full scale output
Non-linearity and hysteresis (BSL)
5
Thermal effects
(0 to 70°C)
6
Offset
144SM070...,144SU01...
144SM350...,144SU05...
all others
0
5.0
5.0
0.1
0.6
0.2
0.15
0.2
0.1
1
70
0.05
0.03
0.05
5.1
5.1
0.5
3.0
1.0
0.6
1.0
%FSO
V
Span
Long term stability
7
Response time (10 to 90 %)
Power consumption (no load)
Power supply rejection
Offset
Span
ms
mW
%FSO/V
2/5
January 2005 / 005
www.sensortechnics.com
144S...-PCB Series
Signal conditioned precision pressure transducers
PERFORMANCE CHARACTERISTICS
BAROMETRIC DEVICES
8
(unless otherwise noted V
S
= 8 V, R
L
> 100 kΩ, t
amb
= 25°C)
Ω
Characteristics
Operating pressure ranges
3
Offset calibration at lowest specified pressure
Full scale output
Non-linearity and hysteresis
5
Long term stability
7
Thermal effects (-10 to 60°C)
9
Response time (10 to 90 %)
Power consumption (no load)
Power supply rejection
Offset
Span
144SC0811BARO
144SC1216BARO
Min.
800
12
-0.05
4.95
Typ.
Max.
1100
16
Unit
mbar
p si a
V
%FSO
%FSO/10°C
ms
mW
%FSO/V
0
5.0
0.05
0.1
0.05
1
70
0.05
0.03
0.05
5.05
0.1
0.3
Specification notes:
1.
2.
3.
4.
5.
6.
7.
8.
Proof pressure is the maximum pressure which may be applied without causing damage to the sensing element.
The output signal of all 144S...D-PCB devices is proportional to the pressure applied to port B, relative to port A, e.g. the
output signal increases when vacuum is applied to port A relative to port B.
The output signal of all 144S...A-PCB and 144SC...BARO devices is proportional to the pressure applied to port A.
Full scale span is the algebraic difference between the positive full scale output and the zero pressure offset.
Non-linearity refers to the
Best Straight Line
fit measured for offset pressure, full scale pressure and 1/2 full scale pressure.
Thermal effects tested and guaranteed from 0 to 70°C relative to 25°C. All specifications shown are relative to 25°C.
Change in output after one year or 1 million pressure cycles.
These devices are factory calibrated at sea level. When used at other altitudes the output signal differs from the reading
expected when comparing to the pressure given from your local weather station. The weather station always reports the pressure
compared to sea level. On that the output signal of the transducer will change 65mV/0.052 psi per 100 feet e.g. 19.7mV/1.18
mbar per 10 m change in altitude. The output signal can be adjusted to sea level reading by turning the offset trimmer.
Thermal effects refer to the combined effects of offset and sensitivity shifts, this is true from -10 to 60°C relative to 25°C.
9.
January 2005 / 005
3/5
www.sensortechnics.com
144S...-PCB Series
Signal conditioned precision pressure transducers
ORDERING INFORMATION
Operating Pressure
Differential/gage devices
0...70 mbar
0...350 mbar
0...1 bar
0...2 bar
0...5 bar
0...10 bar
Absolute devices
0...1 bar
0...2 bar
0...5 bar
Differential/gage devices
0...1 psi
0...5 psi
0...15 psi
0...30 psi
0...100 psi
0...150 psi
Absolute devices
0...15 psi
0...30 psi
0...100 psi
Barometric devices
12...16 psia
800...1100 mbar
Part Number
144SM070D-PCB
144SM350D-PCB
144SB001D-PCB
144SB002D-PCB
144SB005D-PCB
144SB010D-PCB
144SB001A-PCB
144SB002A-PCB
144SB005A-PCB
144SU01D-PCB
144SU05D-PCB
144SU15D-PCB
144SU30D-PCB
144SU100D-PCB
144SU150D-PCB
144SU15A-PCB
144SU30A-PCB
144SU100A-PCB
144SC1216BARO
144SC0811BARO
Custom calibrations available
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of
the application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights
of others.
January 2005 / 005
5/5
www.sensortechnics.com