Model 1700 Series
Non-Contact/Amplified
Output Torque Transducer
DESCRIPTION
These transducers are suitable for laboratory applications as
well as industrial evironments because of their compact size
and multiple mounting options. The contactless transmission
of supply voltage and measuring signal enables continuous
operation with low maintenance.
FEATURES
•
•
•
•
•
•
•
0.02 Nm to 1500 Nm capacity range
0.25 % accuracy
Built-in instrumentation amplifier (±10 V output) (standard)
Speed sensor or angle encoder option
Contactless
Compact size
Wide application range
Model 1700 Series
PERFORMANCE SPECIFICATIONS
Characteristic
Torque range
MECHANICAL SPECIFICATIONS
Characteristic
Mechanical overload
Measure
0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20,
50, 100, 200, 300, 500, 1000, 1500 Nm
Measure
200 % of rated output
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Temperature, operating
Temperature, compensated
Temperature effect, zero
Temperature effect, span
SPEED SENSING (RPM) (S) OPTION
Characteristic
Speed max.
Output
Internal pull-up
External pull-up
I max.
Pulses/rev.
Measure
0 °C to 60 °C [32 °F to 140 °F]
5 °C to 45 °C [41 °F to 113 °F]
0.02 %/°K
0.01 %/°K
Measure
10000 RPM
Open collector
100 kOhm (5 V level)
24 V max.
20 mA
60
ELECTRICAL SPECIFICATIONS
Characteristic
Supply voltage
Current consumption
Rise time
Limit frequency (3dB)
Voltage output
Internal resistance
Ripple
Overall accuracy
Measure
12 Vdc ±10 %
less than 200 mA
2 ms*
200 Hz
0 V to ±10 V
100 ohm
<100 mVpp
<0.25 %
ANGLE (E) OPTION
Characteristic
Speed max.
Pulses/rev.
Resolution
Phase shift
Measure
3000 RPM
360
1°
Quadrature
TECHNICAL SPECIFICATIONS
Model Range
(Nm)
Spring
constant
C
(Nm/rad)
16
16
35
35
45
90
135
960
2115
3955
17 x 10
3
30 x 10
67 x 10
3
* Standard rise time is 2 ms for all transducers 1700 to 1706, in
combination with limit frequency 200 Hz.
Mass
moment
of
inertia
(g•cm
2
)
7
7
10
10
10
10
10
100
100
100
775
785
810
840
9835
10140
10380
Maximum
permissible
axial load
(N)
35
35
140
140
160
210
210
1200
1300
1300
1800
1800
1800
1800
4150
4150
4150
Maximum
permissible
radial load
(N)
30
30
3
3
4
7
13
15
30
60
125
215
450
650
650
1275
1700
1700
1700
1701
1701
1701
1701
1701
1702
1702
1702
1703
1703
1703
1703
1706
1706
1706
0 to 0,02
0 to 0,05
0 to 0,1
0 to 0,2
0 to 0,5
0 to 1
0 to 2
5
10
20
50
100
200
300
500
1000
1500
54 x 10
3
3
260 x 10
3
387 x 10
3
449 x 10
3
2 Honeywell • Sensing and Control
Model 1700 Series
Non-Contact/Amplified Output Torque Transducer
Warranty.
Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective.
The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
WARNING
PERSONAL INJURY
• DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
Find out more
Honeywell serves its customers
through a worldwide network of
sales offices, representatives
and distributors. For application
assistance, current specifica-
tions, pricing or name of the
nearest Authorized Distributor,
contact your local sales office.
To learn more about Honeywell’s
test and measurement products,
call
+1-614-850-5000,
visit
http://measurementsensors.
honeywell.com,
or e-mail inqui-
ries to
info.tm@honeywell.com
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this datasheet is for
reference only. DO NOT USE this document as product
installation information.
• Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422
www.honeywell.com
008779-3-EN IL50 GLO
January 2011
Copyright © 2011 Honeywell International Inc. All rights reserved.