Features
Regulated
Converter
Description
•
•
•
•
•
•
•
•
Long 5 year warranty
2MOPP/250VAC
Suitable for built in Class II applications
Wide input voltage range (85-264VAC)
Low leakage current (<75µA)
5000m operation
Active power factor correction
Connector set available
AC/DC Converter
RACM100
100 Watt
Enclosed &
Open Frame
Case Style
Single Output
The RACM100 is a compact 3” x 2” high efficiency AC/DC power supply with 2xMOPP 3rd Ed. safety
approval for medical applications. The range has now been extended to include open frame models
(/OF suffix). Like the original enclosed versions, the RACM100/OF series are space-saving universal
input voltage power supplies (85-264VAC), with 4kVAC isolation, PFC, no minimum load and can be
used at ambient temperatures of between -25°C and +85°C. The 12V, 15V, 24V or 48V output voltages
are fully protected and have tolerances of less than ±0.2% over the entire input voltage range and less
than ±0.5% over the entire load range. The RACM100/OF series is certified to medical safety standard
IEC/ES/EN-60601-1 3rd Edition and feature BF rated outputs with less than 75µA leakage current.
It has a built-in Class B EMI filter and comes with a five year warranty.
Selection Guide
Part
Number
RACM100-12S
(1)
RACM100-15S
(1)
RACM100-24S
(1)
RACM100-48S
(1)
Input
Voltage Range
[VAC]
85-264
85-264
85-264
85-264
Output
Voltage
[VDC]
12
15
24
48
Output
Current
[A]
8.34
6.67
4.17
2.09
Input Power
@ No Load
[W]
0.3
0.3
0.3
0.3
Efficiency Max. Capacitive
typ.
Load
(1)
[%]
[µF]
91
92
92
91
6950
4450
1750
430
Notes:
Note1: Max Cap Load is tested at minimum input and full resistive load
Model Numbering
RACM100- __ S/ __
Output Power
Output Voltage
Notes:
Note2:
Case Style
(2)
Single
2MOPP
250VAC
without suffix, standard enclosed case
add suffix “/OF” for open frame style
E314885
Examples:
RACM100-12S
RACM100-24S/OF
= 12Vout, standard enclosed case
= 24Vout, open frame style
IEC/EN60601 certified
ANSI/AAMI ES60601 certified
EN55011 certified
CISPR11
FCC Part 15
www.recom-power.com
REV.: 9/2018
MED-1
RACM100
AC/DC Converter
Specifications
(measured @ Ta= 25°C, 250VAC, full load and after warm-up)
BASIC CHARACTERISTICS
Parameter
Input Voltage
Input Current
Inrush Current
No load Power Consumption
Input Frequency Range
Output Voltage Trimming
Minimum Load
Power Factor
Start-up Time
Rise Time
Hold up Time
Internal Operating Frequency
Output Ripple and Noise
(measured @ 20MHz BW)
Series
Condition
Min.
85VAC
120VDC
115VAC, fulll load
230VAC, full load
cold start, 230VAC
AC Input
47Hz
±10.0%
0%
0.95
1s
20ms
115VAC, full load
12VDC, with 10µF/25V MLCC
15VDC, with 10µF/25V MLCC
24VDC, with 1µF/50V MLCC
48VDC, with 0.1µF/100V MLCC
16ms
60kHz
120mVp-p
150mVp-p
160mVp-p
340mVp-p
Typ.
Max.
264VAC
370VDC
1.15A
0.55A
60A
0.11W
63Hz
Efficiency vs. Load
100
90
80
100
90
80
Efficiency vs. Input Voltage
Efficiency [%]
Efficiency [%]
70
60
50
40
30
20
10
0
10
20
115VAC
230VAC
30
40
50
60
70
80
90
100
70
60
50
40
30
20
10
0
85
105
50% load
100% load
125
145
165
185
205
225
245
264
Output Load [%]
Input Voltage [VAC]
1
0.8
Line Derating
100
90
Power Factor vs. Load
Output Load [%]
Power Factor [ ]
80
70
60
50
40
30
20
85
105 125
100
145
165
185
205
225
245
264
0.6
0.4
0.2
0
30
115 VAC
230 VAC
40
50
60
70
80
90
100
Input Voltage [VAC]
REV.: 9/2018
Output Load [%]
MED-2
www.recom-power.com
RACM100
AC/DC Converter
Specifications
(measured @ Ta= 25°C, 250VAC, full load and after warm-up)
REGULATIONS
Parameter
Output Accuracy
Line Regulation
Load Regulation
Transient Peak Deviation
Transient Recovery Time
Series
Condition
230VAC, full load
low line to high line, full load
0% to 100% load
10% to 100% load
load step from 50% - 75% change at 2.5A/µs
load step from 50% - 75% change at 2.5A/µs
Value
±1.0%
±0.2%
0.5% max.
0.4% max.
3.0% Vout max.
500µs typ.
PROTECTIONS
Parameter
Input Fuse
Short Circuit Protection (SCP)
Over Load Protection (OLP)
Over Voltage Protection (OVP)
Isolation Voltage
(5)
Condition
internal line and neutral
% of Iout rated (Hiccup)
% of Vout nominal (Latch off)
tested for 1 minute
500VDC
264VAC
working voltage 250VAC/continuous
clearance
creepage
Notes:
Note5: For repeat Hi-Pot testing, reduce the time and/or the test voltage
I/P to O/P
I/P to Case
O/P to Case
Value
T3.15A / 250VAC, slow blow type
continuous, auto-recovery
115% min. / 150% max.
115% min. / 135% max.
4kVAC
1.5kVAC
1.5kVAC
100MW min.
reinforced
75µA max.
2MOPP
built-in power supply
>8.0mm
>8.0mm
Isolation Resistance
Insulation Grade
Leakage Current
Means of Protection
Medical Device Classification
Internal
ENVIRONMENTAL
Parameter
Operating Temperature Range
Temperature Coefficient
Operating Altitude
Operating Humidity
Pollution Degree
Thermal Shock
Shock
Vibration
MTBF
according to MIL-HDBK-217F, full load, +25°C
continued on next page
www.recom-power.com
REV.: 9/2018
MED-3
Condition
refer to derating graph
full load, 230VAC
enclosed
open frame
Value
-25°C to +85°C
-25°C to +60°C
-25°C to +55°C
±0.02%/K
5000m max.
non-condensing
5% to 95% RH
PD2
MIL-STD-810F
IEC60068-2-27
IEC60068-2-6
790.3 x 10
3
hours
RACM100
AC/DC Converter
Specifications
(measured @ Ta= 25°C, 250VAC, full load and after warm-up)
Derating Graph
(@ Chamber and natural convection 0.1m/s)
Series
/OF version
100
90
80
Enclosed
100
90
80
Output Load [%]
60
50
40
30
20
10
115VAC
230VAC
10 20 30 40 50 60 70 80 90 100 110
55
85
Output Load [%]
70
70
60
50
40
30
20
10
115VAC
230VAC
10 20 30 40 50 60 70 80 90 100 110
85
55
0
-25 -20 -10 0
0
-25 -20 -10 0
Ambient Temperature [°C]
Ambient Temperature [°C]
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
Medical Electric Equipment, General Requirements for Safety and Essential Performance
Medical Electric Equipment, General Requirements for Safety and Essential Performance
(CB Scheme)
Information Technology Equipment - General Requirements for Safety (LVD)
Information Technology Equipment - General Requirements for Safety
EAC
RoHs2+
Report / File Number
E314885
1408016004
TW1708008-001
RU-AT.49.09571
Standard
CAN/CSA-C22.2 No. 60601-1:14
ANSI/AAMI ES60601-1:2005 + A2:2010
IEC60601-1:2005 + C2:2007, 3rd Edition
EN60601-1:2006
EN60950-1:2006 + A2:2013
IEC60950-1:2005, 2nd Edition + A2:2013
TP TC 004/2011 TP TC 004/2011
RoHS-2011/65/EU + AM-2015/863
EMC Compliance (Medical)
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
Industrial, scientific and medical equipment – Radio frequency disturbance characteristics –
Limits and methods of measurement
Industrial, scientific and medical equipment - Radio frequency disturbance characteritics -
Limits and methods of measurement
ESD Electrostatic discharge immunity test
Radiated, radio-frequency, electromagnetic field immunity test
Conditions
Standard / Criterion
EN60601-1-2:2015
EN55011:2009 + A1:2010
Class B Conducted, Class A Radiated
CISPR11:2009 + A1:2010
Class B Conducted, Class A Radiated
Air ±15kV; Contact ±8kV
10V/m (80-2700MHz)
27V/m (385MHz)
28V/m (450MHz)
AC Power Port: ±2kV
AC Port:
L-N= ±1kV
L-GND= ±2kV
6Vr.m.s
50Hz, 30A/m
Dips: >95%; 30%;
Interruptions >95%
IEC61000-4-2:2008
IEC61000-4-3:2006 + A2:2010
IEC61000-4-4:2012
IEC61000-4-5:2005
IEC61000-4-6:2013
IEC61000-4-8:2009
IEC61000-4-11:2004
EN61000-3-2:2005 + A2:2009, Class D
EN61000-3-3:2013
Fast Transient and Burst Immunity
Surge Immunity
Immunity to conducted disturbances, induced by radio-frequency fields
Power Frequency Magnetic Field
Voltage Dips and Interruptions
Limits of Harmonic Current Emissions
Limits of Voltage Fluctuations and Flicker
continued on next page
www.recom-power.com
REV.: 9/2018
MED-4
RACM100
AC/DC Converter
Specifications
(measured @ Ta= 25°C, 250VAC, full load and after warm-up)
EMC Compliance (Industrial)
Limitations on the amount of electromagnetic intererence allowed from digital & electronic devices
Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and
Electronic Equipment in the Range of 9 kHz to 40 GHz
Electromagnetic compatibility of multimedia equipment – Emission Requirements
Information technology equipment - Immunity characteristics - Limits and methods of
measurement
ESD Electrostatic discharge immunity test
Radiated, radio-frequency, electromagnetic field immunity test
Fast Transient and Burst Immunity
Surge Immunity
Immunity to conducted disturbances, induced by radio-frequency fields
Power Frequency Magnetic Field
Voltage Dips and Interruptions
Limits of Harmonic Current Emissions
Limits of Voltage Fluctuations and Flicker
Air ±8kV; Contact ±6kV
3V/m (80-1000MHz)
20V/m (80-1000MHz)
3V/m (1-2.5GHz)
10V/m (1-2.5GHz
DC Port: ±2kV
DC Port: ±1kV
DC Power Port 3V + 20V
50Hz/60Hz 1A/m
50Hz/60Hz 10A/m
Dips: >95%; 60%; 30%
Interruptions >95%
Series
Conditions
Standard / Criterion
47CFR FCC Part 15 Subpart B, Class B
ANSI C63.4:2014
EN55032:2015+AC:2013, Class B
EN55024:2010+A1:2015
IEC61000-4-2:2008, Criteria A
IEC61000-4-3:2006 + A2:2010, Criteria A
IEC61000-4-4:2012, Criteria A
IEC61000-4-5:2014, Criteria A
IEC61000-4-6:2013, Criteria A
IEC61000-4-8:2009, Criteria A
IEC61000-4-11:2004, Criteria A
IEC61000-4-11:2004, Criteria B
EN61000-3-2:2014, Class D
EN61000-3-3:2013
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Material
Dimension (LxWxH)
Weight
Type
enclosed
enclsoded
open frame
enclosed
open frame
continued on next page
Value
aluminum
91.4 x 62.0 x 39.2mm
76.2 x 50.8 x 32.0mm
210g
150g
www.recom-power.com
REV.: 9/2018