A high performance AC to DC external power supply family designed for medical applications. The ME30A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels
which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC
*
. The ME30A Series
models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +70°C,
delivering full rated output power up to +40°C and applicable output power derating at 70°C. These models are available
in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at
a variety of output connectors.
*
Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME30A0503F01
ME30A0903F01
ME30A1203F01
ME30A1503F01
ME30A1803F01
ME30A2403F01
ME30A4803F01
ME30A0503N01
ME30A0903N01
ME30A1203N01
ME30A1503N01
ME30A1803N01
ME30A2403N01
ME30A4803N01
Volts
5.0V
9.0V
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
12.0V
15.0V
18.0V
24.0V
48.0V
Output
Current
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
Output
Power
20W
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
Ripple &
Noise¹
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Overvoltage
Trip Range
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
Output
Connector
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Input
Configuration
Class I
Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II
Desktop,
IEC60320 C8
Receptacle
www.slpower.com
ME30 Rev 2.1
9-NOV-17
1
ME30
20 to 30W Single Output External Power
Medical Series
Model Selection
(continued)
Model
Number
ME30A0503Q01
ME30A0903Q01
ME30A1203Q01
ME30A1503Q01
ME30A1803Q01
ME30A2403Q01
ME30A4803Q01
ME30A0503B01
ME30A0903B01
ME30A1203B01
ME30A1503B01
ME30A1803B01
ME30A2403B01
ME30A4803B01
ME30A0503C01
ME30A0903C01
ME30A1203C01
ME30A1503C01
ME30A1803C01
ME30A2403C01
ME30A4803C01
Notes:
Volts
5.0V
9.0V
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
12.0V
15.0V
18.0V
24.0V
48.0V
Output
Current
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
Output
Power
20W
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
Ripple &
Noise¹
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
Overvoltage
Trip Range
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
Output
Connector
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Input
Configuration
Class II
Desktop,
IEC60320 C18
Receptacle
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II Wall-
Plug,
Interchangeabl
e Blades
(North
American Blade
included)²
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II Wall-
Plug, Fixed
North American
Blades³
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME30B1203F01).
General Specifications
Parameter
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
(Input to Ground)
Patient Leakage
Current
(Output to Earth)
Efficiency
Output Power
No Load Input
Power
Ripple and Noise
Output Voltage
Transient
Response
Regulation
Specification
100-240Vac, ±10%, 47-63Hz, 1∅
115Vac: 1.2A, 230Vac:0.6A
264Vac, cold start: will not exceed 40A
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
<500μA@264Vac, 60Hz, NC
<1mA@264Vac, 60Hz, SFC
<100μA@264Vac, 60Hz, NC
<500µA@264Vac, 60Hz, SFC
>87%, typical
20 to 30W continuous – See models chart for
specific voltage model ratings.
<0.1W per DoE Efficiency Level VI
Requirements
See models chart on pg 1.
See models chart on pg 1.
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load,
∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
See models chart on pg 1.
Parameter
Turn On Time
Hold-up Time
Overtemperature
Protection
Overload Protection
Short Circuit
Protection
Overvoltage
Protection
Isolation
Safety Standards
Operating
Temperature
Storage Temperature
Altitude
Relative Humidity
Drop Test
Specification
Less than 700mS @115Vac, full load
20mS min., at full Load, 100Vac input
Will shutdown upon an overtemperature
condition, auto-recovery.
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
Hiccup mode, see models chart for trip
ranges.
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
EN/IEC/UL60601-1-1, 3rd edition
-20°C to +70°C. See curve for derating
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
1.4m from table top to wooden platform, 6
faces.
www.slpower.com
ME30 Rev 2.1
9-NOV-17
2
ME30
20 to 30W Single Output External Power
Medical Series
General Specifications
(continued)
Parameter
Specification
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of
10 sweeps / axes, 3 axes
See outline drawings
Parameter
Specification
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
>8 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient
25°C at 24 hrs per day, 365 days/year, 6
power up cycles per day.
Vibration
Shock
Dimensions
MTBF
Weight
250g
E-Cap Life
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
EMI/EMC Compliance
Parameter
Conducted Emissions:
Radiated Emissions:
Common Mode Noise:
Electro-Static Discharge (ESD)
Immunity on Power ports:
Radiated RF EM Fields Susceptibility
Electrical Fast Transients (EFT) /Bursts:
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Specification
EN55011/CISPR11 Class B, FCC Part 15.107, Class B:
>
6db margin typ, at 115 and 230Vac
EN55022/CISPR11 Class B, FCC Part 15.109, Class B: >3db margin typ, at 115 and 230Vac
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4
th
Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4
th
Edition, Table 5.
EN55024/IEC1000-4-8, Level 4: 30A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
-- 60% dip for 100mS, Criteria B
-- 30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
EN55011/EN61000-3-2, Class A
EN61000-3-3
Voltage Interruptions, Dips, Sags &
Surges
Harmonic Current Emissions
Flicker Test
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Note: Performance criteria are based are defined as following:
A – Normal performance during and after the test
B – Temporary degradation, self-recoverable
C – Temporary degradation, operator intervention required to recover the operation
D – Permanent damage
www.slpower.com
ME30 Rev 2.1
9-NOV-17
3
ME30
20 to 30W Single Output External Power
Medical Series
Output Power Derating
Output power is derated above 40°C as follows, for operation over the entire AC input range (90-264Vac).
35
30
Output Power - Watts
25
20
15
10
5
0
-40
-30
-20
-10
0
10
20
30
40
Ambient Temperature ͦ
C
50
60
70
80
5V
12V-48V
Mechanical Drawing
54
55
33.6
95.5
1450±50
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Notes:
1. All dimensions in mm.
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
www.slpower.com
ME30 Rev 2.1
9-NOV-17
1450±50
Cable length for 5V and 9V models is 1200 +/- 50mm
96
Cable length for 5V and 9V models is 1200 +/- 50mm
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
4
ME30
20 to 30W Single Output External Power
Medical Series
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
These are the most common standard connectors. SL Power has the capability to incorporate any non-standard output connector. All output
connectors are limited by wattage range and application type. The SL Power applications team is available to provide professional support and