First Sensor PIN PD Data Sheet
Part Description PS100-6b THD
Order # 5000025
Features
• 100 mm² PIN detector
• High sensitivity
• High shunt resistance
• Blue-green enhanced
Description
Blue-green enhanced square active
area PIN photodiode with 100 mm²
active area. Non-hermetic ceramic
carrier package with silicone potting.
Clear glass window on special
request.
Application
• Precision photometry
• Analytical instruments
• Medical equipment
• Fluorescence detector
RoHS
2011/65/EU
Absolute maximum ratings
Symbol
T
STG
T
OP
V
max
I
PEAK
Parameter
Storage temp
Operating temp
Max reverse voltage
Peak DC current
Min
-20
-20
Max
80
60
50
10
Unit
°C
°C
V
mA
Spectral response (23 °C)
0,70
0,60
0,50
Responsivity (A/W)
0,40
0,30
0,20
0,10
0,00
350
450
550
650
750
850
950
1050
Schematic
PIN 2
PIN 1
Wavelength (nm)
Electro-optical characteristics @ 23 °C
Symbol
Characteristic
Active area
Active area
Dark current
Capacitance
Responsivity
t
R
Rise time
Shunt Resistance
N.E.P.
Breakdown voltage
Test Condition
Min
Typ
Max
10000 x 10000
100
1.0
720
275
0.22
0.37
200
50
8 E-14
Unit
µm
mm²
nA
pF
pF
A/W
A/W
ns
MΩ
W/√Hz
V
I
D
C
V
BR
V
R
= 5 V
V
R
= 0 V
V
R
= 5 V
λ = 410 nm
λ = 550 nm
V
R
= 5 V; λ = 410 nm; R
L
= 50 Ω
V
R
= 10 mV
V
R
= 5 V; λ = 410 nm
I
R
= 2 µA
50
Rev. 30/01/2018
subject to change without notice
www.first-sensor.com
contact@first-sensor.com
Page 1/2
ata Sheet
First Sensor PIN PD Data Sheet
Part Description PS100-6b THD
Order # 5000025
00-6b THD
# 5000025
Quantum efficiency (23 °C)
100
Capacitance as fct of reverse bias (23 °C)
800
700
Quantum efficiency (%)
80
600
Capacitance (pF)
60
500
400
300
200
100
40
20
0
350
450
550
650
750
850
950
1050
0
0
5
10
15
20
25
30
35
40
45
50
Wavelength (nm)
Reverse bias (V)
Dark current as fct of bias (23 °C)
1,0E-08
Dark current (A)
1,0E-09
1,0E-10
1,0E-11
0
5
10
15
20
25
30
Reverse bias [V]
Package dimension:
Small quantities: Foam pad, boxed (12 cm x 16.5 cm)
Handling:
Please refer to document "Instructions for handling and processing"
Disclaimer:
Due to our strive for continuous improvement, specifications are subject to change within our PCN policy according to JESD46C.
Rev. 30/01/2018
subject to change without notice
www.first-sensor.com
contact@first-sensor.com
Page 2/2