1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
EMMA120250
X
-
XX
-
IC
-
CXX
Base Number
Reserved for Custom
Configurations
example of 12 Vdc, 2.5 A
DC Plug Type:
Factory
See output plug options
Designation
on page 3 for more details
Blades:
“blank” = North American, European, United Kingdom,
Australian
N = North American blade included
E = European blade included
B = United Kingdom blade included
A = Australian blade included
K = No blades included
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
EMMA 30W
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017│
page
2 of 4
INPUT
parameter
voltage
frequency
input current
inrush current
no load power consumption
at 240 V ac, cold start
conditions/description
min
90
47
typ
max
264
63
0.8
100
0.3
units
Vac
Hz
A
A
W
OUTPUT
parameter
line regulation
1
load regulation
2
voltage accuracy
hold-up time
switching frequency
temperature coefficient
Note:
1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
conditions/description
5 Vdc output
9 Vdc output
all other outputs
at 115 Vac
min
typ
±1
±6
±3
±2
±2
10
70
±0.05
max
units
%
%
%
%
%
ms
kHz
%/°C
PROTECTIONS
parameter
over voltage protection
short circuit protection
conditions/description
TVS component to clamp
continuous, auto restart
SAFETY & COMPLIANCE
parameter
isolation voltage
safety approvals
EMI/EMC
leakage current
MTBF
RoHS
as per MIL-HDBK-217F, 115 Vac, 25 °C
2011/65/EU
200,000
conditions/description
input to output
medical IEC 60601-1, EN 60601-1, UL 60601-1 4th edition
EN 55011 Class B, FCC CRF47 Part 18 Class B, EN 60601-1-2, EN 61000-3-(2,3),
IEC 61000-4-(2,4,5,6,8,11)
0.1
mA
hours
min
typ
max
5,656
units
Vdc
ENVIRONMENTAL
parameter
operating temperature
storage temperature
humidity
non-condensing
conditions/description
min
0
-20
typ
max
40
85
93
units
°C
°C
%
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
EMMA 30W
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017│
page
3 of 4
MECHANICAL
parameter
dimensions
ø3.5 +0/-0.1
conditions/description
108.67 x 61.98 x 36.70 (4.278 x 2.440 x 1.445 inch)
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
1 Introduction
In the mid-1960s, American scientist Maas conducted extensive experimental research on the charging process of open-cell batteries and proposed an acceptable charging curve for ...[详细]