5390
The 5300 Series handheld DMM is the most advanced digital mul-
timeter available in this price range. These meters incorporate a
special 50,000 count integrated circuit for high-accuracy (0.025%)
basic accuracy, high-reliability design. Measure AC Volts (True RMS
AC, AC + DC), Resistance, Capacitance and Frequency, in addition
to dB, Duty Cycle, Pulse Width, Power Line Disturbances and
Temperature. These meters also feature fuse and battery access
without breaking a calibration seal, great for ISO facilities. All mod-
els offer Peak Hold, Hold, Relative Mode and Backlight.
■Precision
50,000 count instruments
■0.025%
Basic DCV accuracy (Models 5380 and 5390)
■True
RMS reading
■Large
LCD display provides easy readability, all test
parameters in clear view.
■Best
value of any DMM
■CE
safety approved
5380
5370
5360
Features
5390
Basic Functions
Basic Accuracy
Voltage (Best Resolution)
Current (Best Resolution)
Frequency
Temperature Probe
Duty Cycle
Event Counter
Backlight
√
√
√
√
5380
5370
models
5360
0.1%
750VAC/1000VDC
(100µV)
10AAC/DC
(1µA)
0.025%
0.05%
750VAC/1000VDC
(10µV)
10AAC/DC
(10nA)
√
√
√
√
√
√
Specifications
5390
AC VOLTAGE
(Auto/Manual Ranging, True RMS)
Modes: AC or AC+DC, selectable
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
Ranges
DC to
1 kHz to
1kHz
4 kHz
500 mV
0.3%R+30D
1%R+3D
5V
0.3%R+30D
1%R+3D
50V
0.3%R+30D
1%R+30D
500V
0.3%R+30D
1%R+30D
750V
0.3%R+30D
50 kHz to 100 kHz (5V, 50 V, 500 V ranges): ±6% FS typical
Crest Factor: 6
models
5380
5370
4 kHz to
10 kHz
5%R+30D
2%R+30D
2%R+30D
2%R+30D
Not specified
10 kHz to
30 kHz
30 kHz to
50 kHz
Not specified
3%R+30D
3%R+30D1 mV
3%R+30D10 mV
Resolution
10µV
100 µV
10 MΩ/100 pF
10 MΩ/100 pF
100 mV
2%R+30D
2%R+30D
2%R+30D
Input
Impedance
10 MΩ/100 pF
or 1 GΩ
11 MΩ/100 pF
10 MΩ/100pF
model
5360
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Ranges
DC to 1kHz
1 kHz to
5 kHz
500 mV
1%R+3D
4%R+3D
5V
1%R+3D
1%R+3D
50V
1%R+3D
1%R+3D
500V
1%R+3D
1%R+3D
750V
1%R+3D
Crest Factor: 3
5 kHz to
10 kHz
2%R+30D
2%R+3D
2%R+3D
10kHz to
20 kHz
Not specified
2%R+3D
2%R+3D
2%R+3D
Not specified
20 kHz to
30 kHz
3%R+3D
3%R+3D
3%R+3D
Resolution
100µV
1m V
10m√
100 mV
1V
Input
Impedance
10 MΩ/100 pF
11 MΩ/100 pF
10 MΩ/100 pF
10 MΩ/100 pF
10 MΩ/100pF
28
Specifications
5390
5380
5370
models
5360
Unless otherwise stated, accuracy specifications apply from 18°C to 28°C. Specifications stated as n%R+nD, mean ±(n% of reading +n least significant digits). Accuracy specifications stated as
n%FS, mean ±(n% of Full Scale)
DC VOLTAGE (Auto Manual Ranging)
Ranges
500 mV
5V 0.025%R+2D
50V
500V
1000V
*(10 MΩ only on 5360)
AC/DC VOLTAGE PROTECTION
Overvoltage Protection
Transient Protection
DC CURRENT (Auto/Manual Ranging)
Ranges
500µA
5 mA
50mA
500 mA
10 A
Input
Impedance
10 MΩ or
1 GΩ*
10 MΩ
10 MΩ
10 MΩ
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
0.025%R+2D
Accuracy
0.05%R+2D
0.05%R+2D
0.05%R+2D
0.05%R+2D
0.05%R+2D
0.1%R+2D
0.1%R+2D
0.1%R+2D
0.2%R+2D
0.3%R+2D
Resolution
10 µV
11 MΩ
1 mV
10 mV
100 mV
±1100 V DC + AC Peak
6kV for 10 µs
Max Burden
Voltage
700 mV
700 mV
700 mV
1.5 V
500 mV
0.2%R+5D
0.2%R+2D
0.05%R+2D
0.2%R+2D
0.5%R+2D
Accuracy
Not Available
0.2%R+2D
0.2%R+2D
0.2%R+2D
0.3%R+2D
Resolution
10 nA
100 nA
1 µA
10 µA
1 mA
5390
5380
models
5370
20 kHz to
30 kHz
3.0%R+50D
2.0%R+50D
2.0%R+50D
Input
Impedance
700 mV
700 mV
700 mV
1.5 V
500 mV
5360
model
AC CURRENT (Auto/Manual Ranging, True RMS) AC or AC + DC
Accuracy at greater than 10% of range, 50,000 counts at Full Scale:
1 kHz to
10 kHz to
Ranges
DC to 1kHz
10 kHz
20 kHz
500 µA
0.75%R+50D
0.75%R+50D 1.0%R+50D
5 mA
0.6%R+50D
0.6%R+50D
1.0%R+50D
50 mA
0.6%R+50D
0.6%R+50D
1.0%R+50D
500 mA
0.7%R+50D
0.7%R+50D
Not specified
10 A
1% of FS
Not specified
Crest Factor: 6
AC CURRENT
Accuracy at greater than 10% of range, 5,000 counts at Full Scale:
Resolution
10 nA
100 nA
1 µA
10 µA
1 mA
Ranges
5 mA
50 mA
500 mA
10 A
Crest Factor: 3
DC to 5 kHz
1.0%R+3D
1.0%R+3D
1.0%R+3D
1% of FS
Resolution
1 µA
10 µA
100 mA
10 mA
Max Burden Voltage
700 mV
700 mV
1.5 V
500 mV
5390
Ranges
500
Ω
5 kΩ
50 kΩ
500 kΩ
5 MΩ
50 MΩ
5380
5370
models
5360
Resolution Test Current
10 mΩ
1 mA
100 mΩ
100 µA
1
Ω
10 µA
10
Ω
1 µA
100
Ω
100 µA
1 kΩ
10 µA
5390
Range
50 nF
500 nF
5 pF
50 µF
500 µF
5000 µF
50 mF
Accuracy
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
5380
5370
models
5360
RESISTANCE (Auto/Manual Ranging)
Accuracy
0.07%R+5D
0.07%R+2D
0.07%R+2D
0.07%R+2D
0.3%R+2D
1.0%R+2D
0.1%R+5D
0.1%R+2D
0.1%R+2D
0.1%R+2D
0.3%R+2D
1.0%R+2D
CAPACITANCE (Auto/Manual Ranging)
Max
Measurement Measurement
Resolution
Current
TIme
10 pF
100 nA
0.5 sec
100 pF
1 µA
0.5 sec
1 nF
10 µA
0.5 sec
10 nF
100 µA
0.5 sec
100 nF
1 mA
1.5 sec
1 µF
1 mA
3 sec/ mF
10 µF
1mA
3 sec/ mF
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
1.0%R+2D
2.0%R+2D
2.0%R+2D
5390 5380 5370 5360
FREQUENCY (Autoranging)
Range: 0.6 Hz to 500 kHz
Accuracy: ±(0.03% rdg+1 digit)
Ranges
0.6Hz to 5kHz
5 kHz to 50 kHz
50 kHz to 500 kHz
500 mV
(5390, 5370)
2% of FS
5% of FS
Not specified
5V to 500 V
500 µA to 500 mA
2% of FS
5% of FS
10% of FS
models
Accessories
SUPPLIED:
Test Leads, Spare Fuses, Instruction Manual
OPTIONAL:
TL-130A
TP-5300
AK-5350
AK-5300
750 V
100 V
250 V
Not specified
10 A
(5390, 5370)
2A
Not specified
General Purpose DMM Kit
Platinum Temperature Probe (for 5370)
Calibration Software (AK-5300 Hardware required)
Communications Hardware & LabView
®
drivers
Common Features
CONTINUITY - Threshold: 10Ω to 20Ω
DIODE TEST - Measures forward voltage drop of diode 0 to 1.999V
dB MEASUREMENT - 5390 and 5370 - Measures value relative to a reference from 1 to 9999
Ω,
5380 at 600Ω
PEAK HOLD - Selection: Peak + or Peak- (surge or sag) Acq. Time: <1rms
DUTY CYCLE: 5390, 5370, 5360-Modes: Selectable %+ ,or %- Min
EVENT COUNTER (Model 5390) - Range: 0 to 99999 (Totalize)
MIN PULSE DURATION: 2 µs
PULSE WIDTH (Model 5390) Modes: Selectable +, - Min. Pulse Width: 20µs Max Pulse Period: 12.5 µs
TEMPERATURE (Model 5370) Using Model TP-5300 Probe (Optional)
RANGE: -200° to +800°C
Computer Interface: RS232 capability built-in. Requires AK-5300
GENERAL
Display: 50,000 count resolution LCD, 0.55" (14mm) digit height,
34 segment analog bargraph
Measurement Rate: 2/sec., 20/sec - bargraph
Power: 9V (NEDA 1604) Battery Life: 500 hrs
Dimensions: 1.57 x 3.23 x 7.44" (40 x 82 x 189mm)
Weight: 14 oz (397g)
Three Year Warranty
For your nearest distributor, call 800-462-9832 or www.bkprecision.com
29