电子工程世界电子工程世界电子工程世界

产品描述

搜索

MDO3034

器件型号:MDO3034
器件类别:热门应用    测试测量   
厂商名称:Tektronix
厂商官网:https://www.tek.com/
下载文档

器件描述

Benchtop Oscilloscopes EMBEDDED (I2C,SPI) SERIAL BUS ANALYSIS

参数
产品属性属性值
产品种类:
Product Category:
Benchtop Oscilloscopes
制造商:
Manufacturer:
Tektronix
RoHS:YES
Number of Channels:4 Channel
Bandwidth:350 MHz
Sampling Rate:2.5 GS/s
Display Type:Color WVGA
Input Voltage:300 V
Equipment Type:Analog Oscilloscopes
商标:
Brand:
Tektronix
Features:6-in-1 integrated customizable oscilloscope that includes a spectrum analyzer, arbitrary function generator, logic analyzer, protocol analyzer and digital voltmeter/counter
接口类型:
Interface Type:
USB
系列:
Series:
MDO3000
工厂包装数量:
Factory Pack Quantity:
1
单位重量:
Unit Weight:
9.259 lbs

MDO3034器件文档内容

Mixed Domain Oscilloscopes

MDO3000 Series Datasheet

                                                                                Logic Analyzer (Optional)

                                                                                16 digital channels

                                                                                10 M record length on all channels

                                                                                121.2 ps timing resolution

                                                                                Protocol Analyzer (Optional)

                                                                                Serial bus support for I2C, SPI, RS-232/422/485/UART, USB 2.0,

                                                                                CAN, LIN, FlexRay, MIL-STD-1553, and Audio standards

                                                                                Digital Voltmeter (Free with product registration)

                                                                                4-digit AC RMS, DC, and AC+DC RMS voltage measurements

                                                                                5-digit frequency measurements

                                                                                Key features

Today's integrated designs need an oscilloscope that is just as integrated -    FastAcq™ high-speed waveform capture rate finds elusive signal

such as the MDO3000 Mixed Domain Oscilloscope (MDO) Series. It is the           anomalies quickly

ultimate 6-in-1 integrated oscilloscope that includes an integrated spectrum    Wave Inspector® Controls provide easy navigation and automated

analyzer, arbitrary function generator, logic analyzer, protocol analyzer, and  search of waveform data

digital voltmeter/counter. The MDO3000 is completely customizable and

fully upgradeable. Add the instruments and performance you need now - or        33 automated measurements and waveform histograms for simplified

later.                                                                          waveform analysis

Key performance specifications                                                  TekVPI® probe interface directly supports active, differential, and

                                                                                current probes for automatic scaling and units

Oscilloscope                                                                    9 in. (229 mm) WVGA widescreen color display

        2 and 4 analog channel models                                           Small footprint and lightweight – Only 5.8 in. (147 mm) deep and 9.2 lb.

        1 GHz, 500 MHz, 350 MHz, 200 MHz, 100 MHz bandwidth models              (4.2 kg)

        Bandwidth is upgradable (up to 1 GHz)                                   Spectral Analysis

        Up to 5 GS/s sample rate                                                Dedicated front-panel controls for commonly performed tasks

        10 M record length on all channels                                      Automated peak markers identify frequency and amplitude of

        >280,000 wfm/s maximum waveform capture rate                            spectrum peaks

        Standard passive voltage probes with 3.9 pF capacitive loading          Manual markers enable non-peak measurements

        and 1 GHz, 500 MHz, or 250 MHz analog bandwidth                         Trace types include: Normal, Average, Max Hold, and Min Hold

Spectrum Analyzer                                                               Spectrogram display enables easy observation and insight into

        Frequency range                                                         slowly changing RF phenomena

        Standard: 9 kHz - oscilloscope bandwidth                                Automated measurements include: Channel Power, Adjacent

        Optional: 9 kHz - 3 GHz                                                 Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW)

        Ultra-wide capture bandwidth up to 3 GHz                                Arbitrary Function Generation

Arbitrary Function Generator (Optional)                                         Generate predefined signals to quickly simulate missing devices in

        13 predefined waveform types                                            designs

        50 MHz waveform generation                                              Capture signals on analog or digital inputs, transfer to the arbitrary

        128 k arbitrary generator record length                                 edit memory, and replicate out from the AFG

        250 MS/s arbitrary generator sample rate                                Add noise to any signal to easily perform margin testing

                                                                                                                                    www.tektronix.com     1
Datasheet

   Mixed signal design and analysis                                   Oscilloscope

   Automated triggering, decode, and search on parallel buses         At the core of the MDO3000 Series is a world-class oscilloscope, offering

   Multichannel setup and hold triggering                             comprehensive tools that speed each stage of debug – from quickly

   MagniVu™ high-speed acquisition provides 121.2 ps fine timing      discovering anomalies and capturing them, to searching your waveform

   resolution on digital channels                                     record for events of interest and analyzing their characteristics and your

                                                                      device’s behavior.

   Protocol Analysis

   Trigger, decode and automatically search on packet-level content   Digital phosphor technology with FastAcq™ high-

   on most common embedded design serial bus standards.               speed waveform capture

   Export protocol decode tables for use in documenting results       To debug a design problem, first you must know it exists. Every design

   Digital Voltmeter and Frequency Counter                            engineer spends time looking for problems in their design, a time-

   Quickly validate voltage or frequency measurements at a glance     consuming and frustrating task without the right debug tools.

   Graphical readout provides information on stability of the         Digital phosphor technology provides you with fast insight into the real

   measurement                                                        operation of your device. Its fast waveform capture rate – greater than

   Fully upgradeable                                                  280,000 wfm/s with FastAcq – gives you a high probability of quickly seeing

                                                                      the infrequent problems common in digital systems: runt pulses, glitches,

   Add functionality, increase bandwidth or spectrum analyzer         timing issues, and more.

   frequency range over time as your needs change or budget allows    To further enhance the visibility of rarely occurring events, intensity grading

Optional application support                                          is used to indicate how often rare transients are occurring relative to normal

                                                                      signal characteristics. There are four waveform palettes available in

   Power analysis                                                     FastAcq acquisition mode.

   Limit and mask testing                                             The Temperature palette uses color-grading to indicate frequency of

                                                                      occurrence with hot colors like red/yellow indicating frequently

                                                                      occurring events and colder colors like blue/green indicating rarely

Need more performance?                                                occurring events.

                                                                      The Spectral palette uses color-grading to indicate frequency of

Need more input frequency range on the spectrum analyzer?             occurrence with colder colors like blue indicating frequently occurring

                                                                      events and hot colors like red indicating rarely occurring events.

Need to analyze analog, digital, and RF simultaneously?               The Normal palette uses the default channel color (like yellow for

Need more record length or a larger display?                          channel one) along with gray-scale to indicate frequency of occurrence

                                                                      where frequently occurring events are bright.

Consider the MDO4000B Series oscilloscopes www.tektronix.com/MDO4000  The Inverted palette uses the default channel color along with gray-

                                                                      scale to indicate frequency of occurrence where rarely occurring events

                                                                      are bright.

                                                                      These color palettes quickly highlight the events that over time occur more

                                                                      often or, in the case of infrequent anomalies, occur less often.

                                                                      Infinite or variable persistence choices determine how long waveforms stay

                                                                      on the display, helping you to determine how often an anomaly is occurring.

   ► 3 GHz and 6 GHz integrated spectrum analyzers

   ► Simultaneous acquisition of analog, digital and RF

   ► 20 M record length

   ► 10.4 inch XGA display

2  www.tektronix.com
                                                                                       Mixed Domain Osciloscopes -- MDO3000 Series

Digital phosphor technology with FastAcq enables greater than 280,000 wfms/s

waveform capture rate and real-time color-intensity grading.

Triggering

Discovering a device fault is only the first step. Next, you must capture the   Wave Inspector controls provide unprecedented efficiency in viewing, navigating, and

event of interest to identify root cause. To enable this, the MDO3000           analyzing waveform data. Zip through your long record by turning the outer pan control

contains over 125 trigger combinations providing a complete set of triggers     (1). Get details from the beginning to end in seconds. See something of interest and want

- including runt, logic, pulse width/glitch, setup and hold violation, serial   to see more details? Just turn the inner zoom control (2).

packet, and parallel data - to help quickly locate your event of interest. And

with up to a 10 M record length, you can capture many events of interest,       Zoom and pan

even thousands of serial packets, in a single acquisition for further analysis

while maintaining high resolution to zoom in on fine signal details.            A dedicated, two-tier front-panel control provides intuitive control of both

                                                                                zooming and panning. The inner control adjusts the zoom factor (or zoom

                                                                                scale); turning it clockwise activates zoom and goes to progressively higher

                                                                                zoom factors, while turning it counterclockwise results in lower zoom

                                                                                factors and eventually turning zoom off. No longer do you need to navigate

                                                                                through multiple menus to adjust your zoom view. The outer control pans

                                                                                the zoom box across the waveform to quickly get to the portion of waveform

                                                                                you are interested in. The outer control also utilizes force-feedback to

                                                                                determine how fast to pan on the waveform. The farther you turn the outer

                                                                                control, the faster the zoom box moves. Pan direction is changed by simply

                                                                                turning the control the other way.

                                                                                User marks

                                                                                Press the Set Mark front-panel button to place one or more marks on the

Over 125 trigger combinations make capturing your event of interest easy.       waveform. Navigating between marks is as simple as pressing the

                                                                                Previous (←) and Next (→) buttons on the front panel.

Wave Inspector® waveform navigation and automated                               Search marks

search

With long record lengths, a single acquisition can include thousands of         The Search button allows you to automatically search through your long

screens of waveform data. Wave Inspector®, the industry’s best tool for         acquisition looking for user-defined events. All occurrences of the event are

waveform navigation and automated search, enables you to find events of         highlighted with search marks and are easily navigated to, using the front-

interest in seconds.                                                            panel Previous (←) and Next (→) buttons. Search types include edge,

                                                                                pulse width/glitch, timeout, runt, logic, setup and hold, rise/fall time, parallel

                                                                                bus, and I2C, SPI, RS-232/422/485/UART, USB 2.0, CAN, LIN, FlexRay,

                                                                                MIL-STD-1553, and Audio packet content. A search mark table provides a

                                                                                tabular view of the events found during the automated search. Each event

                                                                                is shown with a time stamp, making timing measurements between events

                                                                                easy.

                                                                                                                                            www.tektronix.com              3
Datasheet

                                                                                         Waveform analysis

                                                                                         Verifying that your prototype’s performance matches simulations and meets

                                                                                         the project’s design goals requires analyzing its behavior. Tasks can range

                                                                                         from simple checks of rise times and pulse widths to sophisticated power

                                                                                         loss analysis and investigation of noise sources.

                                                                                         The oscilloscope offers a comprehensive set of integrated analysis tools

                                                                                         including waveform- and screen-based cursors, automated measurements,

                                                                                         advanced waveform math including arbitrary equation editing, FFT

                                                                                         analysis, waveform histograms, and trend plots for visually determining how

                                                                                         a measurement is changing over time.

Search step 1: You define what you would like to find.

                                                                                         Automated measurement readouts provide repeatable, statistical views of waveform

                                                                                         characteristics.

Search step 2: Wave Inspector automatically searches through the record and marks

each event with a hollow white triangle. You can then use the Previous and Next buttons

to jump from one event to the next.

                                                                                         Each measurement has help text and graphics associated with it that help explain how

                                                                                         the measurement is made.

                                                                                         Waveform histograms show visually how waveforms vary over time.

                                                                                         Horizontal waveform histograms are especially useful for gaining insight

Search step 3: The Search Mark table provides a tabular view of each of the events       into how much jitter is on a clock signal, and what the distribution of that

found by the automated search. Each event is shown with a time stamp making timing       jitter is. Vertical histograms are especially useful for gaining insight into how

measurements between events easy.                                                        much noise is on a signal, and what the distribution of that noise is.

                                                                                         Measurements taken on a waveform histogram provide analytical

                                                                                         information about the distribution of a waveform histogram, providing

                                                                                         insight into just how broad a distribution is, the amount of standard

                                                                                         deviation, the mean value, etc.

4  www.tektronix.com
                                                                                             Mixed Domain Osciloscopes -- MDO3000 Series

Waveform histogram of a rising edge showing the distribution of edge position (jitter) over  Viewing an NTSC full color bar signal image. Video picture mode contains automatic

time. Included are numeric measurements made on the waveform histogram data.                 contrast and brightness settings as well as manual controls.

Video design and development                                                                 Power analysis (optional)

Many video engineers have remained loyal to analog oscilloscopes,                            Ever increasing consumer demands for longer battery-life devices and for

believing the intensity gradations on an analog display are the only way to                  green solutions that consume less power require power-supply designers to

see certain video waveform details. The fast waveform capture rate of the                    characterize and minimize switching losses to improve efficiency. In

MDO3000, coupled with its intensity-graded view of the signal, provides the                  addition, the supply’s power levels, output purity, and harmonic feedback

same information-rich display as an analog oscilloscope, but with much                       into the power line must be characterized to comply with national and

more detail and all the benefits of digital scopes.                                          regional power quality standards. Historically, making these and many

Standard features such as IRE and mV graticules, holdoff by fields, video                    other power measurements on an oscilloscope has been a long, manual,

polarity, HDTV and custom (nonstandard) video triggers, and an Autoset                       and tedious process. The MDO3000’s optional power analysis tools greatly

smart enough to detect video signals, make these the easiest to use                          simplify these tasks, enabling quick, repeatable and accurate analysis of

oscilloscopes on the market for video applications. And with high                            power quality, switching loss, harmonics, safe operating area (SOA),

bandwidth, four analog inputs, and a built-in 75 Ω input termination (not                    modulation, ripple, and slew rate (di/dt, dv/dt). Completely integrated into

available on 1 GHz models), the oscilloscope provides ample performance                      the oscilloscope, the power analysis tools provide automated, repeatable

for analog and digital video use. There is even a video picture mode                         power measurements with a touch of a button.

enabling you to see the picture of the video signal you are viewing – for

NTSC and PAL signals.

                                                                                             Power Quality measurement table. Automated power measurements enable quick and

                                                                                             accurate analysis of common power parameters.

Viewing an NTSC video signal. Notice the intensity-graded view provided by the

MDO3000's ability to represent time, amplitude, and distribution over time.

                                                                                                                                                           www.tektronix.com     5
Datasheet

Limit/Mask testing (optional)                                                         TekVPI® probe interface

A common task during the development process is characterizing the                    The TekVPI probe interface sets the standard for ease of use in probing. In

behavior of certain signals in a system. One method, called limit testing, is         addition to the secure, reliable connection that the interface provides,

to compare a tested signal to a known good or "golden" version of the                 TekVPI probes feature status indicators and controls, as well as a probe

same signal with user-defined vertical and horizontal tolerances. Another             menu button right on the comp box itself. This button brings up a probe

common method, called mask testing, is to compare a tested signal to a                menu on the oscilloscope display with all relevant settings and controls for

mask, looking for where a signal under test violates the mask. The                    the probe. The TekVPI interface enables direct attachment of current

MDO3000 Series offers both limit and mask testing capability useful for               probes without requiring a separate power supply. TekVPI probes can be

long-term signal monitoring, characterizing signals during design, or testing         controlled remotely through USB, GPIB, or LAN, enabling more versatile

on a production line. Tailor a test to your specific requirements by defining         solutions in ATE environments. The TekVPI inputs provide up to 25 W of

test duration in number of waveforms or time, a violation threshold that              power to the front panel connectors from the internal power supply.

must be met before considering a test a failure, counting hits along with

statistical information, and actions upon violations, test failure, and test

complete. Whether specifying a mask from a known good signal or from a

custom mask, conducting pass/fail tests in search of waveform anomalies

such as glitches has never been easier.

                                                                                      TekVPI probe interface simplifies connecting your probes to the oscilloscope.

Limit Test showing a mask created from a golden waveform and compared against a live  Remote connectivity and instrument control

signal. Results showing statistical information about the test are displayed.         Exporting data and measurements is as simple as connecting a USB cable

Standard passive voltage probes                                                       from the oscilloscope to your PC. Key software applications –

                                                                                      OpenChoice® Desktop, and Microsoft Excel and Word toolbars – are

The MDO3000 Series include passive voltage probes with industry best                  included standard with each oscilloscope to enable fast and easy direct

capacitive loading of only 3.9 pF. The included TPP probes minimize the               communication with your Windows PC.

impact on devices under test and accurately deliver signals to the                    The included OpenChoice Desktop enables fast and easy communication

oscilloscope for acquisition and analysis. The following table shows which            between the oscilloscope and your PC through USB or LAN for transferring

TPP probes come standard with each MDO3000 model.                                     settings, waveforms, and screen images.

MDO3000 model                            Included probe                               The embedded e*Scope® capability enables fast control of the oscilloscope

MDO3012, MDO3014,                        TPP0250: 250 MHz, 10x passive voltage        over a network connection through a standard web browser. Simply enter

MDO3022, MDO3024                         probe. One per analog channel                the IP address or network name of the oscilloscope and a web page will be

MDO3032, MDO3034,                        TPP0500B: 500 MHz, 10x passive               served to the browser. Transfer and save settings, waveforms,

MDO3052, MDO3054                         voltage probe. One per analog channel        measurements, and screen images or make live control changes to settings

MDO3102, MDO3104                         TPP1000: 1 GHz, 10x passive voltage          on the oscilloscope directly from the web browser.

                                         probe. One per analog channel

6  www.tektronix.com
                                                                                             Mixed Domain Osciloscopes -- MDO3000 Series

e*Scope in a web browser showing the display of an MDO3000. Use e*Scope to quickly

document your work by saving screen images, waveforms, or setups for later use.

Spectrum Analyzer

The MDO3000 is the first oscilloscope in its class to include an integrated

spectrum analyzer. Each oscilloscope includes a spectrum analyzer with a

frequency range of 9 kHz up to the analog bandwidth of the instrument. The

spectrum analyzer frequency range of each instrument can be upgraded

from 9 kHz to 3 GHz (option MDO3SA), enabling spectral analysis on most

consumer wireless standards.

Fast and accurate spectral analysis                                                 Key spectral parameters are adjusted quickly with the dedicated front-panel menus and

                                                                                    keypad.

When using the spectrum analyzer input, the MDO3000 Series display                  Intelligent, efficient markers

becomes a full-screen Frequency Domain view.

Key spectral parameters such as Center Frequency, Span, Reference                   In a traditional spectrum analyzer, it can be a very tedious task to turn on

Level, and Resolution Bandwidth are all adjusted quickly and easily using           and place enough markers to identify all your peaks of interest. The

the dedicated front-panel menus and keypad.                                         MDO3000 Series makes this process far more efficient by automatically

                                                                                    placing markers on peaks that indicate both the frequency and the

                                                                                    amplitude of each peak. You can adjust the criteria that the oscilloscope

                                                                                    uses to automatically find the peaks.

                                                                                    The highest amplitude peak is referred to as the reference marker and is

                                                                                    shown in red. Marker readouts can be switched between Absolute and

                                                                                    Delta readouts. When Delta is selected, marker readouts show each peak's

                                                                                    delta frequency and delta amplitude from the reference marker.

                                                                                    Two manual markers are also available for measuring non-peak portions of

                                                                                    the spectrum. When enabled, the reference marker is attached to one of

                                                                                    the manual markers, enabling delta measurements from anywhere in the

                                                                                    spectrum. In addition to frequency and amplitude, manual marker readouts

                                                                                    also include noise density and phase noise readouts depending on whether

                                                                                    Absolute or Delta readouts are selected. A "Reference Marker to Center"

MDO3000 frequency domain display.                                                   function instantly moves the frequency indicated by the reference marker to

                                                                                    center frequency.

                                                                                                                           www.tektronix.com                               7
Datasheet

                                                                                           Ultra-wide capture bandwidth

                                                                                           Today's wireless communications vary significantly with time, using

                                                                                           sophisticated digital modulation schemes and, often, transmission

                                                                                           techniques that involve bursting the output. These modulation schemes can

                                                                                           have very wide bandwidth as well. Traditional swept or stepped spectrum

                                                                                           analyzers are ill equipped to view these types of signals as they are only

                                                                                           able to look at a small portion of the spectrum at any one time.

                                                                                           The amount of spectrum acquired in one acquisition is called the capture

                                                                                           bandwidth. Traditional spectrum analyzers sweep or step the capture

                                                                                           bandwidth through the desired span to build the requested image. As a

                                                                                           result, while the spectrum analyzer is acquiring one portion of the spectrum,

                                                                                           the event you care about may be happening in another portion of the

                                                                                           spectrum. Most spectrum analyzers on the market today have 10 MHz

Automated peak markers identify critical information at a glance. As shown here, the five  capture bandwidths, sometimes with expensive options to extend that to

highest amplitude peaks that meet the threshold and excursion criteria are automatically   20, 40, or even 160 MHz in some cases.

marked along with the peak's frequency and amplitude.                                      In order to address the bandwidth requirements of modern RF, the

Spectrogram                                                                                MDO3000 Series provides up to 3 GHz of capture bandwidth. The

                                                                                           spectrum is generated from a single acquisition, thus guaranteeing you'll

The MDO3000 Series includes a spectrogram display which is ideal for                       see the events you're looking for in the frequency domain.

monitoring slowly changing RF phenomena. The x-axis represents

frequency, just like a typical spectrum display. However, the y-axis

represents time, and color is used to indicate amplitude.

Spectrogram slices are generated by taking each spectrum and "flipping it

up on its edge" so that it's one pixel row tall, and then assigning colors to

each pixel based on the amplitude at that frequency. Cold colors (blue,

green) are low amplitude and hotter colors (yellow, red) are higher

amplitude. Each new acquisition adds another slice at the bottom of the

spectrogram and the history moves up one row. When acquisitions are

stopped, you can scroll back through the spectrogram to look at any

individual spectrum slice.

                                                                                           Spectral display of a bursted communication both into a device through Zigbee at

                                                                                           900 MHz and out of the device through Bluetooth at 2.4 GHz, captured with a single

                                                                                           acquisition.

                                                                                           Spectrum traces

                                                                                           The MDO3000 Series spectrum analyzer offers four different traces or

                                                                                           views including Normal, Average, Max Hold, and Min Hold.

Spectrogram display illustrates slowly moving RF phenomena. As shown here, a signal

that has multiple peaks is being monitored. As the peaks change in both frequency and

amplitude over time, the changes are easily seen in the Spectrogram display.

8  www.tektronix.com
                                                                             Mixed Domain Osciloscopes -- MDO3000 Series

                                                                             RF probing

                                                                             Signal input methods on spectrum analyzers are typically limited to cabled

                                                                             connections or antennas. But with the optional TPA-N-VPI adapter, any

                                                                             active, 50 Ω TekVPI probe can be used with the spectrum analyzer on the

                                                                             MDO3000 Series. This enables additional flexibility when hunting for noise

                                                                             sources and enables easier spectral analysis by using true signal browsing

                                                                             on a spectrum analyzer input.

                                                                             In addition, an optional preamplifier accessory assists in the investigation of

                                                                             lower-amplitude signals. The TPA-N-PRE preamplifier provides 10 dB

                                                                             nominal gain across the 9 kHz – 3 GHz frequency range.

Normal, Average, Max Hold, and Min Hold spectrum traces

RF measurements

The MDO3000 Series includes three automated RF measurements -

Channel Power, Adjacent Channel Power Ratio, and Occupied Bandwidth.

When one of these RF measurements is activated, the oscilloscope

automatically turns on the Average spectrum trace and sets the detection

method to Average for optimal measurement results.

                                                                             The optional TPA-N-VPI adapter enables any active, 50 Ω TekVPI probe to be

                                                                             connected to the RF input.

                                                                             Arbitrary Function Generator (optional)

                                                                             The MDO3000 contains an optional integrated arbitrary function generator

                                                                             (option MDO3AFG), perfect for simulating sensor signals within a design or

                                                                             adding noise to signals to perform margin testing.

Automated Channel Power measurement                                          The integrated function generator provides output of predefined waveforms

                                                                             up to 50 MHz for sine, square, pulse, ramp/triangle, DC, noise, sin(x)/x

Advanced RF analysis                                                         (Sinc), Gaussian, Lorentz, exponential rise/fall, Haversine and cardiac.

The MDO3000 can save the baseband I and Q data from spectrum                 The arbitrary waveform generator provides 128 k points of record for

analyzer acquisitions to a .TIQ file. These files can then be imported into  storing waveforms from the analog input, a saved internal file location, a

Tektronix SignalVu-PC software for general purpose modulation and pulse      USB mass storage device, or from an external PC. Once a waveform is in

analysis or RSAVu for analysis of commercial wireless standards.             the edit memory of the arbitrary waveform generator, it can be modified via

                                                                             an on-screen editor and then replicated out of the generator. The

                                                                             MDO3000 is compatible with Tektronix’ ArbExpress PC-based waveform

                                                                             creation and editing software, making creation of complex waveforms fast

                                                                             and easy. Transfer waveform files to your MDO3000 edit memory via USB

                                                                             or LAN or using a USB mass storage device to be output from the AFG in

                                                                             the oscilloscope.

                                                                                                                                 www.tektronix.com            9
Datasheet

                                                                                  Color-coded digital waveform display

                                                                                  Color-coded digital traces display ones in green and zeros in blue. This

                                                                                  coloring is also used in the digital channel monitor. The monitor shows if

                                                                                  signals are high, low, or are transitioning so you can see channel activity at

                                                                                  a glance without having to clutter your display with unneeded digital

                                                                                  waveforms.

                                                                                  The multiple transition detection hardware shows you a white edge on the

                                                                                  display when the system detects multiple transitions. White edges indicate

                                                                                  that more information is available by zooming in or acquiring at faster

                                                                                  sampling rates. In most cases zooming in will reveal the pulse that was not

                                                                                  viewable with the previous settings. If the white edge is still present after

                                                                                  zooming in as far as possible, this indicates that increasing the sample rate

                                                                                  on the next acquisition will reveal higher frequency information than the

Waveform type selection in the integrated AFG.                                    previous settings could acquire.

                                                                                  You can group digital waveforms and enter waveform labels by using a

                                                                                  USB keyboard. By simply placing digital waveforms next to each other, they

                                                                                  form a group.

Arbitrary waveform editor showing the point-by-point editor.

Logic Analyzer (optional)                                                         With color-coded digital waveform display, groups are created by simply placing digital

The logic analyzer (option MDO3MSO) provides 16 digital channels which            channels together on the screen, allowing digital channels to be moved as a group.

are tightly integrated into the oscilloscope's user interface. This simplifies

operation and makes it possible to solve mixed-signal issues easily.              Once a group is formed, you can position all the channels contained in that

                                                                                  group collectively. This greatly reduces the normal setup time associated

                                                                                  with positioning channels individually

                                                                                  MagniVu™ high-speed acquisition

                                                                                  The main digital acquisition mode on the MDO3000 Series will capture up

                                                                                  to 10 M at 500 MS/s (2 ns resolution). In addition to the main record, the

                                                                                  MDO3000 provides an ultra high-resolution record called MagniVu which

                                                                                  acquires 10,000 points at up to 8.25 GS/s (121.2 ps resolution). Both main

                                                                                  and MagniVu waveforms are acquired on every trigger and can be switched

                                                                                  between in the display at any time, running or stopped. MagniVu provides

                                                                                  significantly finer timing resolution than comparable oscilloscopes on the

                                                                                  market, instilling confidence when making critical timing measurements on

The MDO3000 with MDO3MSO option provides 16 integrated digital channels enabling  digital waveforms.

you to view and analyze time-correlated analog and digital signals.

10  www.tektronix.com
                                                                                         Mixed Domain Osciloscopes -- MDO3000 Series

                                                                                         Serial Protocol Triggering and Analysis

                                                                                         (optional)

                                                                                         On a serial bus, a single signal often includes address, control, data, and

                                                                                         clock information. This can make isolating events of interest difficult.

                                                                                         Automatic trigger, decode, and search on bus events and conditions gives

                                                                                         you a robust set of tools for debugging serial buses.

The MagniVu high-resolution record provides 121.2 ps timing resolution, enabling you to

take critical timing measurements on your digital waveforms.

P6316 MSO probe

This unique probe design offers two eight-channel pods, simplifying the

process of connecting to the device-under-test. When connecting to square

pins, the P6316 can connect directly to 8×2 square pin headers spaced on                 Triggering on a specific address and data packet going across an I2C bus. The yellow

tenth-inch centers. When more attachment flexibility is required, you can                waveform is clock and the blue waveform is the data. A bus waveform provides decoded

use the included flying lead sets and grabbers to clip onto surface mount                packet content including Start, Address, Read/Write, Data, and Stop.

devices or test points. The P6316 offers outstanding electrical

characteristics applying only 8 pF of capacitive loading with 101 kΩ input               Serial triggering

impedance.                                                                               Trigger on packet content such as start of packet, specific addresses,

                                                                                         specific data content, unique identifiers, etc. on popular serial interfaces

                                                                                         such as I2C, SPI, RS-232/422/485/UART, USB2.0, CAN, LIN, FlexRay,

                                                                                         MIL-STD-1553, and I2S/LJ/RJ/TDM.

                                                                                         Bus display

                                                                                         Provides a higher-level, combined view of the individual signals (clock,

                                                                                         data, chip enable, etc.) that make up your bus, making it easy to identify

                                                                                         where packets begin and end and identifying sub-packet components such

                                                                                         as address, data, identifier, CRC, etc.

                                                                                         Bus decoding

                                                                                         Tired of having to visually inspect the waveform to count clocks, determine

                                                                                         if each bit is a 1 or a 0, combine bits into bytes, and determine the hex

                                                                                         value? Let the oscilloscope do it for you! Once you’ve set up a bus, the

                                                                                         MDO3000 Series will decode each packet on the bus, and display the value

The P6316 MSO probe offers two eight-channel pods to simplify connecting to your         in hex, binary, decimal (USB, LIN, FlexRay, and MIL-STD-1553 only),

device.                                                                                  signed decimal (I2S/LJ/RJ/TDM only), or ASCII (USB, MIL-STD-1553 and

                                                                                         RS-232/422/485/UART only) in the bus waveform.

                                                                                                                                                www.tektronix.com              11
Datasheet

Serial bus technologies supported by the MDO3000                                      Search (serial triggering)

Technology                         Trigger, Decode,    Order product                  Serial triggering is very useful for isolating the event of interest, but once

                                   Search                                             you’ve captured it and need to analyze the surrounding data, what do you

                                                                                      do? In the past, users had to manually scroll through the waveform

Embedded           I2C             Yes                 MDO3EMBD                       counting and converting bits and looking for what caused the event. You

                   SPI             Yes                 MDO3EMBD                       can have the oscilloscope automatically search through the acquired data

Computer           RS232/422/485,  Yes                 MDO3COMP                       for user-defined criteria including serial packet content. Each occurrence is

                   UART                                                               highlighted by a search mark. Rapid navigation between marks is as simple

USB                USB LS, FS, HS  Yes (trigger on LS  MDO3USB                        as pressing the Previous (←) and Next (→) buttons on the front panel.

                                   and FS only; HS

                                   decode only on                                     Digital Voltmeter (DVM) and Frequency

                                   1 GHz models)                                      Counter

Automotive         CAN             Yes                 MDO3AUTO

                   LIN             Yes                 MDO3AUTO                       The MDO3000 contains an integrated 4-digit digital voltmeter (DVM) and 5-

                   FlexRay         Yes                 MDO3FLEX                       digit frequency counter. Any of the analog inputs can be a source for the

Military and       MIL-STD-1553    Yes                 MDO3AERO                       voltmeter, using the same probes that are already attached for general

aerospace                                                                             oscilloscope usage. The easy-to-read display offers you both numeric and

                                                                                      graphical representations of the changing measurement values. The

Audio              I2S             Yes                 MDO3AUDIO                      display also shows minimum, maximum, and average values of the

                   LJ, RJ          Yes                 MDO3AUDIO                      measurement as well as the range of values measured over the previous

                   TDM             Yes                 MDO3AUDIO                      five second interval. The DVM and frequency counter is available on any

                                                                                      MDO3000 and is activated when you register your product.

Event table

In addition to seeing decoded packet data on the bus waveform itself, you

can view all captured packets in a tabular view much like you would see in

a software listing. Packets are time stamped and listed consecutively with

columns for each component (Address, Data, etc.). You can save the event

table data in .CSV format.

                                                                                      A DC measurement value is shown with a five second variation along with minimum,

                                                                                      maximum, and average voltage values. The frequency of the waveform is also shown.

Event table showing decoded identifier, DLC, DATA, and CRC for every CAN packet in a

long acquisition.

12  www.tektronix.com
                                                                                        Mixed Domain Osciloscopes -- MDO3000 Series

Designed to fit into your work environment                                              Compact form factor

                                                                                        With the compact, portable form factor, you can easily move the

                                                                                        oscilloscope between labs. And with a depth of just 5.8 inches (147 mm), it

                                                                                        saves you valuable space on your test bench. The MDO3000 has all the

                                                                                        tools you'll need for everyday debug tasks, all in a single instrument.

The MDO3000 is designed to fit into your work environment. With six instruments in one

small, portable package, the MDO3000 offers a unique combination of debug tools

without taking up valuable space on your bench.

Large high-resolution display                                                           The MDO3000 Series compact form factor frees up valuable space on your bench or

                                                                                        desktop while making sure you will always have the debug tools you need.

The MDO3000 Series features a 9 inch (229 mm) wide-screen, high-

resolution (800 × 480 WVGA) display for seeing intricate signal details.

Connectivity

The MDO3000 contains a number of ports which can be used to connect

the instrument to a network, directly to a PC, or other test equipment.

Front and rear USB host ports enable easy transfer of screen shots,

instrument settings, and waveform data to a USB mass storage device.

A USB keyboard can also be attached to a USB host port for data

entry.

Rear USB device port is useful for controlling the oscilloscope remotely

from a PC or for printing directly to a PictBridge®-compatible printer.

The standard 10/100 Ethernet port on the rear of the instrument

enables easy connection to networks, provides network and e-mail

printing, and provides LXI Core 2011 compatibility.

A video out port on the rear of the instrument allows the display to be

exported to an external monitor or projector.

                                                                                                             www.tektronix.com                                           13
Datasheet

Specifications

All specifications apply to all models unless noted otherwise.

                               MDO3012   MDO3014                MDO3022   MDO3024          MDO3032          MDO3034     MDO3052   MDO3054   MDO3102  MDO3104

    Analog channels            2         4                      2         4                2                4           2         4         2        4

    Analog channel bandwidth   100 MHz   100 MHz                200 MHz   200 MHz          350 MHz          350 MHz     500 MHz   500 MHz   1 GHz    1 GHz

    Rise time                  4 ns      4 ns                   2 ns      2 ns             1.14 ns          1.14 ns     800 ps    800 ps    400 ps   400 ps

    (10 mV/div setting with

    50Ω input termination)

    Sample rate (1 ch)         2.5 GS/s  2.5 GS/s               2.5 GS/s  2.5 GS/s         2.5 GS/s         2.5 GS/s    2.5 GS/s  2.5 GS/s  5 GS/s   5 GS/s

    Sample rate (2 ch)         2.5 GS/s  2.5 GS/s               2.5 GS/s  2.5 GS/s         2.5 GS/s         2.5 GS/s    2.5 GS/s  2.5 GS/s  5 GS/s   5 GS/s

    Sample rate (4 ch)         -         2.5 GS/s               -         2.5 GS/s         -                2.5 GS/s    -         2.5 GS/s  -        2.5 GS/s

    Record length (1 ch)       10 M      10 M                   10 M      10 M             10 M             10 M        10 M      10 M      10 M     10 M

    Record length (2 ch)       10 M      10 M                   10 M      10 M             10 M             10 M        10 M      10 M      10 M     10 M

    Record length (4 ch)       -         10 M                   -         10 M             -                10 M        -         10 M      -        10 M

    Digital channels with      16        16                     16        16               16               16          16        16        16       16

    MDO3MSO option

    Arbitrary Function         1         1                      1         1                1                1           1         1         1        1

    Generator outputs with

    MDO3AFG option

    Spectrum analyzer          1         1                      1         1                1                1           1         1         1        1

    channels

    Standard spectrum          9 kHz -   9 kHz -                9 kHz -   9 kHz -          9 kHz -          9 kHz -     9 kHz -   9 kHz -   9 kHz -  9 kHz -

    analyzer frequency range   100 MHz   100 MHz                200 MHz   200 MHz          350 MHz          350 MHz     500 MHz   500 MHz   1 GHz    1 GHz

    Optional spectrum          9 kHz -   9 kHz -                9 kHz -   9 kHz -          9 kHz -          9 kHz -     9 kHz -   9 kHz -   9 kHz -  9 kHz -

    analyzer frequency range   3 GHz     3 GHz                  3 GHz     3 GHz            3 GHz            3 GHz       3 GHz     3 GHz     3 GHz    3 GHz

    with MDO3SA option

Vertical system analog channels

    Hardware bandwidth limits

    ≥350 MHz models                      20 MHz or 250 MHz

    100 MHz and 200 MHz models           20 MHz

    Input coupling                       AC, DC

    Input impedance                      1 MΩ ±1%, 50 Ω ±1%, 75 Ω ±1%; 75          Ω  not  available on  1  GHz models

    Input sensitivity range

    1 MΩ                                 1 mV/div to 10 V/div

    50 Ω, 75 Ω                           1 mV/div to 1 V/div

    Vertical resolution                  8 bits (11 bits with Hi Res)

    Maximum input voltage

    1 MΩ                                 300 VRMS CAT II with peaks ≤ ±425 V

    50 Ω, 75 Ω                           5 VRMS with peaks ≤ ±20 V

14  www.tektronix.com
                                                                                     Mixed Domain Osciloscopes -- MDO3000                  Series

Vertical system analog channels

DC gain accuracy              ±1.5% for 5 mV/div and above, derated at 0.10%/°C above 30 °C

                              ±2.0% for 2 mV/div, derated at 0.10%/°C above 30 °C

                              ±2.5% for 1 mV/div, derated at 0.10%/°C above 30 °C

                              ±3.0% for variable gain, derated 0.10%/°C above 30 °C

Channel-to-channel isolation  Any two channels at equal vertical scale ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth

Offset range                  Volts/div setting                      Offset range

                                                                     1 M Ω input             50 Ω, 75 Ω input

                              1 mV/div to 50 mV/div                  ±1 V                    ±1 V

                              50.5 mV/div to 99.5 mV/div             ±0.5 V                  ±0.5 V

                              100 mV/div to 500 mV/div               ±10 V                   ±10 V

                              505 mV/div to 995 mV/div               ±5 V                    ±5 V

                              1 V/div to 5 V/div                     ±100 V                  ±5 V

Vertical system digital channels

(Requires MDO3MSO option)

Input channels                16 digital (D15 to D0)

Thresholds                    Threshold per set of 8 channels

Threshold selections          TTL, CMOS, ECL, PECL, User-defined

User-defined threshold range  -15 V to +25 V

Maximum input voltage         -20 V to +30 V

Threshold accuracy            ±[100 mV + 3% of threshold setting]

Input dynamic range           50 Vp-p (threshold setting dependent)

Minimum voltage swing         500 mV

Input resistance              101 kΩ

Probe loading                 8 pF

Vertical resolution           1 bit

                                                                                                     www.tektronix.com                     15
Datasheet

Horizontal system analog channels

    Time base range

    1 GHz models                       400 ps/div to 1000 s/div

    ≤ 500 MHz models                   1 ns/div to 1000 s/div

    Maximum duration at highest

    sample rate (all/half channels)

    1 GHz models                       4/2 ms

    ≤ 500 MHz models                   4/4 ms

    Time-base delay time range         -10 divisions to 5000 s

    Channel-to-channel deskew range    ±125 ns

    Time base accuracy                 ±10 ppm over any ≥1 ms interval

Horizontal system digital channels

(Requires MDO3MSO option)

    Maximum sample rate (Main)         500 MS/s (2 ns resolution)

    Maximum record length (Main)       10 M

    Maximum sample rate (MagniVu)      8.25 GS/s (121.2 ps resolution)

    Maximum record length (MagniVu     10k centered on the trigger

    Minimum detectable pulse width     2 ns

    (typical)

    Channel-to-channel skew (typical)  500 ps

    Maximum input toggle rate          250 MHz (Maximum frequency sine wave that can accurately be reproduced as a logic square wave. Requires the use of a short

                                       ground extender on each channel. This is the maximum frequency at the minimum swing amplitude. Higher toggle rates can be

                                       achieved with higher amplitudes.)

Spectrum analyzer          input

    Capture bandwidth                  MDO3012, MDO3014 models: 100 MHz

                                       MDO3022, MDO3024 models: 200 MHz

                                       MDO3032, MDO3034 models: 350 MHz

                                       MDO3052, MDO3054 models: 500 MHz

                                       MDO3102, MDO3104 models: 1 GHz

                                       All models: 3 GHz with option MDO3SA

    Span                               MDO3012, MDO3014 models: 9 kHz – 100 MHz

                                       MDO3022, MDO3024 models: 9 kHz – 200 MHz

                                       MDO3032, MDO3034 models: 9 kHz – 350 MHz

                                       MDO3052, MDO3054 models: 9 kHz – 500 MHz

                                       MDO3102, MDO3104 models: 9 kHz – 1 GHz

                                       All models: 9 kHz – 3 GHz with option MDO3SA, in a 1-2-5 sequence

16  www.tektronix.com
                                                                                         Mixed Domain   Osciloscopes  -- MDO3000 Series

Spectrum analyzer input

Resolution bandwidth            20 Hz - 150 MHz in a 1-2-3-5 sequence

Reference level                 -130 dBm to +20 dBm in steps of 5 dBm

Vertical scale                  1 dB/div to 20 dB/div in a 1-2-5 sequence

Vertical position               -100 divs to +100 divs (displayed in dB)

Vertical units                  dBm, dBmV, dBµV, dBµW, dBmA, dBµA

Displayed average noise level

(DANL)

9 kHz - 50 kHz                  < -109 dBm/Hz (< -113 dBm/Hz typical)

50 kHz – 5 MHz                  < -126 dBm/Hz (< -130 dBm/Hz typical)

5 MHz - 2 GHz                   < -138 dBm/Hz (< -142 dBm/Hz typical)

2 GHz – 3 GHz                   < -128 dBm/Hz (< -132 dBm/Hz typical)

DANL with TPA-N-PRE preamp      Preamp set to "Auto", and Reference Level set to -40 dB

attached

9 kHz - 50 kHz                  < -117 dBm/Hz (< -121 dBm/Hz typical)

50 kHz – 5 MHz                  < -136 dBm/Hz (< -140 dBm/Hz typical)

5 MHz - 2 GHz                   < -148 dBm/Hz (< -152 dBm/Hz typical)

2 GHz – 3 GHz                   < -138 dBm/Hz (< -142 dBm/Hz typical)

Spurious response

2nd harmonic distortion         < -55 dBc (< -60 dBc typical)

(>100 MHz)

3rd harmonic distortion         < -53 dBc (< -58 dBc typical)

(>100MHz)

2nd order intermodulation       < -55 dBc (< -60 dBc typical)

distortion (>15 MHz)

3rd order intermodulation       < -55 dBc (< -60 dBc typical)

distortion (>15 MHz)

Residual response               < -78 dBm (≤ -15 dBm reference level and RF input terminated with 50Ω)

At 2.5 GHz                      <-67 dBm

At 1.25 GHz                     <-76 dBm

Crosstalk to spectrum analyzer

from oscilloscope channels

≤800 MHz input frequencies      < -60 dB from ref level (typical)

>800 MHz - 2 GHz input          < -40 dB from ref level (typical)

frequencies

Phase noise at 1 GHz CW

10 kHz                          < -81 dBc/Hz, < -85 dBc/Hz (typical)

100 kHz                         < -97 dBc/Hz, < -101 dBc/Hz (typical)

1 MHz                           < -118 dBc/Hz, < -122 dBc/Hz (typical)

                                                                                                                      www.tektronix.com  17
Datasheet

Spectrum analyzer input

    Level measurement uncertainty   Reference level 10 dBm to -15 dBm. Input level ranging from reference level to 40 dB below reference level. Specifications

                                    exclude mismatch error.

    18 °C - 28 °C                   < ±1.2 dBm (< ±0.6 dBm typical)

    Over operating range            < ±2.0 dBm

    Level measurement uncertainty   Preamp mode set to “Auto”. Reference level 10 dBm set to -40dBm. Input level ranging from reference level to 30 dB below

    with TPA-N-PRE preamp attached  reference level. Specifications exclude mismatch error.

    18 °C - 28 °C                   < ±1.5 dBm (typical) either preamp state

    Over operating range            < ±2.3 dBm either preamp state

    Frequency measurement accuracy  ±(([Reference Frequency Error] x [Marker Frequency]) + (span/750 + 2)) Hz; Reference Frequency Error = 10ppm (10 Hz / MHz)

    Maximum operating input level

    Average continuous power        +20 dBm (0.1 W)

    DC maximum before damage        ±40 V DC

    Maximum power before            +33 dBm (2 W)

    damage (CW)

    Maximum power before            +45 dBm (32 W) (<10 µs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm)

    damage (pulse)

    Maximum operating input level

    with TPA-N-PRE preamp attached

    Average continuous power        +20 dBm (0.1 W)

    DC maximum before damage        ±20 V DC

    Maximum power before            +30 dBm (1 W)

    damage (CW)

    Maximum power before            +45 dBm (32 W) (<10 μs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm)

    damage (pulse)

    Frequency domain trace types    Normal, Average, Max Hold, Min Hold

    Detection methods               +Peak, -Peak, Average, Sample

    Automatic markers               One to eleven peaks identified based on user-adjustable threshold and excursion values

    Manual markers                  Two manual markers indicating frequency, amplitude, noise density, and phase noise

    Marker readouts                 Absolute or Delta

    FFT windows                     FFT window                                               Factor

                                    Kaiser                                                   2.23

                                    Rectangular                                              0.89

                                    Hamming                                                  1.30

                                    Hanning                                                  1.44

                                    Blackman-Harris                                          1.90

                                    Flat-Top                                                 3.77

18  www.tektronix.com
                                                                                                  Mixed Domain Osciloscopes -- MDO3000 Series

Trigger system

Trigger modes                  Auto, Normal, and Single

Trigger coupling               DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)

Trigger holdoff range          20 ns to 8 s

Trigger sensitivity (typical)  Edge type, DC coupled

                               Trigger source                                                     Sensitivity

                               Any analog channel input                                           For 1 mV/div to 4.98 mV/div; 0.75 div from DC to 50 MHz,

                                                                                                  increasing to 1.3 div at instrument bandwidth

                                                                                                  ≥ 5mV/div: 0.40 div from DC to 50 MHz

                               Aux In (External); available on two-channel instruments only       200 mV from DC to 50 MHz, increasing to 500 mV at 200 M

                               Line                                                               Fixed

Trigger level ranges

Any input channel              ±8 divisions from center of screen, ±8 divisions from 0 V when vertical LF reject trigger coupling is selected

Aux In (External)              ±8 V

Line                           The line trigger level is fixed at about 50% of the line voltage.

Trigger frequency readout      Provides 6-digit frequency readout of triggerable events.

Trigger types

Edge                           Positive, negative, or either slope on any channel. Coupling includes DC, AC, HF reject, LF reject, and noise reject.

Sequence (B-trigger)           Trigger Delay by Time: 8 ns to 8 s. Or Trigger Delay by Events: 1 to 4,000,000 events. Not available when “Either” edge is

                               selected.

Pulse Width                    Trigger on width of positive or negative pulses that are >, <, =, ≠, or inside/outside a specified period of time.

Timeout                        Trigger on an event which remains high, low, or either, for a specified time period (4 ns to 8 s).

Runt                           Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.

Logic                          Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock

                               to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for all input channels defined as High, Low, or

                               Don’t Care.

Setup and Hold                 Trigger on violations of both setup time and hold time between clock and data present on any of the analog and digital input

                               channels.

                               Setup and hold trigger type                                        Description

                               Setup Time Range                                                   -0.5 ns to 1.024 ms

                               Hold Time Range                                                    1.0 ns to 1.024 ms

                               Setup + Hold Time Range                                            0.5 ns to 2.048 ms

Rise/Fall Time                 Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative, or either and time range is

                               4.0 ns to 8 s.

Video                          Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals.

                               480p/60, 576p/50, 720p/30, 720p/50, 720p/60, 875i/60, 1080i/50, 1080i/60, 1080p/24, 1080p/24sF, 1080p/25, 1080p/30,

                               1080p/50, 1080p/60

                               Custom bi-level and tri-level sync video standards.

I2C (optional)                 Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I2C buses up to 10 Mb/s.

SPI (optional)                 Trigger on SS active, Start of Frame, MOSI, MISO, or MOSI and MISO on SPI buses up to 50.0 Mb/s.

RS-232/422/485/UART            Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error up

(optional)                     to 10 Mb/s.

                                                                                                                                                   www.tektronix.com  19
Datasheet

Trigger system

    USB: Low speed (optional)     Trigger on Sync Active, Start of Frame, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet,

                                  Handshake Packet, Special Packet, Error.

                                  Token packet trigger - Any token type, SOF, OUT, IN, SETUP; Address can be specified for Any Token, OUT, IN, and SETUP

                                  token types. Address can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular value, or inside or outside of a range. Frame

                                  number can be specified for SOF token using binary, hex, unsigned decimal and don't care digits.

                                  Data packet trigger - Any data type, DATA0, DATA1; Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular data

                                  value, or inside or outside of a range.

                                  Handshake packet trigger - Any handshake type, ACK, NAK, STALL.

                                  Special packet trigger - Any special type, Reserved

                                  Error trigger - PID Check, CRC5 or CRC16, Bit Stuffing.

    USB: Full speed (optional)    Trigger on Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special

                                  Packet, Error.

                                  Token packet trigger - Any token type, SOF, OUT, IN, SETUP; Address can be specified for Any Token, OUT, IN, and SETUP

                                  token types. Address can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular value, or inside or outside of a range. Frame

                                  number can be specified for SOF token using binary, hex, unsigned decimal and don't care digits.

                                  Data packet trigger - Any data type, DATA0, DATA1; Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a particular data

                                  value, or inside or outside of a range.

                                  Handshake packet trigger - Any handshake type, ACK, NAK, STALL.

                                  Special packet trigger - Any special type, PRE, Reserved.

                                  Error trigger - PID Check, CRC5 or CRC16, Bit Stuffing.

    CAN (optional)                Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data,

                                  End of Frame, Missing ACK, or Bit Stuffing Error on CAN signals up to 1 Mb/s.

                                  Data can be further specified to trigger on ≤, <, =, >, ≥, or ≠ a specific data value. User-adjustable sample point is set to 50% by

                                  default.

    LIN (optional)                Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, Errors such as Sync, Parity, or Checksum

                                  Errors up to 100 kb/s (by LIN definition, 20 kb/s).

    FlexRay (optional)            Trigger on Start of Frame, Type of Frame (Normal, Payload, Null, Sync, Startup), Identifier, Cycle Count, Complete Header Field,

                                  Data, Identifier and Data, End of Frame or Errors such as Header CRC, Trailer CRC, Null Frame, Sync Frame, or Startup Frame

                                  Errors up to 100 Mb/s.

    MIL-STD-1553 (optional)       Trigger on Sync, Word Type 1 (Command, Status, Data), Command Word (set RT Address, T/R, Sub-address/Mode, Data Word

                                  Count/Mode Code, and Parity individually), Status Word (set RT Address, Message Error, Instrumentation, Service Request Bit,

                                  Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance (DBCA), Terminal Flag, and Parity

                                  individually), Data Word (user-specified 16-bit data value), Error (Sync, Parity, Manchester, Non-contiguous data), Idle Time

                                  (minimum time selectable from 2 µs to 100 µs; maximum time selectable from 2 µs to 100 µs; trigger on < minimum, > maximum,

                                  inside range, outside range).

                                  RT Address can be further specified to trigger on =, ≠, <, >, ≤, ≥ a particular value, or inside or outside of a range.

    I2S/LJ/RJ/TDM (optional)      Trigger on Word Select, Frame Sync, or Data. Data can be further specified to trigger on ≤, <, =, >, ≥, ≠ a specific data value, or

                                  inside or outside of a range. Maximum data rate for I2S/LJ/RJ is 12.5 Mb/s. Maximum data rate for TDM is 25 Mb/s.

    Parallel (available when      Trigger on a parallel bus data value. Parallel bus can be from 1 to 20 bits (from the digital and analog channels) in size. Binary and

    option MDO3MSO is installed)  Hex radices are supported.

1   Trigger selection of Command Word will trigger on Command and ambiguous Command/Status words. Trigger selection of Status Word will trigger on Status and ambiguous Command/Status words.

20  www.tektronix.com
                                                                              Mixed Domain Osciloscopes -- MDO3000 Series

Acquisition system

Acquisition modes

     Sample             Acquire sampled values.

     Peak Detect        Captures glitches as narrow as 1.5 ns (1 GHz models), 2.0 ns (500 MHz models), 3.0 ns (350 MHz models), 5.0 ns (200 MHz

                        models), 7.0 ns (100 MHz models) at all sweep speeds

     Averaging          From 2 to 512 waveforms included in average.

     Envelope           Min-max envelope reflecting Peak Detect data over multiple acquisitions. Number of waveforms in the envelope selectable

                        between 1 and 2000 and infinity

     Hi Res             Real-time boxcar averaging reduces random noise and increases vertical resolution.

     Roll               Scrolls waveforms right to left across the screen at sweep speeds slower than or equal to 40 ms/div.

     FastAcq™           FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events, capturing >280,000 wfms/s on

                        1 GHz models and >235,000 wfms/s on 100 MHz – 500 MHz models.

Waveform measurements

Cursors                 Waveform and Screen

Automatic measurements  30, of which up to four can be displayed on-screen at any one time. Measurements include: Period, Frequency, Delay, Rise Time,

(time domain)           Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive

                        Overshoot, Negative Overshoot, Total Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle

                        RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area and Cycle Area.

Automatic Measurements  3, of which one can be displayed on-screen at any one time. Measurements include Channel Power, Adjacent Channel Power

(frequency domain)      Ratio (ACPR), and Occupied Bandwidth (OBW)

Measurement statistics  Mean, Min, Max, Standard Deviation.

Reference levels        User-definable reference levels for automatic measurements can be specified in either percent or units.

Gating                  Isolate the specific occurrence within an acquisition to take measurements on, using either the screen or waveform cursors.

Waveform histogram      A waveform histogram provides an array of data values representing the total number of hits inside of a user-defined region of the

                        display. A waveform histogram is both a visual graph of the hit distribution as well as a numeric array of values that can be

                        measured.

     Sources            Channel 1, Channel 2, Channel 3, Channel 4, Ref 1, Ref 2, Ref 3, Ref 4, Math

     Types              Vertical, Horizontal

Waveform histogram      12, of which up to four can be displayed on-screen at any one time. Waveform Count, Hits in Box, Peak Hits, Median, Max, Min,

measurements            Peak-to-Peak, Mean, Standard Deviation, Sigma 1, Sigma 2, Sigma 3

Waveform math

Arithmetic              Add, subtract, multiply, and divide waveforms.

Math functions          Integrate, differentiate, FFT

FFT                     Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or

                        Blackman-Harris.

                                                                                                                                 www.tektronix.com          21
Datasheet

Waveform math

    Spectrum math                 Add or subtract frequency-domain traces.

    Advanced Math                 Define extensive algebraic expressions including waveforms, reference waveforms, math functions (FFT, Intg, Diff, Log, Exp, Sqrt,

                                  Abs, Sine, Cosine, Tangent, Rad, Deg), scalars, up to two user-adjustable variables and results of parametric measurements

                                  (Period, Freq, Delay, Rise, Fall, PosWidth, NegWidth, BurstWidth, Phase, PosDutyCycle, NegDutyCycle, PosOverShoot,

                                  NegOverShoot, TotalOverShoot, PeakPeak, Amplitude, RMS, CycleRMS, High, Low, Max, Min, Mean, CycleMean, Area,

                                  CycleArea, and trend plots). for example, (Intg(Ch1 - Mean(Ch1)) × 1.414 × VAR1)

Act on Event

    Events                        None, when a trigger occurs, or when a defined number of acquisitions complete (1 to 1,000,000)

    Actions                       Stop acquisition, save waveform to file, save screen image, print, AUX OUT Pulse, remote interface SRQ, e-mail notification, and

                                  visual notification

    Repeat                        Repeat the act on event process (1 to 1,000,000 and infinity)

Video Picture mode

    Sources                       Channel 1, Channel 2, Channel 3, Channel 4

    Video standards               NTSC, PAL

    Contrast and brightness       Manual and automatic

    Field selection               Odd, Even, Interlaced

    Picture location on screen    Selectable X and Y location, width and height adjustment, start line and pixel and line-to-line offset control.

Power measurements (optional)

    Power quality measurements    VRMS, VCrest Factor, Frequency, IRMS, ICrest Factor, True Power, Apparent Power, Reactive Power, Power Factor, Phase Angle.

    Switching loss measurements

    Power loss                    Ton, Toff, Conduction, Total.

    Energy loss                   Ton, Toff, Conduction, Total.

    Harmonics                     THD-F, THD-R, RMS measurements. Graphical and table displays of harmonics. Test to IEC61000-3-2 Class A and MIL-

                                  STD-1399, Section 300A.

    Ripple measurements           VRipple and IRipple.

    Modulation analysis           Graphical display of +Pulse Width, -Pulse Width, Period, Frequency, +Duty Cycle, and -Duty Cycle modulation types.

    Safe operating area           Graphical display and mask testing of switching device safe operating area measurements.

    dV/dt and dI/dt measurements  Cursor measurements of slew rate

22  www.tektronix.com
                                                                                                    Mixed Domain Osciloscopes -- MDO3000 Series

Limit/Mask testing (optional)

Test source                        Limit test: Any Ch1 - Ch4 or any R1 - R4

                                   Mask test: Any Ch1 - Ch4

Mask creation                      Limit test vertical tolerance from 0 to 1 division in 1 m division increments; Limit test horizontal tolerance from 0 to 500 m division in

                                   1 m division increments.

                                   Load custom mask from text file with up to 8 segments.

Mask scaling                       Lock to Source ON (mask automatically re-scales with source-channel settings changes)

                                   Lock to Source OFF (mask does not re-scale with source-channel settings changes)

Test criteria run until            Minimum number of waveforms (from 1 to 1,000,000 and Infinity)

                                   Minimum elapsed time (from 1 second to 48 hours and Infinity)

Violation threshold                From 1 to 1,000,000 and Infinity

Actions on test failure            Stop acquisition, save screen image to file, save waveform to file, print screen image, AUX OUT pulse, set remote interface SRQ

Actions on test complete           AUX OUT pulse, set remote interface SRQ

Results display                    Test status, total waveforms, number of violations, total tests, failed tests, elapsed time, total hits for each mask segment

Arbitrary Function Generator

(Requires MDO3AFG option)

Waveforms                          Sine, Square, Pulse, Ramp/Triangle, DC, Noise, Sin(x)/x (Sinc),  Gaussian,  Lorentz,  Exponential  Rise,  Exponential          Decay,

                                   Haversine, Cardiac, and Arbitrary.

Sine

      Frequency range              0.1 Hz to 50 MHz

      Amplitude range              20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

      Amplitude flatness           ±0.5 dB typical at 1 kHz (±1.5 dB for <20 mVp-p amplitudes)

      Total harmonic distortion    1% into 50Ω

      (typical)                    2% for amplitude < 50 mV and frequencies > 10 MHz

                                   3% for amplitude < 20 mV and frequencies > 10 MHz

      Spurious free dynamic range  -40 dBc (Vp-p ≥ 0.1 V); -30dBc (Vp-p ≤ 0.1 V), 50 Ω load

      (SFDR)

Square / Pulse

      Frequency range              0.1 Hz to 25 MHz

      Amplitude range              20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

      Duty cycle                   10% to 90% or 10 ns minimum pulse, whichever is larger cycle

      Duty cycle resolution        0.1%

      Pulse width minimum          10 ns typical

      Rise/fall time               5 ns typical (10% - 90%)

      Pulse width resolution       100 ps

      Overshoot                    < 2% typical for signal steps greater than 100 mV

      Asymmetry                    ±1% ±5 ns, at 50% duty cycle

      Jitter (TIE RMS)             < 500 ps typical

                                                                                                                                             www.tektronix.com                 23
Datasheet

Arbitrary Function Generator

    Ramp / Triangle

        Frequency range        0.1 Hz to 500 kHz

        Amplitude range        20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

        Variable symmetry      0% to 100%

        Symmetry resolution    0.1%

    DC

        Level range            ±2.5 V into Hi-Z; ±1.25 V into 50 Ω

    Noise

        Amplitude range        20 mVp-p to 5 Vp-p in to Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

        Amplitude resolution   0% to 100% in 1% increments

    Sin(x)/x (Sinc)

        Frequency range        0.1 Hz to 2 MHz

        Amplitude range        20 mVp-p to 3.0 Vp-p into Hi-Z; 10 mVp-p to 1.5 Vp-p into 50 Ω

    Gaussian

        Frequency range        0.1 Hz to 5 MHz

        Amplitude range        20 mVp-p to 2.5 Vp-p into Hi-Z; 10 mVp-p to 1.25 Vp-p into 50 Ω

    Lorentz

        Frequency range        0.1 Hz to 5 MHz

        Amplitude range        20 mVp-p to 2.4 Vp-p into Hi-Z; 10 mVp-p to 1.2 Vp-p into 50 Ω

    Exponential Rise / Decay

        Frequency range        0.1 Hz to 5 MHz

        Amplitude range        20 mVp-p to 2.5 Vp-p into Hi-Z; 10 mVp-p to 1.25 Vp-p into 50 Ω

    Haversine

        Frequency range        0.1 Hz to 5 MHz

        Amplitude range        20 mVp-p to 2.5 Vp-p into Hi-Z; 10 mVp-p to 1.25 Vp-p into 50 Ω

    Cardiac

        Frequency range        0.1 Hz to 500 kHz

        Amplitude range        20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

    Arbitrary

        Memory depth           1 to 128 k

        Amplitude range        20 mVp-p to 5 Vp-p into Hi-Z; 10 mVp-p to 2.5 Vp-p into 50 Ω

        Repetition rate        0.1 Hz to 25 MHz

        Sample rate            250 MS/s

    Frequency accuracy

        Sine wave and ramp     130 ppm (frequency < 10 kHz)

                               50 ppm (frequency ≥ 10 kHz)

        Square wave and pulse  130 ppm (frequency < 10 kHz)

                               50 ppm (frequency ≥ 10 kHz)

        Resolution             0.1 Hz or 4 digits; whichever is larger

24  www.tektronix.com
                                                                               Mixed Domain Osciloscopes -- MDO3000 Series

Arbitrary Function Generator

Amplitude accuracy                 ±[ (1.5% of peak-to-peak amplitude setting) + (1.5% of DC offset setting) + 1 mV ] (frequency = 1 kHz)

DC offset

DC offset range                    ±[2.5 V – (signal amplitude) / 2 ] into Hi-Z; ±[1.25 – (signal amplitude) / 2 ] into 50 Ω

DC offset resolution               1 mV into Hi-Z; 500 uV into 50 Ω

Offset accuracy                    ±[(1.5% of absolute offset voltage setting) + 1 mV]; derated 3 mV for every 10 °C away from 25 °C

ArbExpress®                        The MDO3000 is compatible with ArbExpress® PC-based signal generator waveform creation and editing software. Capture

                                   waveforms on the MDO3000 oscilloscope and transfer them to ArbExpress for editing. Create complex waveforms in ArbExpress

                                   and transfer them to the arbitrary function generator in the MDO3000 for output. To download ArbExpress software, go to

                                   www.tektronix.com/downloads.

Digital voltmeter (DVM)       and  frequency counter

Source                             Channel 1, Channel 2, Channel 3, Channel 4

Measurement types                  AC RMS, DC, AC+DC RMS (reads out in volts or amps); Frequency

Resolution                         ACV, DCV: 4 digits

                                   Frequency: 5 digits

Frequency accuracy                 10 ppm

Measuring rate                     100 times/second; measurements updated on the display 4 times/second

Vertical settings autorange        Automatic adjustment of vertical settings to maximize measurement dynamic range; available         for  any  non-trigger  source

Graphical measurement              Graphical indication of minimum, maximum, current value, and five second rolling range

Software

OpenChoice® Desktop                Enables fast and easy communication between a Windows PC and your oscilloscope using USB or LAN. Transfer and save

                                   settings, waveforms, measurements, and screen images. Word and Excel toolbars automate the transfer of acquisition data and

                                   screen images from the oscilloscope into Word and Excel for quick reporting or further analysis.

IVI driver                         Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI,

                                   MicrosoftNET, and MATLAB.

e*Scope® Web-based interface       Enables control of the oscilloscope over a network connection through a standard web browser. Simply enter the IP address or

                                   network name of the oscilloscope and a web page will be served to the browser. Transfer and save settings, waveforms,

                                   measurements, and screen images or make live control changes to settings on the oscilloscope directly from the web browser.

LXI Core 2011 Web interface        Connect to the oscilloscope through a standard Web browser by simply entering the oscilloscope IP address or network name in

                                   the address bar of the browser. The Web interface enables viewing of instrument status and configuration, status and modification

                                   of network settings, and instrument control through e*Scope Web-based remote control. All Web interaction conforms to LXI Core

                                   2011 specification, version 1.4.

                                                                                                                                                www.tektronix.com     25
Datasheet

Display system

    Display type                     9 in. (229 mm) color display

    Display resolution               800 horizontal × 480 vertical pixels (WVGA)

    Interpolation                    Sin(x)/x

    Waveform styles                  Vectors, Dots, Variable Persistence, Infinite Persistence

    FastAcq. palettes                Temperature, Spectral, Normal, Inverted

    Graticules                       Full, Grid, Solid, Cross Hair, Frame, IRE and mV

    Format                           YT, XY, and simultaneous XY/YT

    Maximum waveform capture rate    >280,000 wfms/s in FastAcq acquisition mode for 1 GHz models

                                     >235,000 wfms/s in FastAcq acquisition mode for 100 MHz – 500 MHz models

                                     >50,000 wfms/s in DPO acquisition mode on all models

Input/output ports

    USB 2.0 high-speed host port     Supports USB mass storage devices, printers and keyboard. One port on front and one port on rear of instrument.

    USB 2.0 device port              Rear-panel connector allows for communication/control of oscilloscope through USBTMC or GPIB (with a TEK-USB-488), and

                                     direct printing to PictBridge-compatible printers.

    Printing                         Print to network printer, PictBridge printer, or to a printer that supports e-mail printing. Note: This product includes software

                                     developed by the OpenSSL Project for use in the OpenSSL Toolkit. (http://www.openssl.org/)

    LAN port                         RJ-45 connector, supports 10/100 Mb/s

    Video out port                   DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector. XGA resolution.

    Auxilliary input                 (Available on two-channel models only)

    Front-panel BNC connector        Input impedance, 1 MΩ

    Maximum input                    300 VRMS CAT II with peaks ≤ ±425 V

    Probe compenstor output voltage  Front-panel pins

    and frequency

    Amplitude                        0 to 2.5 V

    Frequency                        1 kHz

    Auxiliary out                    Rear-panel BNC connector

                                     VOUT (Hi): ≥2.25 V open circuit, ≥0.9 V 50 Ω to ground

                                     VOUT (Lo): ≤0.7 V into a load of ≤4 mA; ≤0.25 V 50 Ω to ground

                                     Output can be configured to provide a pulse out signal when the oscilloscope triggers, a trigger signal from the internal arbitrary

                                     function generator, or an event out for limit/mask testing.

    Kensington-style lock            Rear-panel security slot connects to standard Kensington-style lock.

    VESA mount                       Standard (MIS-D 75) 75 mm VESA mounting points on rear of instrument.

26  www.tektronix.com
                                                                                           Mixed Domain Osciloscopes -- MDO3000           Series

LAN eXtensions for Instrumentation (LXI)

Class                          LXI Core 2011

Version                        V1.4

Power source

Power source voltage           100 to 240 V ±10%

Power source frequency         50 to 60 Hz at 100 to 240 V

                               400 Hz ±10% at 115 V

Power consumption              120 W maximum

Physical characteristics

Dimensions

Height                         203.2 mm (8 in.)

Width                          416.6 mm (16.4 in.)

Depth                          147.4 mm (5.8 in.)

Weight

Net                            4.2 kg (9.2 lb.)

Shipping                       8.6 kg (19 lb.)

Rackmount configuration        5U

Cooling clearance              2 in. (51 mm) required on left side and rear of instrument

EMC, environment, and safety

Temperature

Operating                      -10 ºC to +55 ºC (+14 ºF to 131 ºF)

Nonoperating                   -40 ºC to +71 ºC (-40 ºF to 160 ºF)

Humidity

Operating                      Up to +40 ºC, 5% to 90% relative humidity

                               +40 ºC to +55 ºC, 5% to 60% relative humidity

Nonoperating                   Up to +40 ºC, 5% to 90% relative humidity

                               Above +40 ºC up to +55 ºC, 5% to 60% relative humidity

                               Above +55 ºC up to +71 ºC, 5% to 40% relative humidity, non-condensing

Altitude

Operating                      3,000 meters (9,843 feet)

Nonoperating                   12,000 meters (39,370 feet)

Regulatory

Electromagnetic compatibility  EC Council Directive 2004/108/EC

Safety                         UL61010-1:2004, CAN/CSA-C22.2 No. 61010.1: 2004, Low Voltage Directive 2006/95/EC and EN61010-1:2001, IEC

                               61010-1:2001, ANSI 61010-1-2004, ISA 82.02.01

                                                                                                       www.tektronix.com                  27
Datasheet

Ordering information

Step 1: Choose the MDO3000 base model

MDO3000          family

    MDO3012                        Mixed Domain Oscilloscope with (2) 100 MHz analog channels, and (1) 100 MHz spectrum analyzer input

    MDO3014                        Mixed Domain Oscilloscope with (4) 100 MHz analog channels, and (1) 100 MHz spectrum analyzer input

    MDO3022                        Mixed Domain Oscilloscope with (2) 200 MHz analog channels, and (1) 200 MHz spectrum analyzer input

    MDO3024                        Mixed Domain Oscilloscope with (4) 200 MHz analog channels, and (1) 200 MHz spectrum analyzer input

    MDO3032                        Mixed Domain Oscilloscope with (2) 350 MHz analog channels, and (1) 350 MHz spectrum analyzer input

    MDO3034                        Mixed Domain Oscilloscope with (4) 350 MHz analog channels, and (1) 350 MHz spectrum analyzer input

    MDO3052                        Mixed Domain Oscilloscope with (2) 500 MHz analog channels, and (1) 500 MHz spectrum analyzer input

    MDO3054                        Mixed Domain Oscilloscope with (4) 500 MHz analog channels, and (1) 500 MHz spectrum analyzer input

    MDO3102                        Mixed Domain Oscilloscope with (2) 1 GHz analog channels, and (1) 1 GHz spectrum analyzer input

    MDO3104                        Mixed Domain Oscilloscope with (4) 1 GHz analog channels, and (1) 1 GHz spectrum analyzer input

Standard accessories

Probes

    100 MHz, 200 MHz models        TPP0250, 250 MHz bandwidth, 10X, 3.9 pF. One passive voltage probe per analog channel

    350 MHz, 500 MHz models        TPP0500B, 500 MHz bandwidth, 10X, 3.9 pF. One passive voltage probe per analog channel

    1 GHz models                   TPP1000, 1 GHz bandwidth, 10X, 3.9 pF. One passive voltage probe per analog channel

    Any model with MDO3MSO option  One P6316 16-channel logic probe and accessories

Accessories

    103-0473-00                    N-to-BNC adapter

    063-4526-xx                    Documentation CD

    071-3249-00                    Installation and Safety Instructions, printed manual (translated in English, Japanese, and Simplified Chinese)

    016-2008-xx                    Accessory bag

    -                              Power cord

    -                              OpenChoice® Desktop Software (available on the Documentation CD and for download from www.tektronix.com/downloads.)

    -                              Calibration certificate documenting traceability to National Metrology Institute(s) and ISO9001 quality system registration

28  www.tektronix.com
                                                                                                   Mixed Domain Osciloscopes -- MDO3000 Series

Warranty

Three-year warranty covering all parts and labor, excluding probes.

Step 2: Configure your MDO3000 by adding instrument options

Instrument options

All MDO3000 Series instruments can be preconfigured from the factory with the following options:

MDO3AFG                         Arbitrary function generator with 13 predefined waveforms and arbitrary waveform generation

MDO3MSO                         16 digital channels; includes P6316 digital probe and accessories

MDO3SA                          Increase spectrum analyzer input frequency range to 9 kHz – 3 GHz and capture bandwidth to 3 GHz.

MDO3SEC                         Enhanced instrument security to enable password protected control of turning on/off all instrument ports and instrument firmware

                                update functionality.

Power cord  and  plug  options

Opt. A0                         North America power plug (115 V, 60 Hz)

Opt. A1                         Universal Euro power plug (220 V, 50 Hz)

Opt. A2                         United Kingdom power plug (240 V, 50 Hz)

Opt. A3                         Australia power plug (240 V, 50 Hz)

Opt. A5                         Switzerland power plug (220 V, 50 Hz)

Opt. A6                         Japan power plug (100 V, 110/120 V, 60 Hz)

Opt. A10                        China power plug (50 Hz)

Opt. A11                        India power plug (50 Hz)

Opt. A12                        Brazil power plug (60 Hz)

Opt. A99                        No power cord

Language options

All products are shipped with an Installation and Safety manual that is in English, Japanese, and Simplified Chinese. Full user manuals translated in each language listed

below are included with each product in pdf format on the Documentation CD.

Opt. L0                         English front panel label

Opt. L1                         French front panel overlay

Opt. L2                         Italian front panel overlay

Opt. L3                         German front panel overlay

Opt. L4                         Spanish front panel overlay

Opt. L5                         Japanese front panel overlay

Opt. L6                         Portuguese front panel overlay

Opt. L7                         Simplified Chinese front panel overlay

Opt. L8                         Traditional Chinese front panel overlay

Opt. L9                         Korean front panel overlay

                                                                                                                                   www.tektronix.com                        29
Datasheet

    Opt. L10             Russian front panel overlay

    Opt. L99             No manual, English front panel label

Service options

    Opt. C3              Calibration Service 3 Years

    Opt. C5              Calibration Service 5 Years

    Opt. D1              Calibration Data Report

    Opt. D3              Calibration Data Report 3 Years (with Opt. C3)

    Opt. D5              Calibration Data Report 5 Years (with Opt. C5)

    Opt. G3              Complete Care 3 Years (includes loaner, scheduled calibration, and more)

    Opt. G5              Complete Care 5 Years (includes loaner, scheduled calibration, and more)

    Opt. R5              Repair Service 5 Years (including warranty)

Probes and accessories are not covered by the oscilloscope warranty and service offerings. Refer to the datasheet of each probe and accessory model for its unique warranty

and calibration terms.

Step 3: Select application models and accessories

    Application modules  Application modules are purchased as stand-alone products and can be purchased at the time of initial MDO3000 purchase or at

                         any future time.

                         Application modules have licenses which can be transferred between an application module and an oscilloscope. The license may

                         be contained in the module; allowing the module to be moved from one instrument to another. Or, the license can be contained in

                         the oscilloscope; allowing the module to be removed and stored for safekeeping. The license can be transferred back to the

                         module for use in another MDO3000 oscilloscope. Transferring the license to an oscilloscope and removing the module permits

                         the use of more than two applications simultaneously.

    MDO3AERO             Aerospace Serial Triggering and Analysis Module. Enables triggering on packet-level information on MIL-STD-1553 buses as well

                         as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with

                         time- stamp information.

                         Signal Inputs - Any Ch1 - Ch4, Math, Ref1 - Ref4

                         Recommended Probing - Differential or single ended (only one single-ended signal required)

    MDO3AUDIO            Audio Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2S, LJ, RJ, and TDM audio buses

                         as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables

                         with time-stamp information.

                         Signal Inputs - Any Ch1 - Ch4, any D0 - D15

                         Recommended Probing - Single ended

    MDO3AUTO             Automotive Serial Triggering and Analysis Module. Enables triggering on packet-level information on CAN and LIN buses as well

                         as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with

                         time- stamp information.

                         Signal Inputs – CAN or LIN: Any Ch1 - Ch4, any D0 - D15

                         Recommended Probing - CAN: Single ended or differential; LIN: Single ended

    MDO3COMP             Computer Serial Triggering and Analysis Module. Enables triggering on packet-level information on RS-232/422/485/UART buses

                         as well as analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables

                         with time-stamp information.

                         Signal Inputs - Any Ch1 - Ch4, any D0 - D15

                         Recommended Probing - RS-232/UART: Single ended; RS-422/485: Differential

30  www.tektronix.com
                                                                                                              Mixed Domain Osciloscopes -- MDO3000 Series

MDO3EMBD                                            Embedded Serial Triggering and Analysis Module. Enables triggering on packet-level information on I2C and SPI buses as well as

                                                    analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet decode tables with time-

                                                    stamp information.

                                                    Signal Inputs - I2C or SPI: Any Ch1 - Ch4, any D0 - D15

                                                    Recommended Probing - Single ended

MDO3FLEX                                            FlexRay Serial Triggering and Analysis Module. Enables triggering on packet-level information on FlexRay buses as well as

                                                    analytical tools such as digital views of the signal, bus views, packet decoding, search tools, packet decode tables with time-stamp

                                                    information.

                                                    Signal Inputs - Any Ch1 - Ch4 (and any D0 - D15 when MDO3MSO option is installed; single-ended probing only)

                                                    Recommended Probing - Single ended or differential

MDO3USB                                             USB Serial Triggering and Analysis Module. Enables triggering on packet-level content for low-speed, and full-speed USB serial

                                                    buses. Also enables analytical tools such as digital views of the signal, bus views, packet decoding, search tools, and packet

                                                    decode tables with time-stamp information for low-speed, full-speed, and high-speed USB serial buses.

                                                    Signal Inputs - Low-speed and Full-speed: Any Ch1 - Ch4, any D0 - D15; Low-speed, Full-speed, and High-speed: Any Ch1 - Ch4,

                                                    Math, Ref1 - Ref4

                                                    Note: High-speed decode support only available on 1 GHz models.

                                                    Recommended Probing - Low-speed and Full-speed: Single ended or differential; High-speed: Differential

MDO3PWR                                             Power Analysis Application Module. Enables quick and accurate analysis of power quality, switching loss, harmonics, safe

                                                    operating area (SOA), modulation, ripple, and slew rate (dI/dt, dV/dt).

MDO3LMT                                             Limit and Mask Testing Application Module. Enables testing against limit templates generated from "golden" waveforms and mask

                                                    testing using custom masks.

Recommended accessories

Probes

Tektronix offers  over  100  different  probes  to  meet your application needs. For a comprehensive listing of available probes, please  visit  www.tektronix.com/probes.

TPP0250                                             250 MHz, 10X TekVPI® passive voltage probe with 3.9 pF input capacitance

TPP0500B                                            500 MHz, 10X TekVPI® passive voltage probe with 3.9 pF input capacitance

TPP0502                                             500 MHz, 2X TekVPI® passive voltage probe with 12.7 pF input capacitance

TPP0850                                             2.5 kV, 800 MHz, 50X TekVPI® passive high-voltage probe

TPP1000                                             1 GHz, 10X TekVPI® passive voltage probe with 3.9 pF input capacitance

TAP1500                                             1.5 GHz TekVPI® active single-ended voltage probe

TAP2500                                             2.5 GHz TekVPI® active single-ended voltage probe

TAP3500                                             3.5 GHz TekVPI® active single-ended voltage probe

TCP0020                                             50 MHz TekVPI® 20 Ampere AC/DC current probe

TCP0030A                                            120 MHz TekVPI® 30 Ampere AC/DC current probe

TCP0150                                             20 MHz TekVPI® 150 Ampere AC/DC current probe

TDP0500                                             500 MHz TekVPI® differential voltage probe with ±42 V differential input voltage

TDP1000                                             1 GHz TekVPI® differential voltage probe with ±42 V differential input voltage

TDP1500                                             1.5 GHz TekVPI® differential voltage probe with ±8.5 V differential input voltage

TDP3500                                             3.5 GHz TekVPI® differential voltage probe with ±2 V differential input voltage

THDP0200                                            ±1.5 kV, 200 MHz TekVPI® high-voltage differential probe

                                                                                                                                                            www.tektronix.com             31
Datasheet

    THDP0100                             ±6 kV, 100 MHz TekVPI® high-voltage differential probe

    TMDP0200                             ±750 V, 200 MHz TekVPI® high-voltage differential probe

Accessories

    TPA-N-PRE                            Preamplifier, 12 dB nominal Gain, 9 kHz - 6 GHz

    TPA-N-VPI                            N-to-TekVPI adapter

    119-4146-00                          Near field probe set, 100 kHz - 1 GHz

    119-6609-00                          Flexible monopole antenna

    077-0981-xx                          Service manual (English only)

    TPA-BNC                              TekVPI® to TekProbe™ BNC adapter

    TEK-DPG                              TekVPI Deskew pulse generator signal source

    067-1686-xx                          Power measurement deskew and calibration fixture

    SignalVu-PC-SVE                      Vector Signal Analysis Software

    TEK-USB-488                          GPIB-to-USB adapter

    ACD3000                              Soft transit case (includes front protective cover)

    HCTEK54                              Hard transit case (requires ACD3000)

    RMD3000                              Rackmount kit

    200-5052-00                          Front protective cover

Other RF probes

Contact Beehive Electronics  to  order:  http://beehive-electronics.com/probes.html

    101A                                 EMC probe set

    150A                                 EMC probe amplifier

    110A                                 Probe cable

    0309-0001                            SMA probe adapter

    0309-0006                            BNC probe adapter

32  www.tektronix.com
                                                                                                       Mixed Domain Osciloscopes -- MDO3000 Series

Step 4: Add instrument upgrades in the future

Instrument upgrades

The MDO3000 Series products offer a number of ways to add functionality after the initial purchase. Listed below are the various product upgrades available and the method of

upgrade used for each product.

Post-purchase instrument options  The following products are sold as stand-alone products and can be purchased at any time to add functionality to any MDO3000

                                  product.

MDO3AFG                           Add arbitrary function generator to any MDO3000 Series product.

                                  One-time, permanent upgrade to any model enabled through single-use application module hardware key. The hardware key is

                                  used to enable the feature and then is not required for future use.

MDO3MSO                           Add 16 digital channels; includes P6316 digital probe and accessories .

                                  One-time, permanent upgrade to any model enabled through single-use application module hardware key. The hardware key is

                                  used to enable the feature and then is not required for future use.

MDO3SA                            Increase spectrum analyzer input frequency range to 9 kHz – 3 GHz and capture bandwidth to 3 GHz.

                                  One-time, permanent upgrade to any model enabled through single-use application module hardware key. The hardware key is

                                  used to enable the feature and then is not required for future use.

MDO3SEC                           Add enhanced instrument security to enable password protected control of turning on/off all instrument ports and instrument

                                  firmware update functionality.

                                  One-time, permanent upgrade to any model enabled through software option key. Software option key products require that the

                                  instrument model and serial number be provided at the time of purchase. The software option key is specific to the model and

                                  serial number combination.

                                                                                                                                     www.tektronix.com                         33
Datasheet

    Bandwidth upgrade options  Instrument bandwidth can be upgraded on any MDO3000 Series product after initial purchase. Each upgrade product will increase

                               the analog bandwidth and the spectrum analyzer frequency range to the new bandwidth level. Bandwidth upgrades are purchased

                               based on the combination of the current bandwidth and the desired bandwidth. Software option key products require that the

                               instrument model and serial number be provided at the time of purchase. The software option key is specific to the model and

                               serial number combination. Bandwidth upgrades up to 500 MHz can be performed in the field. Bandwidth upgrades to 1 GHz

                               require installation at a Tektronix service center. The table below shows the various bandwidth upgrade products to purchase

                               based on current and desired bandwidth levels.

                               Model to be upgraded  Bandwidth before upgrade  Bandwidth after upgrade  Order protduct

                               MDO3012               100 MHz                   200 MHz                  MDO3BW1T22

                                                     100 MHz                   350 MHz                  MDO3BW1T32

                                                     100 MHz                   500 MHz                  MDO3BW1T52

                                                     100 MHz                   1 GHz                    MDO3BW1T102

                                                     200 MHz                   350 MHz                  MDO3BW2T32

                                                     200 MHz                   500 MHz                  MDO3BW2T52

                                                     200 MHz                   1 GHz                    MDO3BW2T102

                                                     350 MHz                   500 MHz                  MDO3BW3T52

                                                     350 MHz                   1 GHz                    MDO3BW3T102

                                                     500 MHz                   1 GHz                    MDO3BW5T102

                               MDO3014               100 MHz                   200 MHz                  MDO3BW1T24

                                                     100 MHz                   350 MHz                  MDO3BW1T34

                                                     100 MHz                   500 MHz                  MDO3BW1T54

                                                     100 MHz                   1 GHz                    MDO3BW1T104

                                                     200 MHz                   350 MHz                  MDO3BW2T34

                                                     200 MHz                   500 MHz                  MDO3BW2T54

                                                     200 MHz                   1 GHz                    MDO3BW2T104

                                                     350 MHz                   500 MHz                  MDO3BW3T54

                                                     350 MHz                   1 GHz                    MDO3BW3T104

                                                     500 MHz                   1 GHz                    MDO3BW5T104

                               MDO3022               200 MHz                   350 MHz                  MDO3BW2T32

                                                     200 MHz                   500 MHz                  MDO3BW2T52

                                                     200 MHz                   1 GHz                    MDO3BW2T102

                                                     350 MHz                   500 MHz                  MDO3BW3T52

                                                     350 MHz                   1 GHz                    MDO3BW3T102

                                                     500 MHz                   1 GHz                    MDO3BW5T102

                               MDO3024               200 MHz                   350 MHz                  MDO3BW2T34

                                                     200 MHz                   500 MHz                  MDO3BW2T54

                                                     200 MHz                   1 GHz                    MDO3BW2T104

                                                     350 MHz                   500 MHz                  MDO3BW3T54

                                                     350 MHz                   1 GHz                    MDO3BW3T104

                                                     500 MHz                   1 GHz                    MDO3BW5T104

                               MDO3032               350 MHz                   500 MHz                  MDO3BW3T52

                                                     350 MHz                   1 GHz                    MDO3BW3T102

                                                     500 MHz                   1 GHz                    MDO3BW5T102

                               MDO3034               350 MHz                   500 MHz                  MDO3BW3T54

                                                     350 MHz                   1 GHz                    MDO3BW3T104

                                                     500 MHz                   1 GHz                    MDO3BW5T104

34  www.tektronix.com
                                                                                                              Mixed Domain Osciloscopes -- MDO3000     Series

Model to be upgraded                                                                Bandwidth before upgrade  Bandwidth after upgrade  Order protduct

MDO3052                                                                             500 MHz                   1 GHz                    MDO3BW5T102

MDO3054                                                                             500 MHz                   1 GHz                    MDO3BW5T104

Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.

Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.

                                                                                                                                       www.tektronix.com  35
Datasheet

ASEAN / Australasia (65) 6356 3900                                     Austria 00800 2255 4835*                   Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777

Belgium 00800 2255 4835*                                               Brazil +55 (11) 3759 7627                  Canada 1 800 833 9200

Central East Europe and the Baltics +41 52 675 3777                    Central Europe & Greece +41 52 675 3777    Denmark +45 80 88 1401

Finland +41 52 675 3777                                                France 00800 2255 4835*                    Germany 00800 2255 4835*

Hong Kong 400 820 5835                                                 India 000 800 650 1835                     Italy 00800 2255 4835*

Japan 81 (3) 6714 3010                                                 Luxembourg +41 52 675 3777                 Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90

Middle East, Asia, and North Africa +41 52 675 3777                    The Netherlands 00800 2255 4835*           Norway 800 16098

People's Republic of China 400 820 5835                                Poland +41 52 675 3777                     Portugal 80 08 12370

Republic of Korea 001 800 8255 2835                                    Russia & CIS +7 (495) 6647564              South Africa +41 52 675 3777

Spain 00800 2255 4835*                                                 Sweden 00800 2255 4835*                    Switzerland 00800 2255 4835*

Taiwan 886 (2) 2722 9622                                               United Kingdom & Ireland 00800 2255 4835*  USA 1 800 833 9200

* European toll-free number. If not accessible, call: +41 52 675 3777                                                                                                                                                                     Updated 10 April 2013

For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com.

Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and

price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.

04 Feb 2014                                                                                                                                                                                                                               48W-30020-1

www.tektronix.com
Mouser Electronics

Authorized Distributor

Click to View Pricing, Inventory, Delivery & Lifecycle Information:

Tektronix:

MDO3024     MDO3AERO    MDO3052    MDO3MSO  MDO3FLEX  MDO3LMT   MDO3054  MDO3AFG  MDO3102

MDO3022  MDO3AUTO       MDO3AUDIO  MDO3032  MDO3104   MDO3COMP  MDO3014  MDO3PWR  MDO3SA

MDO3USB  MDO3EMBD       MDO3034  MDO3012
小广播

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet 电子工程

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2020 EEWORLD.com.cn, Inc. All rights reserved