   LABCAL PRO HIGH PRECISION THERMOMETER/
LOGGER FOR THERMOCOUPLE & Pt100 SENSORS

G Reference standard thermometer for Pt100 and thermocouple sensors
G High accuracy 0.01C Pt100, 0.18C thermocouple
G 25 milliKelvin (0.025C) system accuracy with Pt100
G Precise linearisation conformity is realised by a hybrid maths polynomial technique
G Simple, fast digital matching to calibrated sensors, up to 10 points per channel retained in memory
G Additional digital matching to Pt100 sensors using custom constants
G Clear, comprehensive user information on alpha numeric display
G User selectable resolution, 0.001/0.01C for Pt100 inputs
G Smart sensor capability Pt100 for

   corrections stored on probe
G Data logging
G Wide temperature ranges
G Measures Pt100 and thermocouples J, K,

   T, E, R, S, B and N
G Readout directly in C, F, K,  or mV
G Exceptional automatic CJC performance,

   100:1 rejection
G Input circuitry configured to eliminate

   thermal voltages
G External and manual CJ function
G Four sensor input ports
G Channel A, B or A-B reading including

   dissimilar sensor types
G USB serial communications
G USB flash drive port
G Excellent long term stability
G Supplied with PC software
G Programmable analogue (retransmission)

   output
G Front panel data entry and function keys
G Our own design and manufacture
G Rechargeable Lithium Ion battery/mains powered
G CE & RoHS compliant

The Labcal PRO is a highly accurate, portable digital thermometer / logger for metrology and other
exacting laboratory and industrial applications. The Labcal PRO is particularly suitable as a reference
measurement standard for thermal calibration baths in pharmaceutical, medical, food, environmental
testing, R&D and general industrial applications.

                           www.labfacility.co.uk
The Labcal PRO is a third generation instrument developed over a 25 year period from the very
successful Labcal Plus thermometer. Like its predecessors, the Labcal PRO provides laboratory
standard temperature measurement and many user features but provides higher accuracy, higher
resolution and greater versatility to extend the scope of applications.

The Labcal PRO is fully characterised for Pt100 sensors and all major thermocouples J,K,N,T,E,R,S
and B. Exceptionally stable automatic cold junction compensation with a rejection of 100:1 is
incorporated for thermocouple ranges. Alternatively, external or manual referencing can be selected
as required.

Displayed values and user information are indicated on a bright clear OLED screen with diffused
backlighting. Data can be displayed in C,F,K, or mV as required; nine front panel push keys, the
only user controls, are used in conjunction with the display screen. It is this arrangement which
makes for very simple and "friendly" operation.

Resolution is a user selectable 0.001 or 0.01C for Pt100 sensors and 0.1C for thermocouples. All
computations are performed digitally without drift. Overall stability is optimised by utilising only the
highest quality components including high precision, expensive metal film resistor networks.

There are four input ports, two for thermocouples and two for 3 or 4 wire Pt100 sensors; the
instrument automatically recognises 3 or 4 wire configurations.

The Labcal PRO can accept two Pt100 sensors and/or two thermocouples of similar or different
types (J,K,N,T,E,R,S and B). The measured temperature can be displayed directly from one of the
inputs or differentially between similar or different inputs. Differential temperature and the two
individual channel temperatures are displayed simultaneously.

Individual calibrated sensors can have their appropriate calibration values programmed into the
Labcal PRO using either the keypad or the PC software supplied. The PC software also allows
corrections in the form of constants for individual Pt100 sensors to be programmed in. The non-
volatile memory ensures that the values are retained (until such time as they are changed) even after
switch-off. Where the calibration certificate relating to a particular probe states specific  or mV
values at stated temperatures, up to ten such values are entered into the Labcal PRO with their
relevant temperature points using the front panel keypad or the software. The instrument then
digitally self-calibrates to the associated probe over the range embraced in the calibrated values
used; the temperature readout is "corrected" accordingly. Additional manual procedures are not
required to compute precise temperature readings since these are displayed directly; miscalculation
errors are thus eliminated. Very high system accuracies, better than 25 milliKelvins can be achieved
using probe matching.

Corrections which are "programmed in" can be displayed for confirmation. However, initial input,
adjustment, enabling and disabling of corrections are protected by a special key sequence to prevent
accidental or unauthorised changes. Values can be changed by an authorised user at any time.
For Pt100 inputs, SMART probes can be used which incorporate dedicated `correction values` within
the connector of each probe.When a SMART probe is connected to input A or B or both inputs,the
corrections are automatically applied without the need for user intervention.

A common application for the Labcal PRO is the comparison calibration of "working" sensors
against a calibrated reference standard or semi-standard sensor. For example, the sensor under test
in channel B is compared against the reference sensor (with programmed corrections) in channel A.

When applied, the null function corrects the differential temperature readout between two sensors
to zero. For example, the apparent temperature difference between two sensors known to be at the
same actual temperature can be corrected to zero prior to, for example, heat exchanger
experiments.

All four inputs can be scanned and values logged according to parameters set up by the user in the
PC software;the instrument incorporates a real-time clock and on-board memory.

The adjustable contrast OLED display screen provides data readout, user prompts and mode
annunciation. This very important feature ensures a high degree of user friendliness and confidence.
User prompts indicate which buttons to press in the appropriate sequence when selecting
parameters and functions and when setting calibration values. Mode annunciation indicates which
mode of operation is currently selected.

A flash drive USB port is provided to allow data to be stored and/or exported. Firmware update are
also facilitated via this port.

PC software running in WINDOWS is provided as standard; it allows programming of custom
calibration, remote control & measure and logging functions.

SPECIFICATIONS

All values are valid for a nominal 240V 50Hz instrument supply and 20C ambient temperature
( 2C)

General                                                               Thermocouples
                                                                      0.05C base metal (-200C to span)
      Inputs/Ranges/Sensors                                           0.1C noble metal above 200C

Type  Pt100 to IEC 751 (ITS 90 refers).                  Note: Performance of ADC (0.5V resolution ) is superior to
                                                         published thermocouple table best resolution of 1V.
      -200C to +850C
                                                         Stability (vs ambient temperature)
      Ro = 100                                                        Pt100
                                                                      Better than 0.0025C per 1C ambient change
      3 or 4 wire connection with automatic recognition
                                                                      Thermocouples
      (with manual override)                                          Zero: negligible drift excluding CJC effect
                                                                      Scale: < 0.1C in 15C
      Thermocouples to IEC 584
      No isolation exists between inputs.                Warm-up Negligible under normal ambient conditions. Allow 5-10
                                                                      minutes for full stability unless stored at low temperature,
Type  Type S -50 to 1768C                                            then 30 minutes minimum.

      Type R -42 to 1766C           Noble Metal         Cold Junction Compensation
                                                                      Automatic on all thermocouple ranges. Better than 1C
      Type B +254 to 1818C                                           max shift for 10C ambient variation (100:1) or External
                                                                      via Pt100 on channel A or B or Manual input of CJ
      Type J -200 to 1200C          Base Metal                       temperature value.
      Type K -195 to 1370C
      Type E -190 to 995C                               Pt100 Sensor Current
      Type N -160 to 1300C                                           0.5mA
      Type T -193 to 399C
                                                         Resolution of data display
Overall Pt100                                                         Pt100
Accuracy 0.01C 0.0005% of span                                   0.01/0.001 user select for C, K and

      Thermocouples                                      Variable filter
                                                                      Sampling rate selectable between 4 and 64
      i) Base Metal                  ii) Noble Metal                  (measurements averaged per reading).

       0.2C from -50C to 700C 0.3C from 850C to               Thermocouples
                                                                      0.1C all ranges
                                     full scale

       0.4C from -200C to -50C 0.4C below 850C

       0.4C from 700C to          0.7C below 0C

      full scale

      Note: Type B not characterized below +200C

Linearisation Conformity
             Pt100
             Better than 0.01C
Measurement Units                                                       USB Serial Communications
             C, F, K,  or mV as applicable to sensor                               Isolated, 9600 Baud, 8 data, no parity, 1 stop bit. Remote
                                                                                     control and measure.
Measurement Modes
             A, B or A-B. Any combination of sensor types.              PC Software (standard)
                                                                                     Supplied as standard on CD ROM. Remote control and
Custom Calibration                                                                   measure: Log readings to file/Download to
             Up to 10 calibration values can be allocated to Pt100                   PC/Programming corrections
             input A & B and to thermocouples input A & B. Values
             are retained in non-volatile memory until replaced by      Analogue Output (standard)
             user. Alternatively, for Pt100 inputs ITS 90 or IPTS 68                 User programmable, 0 to 5V
             coefficients can be used for custom calibration.
                                                                        Application Note:
Smart Sensor Connection                                                              Inputs are not isolated in the Labcal PRO which is
             Pt100 inputs only. Correction values stored in connector.               primarily designed for laboratory applications and site
                                                                                     calibration of industrial temperature sensors. Probes
Null Function                                                                        connected to the instrument must therefore be isolated
             Corrects differential temperature readout between two                   from high voltage pick-up. Particular care must be taken
             sensors to zero.                                                        when high temperature thermocouples are used with
                                                                                     ceramic (refractory oxide) sheaths; these can become
Sensor Lead                                                                          electrically conductive at elevated temperatures and
             Pt100                                                                   mains pick-up can occur from electrical heaters.

Resistance 5 each lead maximum                                          The instrument should be disconnected from the mains adaptor
                                                                        and operated from its internal batteries in such situations.
        Thermocouples
        300 loop total

Logging 8000 readings                                                   ACCESSORIES AND ORDER CODES

Supply  Internal Lithium Ion rechargeable batteries. Mains 90-
        260V 50/60Hz universal adaptor included. Battery charge
        life up to 12 hours with full charge.                           Labcal PRO is supplied complete with switch mode power supply
                                                                        (90-260V, 50/60Hz), two Pt100 "D" connectors, operating manual
Series Mode Rejection                                                   and PC software.
             60dB @ 50Hz (50mV RMS applied)
                                                                        Precision Pt100 Probes

Common Mode Rejection                                                   Stainless steel probes, 6mm diameter with 2m screened PTFE lead
             30V RMS applied between input and earth produces no        and connector
             measurable effect.
                                                                        LP-L250:       250mm long, maximum 250C
Display 128 x 64 pixel OLED with diffused backlighting.                 LP-H450:       350mm long, maximum 450C
             Wide viewing angle, high contrast (adjustable).
                                                                        UKAS Calibration

Front Panel Controls                                                    LP - UKAS:     UKAS calibration for instrument alone.
             9 membrane push-keys to control all instrument             LP - SYS.CAL:  UKAS calibration of instrument and sensor
             functions                                                                 together at five points.
                                                                        LP - COR.CAL:  UKAS calibration of instrument and sensor
Mechanical/Case                                                                        together at five points, after initial calibration of
             Metal bench top case / adjustable tilt                                    sensor only and programming of corrections.

Dimensions                                                              Accessories
             Overall
             190mm(W) x 70mm(H) x 250mm(D)

Weight 3.0kg approx.                                                    LP - TBLK3:    Terminal Block for connection of 3 wire Pt100s
                                                                        LP - TBLK4:    Terminal Block for connection of 4 wire Pt100s

Input Connections
             2 x Pt100 via D type connectors
             2 x Thermocouple via standard sockets.

                                Specifications may be subject to change.

           LABFACILITY LIMITED www.labfacility.co.uk

SOUTHERN UK & EXPORT DIVISION: Units 5,6 & 7, Block K, Southern Cross Industrial Estate, Shripney Road, Bognor Regis,
  West Sussex PO22 9SE Export Sales: tel: +44(0)1243 871287 fax: +44(0)1243 871281 email: exportsales@labfacility.co.uk
              Southern UK Sales: tel: +44(0)1243 871280 fax: +44(0)1243 871281 email: southernsales@labfacility.co.uk

                  NORTHERN UK DIVISION: Eden Place, Unit 3b Outgang Lane, Dinnington, Sheffield S25 3QT
               Northern UK Sales: tel: +44(0)1909 569446 fax: +44(0)1909 550632 email: northernsales@labfacility.co.uk

                                                                                                                                                                                                         Data Ref: 050111DS/A

LABCAL-PRO器件购买:

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2017 EEWORLD.com.cn, Inc. All rights reserved