AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER

文档预览

3174器件文档内容

AC AUTOMATIC INSULATION/

WITHSTANDING HiTESTER 3174

                                                                   Safety Standards
                                                                   Measuring Instruments

Featuring contact check functionality

The effects of test lead wire breaks, erroneous test results caused by faulty contact, and
fluctuations in test voltage caused by variations in the instrument's supply voltage on
withstanding voltage and insulation resistance testing are well known.
The 3174 (3174-01) AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER is a
low-cost solution featuring contact check functionality as well as a stabilized power supply
to prevent the reduced test reliability that can result from these issues.
To streamline production line test processes, the HiTESTER also features configuration of
test parameters via RS-232C (GP-IB) and reading of parameters from the EXT I/O interface.
2
     2

    Improved Test Reliability

     NEW

Contact check function improves test reliability

The 3174's contact check function lets you detect test lead                                                                                        If a test lead wire were
wire breaks and faulty contact during testing by sensing                                                                                       to break during testing with
measurement issues in real time.                                                                                                               a tester that does not offer
                                                                                                                                               contact check functionality,
                                                                                                                                               defective test pieces would be
                                                                                                                                               judged to be non-defective.

                     Withstanding voltage tester                                                                  Withstanding voltage tester
Voltage output
                                                                       The 3174 improves test            Voltage                                                Faulty contact
                                                                       reliability by monitoring         output
                                                                       voltage to detect lose                                                                     *Break
                                                                       leads during testing                                                                       in lead
                                                                       and faulty wiring during
                                                  Voltage monitor      measurement.                                                            Voltage monitor

                                         Voltmeter                 *Since contact check is performed                                           Voltmeter
                                                                   while testing is in progress, use of
                Contact check                                      contact check functionality does               Contact check
                                                                   not increase cycle time.
                Test current                                                                                            Test current

             Test piece                           Faulty contact                                                  Test piece

    If a test lead were to come lose              *Disconnected
during testing with a measurement                      lead
device that does not offer contact check
functionality, defective test pieces                                                                              Fluorescent display tube
would be judged to be non-defective.
                                                                                                                  The display uses a bright,
                                                                                                                  easy-to-read fluorescent tube.

DANGER lamp

Flashes a warning during testing and
whenever high voltage is present at the
terminals.
The DANGER lamp turns off when
the voltage at the output terminals is
no greater than AC 30 V or DC 60 V.

9613            9614                                                                    Test mode selection
SINGLE-HAND
REMOTE CONTROL  TWO-HAND                                                                Select from three test modes:
(option)                                                                                1.Manual test mode: W (withstand voltage testing) / I (insulation resistance testing)
                REMOTE CONTROL                                                          2.Automatic test mode: WI / IW
NEW
                (option)

                External switch

                Enables start/stop control by means of
                the 9613 SINGLE-HAND REMOTE
                CONTROL or 9614 TWO-HAND
                REMOTE CONTROL.
                 (The 9613 and 9614 are optional.)

    Judgment output at forced stop                                                       Ramp timer function

  The ability to obtain a judgment even after a forced stop increases testing freedom.  The ramp-up initial value, ramp-up and ramp-down time
                                                                                        parameters can be set independently.
NEW
                                                                                         True effective value display
    Continued analytical testing after FAIL judgments
                                                                                         Eliminate the effects of supply voltage fluctuations

                                                                                         Delay timer function

Test pieces can now be analyzed by means of detailed monitoring of the
test current accompanying FAIL judgments.
                                                                                                                                                                            3
                                                                                                                                                    3

Safe, Automated Operation

Convenient                                                                    Safe

Continuous full-auto withstanding voltage and insulation resistance testing   Interlock function

The 3174 lets you set independent test conditions including                   Signal input from an external device such as an automatic
test voltage for withstanding voltage and insulation                          tester can be used to disable output and prevent testing,
resistance testing, and you can perform these tests                           ensuring safety during automatic testing and other uses.
continuously. Press the WI key to automatically perform
withstanding voltage followed by insulation resistance tests,                                                          GP-IB (option)
or press the IW key to automatically perform insulation                                                                     Enables remote control.
resistance followed by withstanding voltage tests.
                                                                                                                          Contact check
       RS-232C
        Enables remote control.                                                                                                      LOW: A break in the lead is determined to have
                                                                                                                                     occurred when the check current is not detected.
               Power cord inlet                                                                                                      HIGH: A break in the lead is determined
                                                                                                                                     to have occurred when the contact check
                                                                                                                                     measurement voltage falls outside the range
                                                                                                                                     defined by the upper and lower thresholds.
                                                                                                                                     (Connectors are locked in place as shown
                                                                                                                                     on the left to prevent the leads from
                                                                                                                                     being disconnected.)
                                                                                                                                             *Requires an additional pair of
                                                                                                                                             9615 HIGH-VOLTAGE TEST LEADS.

                                                                                                   Rear panel voltage output terminals

            Status out relay terminals                                                             The rear panel terminals are connected
                                                                                                   at all times to the front terminals. When
            Maximum input voltage: AC 150 V, DC 30 V                                               embedding the 3174 HiTESTER inside
            Maximum contact current: 0.5 A                                                         an automatic tester, connections can be
                                                                                                   made via the rear panel terminals.
            An OR operation applied to various enabled conditions             9616 WARNING LAMP
            determines whether relay terminal output is turned on. Devices    (option)             Buzzer volume knobs
            such as the 9616 WARNING LAMP can be connected.                   (AC100V,0.1A)
                                                                                                   The 3174 provides separate
            1. H.V.ON     Output voltage generation                                                volume adjustments for the PASS
            2. TEST       Testing in progress                                                      and FAIL judgment buzzers.
            3. PASS       PASS state
            4. FAIL       FAIL state                                                (Includes warning lamp unit only.
            5. INT.LOCK   Interlocked state                                         Refer to the included plans and drill
            6. READY      Ready state                                               installation holes in a base such as a
            7. EXT.CONT.  Under external control                                    metal plate as necessary.)
            8. POWER.ON   When the 3174 is on

   External I/O connector (Signal lines have photocoupler isolation.)

Pin    I/O  Signa                       Function                               Interface specifications
                                                                              E X T I / O : Output signals
1      OUT READY Low in ready state
                                                                                                   Open collector output (with photocoupler isolation)
2      OUT  L-FAIL Low in FAIL state (lower bound)                                                 All active low with a maximum load voltage of DC 30 V
                                                                                                   Maximum output current: DC 100 mA per signal
3      OUT  U-FAIL Low in FAIL state (upper bound)                                                 Output saturation voltage: DC 1.5 V or less

4      OUT  PASS Low in PASS state                                                                 Input signals

5      OUT  TEST Low in test sate                                                                  All active low input (with photocoupler isolation)
                                                                                                   Maximum applied voltage: 30 V
6      OUT  H.V.ON Low when voltage is present at output terminals                                 High level voltage: from DC 15 V to 30 V, or open
                                                                                                   Low level voltage: DC 5 V or less (-6 mA typ.)
7      IN   EXT-E When low, external I/O input signals are enabled
                                                                              E X T S W : Input signals (contact input)
8      IN   START When low, same function as START key
                                                                                                   START, STOP, SW.EN (external switch terminal enable)
9      IN   STOP When low, same function as STOP key
                                                                                                   Output signals
10     IN   INT.LOCK Interlock on open
                                                                                                   LED activation signal (maximum load current: 40 mA)
11     OUT W-MODE Low during withstanding voltage testing

12     OUT I-MODE Low during insulation resistance testing

13     OUT  W-FAIL Low in FAIL state during withstand voltage testing

14     OUT  I-FAIL Low in FAIL state during insulation resistance testing

15-18  IN   ISO.COM Ground inputs for external devices

22-25  IN MEM-0 to 3 Saved test selected pins

27     IN   MEM-E When low, enables memory selected pins

28-29 OUT MODE-0,1 Current test mode
33-36 OUT ISO.DCV Internal DC 15 V power supply (100 mA)
4 AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174

Withstanding Voltage Testing                                                                                         Timers

Test Voltage                                                                                                         Setting range       : 0.3 to 999 s

Output voltage     : AC 0.2 to 5 kV (50/60 Hz), single-range output                                                  Operation           : When set to on: Display counts down from the set time after start.
                                                                                                                                            When set to off: Display indicates time elapsed since start.
Voltage output method : PWM switching method (0 V start; voltage can be
                                 changed while generating output)                                                    Setting resolution/accuracy : 0.1 s (0.3 to 99.9 s) 50 ms
                                                                                                                                                      1 s (100 to 999 s) 0.5 s
Voltage specification method : Digital (setting resolution: 0.01 kV)
                                                                                                                     Ramp Timers (Withstand voltage testing)
Output voltage accuracy : 1.5% of setting 20 V
                                                                                                                     Setting range       : 0.1 to 99.9 s: Ramp-up and ramp-down can be set independently.
Maximum rated output : AC 100 VA (5 kV/20 mA) continuous rating                                                      Operation
                                                                                                                                         : Ramp-up: The output voltage increases linearly from the
Transformer capacity : 100VA                                                                                         Setting resolution     initial voltage to the test voltage over the ramp-up time.
                                                                                                                                            Ramp-down: The output voltage decreases from the set
Voltmeter          : True effective value display                                                                                           voltage to 0 V over the ramp-down time after the test time
                                                                                                                                            elapses, and the display counts down from the set time.
                   Digital meter: AC 0 kV to 5.00 kV                                                                                        *The actual ramp-up waveform varies with the load due to the analog response delay.

                   Accuracy: 1.5% rdg (1,000 V or less, 15 V)                                                                          : 0.1s

Output waveform : Sine wave

Voltage change rate : 15% or less (converges to setting within 1 s during                                            Delay Timers (Insulation resistance testing)

                   change from maximum rated load to no load)                                                        Setting range       : 0.1 to 99.9 s

Distortion factor  : 5% or less (tester impedance during measurement with 5                                          Setting resolution : 0.1s

                   kV output under 40 M load)                                                                        Decision Function

Frequency          : 50/60Hz (0.2%)

Current Detection                                                                                                    Decision method : Window comparison method with upper and lower bound settings (digital specification)

Current measurement range : 0.01 mA to 20 mA (2 ranges)                                                              Decision results    : UPPER-FAIL: The measured current (measured insulation
                                                                                                                                            value) exceeded the specified upper threshold.
Measurement range : 10mA/20mA                                                                                                               PASS: The measured current (measured insulation value) fell within

Indicated value    : True effective value display                                                                                        the range defined by the specified upper and lower thresholds.

Measurement ranges and : 0.00 mA to 10.00 mA, 0.01 mA (10 mA range)                                                                      LOWER-FAIL: The measured current (measured

resolutions        0.0 mA to 20.0 mA, 0.1 mA (20 mA range)                                                                               insulation value) was less than the specified lower bound.

Measurement accuracy : 2% rdg 0.05 mA (10 mA range)                                                                                    UPPER LOWER-FAIL: A testing error occurred, for

                                                                                                                                         example due to a failure to generate the set voltage.

                   2% rdg 0.5 mA (20 mA range)                                                                     Decision processing : Display, buzzer, and EXT I/O signal output is generated according to each decision result.

Insulation Resistance Measurement                                                                                    Setting range       : AC withstanding voltage: 0.1 to 20.0 mA (upper threshold), 0.1 to 19.9 mA (lower threshold)
                                                                                                                     Setting resolution     DC insulation: 0.2 to 2,000 M (500 V) or 1.0 to 2,000 M (1,000 V) for both upper and lower thresholds
Test Voltage
                                                                                                                                         : AC withstanding voltage: 0.1 mA
Rated voltage      : DC 500 V/1,000 V (positive polarity)                                                                                   DC insulation: 0.01 M (0.2 to 2.00 M), 0.1 M (2.10 to
                                                                                                                                            20.0 M), 1 M (21.0 to 200 M), 10 M (210 to 2,000 M)

Unloaded voltage : 1 to 1.2 times rated voltage                                                                      Contact Check

Rated measurement current : 1 to 1.2 mA                                                                              Voltmeter accuracy : Detection method: Average value detection/effective value conversion
                                                                                                                                                      Accuracy: Setting 50 V *Inaccuracy may increase when the waveform is distorted.
Short-circuit current : 4 to 5 mA (500 V), 2 to 3 mA (1,000 V)

Measurement range : 0.2 to 2,000 M (500 V), 0.5 to 2,000 M (1,000 V)                                                 Decision results    : Enables the contact check function (does not increase cycle time).
                                                                                                                                            LOW: A break in the lead is determined to have occurred when the check current is not detected.
Guaranteed accuracy : 0.5 to 999 M (500 V), 1 M to 999 M (1,000 V): 4% rdg                                                                 HIGH: Upper and lower thresholds for the check detection voltage can be set. A
                                                                                                                                            break in the lead is determined to have occurred when the contact check measurement
ranges/accuracies  1,000 M to 2,000 M: 8% rdg                                                                                              voltage falls outside the range defined by the upper and lower thresholds.

Measurement resolution : 0.01 M (0.20 M to 19.9 M)

                   0.1 M (20.0 M to 199.9 M), 1 M (200 M to 2,000 M)

Measured resistance ranges : 2 M, 20 M, 200 M, 2,000 M (500 V)                                                       Voltage setting range : Withstanding voltage testing: 0.20 kV to 5.0 kV (0.01 kV resolution; applies to both
                                                                                                                                                      upper and lower thresholds)Insulation resistance measurement: Upper threshold of
                   4 M, 40 M, 400 M, 2,000 M (1,000 V)                                                                                                600 V and lower threshold of 500 V (during 500 V measurement); upper threshold of
                                                                                                                                                      1,200 V and lower threshold of 1,000 V (during 1,000 V measurement) (both fixed)

General Specifications                                                                                                                         Model

Display            : Fluorescent display tube (digital display)                                                      AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174
                                                                                                                     AC AUTOMATIC INSULATION/WITHSTANDING HiTESTER 3174-01 (GP-IB Model)
Monitor functions : Output voltage, detected current, insulation resistance

Monitor period     : 2 times per second, minimum

Operating temperature range : 0C to 40C, 80% RH or less (non-condensing)

Storage temperature range : -10C to 50C, 90% RH or less (non-condensing)                                                                      Options

Temperature and humidity : 23 5C, 80% RH or less (non-condensing)                                                  ELECTRIC SAFETY TESTING SOFTWARE                            9267 (available soon)
range for guaranteed accuracy (With warm-up period of at least 10 min)                                               SINGLE-HAND REMOTE CONTROL                                  9613
                                                                                                                     TWO-HAND REMOTE CONTROL                                     9614
Guaranteed accuracy term : 1 year                                                                                    HIGH-VOLTAGE TEST LEADS                                     9615
                                                                                                                     WARNING LAMP (AC100V,0.1A)                                  9616
Operating environment : Indoors at elevations of up to 2,000 m at a pollution level of 2                             RS-232C CABLE (9-pin to 9-pin/cross/1.8 m)                  9637
                                                                                                                     RS-232C CABLE (9-pin to 25-pin/cross/1.8 m)                 9638
Supply voltage     : AC 100 to 240 V Designed to tolerate voltage fluctuations of 10% of the rated supply voltage.  GP-IB CONNECTING CABLE (2 m)                                9151-02

Power supply frequency : 50Hz/60Hz

Withstanding voltage : Power supply to chassis: 1.39 kV at 10 mA for 15 s

Maximum rated power : 200VA

Dimensions         : Approx. 320 (W) 155 (H) 395 (D) mm (excluding protruding parts)

Weight             : Approx. 15 kg

Applicable standards : EMC: EN61326 Class A, EN61000-3-2, EN61000-3-3, Safety: EN61010-1

Included accessories : HIGH-VOLTAGE TEST LEADS 9615 (1 each high-voltage
                                 and return), power cord, disconnection prevention plate

                                           HIOKI (Shanghai) Sales & Trading Co., Ltd. :                              DISTRIBUTED BY
                                           1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road
HEAD OFFICE :                              Shanghai, P.R.China POSTCODE: 200021
81 Koizumi, Ueda, Nagano, 386-1192, Japan  TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360
TEL +81-268-28-0562 / FAX +81-268-28-0568  E-mail: info-sh@hioki.cn
E-mail: os-com@hioki.co.jp                 Beijing Office :
                                           A-2602 Freetown, 58 Dong San Huan Nan Road
HIOKI USA CORPORATION :                    Beijing, P.R.China POSTCODE: 100022
6 Corporate Drive, Cranbury, NJ 08512 USA  TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090
TEL +1-609-409-9109 / FAX +1-609-409-9108  E-mail: info-bj@hioki.cn
E-mail: hioki@hiokiusa.com                 Guangzhou Office :
                                           Room 303, Profit Plaza, No.76, West Huangpu Road
                                           Guangzhou, P.R.China POSTCODE: 510623
                                           TEL +86-20-38392673/2676 FAX +86-20-38392679
                                           E-mail: info-gz@hioki.cn

All information correct as of Jun. 4, 2008. All specifications are subject to change without notice.                                                                             3174E2-86B
This datasheet has been downloaded from:
             www.EEworld.com.cn

                 Free Download
           Daily Updated Database
      100% Free Datasheet Search Site
  100% Free IC Replacement Search Site
     Convenient Electronic Dictionary

               Fast Search System
             www.EEworld.com.cn

                                                 All Datasheets Cannot Be Modified Without Permission
                                                                Copyright Each Manufacturing Company
小广播