             3159 INSULATION /

             WITHSTANDING HiTESTER

                                                                        Safety Standards Measuring Instruments

Improved Efficiency

                                                 for Electrical Safety Testing

                                                                        The 3159 INSULATION / WITHSTANDING HiTESTER is a

                                                                        combination insulation resistance as well as voltage endurance tester.

                                                                        It continuously performs insulation testing and voltage endurance

                                                                        testing of electrical equipment and parts, doing both tests in a simpler

                                                                        and more efficient way.  This instrument implements all the applicable

ISO14001                                                                safety standards, and is small, light, and inexpensive.      It also comes

JQA-E-90091                                                             standard with external I/O for automating production lines.

             http://www.hioki.co.jp/

HIOKI company overview, new products, environmental considerations and

other information are available on our website.
1

   One Unit Serves Two Functions --

   Continuous Testing of Insulation and of

   Voltage Endurance

   q Continuous Testing of Insulation                                        q Standards Testing

   and Voltage Endurance                                                     Contains on-board comparator and timer functions for

   In automatic testing mode, the 3159 continuously tests                    determining compliance, thus simplify the testing of all

   either insulation followed by voltage endurance (the                      applicable safety standards.

   ability to "withstand" voltage) (I→W), or voltage                         q Interlock Function

   endurance followed by insulation (W→I).        In manual

   mode it separately performs insulation testing or voltage                 Based on inputs such as a starter signal, enters a state

   endurance testing.                                                        where output is blocked and testing is impossible, to

   q Stores up to 10 Sets of Test Conditions                                 guarantee safety such as during automated testing.

   Stores up to 10 sets of test conditions for each of voltage

   endurance mode and insulation mode, and can quickly

   switch among the test conditions. (Save/Load)

                                                                                                           Fluorescent Display Tube

                                                                                                           The display uses a bright and

                                                                                                           clear fluorescent tube.

   Analog Voltmeter

   The test voltage can be verified

   not only from the digital display

   but also in analog form.  (Only

   for voltage endurance testing.)

   Danger Light

   The  warning     light    flashes

   during testing and whenever

   there is a high voltage between

   the terminals.

   External Switch

   Start/stop may be controlled with

   the 9613  or the 9614.

   (The 9613 and 9614 are options.)                                          Insulation

                                                                             Testing

                                                                             Two Rated Measurement Voltages:

        9613                     9614                                        DC 500 or 1000 V

                                                                 Voltage     Measurement Range:

                                           Endurance Testing                 0.5 MΩ to 2000 MΩ

                    Transformer Capacity : 500VA                             q Discharge Feature

                    Two Test Voltage Ranges :                                The charge on the object being tested is

                    AC 0 to 2.5 or 5.0 kV                                    discharged within the tester.  This discharge

                                                                             function avoids impacting the next test

                    q Voltage Comparator                                     process.

                    In order to prevent test voltage errors, the timer is

                    only activated after the voltage rises to the

                    specified test voltage ± 5% (or ± 50V, whichever

                    is larger).

                    (Depending on options, it may be possible to change the

                    voltage check just before test completion.)
                                                                                                                                                                                 2

Functions for Handling a Wide Range of Situations

                                                                                                                                       Automatically

This tester comes standard with features for automating EXT I/O, RS-232C,

and status out (relay contact output), as well as data management features.

Status Out                                                                                                                               Buzzer Volume

When the output conditions set up by                                                                                                     Adjustment

the dip switches are satisfied (OR                                                                                                       To indicate "pass" or "fail".

condition), there is relay contact

output.

1. H.V.ON     Output voltage generation

2. TEST       Testing in progress                                                                                                        Rear Panel Voltage

3. PASS       Passed                                                                                                                     Output Terminals

4. FAIL       Failed                                                                                                                     These are normally

5. INT.LOCK   Interlocked                                                                                                                connected to the front

                                                                                                                                         terminals.

6. READY      Ready

7. EXT.CONT.  Under external control                                                                  EXT I/O (Open Collector, Photocoupler Insulation)

8. POWER ON   Powers the 3159 on

                                                                                                      EXT I/O   Output Signals

                                RS-232C                                                                  Pin            Signal            Function

                                                                                                      Number            Name

                                Allows automatic testing and reading of                                  1              READY    LO when in "ready state"

                                test results from a personal computer.                                   2              L-FAIL   LO when in "fail state" for the lower bound

                                                                                                         3              U-FAIL   LO when in "fail state" for the upper bound

                                                                                                         4              PASS     LO when in "pass state"

                      EXT I/O       Input Signals                                                        5              TEST     LO when in "test state"

                           Pin        Signal                                                             6              H.V.ON   LO when a voltage is being generated to

                      Number          Name         Function                                                                      the output terminals

                                                                                                         11             W-MODE   LO during voltage endurance

                           7          EXT-E        When LO, the external I/O input signal is effect.                             ("withstanding") testing

                           8          START        When LO, it functions as a "Start" key.               12             I-MODE   LO during insulation testing

                           9          STOP         When LO, it functions as a "Stop" key.                13             W-FAIL   LO when in "fail state" for voltage

                                                                                                                                 endurance testing

                           10       INT.LOCK       Interlocked when open.                                14             I-FAIL   LO when in "fail state" for insulation testing

                      15 to 18      ISO.COM        Ground inputs for external devices.                33 to 36          ISO.DCV  15 V (0.1 A) outputs

Wide Range of

           Functions

1. Pass Hold Function (0: No Hold, 1: Hold)                                                           ➀  ➁      ➂       ➃        ➄  ➅  ➆  ➇            ➈       ➉

The pass state is held when it occurs.             This is convenient

for verifying the decision value.

2. Fail Hold Function (0: No Hold, 1: Hold)                                                           Various functions can be specified with the SHIFT + STOP keys.

The fail state is held when it occurs.             This is convenient

for temporarily stopping the test process.

                                                                                        8. Interlock Function (0: Not Specified, 1: Specified)

3. Hold State (0: No Hold, 1: Hold)                                                                   This specifies whether the interlock terminal for

This saves the state when the stop key is pressed during                                              external I/O should be effective.

a test in order to unconditionally end the test.

                                                                                        9. Voltage Comparison Time (0: Start of Test, 1: End of Test)

4. Momentary Out (0: Not Specified, 1: Specified)                                                     When the voltage comparator is on during voltage

This function outputs a voltage only when the start key                                               endurance testing, this specifies whether the comparison

is being pressed.     The start key is effective both as EXT                                          should be done at the start or the end of the test.

SW and external I/O.

                                                                                        10. Insulation Resistance Measurement Range

5. Double Action (0: Not Specified, 1: Specified)                                                     (0: Fixed Range, 1: Automatic Range)

This function starts the test if the start key is pressed less                                        This specifies whether the measurement range for the

than about 0.5 seconds after the stop key.                                                            insulation resistance test should be a fixed range or an

                                                                                                      automatic range.

6. Fail Mode (0: Not Specified, 1: Specified)

This function means that hold state can be released only                                11. Insulation Resistance Test End Mode

by the stop key on the main body.                                                                     0 : Test for the specified time

                                                                                                      1 : Stop when "pass" is determined

7. RS Command "START" (0: Not Specified, 1: Specified)                                                2 : Stop when "fail" is determined

This specifies whether the RS command "START"                                                         This specifies the method of ending the insulation

should be effective.                                                                                  resistance test.
3

   3159 Specifications

   s Voltage Endurance Testing                                                                           s Decision Function

   [ Test Voltage ]                                                                                      Decision method           :  Window comparison method (digital specification).

                                                                                                         Decision results          :  UPPER-FAIL:

   Output voltage             :  Two ranges: AC 0 to 2.5 or 5.0 kV                                                                    The measured current (insulation resistance value)

   Voltage output method      :  Zero input switch                                                                                    exceeded the specified upper bound.

   Transformer capacity       :  500 VA (rating: 30 minutes)                                                                          PASS:

   Voltage adjustment         :  Manual adjustment                                                                                    The measured current (insulation resistance value)

   method                                                                                                                             was between the specified upper and lower bounds

   Voltmeter                  :  Average value rectified root-mean-square display                                                     and the specified time elapsed

                                 Digital: AC 0.00 kV to 5.00 kV (full scale)                                                          LOWER-FAIL:

                                 Accuracy: ± 1.5% f.s.                                                                                The measured current (insulation resistance value)

                                 Analog: AC 0 to 5 kV (full scale)                                                                    was less than the specified lower bound.

                                 Accuracy: ± 5% f.s                                                      Decision processing       :  For each decision result, output the display portion,

   Waveform                   :  Same as the power supply waveform                                                                    the buzzer sound, and EXT I/O signal

   Frequency                  :  Same as the power supply frequency                                      Specification ranges      :  Voltage endurance testing :

                                                                                                                                      0.1 mA to 120 mA (upper bound) / 0.1 mA to 119 mA (lower bound)

   [ Current Detection ]                                                                                                              Insulation testing :

   Current measurement        :  0.01 mA to 120 mA                                                                                    0.2 MΩ to 2000 MΩ (same for the upper and lower bounds).

   range                                                                                                 Specification resolution  :  Voltage endurance testing :

   Indicated valueange        :  Average value rectified root-mean-square display (digital)                                           0.1 mA (0.1 mA to 9.9 mA) / 1 mA (10 to 120 mA)

   Measurement                :  0.01 mA (2 mA or 8 mA range)                                                                         Insulation testing :

   resolution                    0.1 mA (32 mA range)                                                                                 0.01 MΩ (0.2 MΩ to 2 MΩ), 0.1 MΩ (2.1 MΩ to 20 MΩ)

                                 1 mA (120 mA range)                                                                                  1 MΩ (21 MΩ to 200 MΩ), 10 MΩ (210 MΩ to 2000 MΩ)

   Measurement                :  ( ± 3% f.s. + 20 µA) over the entire range                              s Timer

   accuracy                      (Assumes a power supply waveform distortion ratio                       Specification range       :  0.5 to 999 sec

                                 of 5% maximum.)                                                         Action                    :  When ON is specified:

   s Insulation Resistance Testing                                                                                                    After starting, a countdown from the specified time is displayed

                                                                                                                                      When OFF is specified:

   [ Test Voltage and Measurement Range ]                                                                                             Displays the elapsed time since the start.

                                                                                                         Specification resolution  :  0.1 sec. (0.5 to 99.9 sec.) ± 50 msec

   Rated voltage              :  DC 500 V or 1000 V                                                      and accuracy                 1 sec. (100 to 999 sec.) ± 0.5 sec

   Unloaded voltage           :  From 1 to 1.2 times the rated voltage                                   Nondeterministic          :  0.5 sec. (Mask time until the determination begins

   Rated measured             :  1 mA to 1.2 mA                                                          interval                     during insulation resistance testing.)

   current                                                                                               s Interfaces

   Short circuit current      :  4 mA to 5 mA (500 V) / 2 mA to 3 mA (1000 V)

   Measurement range          :  0.5 MΩ to 999 MΩ (500 V), 1 MΩ to 999 MΩ                                EXT I/O                   :  Output signal:

   and accuracy                  (1000 V) / ± 4% rdg                                                                                  Open collector output

                                 1000 MΩ to 2000 MΩ / ± 8% rdg.                                                                       Maximum loaded voltage DC 30 V

   Measured resistance        :  2 MΩ, 20 MΩ, 200 MΩ, 2000 MΩ (500 V)                                                                 Maximum output current DC 100 mA / 1 signal

   range                         4 MΩ, 40 MΩ, 400 MΩ, 2000 MΩ (1000 V)                                                                Input signal:

                                                                                                                                      Active low input / maximum applied voltage DC 30 V

                                                                                                         EXT SW                    :  Input signal (contact point input)

                                                                                                                                      START, STOP, SW.EN (front terminal switch is effective)

   s General Specifications                                                                              RS-232C                   :  Start-stop synchronization, full duplex, 9600 bps

   Display                    : Fluorescent display tube (digital display)                               Voltage endurance         : AC 1.35 kV, 10 mA, for one minute, between power

   Monitor functions          : Output voltage, detected current, measured resistance                                                 supply and chassis

   Monitor period             : 2 times/sec. minimum                                                     Maximum rated             : 800 VA

   Operating temperature      : 0 ˚C to 40 ˚C, 80% rh maximum (no condensation)                          power

   range                                                                                                 Dimensions                : Approx. 320 (W) × 155 (H) × 330 (D) mm (not

   Storage temperature        : -10 ˚C to 50 ˚C, 90% rh maximum (no condensation)                                                     including protrusions.)

   range                                                                                                 Mass                      : Approx. 18 kg (3159), Approx. 20.5 kg (3159-01),

   Temperature and            : 23 ˚C ± 5 ˚C, 80% rh maximum (no condensation),                                                       Approx. 21.5 kg (3159-02 to 3159-04)

   humidity range for            after warming up for at least 5 minutes                                 Applicable standards      : EMC:      EN 61326-1:1997 + A1:1998 class A

   assured accuracy                                                                                                                   Safety:  EN 61010-1:1993 + A1:1995

   Operating locations        : Indoor, altitude 2000 m maximum                                                                       Contamination level 2, overvoltage category

   Power supply voltage       : AC 100 V (3159), AC 120 V (3159-01),                                                                  (expected overvoltage category 2500V)

                                 AC 220 V (3159-02), AC 230 V (3159-03),                                 Standard accessories      : 9615 H.V. test lead (high voltage side and return, one

                                 AC 240 V (3159-04),                                                                                  each), power cord, extra fuse

   Power supply               : 50 Hz to 60 Hz                                                           Options                   : 9613 REMOTE CONTROL BOX (SINGLE),

   frequency                                                                                                                          9614 REMOTE CONTROL BOX (DUAL),

                                                                                                                                      9616 WARNING LAMP

   3159           INSULATION / WITHSTANDING HiTESTER (100V)                                              s Options

   3159-01 INSULATION / WITHSTANDING HiTESTER (120V)                                                     9613      REMOTE          CONTROL            BOX (SINGLE)

   3159-02 INSULATION / WITHSTANDING HiTESTER (220V)                                                     9614      REMOTE          CONTROL            BOX (DUAL)

                                                                                                         9616 WARNING              LAMP

   3159-03 INSULATION / WITHSTANDING HiTESTER (230V)                                                     9637 RS-232C CABLE (1.8 m) (9pin-9pin/Cross)

   3159-04 INSULATION / WITHSTANDING HiTESTER (240V)                                                     9638 RS-232C CABLE (1.8 m)                         (9pin-25pin/Cross)

                                                                    DISTRIBUTED BY

   HEAD OFFICE :

   81 Koizumi, Ueda, Nagano, 386-1192, Japan

   TEL +81-268-28-0562 / FAX +81-268-28-0568

   E-mail: os-com@ hioki.co.jp

   HIOKI USA CORPORATION :

   6 Corporate Drive, Cranbury, NJ 08512 USA

   TEL +1-609-409-9109 / FAX +1-609-409-9108

   E-mail: hioki@ hiokiusa.com

   All information correct as of Mar. 9, 2001. All specifications are subject to change without notice.

   s Inter net HIOKI website     http://www.hioki.co.jp/                                                                                                       F3159E1-13B-05P    Printed in Japan
This datasheet has been downloaded from:

datasheet.eeworld.com.cn

Free Download

Daily Updated Database

100% Free Datasheet Search Site

100% Free IC Replacement Search Site

Convenient Electronic Dictionary

Fast Search System

www.EEworld.com.cn

All Datasheets Cannot Be Modified Without Permission