    3156 LEAK CURRENT HiTESTER

                                                                                                         Safety Standards
                                                                                                         Measuring Instruments

       The Long-Awaited Bench-type Model Is Here !

    Leakage Current Measurement Essential
    for Electrical Safety

                                                         Because of the risks it poses, many standards stipulate leak
                                                         current test methods, the performance of test equipment, limits
                                                         for leak current, and other factors related to leak current. The
                                                         bench-type 3156 LEAK CURRENT HiTESTER by itself, without
                                                         any additional equipment, is capable of performing tests that
                                                         comply with standards for a wide range of equipment, from
                                                         general purpose electrical equipment to medical electrical
                                                         equipment. The 3156 also is equipped with functions and an
                                                         interface that support embedded test lines.

          http://www.hioki.co.jp/

HIOKI company overview, new products, environmental considerations and
other information are available on our website.
1

   Automatically compatible with Networks

   stipulated by IEC/UL/JIS standards

   In order to avoid dangers such as electric shock,                               Main Features
   electrical equipment is designed so that those
   portions that are likely to be touched by people              s Standard support for a variety of Networks
   are insulated from the power source. However,
   the insulation resistance is not infinite, so there is        The 3156 is engineered to automatically support the
   always some leak current present. Furthermore,                "Networks" (human simulated resistance) stipulated by
   the extent of the leak current changes over time              the various international safety standards. Switch
   as the insulation deteriorates. To determine their            between Networks or use the filters using a PC.
   safety level, the 3156 LEAK CURRENT HiTESTER
   can prove to be highly valuable, both as a tester             s Measurement that complies with standards
   on the production line as well as a maintenance
   and inspection tool.                                          The 3156 by itself can measure leak current in compliance
                                                                 with IEC, UL, and JIS standards.

   s Automatic measurement function                              s RS-232C, GP-IB, and EXT I/O interfaces

   The 3156 is capable of automatically measuring power              standard

   supply polarity switching as well as the normal state/single  Because the 3156 is equipped with several standard

   failure state, and can display the maximum values. The        interfaces, it is easy to set up for automatic inspection on

   measurement time and the wait time can be set by the user.    a production line, etc. In addition, the RS-232C port can

   s Stores data for 100 units                                   be used to connect the optional 9442 Printer for easy
                                                                 print-outs.

   The 3156 can store data such as equipment name, control       s Save up to 30 sets of measuring conditions
   number, ground class, and measured values for up to 100

   pieces of equipment being tested.                             The 3156 can save and load up to 30 different sets of

   s Power supply separation                                     measuring conditions, and can switch between different
                                                                 sets quickly.

   The power supply for the 3156 is separate from the line

   power supply of the equipment being tested, preventing             Simple operation in an interactive format
   damage to equipment due to wrong supply voltage being
   input.                                                            The 3156 is equipped with a touch panel that can
                                                                     be used to make settings simply by touching the

   110% voltage application terminals                                items to be selected on the display, making
                                                                     operation even more simple.

   The 110% voltage application terminals are used when testing

   medical equipment. It outputs on a 1:1 basis the voltage that is

   supplied to the line power supply for the equipment being

   tested. The polarity of the connectors can be switched.

                                                                 Measurement mode

   Breaker for equipment being tested                            General       q Leakage current between enclosure and line
                                                                 equipment     q Leakage current between enclosure and earth
   Since the 3156 is a bench-type unit that is suited                          q Leakage current between enclosure and enclosure
   embedded test lines, it is equipped with a terminal           Medical       q Earth leakage current
   board and breaker on the front panel, allowing                equipment     q Patient leakage current I
   test equipment to be easily connected to the 3156                           q Patient leakage current II
   even when it is rack mounted.                                               q Patient leakage current III
                                                                               q Patient auxiliary current
                                                                                                                                                        2

Based on the settings for the Network and
the equipment being tested, the 3156 displays
the appropriate measurement mode. The
state of the settable keys can be seen at a
glance. All settings can be made by selecting
the appropriate items or values.

                                                [Initial screen]

s Wiring check functions                        s Support for proprietary networks

The 3156 performs polarity                      When making measurements on a proprietary network that is
checks and VA checks.                                                                  not built into the 3156, the
                                                                                       3156 can be used as a true
                                                                                       effective value ammeter in the
                                                                                       1MHz band.

[System screen]                                                                                s Browsing saved data

     Maximum value display                                                                                    The 3156 allows the user to
                                                                                                              browse all saved data. Saved
Displays the power supply failure status                                                                      data can also be printed out
and the maximum value for the leak current                                                                    or downloaded to a computer.
as it fluctuates due to the operation of the
equipment being tested.                                                                       Result of evaluation against tolerance setting
                                                                    Current measured value
      Power supply polarity/equipment
      status/measured current                                                                  Data storage

                Allowable value                                                                                  Measurement data: For up to 100 units
                                                                                                                 Measuring conditions: Up to 30 sets
The upper limit for tolerance is set according
to the standard. The set value can be changed                            [Measurement screen]
freely, if necessary.
                                                Single failure setting

                                                The single failure status can be set for
                                                individual items in automatic testing, with
                                                continuous testing performed automatically
                                                only for the items for which the single failure
                                                setting was made. In addition, the wait time
                                                and measurement time can also be set.
3

              Leak current measurement
                 for general equipment

   The network (human simulated resistance) is                       IEC standard
   specified separately in each standard, and when
   taking measurements a network that is suited to                         [ IEC 60990, others ]
   the standards is needed. The 3156 is pre-engineered

   with basic Networks that correspond to each

   standard.                             UL standard

                                         [ UL471, UL1310, UL1437, others ]

       JIS standard                                     The standard numbers that are listed are only examples.
                                                        The networks listed below can be used with all applicable
                  [ JIS B8561, others ]                 standards.

   For JIS standards                     Network types

                                         For IEC60990s                      For UL standards

                                         1.5k           0.22F              1.5k
                                         0.1%          1%                 0.1%

              1k                                                                   0.15F
              0.1%                                                                  2%

                                                         500                Basic measuring element
                                                         0.1%
   Basic measuring element               Basic measuring element

              10k0.1%                                                      General purpose 1

              1k         11.22nF         1.5k           0.22F1%
              0.1%      1%             0.1%            10k0.1%

                    579                   500           0.022F                                    1k
                    1%                  0.1%              1%                                    0.1%
                                                                                 Basic measuring element
              Filter on                                 Filter 1 on
                                                                            General purpose 2

                                         1.5k           0.22F1%                             2k
                                         0.1%           10k0.1%                             0.1%
                                                                            Basic measuring element
                                             500             20k
                                            0.1%            0.1%

                                         0.0062F1% 0.0091F1%
                                                 Filter 2 on
                                                                                                                                                        4

Leak current measurement
  for medical equipment

Medical equipment leak current includes not only

ground leak current and exterior leak current, but                                               IEC standard
also patient leak current (I to III) and patient

measured current. The 3156 can measure all of                                   [ IEC 60601-1 ('88) + am1 ('91) + am2 ('95) ]

these different types of leak current without any

additional equipment.

                                          UL standard

JIS standard                              [ UL 2601-1, others ]

   [ JIS T 0601-1, others ]                                                                                   The 3156 does not have a built-in
                                                                                                                      insulated transformer. When
                                                                                                                                             measuring

                                                                    medical equipment, use a step-up insulated transformer or other
                                                                    device to provide 110% of the rated voltage as the line power
                                                                    supply for the equipment being tested.
                                                                    The standard numbers that are listed are only examples. The

                       Leak current types and Network types

                                                                                                 For medical equipment

Medical equipment leak current types

Faulty medical equipment                  Faulty medical equipment

Signal input section   Medical equipment               Patient leakage                                         1k
                                                       current III                                             0.1%
                                                                                                 Basic measuring element
                                                       Patient leakage
                                                       current II

                                                                                Patient leakage         10k0.1%
                                                                                current I

Earth leakage current

                                                                                                 1k     0.015F
                                                                                                 0.1%     1%

                      Enclosure leakage                                                                 Filter on
                      current

Earth leakage current
  A current flowing from the power supply through the interior or surface of
  the insulation to the protective ground line.

Enclosure leakage current
A current flowing from a part of the enclosure which during normal use can be
  touched by an operator or patient either to ground or to another part of the
  enclosure, through an external conductor other than the protective ground.

Patient leakage current
  A current flowing from the device through the patient to ground.

Patient auxiliary current
  A current flowing between contacts which during
  normal use are connected through the patient, with
  no intended physiological effect.
5

    Separation of the power supply for the 3156 and                                               GP-IB RS-232C EXT I/O
    the line for the equipment being tested

   The power supply voltage for the 3156 does not need to
   be changed, even if the supply voltage for the equipment
   being tested is different. The difference can be managed
   by controlling the line supply voltage for the equipment
   being tested. This helps prevent malfunctions due to the
   wrong voltage being input for the 3156 power supply.

                                                                                       Main unit (3156)              Line power supply
                                                                                       power supply                  100 to 240V (50/60Hz)
                                                                                                                     1500VA
   External control through EXT I/O

   The start of measurement and the loading of the measuring                                             Contents of EXT I/O
   conditions can be controlled externally. The evaluation
   results, test signals, etc., can also be output externally,                         q Outputs
   permitting the construction of an automated line.
                                                                                       TEST      : Continues to output a low signal during automatic measurement

                                                                                       MEAS      : Outputs the count when measuring multiple items during
                                                                                                   automatic measurement
   Input signal      Active low signal

   Maximum                                                                             PASS      : Outputs the evaluation result "PASS" for individual items
   applied voltage                                                                                during automatic measurement
                     Voltage input to the EXT.DCV terminal

                                                                                       FAIL      : Outputs the evaluation result "FAIL" for individual items
                                                                                                  during automatic measurement
   High level        Voltage input to the EXT.DCV
                     terminal, or open                                                 T-FAIL    : Continues output even if a FAIL occurs during automatic testing.

   Low level         0.3V DC or less                                                   NT.DCV : Internal 5V DC output (not insulated from internal circuitry)

   Output signal     Open collector output                                             INT.GND : Internal GND output (identical to rack grounded GND level)

   Maximum load      24V DC (when not using the                                        q Inputs
   voltage           EXT.DCV terminal)
   Maximum           60mA DC per signal                                                START : Starts automatic measurement when low
   output current    (when low level)
                                                                                       STOP      : Forcibly ends measurement when low

                                                                                       LOAD (0 to 4): Loads a saved panel (30 panels)

   In addition to outputting evaluation results for each                               0ADJ      : Zero adjust signal (valid only in ground leak current mode)
   measurement item, the T-FAIL output is provided for
   continued output after one FAIL during automatic testing.                           EXT.DCV : External power supply input (5V to 24V DC)

                                                                                       EXT.COM : External COM input

           External control through a PC                                                                             RS-232C
                                                                                                                      GP-IB
   Includes RS-232C interface and GP-IB interface as
   standard. These can be used to control the functions of
   the 3156 through a PC, and to collect measurement
   values.

   s RS-232C interface

   Transfer method: Start-stop synchronization, full-duplex

   Transfer speed : 9600 fixed

   Data length : 8 bits

   Stop bit      :1

   Parity bit    : None                                                                s GP-IB interface

   Delimiter     : CR+LF                                                               Compliant standards: IEEE-488.1 1987,

   Handshaking : None                                                                                    Refer to IEEE-488.2 1987

   XON/XOFF      : Not used                                                            Connector         : D-sub 24 pin female

   Connector     : D-sub 9-pin male

                     Data printout

   Data can be printed out on the optional 9442 Printer by                                                   Printout contents
   using the RS-232C interface connector. This is useful                               (The printed items can be selected from among the items listed below.)
   for attaching inspection data, etc.
                                                                                       q Measurement date            q Maximum value
   9442 Printer           Printing method: Thermal serial dot printing                 q Equipment name              q Evaluation result
   (Optional)             Paper width/printing speed: 112 mm/52.5cps                   q Control number              q Measured current
                          Power supply: 9443 AC Adapter, or NiMH                       q Class (mounting section)
                          battery provided (charge with the 9443; can print            q Network                        (AC, DC, AC+DC, AC peak)
                          approximately 3000 characters with a full charge)            q Measurement mode
                          Dimensions: Approx. 160(W)66.5(H)170 (D) mm                q Filter setting              q Power supply polarity
                          Mass: Approx. 580g                                           q Tolerance                      (normal phase, reverse phase)

   * Connecting the 9442 requires either the 9444 Connecting Cable or the AC adapter.                                q Equipment status
                                                                                                                        (normal, not grounded)
                                                                                                                                                                    6

3156 Specifications
s Leak current measurement

Measurement mode : Earth leakage current / Leakage current between                 Measurement          : Indication of a current value calculated based
                                                                                   system                the measured drop in voltage caused by simulated
                    enclosure and earth / Leakage current between

                    enclosure and enclosure / Leakage current between                                   resistance of the human body.

                    enclosure and line / Patient leakage current I /                                    Measurement of true effective value.

                    Patient leakage current II / Patient leakage current                                The measurement section: chassis-grounded and

                    III / Patient auxiliary current                                                     floating.

Target current : DC, AC, AC+DC, AC peak                                            A/D conversion system :   system (20 bits)

Allowable value : 25 mA max. (DC/AC/AC+DC mode)                                    Input resistance : 1 M 1% (single-ended input)
measurement current 75 mA max. (ACpeak mode)
                                                                                                        Excluding voltmeter section, simulated resistance

Measurement range : DC/AC/AC+DC mode:                                                                   of the human body (current detection circuit)

                    50 A/500 A/5 mA/25 mA                                        Input capacity : 200 pF or lower (between terminals T1 and T2)

                    AC peak mode:                                                                       (f = 100 kHz, with network circuit isolated)

                    500 A/1 mA /10 mA/75 mA                                       Grounding capacity : 200 pF or lower(between terminals T1/T2 and chassis)

Range switch        : Auto range/Hold range                                        CMRR                 : 60 dB or higher, at 60 Hz, 10 kHz

Trigger system : Manual: Automatic generation of internal trigger,                 (between terminals T1/T2 40 dB or higher, at 100 kHz, 1 MHz

                    and free-run measurement                                       and chassis)

                    Automatic: measurement started by external                     s Network (human simulated resistance)

                    start signal                                                   q For medical electrical equipment

                    1. Started by pressing Start key on the

                    operation screen.                                                                   : Basic measurement element: 1 k

                    2. Asserted by the START terminal on                                                Filter: 10 k + 15 nF

                    the EXT I/O connector.                                         q For IEC 60990 : Basic measurement element: 1.5 k + 500

                    3. Started by the :START interface                                                  Filter 1: 10 k + 22 nF

                    command.                                                                            Filter 2: 10 k + (20 k+ 6.2 nF) // 9.1 nF

Measurement speed : Trigger-system manual measurement: 100 1 ms                   q For JIS            : Basic measurement element: 1 k

                    (indication of moving average of 16 measurements)                                   Filter: 10 k + 11.22 nF + 579

                    Trigger-system automatic measurement:                          q For UL             : Basic measurement element: 1.5 k // 0.15 F

                    100 1 ms (min.) - measurement setting time                    q General-purpose 1: Basic measurement element : 1 k

Measurement         : Terminal T1, terminal T2 (with built-in fuse holder)         q General-purpose 2: Basic measurement element : 2 k
terminals            Terminal T3 (110% voltage application terminal)

s Functions

110% voltage        : Equipped with a voltage output terminal (T3) that applies    Allowable value      : Allowable value: Sets the upper-limit current value
                                                                                   judgement function    Judgement: PASS measurement value  upper-limit value
application function 110% power supply voltage between the functionally
                                                                                   Mode selection                       FAIL measurement value > upper-limit value
                    insulated signal input/output section (or Type F applied       function              Processing: Indication, buzzer, judgement output from

                    part) and ground.                                              Beep sound setting                  EXT I/O
                                                                                                        : Current measurement function
                    Output ON/OFF selection                                        Save/load function
                                                                                                           Unit of current measurement: Auto/ mA, fixed
                    Output impedance: 22.5 1 k                                    Data save function    Voltage measurement function
                                                                                                          Isolates the internal network for using the product as a
                    Applied when positive phase (to input power supply voltage)   Clock function         voltmeter between terminals T1 and T2.
                                                                                                          Maximum measurement voltage: 25 V
                    Applied when negative phase (to input power supply voltage)   Data backup          : Allowable value judgement
                                                                                   function              Key input
                    Automatic switching function (automatic measurement function)  Backlight automatic   T3 (110% voltage application terminal) output
                                                                                   OFF function          Line voltage output from T2
Wiring check function : Polarity check/VA check                                    Self-test function   : 30 panels for saving the following setting data
                                                                                   Language setting      (measurement mode, network, equipment name, control
Automatic measurement : Setting of measuring time.                                 System reset          number, grounding class, applied part, measurement
                                                                                                         range, filter, target current, allowable value setting,
function            Setting of delay (wait) time for changing setting conditions.                        malfunction condition setting, power supply polarity
                                                                                                         switching, automatic measurement items, automatic
                    The power supply polarity and equipment status are                                   measuring time, measurement delay time)
                                                                                                        : Saved content: Sample equipment information (equipment
                    automatically switched during measurement.                                           name, control number), measurement data, date
                                                                                                         Memory capacity: Data on up to 100 units
Application line    : Use of T2 and internal contact/Use of T1 and T2                                   : Auto calendar, automatic leap-year adjustment, 24-
                                                                                                         hour clock
selection function                                                                                       Clock accuracy: Deviation of about 4 minutes a month
                                                                                                        : SRAM (test condition data), RTC
Ground fault        : Pre-check of current value between connection terminals to                         Backup battery life: 4 years (reference value at 25C)
                                                                                                        : Constant ON/Auto OFF
prevention function prevent a ground fault. Only effective for leakage current                           1 min. to 30 min., in 1-min. increments
                                                                                                        : MEM (internal RAM)/KEY (6 6 matrix touch panel)/
                    measurement between enclosure and line.                                              LCD (front LCD panel)/LED /Buzzer
                                                                                                        : Japanese or English
Setting of single-  : Setting of malfunction mode for power line for                                   : Clears all data including measurement conditions and
fault condition        sample equipment.                                                                 measurement data.
                                                                                                         Clears all saved measurement data.
                    1. Disconnection of one wire in power line (neutral side)                            Clears all saved condition setting data including panels.

                    2. Disconnection of protective earth conductor

                    Application of 110% voltage for simulated connection of

                    malfunctioning equipment.

                    Positive phase/negative phase

                    Selection of application line for leakage current

                    measurement between enclosure and line.

Power line for sample : Positive/negative phase
equipment (switching (Automatic switching possible when using automatic
power supply polarity) measurement function)

Setting of measuring : Setting range: 1 sec. to 5 min., in 1-sec. increments

time                Effective only in automatic measurement

Measurement delay : Setting range: 1 sec. to 30 min., in 1-sec. increments

(setting) function  Wait time from the completion of measurement to

                    power supply disconnection

                    Wait time from switching power supply polarity to

                    the start of measurement

                    Wait time for operations other than switching polarity

Maximum value hold : Effective in all measurement modes
function
7

   s Accuracy (Current measurement )                                                                             q Measurement of DC

    s Operating temperature and humidity for guaranteed accuracy: 23 5 C, 80%rh or lower (no condensation)    Range             Guaranteed Measurement         Accuracy
    s Temperature coefficient: 0.1 x basic accuracy x (T-23) weighted --- operating temperature T (C)                           accuracy range resolution   (0.2%rdg.+3dgt.)
    s Warm-up time: 30 min.
    s Value calculated based on voltage detected at terminals of Network having non-inductive resistance of 1 k  25.00 mA 4 mA or more 10 A                     1.0%f.s.

      (theoretical value).
    s Measurements in voltmeter mode conform to the following accuracy level.

   q Measurement of AC*1, AC+DC

                Guaranteed Measurement                             Accuracy                                      5.000 mA 400 A or more 1 A
              accuracy range resolution       DC < f  100kHz 100kHz < f  1MHz
   Range                                                                                                         500.0 A 40 A or more 0.1 A

   25.00 mA 4 mA or more 10 A                                                                                   50.00 A 4 A or more 0.01 A
   5.000 mA 400 A or more 1 A
   500.0 A 40 A or more 0.1 A              (2.0%rdg.+6dgt.)         (2.0%rdg.+10dgt.)                       q Power supply voltage monitor accuracy
   50.00 A 4 A or more 0.01 A
                                                 2.0%f.s.                         2.0%f.s.                      Range            Guaranteed Measurement                       Accuracy
                                                                                                                 300.0 V         accuracy range resolution

   q Measurement of AC Peak*2                                                                                                    85 V or more*3 0.1 V (2.0%rdg.+10dgt.)

   Range        Guaranteed Measurement                      Accuracy                                             q Current consumption monitor accuracy
              accuracy range resolution
                                                                                                                       (measurement method: average value response, effective value conversion)

                                              20Hz < f  1kHz            1kHz < f  10kHz                          Range             Guaranteed Measurement         Accuracy
                                                                                                                  16 A           accuracy range resolution   (2.0%rdg.+5dgt.)
   75.0 mA 8mA or more 100 A                 (2.0%rdg.+2dgt.)         (5.0%rdg.+10dgt.)
   10.00 mA 0.8mA or more 10 A                                                                                                  0.5A or more*4 0.1 A
   1.000 mA 100 A or more 1 A
   500.0 A 40 A or more 0.1 A                 2.5%f.s.                         5.0%f.s.                     *1: Weighted with frequency characteristic (fc = 4 Hz) of high-pass filter.
                                                 4.0%f.s.                                                       *2: Cannot be set when network A , B and C(filter off) is selected.
                                                                                                                 *3: "less than 80 V" is indicated for a value less than 80 V.
                                                                                                                 *4: "less than 0.5 A" is indicated for a value less than 0.5 A.

   s General Specifications

   Display section : 320 240 dot matrix LCD (with backlight)                                            Applicable Standards : EMC: EN61326:1997+A1:1998+A2:2001

   Operation section : 6 6 matrix touch panel                                                                                            EN61000-3-2:2000

   Operating temperature : 0 to 40 C , 80%rh or less (no condensation)                                                                    EN61000-3-3:1995+A1:2001
   and humidity
                                                                                                                                 Safety: EN61010-1:2001

   Storage temperature : -10 to 50 C, 80%rh or less (no condensation)                                                                     EN61010-2-031:1994 Pollution Degree 2
   and humidity
                                                                                                                                           Terminals T1, T2: Measurement category II
   Recommended         : 6 months
                                                                                                                                           (Anticipated Transient Overvoltage: 2.5 kV)
   Calibration Period
                                                                                                                                           Terminal T3: Measurement category I
   Operating Environment : Indoors, <2000 m ASL
                                                                                                                                           (Anticipated Transient Overvoltage: 1.5 kV)
   Power source for : 100, 120, 220, 240 VAC (default setting)
                                                                                                          Effect of conducted : 16%f.s. at 3 V
   main unit           Rated power source frequency: 50/60 Hz
                                                                                                          radio-frequency
                       Rated power: 30 VA                                                                                        (typical value for measurement in AC 500 mA range)
                                                                                                          electromagnetic field

   Power line for sample : 100 to 240 VAC                                                                 Accessories            : 9170 TEST LEAD 2 sets, 9195 ENCLOSURE

   equipment and power Rated power source frequency: 50/60 Hz                                                                    PROBE 1, 9399 CARRYING CASE 1, Alligator
   output
                       Rated power: 1,500 VA                                                                                     clips 3 (2 red, 1 black), AC power cord 2 (for

   Power output maximum : 25 mA                                                                                                  main unit, for power line of sample equipment),
   allowable leakage current
                                                                                                                                 Spare fuse 2 (for main unit power supply 250V
   Withstand voltage : [All power supply terminals] - [Protective earthing]
                                                                                                                                 T0.1AL, for measurement operation 250V
                       1.35 kV AC (5 mA), 1 min.
                                                                                                                                 T32mAL)
                       [All measurement terminals] - [All power supply terminals]
                                                                                                          Dimensions             : Approx. 320W 110H 263D mm
                       2.30 kV AC (5 mA), 1 min
                                                                                                          Mass                   : Approx. 4.0kg
                       [All measurement terminals] - [Control circuit]

                       2.30 kV AC (5 mA), 1 min

   3156 LEAK CURRENT HiTESTER                                                      9195 ENCLOSURE PROBE                    9170 TEST LEAD  9399 CARRYING CASE 9686 CARRYING CASE
   q Options
                                                                                                          (Standard accessory) (Standard accessory/2 set) (Standard accessory)                (Option)
   9637 RS-232C CABLE (9-pin to 9-pin., crossing cable,1.8m)
   9638 RS-232C CABLE (9-pin to 25-pin., crossing cable,1.8m)                                             9444 CONNECTION CABLE (for printer)
   9151-02 GP-IB CABLE (2 m)                                                                              1196 RECORDING PAPER (for printer)
   9151-04 GP-IB CABLE (4 m)                                                                              9686 CARRYING CASE (with casters)
   9442 PRINTER                                                                                           9267 SAFETY TEST DATA MEASUREMENT SOFTWARE
   9443-01 AC ADAPTER (for printer, for use in Japan)
   9443-02 AC ADAPTER (for printer, for use in EU)
   9443-03 AC ADAPTER (for printer, for use in U.S.)

                                                                                                                 DISTRIBUTED BY

   HEAD OFFICE :                              Shanghai Representative Office :
   81 Koizumi, Ueda, Nagano, 386-1192, Japan  1704 Shanghai Times Square Office
   TEL +81-268-28-0562 / FAX +81-268-28-0568  93 Huaihai Zhong Road
   E-mail: os-com@ hioki.co.jp                Shanghai, 200021, P.R.China
                                              TEL +86-21-6391-0090, 0092
   HIOKI USA CORPORATION :                    FAX +86-21-6391-0360
   6 Corporate Drive, Cranbury, NJ 08512 USA  E-mail: hioki-sh@81890.net
   TEL +1-609-409-9109 / FAX +1-609-409-9108
   E-mail: hioki@ hiokiusa.com

   All information correct as of Jun. 20, 2003. All specifications are subject to change without notice.                                          3156E2-36M-05P Printed in Japan
This datasheet has been downloaded from:
             www.EEworld.com.cn

                 Free Download
           Daily Updated Database
      100% Free Datasheet Search Site
  100% Free IC Replacement Search Site
     Convenient Electronic Dictionary

               Fast Search System
             www.EEworld.com.cn

                                                 All Datasheets Cannot Be Modified Without Permission
                                                                Copyright Each Manufacturing Company
小广播

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet 电子工程

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2020 EEWORLD.com.cn, Inc. All rights reserved