3140R

器件型号:3140R
器件类别:测试和测量    万用表与电压表   
文件大小:2626.99KB,共2页
厂商名称:FLUKE [FLUKE]
下载文档

器件描述

clamp multimeters & accessories 400a current clamp

参数

Manufacturer: Fluke
Product Category: Clamp Multimeters & Accessories
Brand: Fluke
Part # Aliases: 2650136

3140R器件文档内容

Maintenance                                                                                                                                                                                                           

                                XW Warning                                                     3140R
      To avoid possible electric shock or personal
      injury:                                                                             AC Current Clamp
       Before each use, inspect the Current Clamp.
                                                                                                                 Instruction Sheet
            Look for cracks or missing portions of the
            clamp housing and output cable insulating                    Introduction
            cover and for loose or weakened
            components. Pay particular attention to the                  The 3140R (hereafter called the "Current Clamp") is compatible
            insulation surrounding the clamp jaws.                       with Fluke measurement instruments that can accept a 2.0 V
       Do not use a damaged Current Clamp. If a                         rms output with a full-scale input to the clamp of
            clamp is damaged, tape it shut to prevent                    400 A rms (5 mV/A).
            unintended operation. A damaged clamp
            under warranty will be promptly repaired or                  Contacting Fluke
            replaced (at Fluke's discretion) and returned
            at no extra charge.                                          To contact Fluke, call one of the following telephone numbers:

If the Current Clamp does not work or perform properly, use                   USA: 1-888-44-FLUKE (1-888-443-5853)
the following steps to help isolate the problem:                              Canada: 1-800-36-FLUKE (1-800-363-5853)
                                                                              Europe: +31 402-675-200
1. Inspect the jaw mating surface for cleanliness. If any                     Japan: +81-3-3434-0181
      foreign material is present, the jaw will not close properly            Singapore: +65-738-5655
      and measurement errors will result.                                     Anywhere in the world: +1-425-446-5500

2. Verify that the function selection and range on the                        USA Service: 1-888-99-FLUKE (1-888-993-5853)
      measurement instrument are correct and adjusted to the
      sensitivity of the Current Clamp.                                  Or, visit Fluke's Web site at www.fluke.com.
                                                                         To register your product, visit register.fluke.com
Cleaning
                                                                         Safety Information
Periodically wipe the case with a damp cloth and mild
detergent.                                                                               XW Read First: Safety Information
                                                                               To ensure safe operation and service of the
                                  W Caution                                    current clamp, follow these instructions:
                                                                               Read the operating instructions before use
      To avoid damaging the Current Clamp, do not use
      abrasives or solvents to clean the clamp.                                      and follow all safety instructions.
                                                                               Use the Current Clamp only as specified in
Open the jaws and wipe the magnetic pole pieces with a lightly
oiled cloth. Do not allow rust or corrosion to form on the                           the operating instructions, otherwise the
magnetic core ends.                                                                  clamp's safety features may not protect you.
                                                                               Adhere to local and national safety codes.
LIMITED WARRANTY AND LIMITATION OF LIABILITY                                         Individual protective equipment must be used
                                                                                     to prevent shock and arc blast injury where
This Fluke product will be free from defects in material and                         hazardous live conductors are exposed.
workmanship for one year from the date of purchase. This warranty               Do not hold the Current Clamp anywhere
does not cover fuses, disposable batteries, or damage from accident,                 beyond the tactile barrier, see Figure 1.
neglect, misuse, alteration, contamination, or abnormal conditions of          Before each use, inspect the Current Clamp.
operation or handling. Resellers are not authorized to extend any other              Look for cracks or missing portions of the
warranty on Fluke's behalf. To obtain service during the warranty                    clamp housing or output cable insulation.
period, contact your nearest Fluke authorized service center to obtain               Also look for loose or weakened components.
return authorization information, then send the product to that Service              Pay particular attention to the insulation
Center with a description of the problem.                                            surrounding the jaws.
                                                                               Never use the clamp on a circuit with voltages
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER                                          higher than 1000 V CAT III or 600 V CAT IV.
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR                                          CAT III equipment is designed to protect
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL                                     against transients in equipment in fixed-
DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY.                                       equipment installations, such as
Since some states or countries do not allow the exclusion or limitation                    distribution panels, feeders and short
of an implied warranty or of incidental or consequential damages, this                     branch circuits, and lighting systems in
limitation of liability may not apply to you.                                              large buildings.
                                                                                     CAT IV equipment is designed to protect
Fluke Corporation       Fluke Europe B.V.                                                  against transients from the primary
                                                                                           supply level, such as an electricity meter
P.O. Box 9090           P.O. Box 1186                                                      or an overhead or underground utility
                                                                                           service.
Everett, WA 98206-9090  5602 BD Eindhoven                                       Use extreme caution when working around
                                                                                     bare conductors or bus bars. Contact with the
U.S.A.                  The Netherlands                                              conductor could result in electric shock.
                                                                               Use caution when working with voltages
11/99                                                                                above 60 V dc or 30 V ac. Such voltages pose
                                                                                     a shock hazard.
PN 2648112
April 2006

2005 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies.
Symbols                                                           General Specifications

, Application around and removal from hazardous live             Output Cable Length: 2.5 m

          conductors is permitted.                                Maximum Conductor Size: 32 mm
                                                                  Storage Temperature: -20 C to 60 C
T Product is protected by double insulation.
W Risk of Danger. Important information. See Instruction Sheet.  Operating Temperature: 0 C to 50 C
X Hazardous voltage.
                                                                  Relative Humidity: 10 C to 30 C: 95 %
) Conforms to relevant Canadian Standards Association                                       30 C to 40 C: 75 %

          directives.                                                                 40 C to 50 C: 45 %

P Conforms to relevant European Union directives.                Temperature Coefficient: 0.01 % X (specified accuracy)/ C
                                                                  (< 18 C or > 28 C)
Safety Specifications
                                                                  Altitude: Operating: 2000 m; 2000 m to 4000 m, derate
Category Rating: CAT III 1000 V and CAT IV 600 V per              category rating to 1000 V CAT II/600 V CAT III, Non-operating:
EN61010-1, Pollution Degree 2                                     12000 m
EMC: EN 61326-1, FCC for emission and immunity
                                                                  Dimensions: 150 x 70 x 30 mm
): Tested to US and Canadian standards for compliance to
                                                                  Weight: 114 g
UL61010-1 and CAN/CSA C22.2 No. 101.1:2004
                                                                  Measurement Considerations
P: IEC 61010-1 2nd Edition IEC 61010-02-032
                                                                  Observe the following guidelines when making measurements,
Electrical Specifications                                         refer to Figure 1:

Reference Conditions: 23 5 C, 20 to 75 % RH; conductor          Center the conductor inside the Current Clamp jaw.
centered in jaw opening; no DC component; no adjacent
conductor.                                                         Make sure the clamp is perpendicular to the conductor.

Output: 5 mV/A                                                    For optimal reading, make sure the conductor is positioned
                                                                        between the alignment marks on the jaws of the Current
Range           Frequency         Accuracy       Crest                  Clamp.
                                                 Factor
                            (% of reading + A)                   Avoid taking measurements close to other current-carrying
                                                                        conductors.

                45 Hz to                                          Operation
                400 Hz
2 A to 400 A                                      1.41            To use the Current Clamp:
2 A to 300 A                                       4
                           2 + 0.04                               1.  Connect the clamp cable to the measurement

                                                                      instrument's current inputs.

                                                                  2.  Make sure the measurement instrument is set to the

300 A to 400 A  45 Hz to                            3                 proper range.
300 A to 400 A   60 Hz
                                                 > 3 and          3.  Connect the current clamp jaws around the conductor
                                                    4
                           9 + 0.04                                   to be measured.

                                                                                      XW Warning

                           Phase Shift                                To avoid shock or personal injury, keep fingers
                                                                      behind the tactile barrier, see Figure 1.

  2 A to 5 A    45 Hz to   4                                                                                      Load Direction Arrow
5 A to 40 A    400 Hz     3                                                                                                   Alignment
40 A to 400 A              2                                                                                                   Marks

Typical Bandwidth: 5 Hz to 20 kHz                                     Single Current                                       Tactile Barrier
                                                                      Carrying
Working Voltage: 1000 V ac rms, in compliance with                    Conductor
EN61010

Common Mode Voltage: 1000 V ac rms from earth ground, in
compliance with EN61010

Input Load Impedance (of host instrument): > 1 M  in
parallel with up to 47 pF

Maximum Non-destructive Current: 1000 A

Duty Cycle: 2 A to 400 A continuous

Influence of Adjacent Conductor: < 9.0 mA/A

Influence of Conductor Position in Jaw Opening: 1.0 % of
reading + 0.05 A

                                                                                                                   Connector

                                                                                                                             bbu01.eps

                                                                                      Figure 1.3140R Setup
This datasheet has been downloaded from:
            datasheet.eeworld.com.cn

                 Free Download
           Daily Updated Database
      100% Free Datasheet Search Site
  100% Free IC Replacement Search Site
     Convenient Electronic Dictionary

               Fast Search System
             www.EEworld.com.cn

                                                 All Datasheets Cannot Be Modified Without Permission
                                                                Copyright Each Manufacturing Company

3140R器件购买:

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2017 EEWORLD.com.cn, Inc. All rights reserved