电子工程世界电子工程世界电子工程世界

型号

产品描述

搜索

179-NISTW/DATA

器件型号:179-NISTW/DATA
器件类别:测试和测量    万用表与电压表   
文件大小:528.71KB,共4页
厂商名称:FLUKE [FLUKE]
标准:
下载文档

器件描述

digital multimeters trms multimeter calibtrated W/data

参数

Manufacturer: Fluke
Product Category: Digital Multimeters
RoHS: Yes
Product: Multimeters
Type: Hand Held
Accuracy: +/- 1 %
Voltage Range: 0.1 mVAC to 1000 VAC
Resistance Range: 0.1 Ohms to 50 Mohms
Capacitance Range: 1 nF to 9999 uF
Display Count: 6000
Frequency: 2 Hz to 50 kHz
Ranging: Auto, Manual
True RMS: True RMS
Data Hold: Data Hold
AC Supply Voltage: 1000 V
Brand: Fluke
Current Rating: 10 mA
Series: 170
Voltage Rating: 1000 VAC
Part # Aliases: 1256480

179-NISTW/DATA器件文档内容

Page 1 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
2000 Fluke Corporation
Rev. A-11/2000

                  Fluke 170 Series
True-rms Digital Multimeter

                Extended Specifications

Model Differences

(all other specifications are the same for each model)

Feature                            175             Model                 179
                                  0.15%             177                    
Backlight                                                                  
Temperature Measurement                              
Basic dc voltage accuracy                                               0.09%
                                                   0.09%

Nominal Specifications

Function                                                     Absolute Range or Description
AC Voltage, True-rms              0.1 mV to 1000V (1 kHz)
DC Voltage
Continuity                        0.1 mV to 1000V
                                  Beeper guaranteed on < 25 , guaranteed off > 250 ;
Resistance                        detects opens or shorts of 250 s or longer.
Diode Test
Capacitance                       0.1 to 50.00 M
AC Current, True-rms
DC Current                        2.400V
Frequency
Temperature (179 Only)            1 nF to 9999 uF
Basic dc voltage accuracy
Basic ac voltage accuracy         0.01 mA to 10.00A (20.00 A over-range for 30 seconds)

                                  0.01 mA to 10.00A (20.00 A over-range for 30 seconds)

                                  2 Hz to 50 kHz

                                  -40 C to +400 C; -40 F to +752 F

                                  0.15% (175)                           0.09% (177 & 179)

                                  1.0%

Features                          Description

  Feature                         6000 counts, updates 4 x second
  Digital Display                 33 segments, updates 40 x second
  Analog Bargraph Display
  Backlight (177 & 179 only)      Automatically turns off after 2 minutes to save battery life
  HOLD & Auto HOLD                The timeout feature can be disabled with a power-up option

  MIN MAX AVG                     HOLD: freezes the display at the push of a button
  Manual or auto ranging
  Fast continuity/open detection  Auto HOLD: Display holds present reading until it detects new stable input,
  Test lead alert                 then the meter beeps and displays new reading
  Power-up options
  Closed-case calibration         Minimum, maximum, and average reading memory
  Probe holders                   In auto range, the meter selects the range with the best resolution for
                                  the present measurement value
  Battery access door             The beeper sounds with a stretched pulse for opens or shorts as
  High-impact overmold case       brief as 250 s
                                  The message "LEAd" appears briefly on the display when the rotary switch
                                  is moved to or from any A (Amps) position

                                  (1) Turn on all LCD segments, (2) Disable beeper, (3) Disable sleep mode,
                                  (4) Enable smoothing, (5) Disable backlight timeout (Models 177 & 179 only)

                                  No internal adjustments needed

                                  The instrument comes with built-in probe holders for probe storage and
                                  for convenience when making measurements

                                  Battery replacement without voiding calibration

                                  Integrated overmolded protection provides superior impact protection for
                                  your meter
Page 2 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
2000 Fluke Corporation
Rev. A-11/2000

General Specifications

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 C to 28 C, with relative
humidity of 0 % to 75 %. Accuracy specifications take the form of: ( [ % of Reading ] + [ Counts ] )

Maximum voltage between any     1000V DC or AC RMS
terminal and earth ground

Surge Protection                8 kV peak per IEC 61010

Fuse for mA inputs              440 mA, 1000 V FAST Fuse

Fuse for A input                11A, 1000V FAST Fuse

Display                         Digital: 6,000 counts, updates 4/sec
                                Bar Graph: 33 segments, updates 40/sec
                                Frequency: 9,999 counts
                                Capacitance: 9,999 counts

Altitude                        Operating: 2000 m; Storage: 12000 m
Temperature
                                Operating: -10 C to +50 C
                                Storage: -30 C to +60 C

Temperature coefficient         0.1 X (specified accuracy / C)
                                (< 18 C or > 28 C)

Electromagnetic Compatibility   In an RF field of 3 V/M, accuracy = specified accuracy except in
(EN 61326-1:1997)               temperature: specified accuracy 5 C, 9 F

Relative Humidity               0 % to 90 % @ 0 C to 35 C;
                                0 % to 70 % @ 36 C to 50 C

Relative Humidity in            0 % to 80 % @ 0C to 35C;
50 M Range                      0 % to 70 % @ 36 C to 50 C

Battery Life                    Alkaline: ~200 hrs typical

Size, with Holster (H x W x L)  4.3 cm x 9 cm x 18.5 cm

Weight                          420g

Safety Compliances              ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage
                                Category III, 600 V Overvoltage Category IV

Certifications                  CSA, TV (EN61010),
                                Australian (N10140)
Page 3 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
2000 Fluke Corporation
Rev. A-11/2000

Detailed Specifications

                                                               Accuracy ( [ % of Reading ] + [ Counts ] )

Function                     Range 1              Resolution Model 175       Model 177     Model 179

AC Volts 2                   600.0 mV             0.1 mV       1.0 % + 3     1.0 % + 3     1.0 % + 3
                             6.000V               0.001V
DC mV                        60.00V               0.01V        (45 Hz to 500 Hz) (45 Hz to 500 Hz) (45 Hz to 500 Hz)
DC Volts                     600.0V               0.1V
                             1000V                1V           2.0 % + 3     2.0 % + 3     2.0 % + 3

                             600.0 mV             0.1 mV       (500 Hz to 1 kHz) (500 Hz to 1 kHz) (500 Hz to 1 kHz)

                             6.000V               0.001V       0.15 % + 2    0.09 % + 2    0.09 % + 2
                             60.00V               0.01V
                             600.0V               0.01V        0.15 % + 2    0.09 % + 2    0.09 % + 2

                             1000V                1V           0.15 % + 2    0.1 % + 2     0.1 % + 2

Continuity                   600                  1            Meter beeps at < 25 , beeper turns off at > 250 ;

                                                               detects opens or shorts of 250 ms or longer.

Ohms                         600.0                0.1          0.9 % + 2     0.9 % + 2     0.9 % + 2
                             6.000 k              0.001 k      0.9 % + 1     0.9 % + 1     0.9 % + 1
                             60.00 k              0.01 k       0.9 % + 1     0.9 % + 1     0.9 % + 1
                             600.0 k              0.1 k        0.9 % + 1     0.9 % + 1     0.9 % + 1
                             6.000 M              0.001 M      0.9 % + 1     0.9 % + 1     0.9 % + 1
                             50.00 M              0.01 M       1.5 % + 3     1.5 % + 3     1.5 % + 3

Diode test                   2.400V               0.001V       1%+2

Capacitance                  1000 nF              1 nF         1.2 % + 2     1.2 % + 2     1.2 % + 2
                             10.00 F             0.01 F      1.2 % + 2     1.2 % + 2     1.2 % + 2
                             100.0 F             0.1 F       1.2 % + 2     1.2 % + 2     1.2 % + 2
                             9999 F 3            1 F         10 % typical  10 % typical  10 % typical

AC Amps                      60.00 mA             0.01 mA      1.5 % + 3     1.5 % + 3     1.5 % + 3
(True-rms)                   400.0 mA             0.1 mA
                             (600 mA for 18 hrs)  0.001A
(45 Hz to 1 kHz)             6.000A
                             10.00A               0.01A
                             (20A for 30s)

DC Amps                      60.00 mA             0.01 mA      1.0 % + 3     1.0 % + 3     1.0 % + 3
                             400.0 mA             0.1 mA
                             (600 mA for 18 hrs)
                             6.000A               0.001A
                             10.00A               0.01A
                             (20A for 30s)

Hz                           99.99 Hz             0.01 Hz      0.1 % + 1     0.1 % + 1     0.1 % + 1
(AC- or DC-                  999.9 Hz             0.1 Hz
                             9.999 kHz            0.001 kHz
coupled,                     99.99 kHz            0.01 kHz
V or A 4, 5 input )

Temperature                  -40 C to +400 C 0.1 C          NA            NA            1 % + 1.0 C

                             -40 F to +752 F    0.1 F                                   1 % + 1.8 F

MIN MAX AVG                  For DC functions, accuracy is the specified of the measurement function 12 counts for changes

                             longer than 275 ms in duration.
                             For AC functions, accuracy is the specified of the measurement function 40 counts for changes
                             longer than 1.2 s in duration.

1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.                                    Fluke. Keeping your world
2. Crest factor of  3 at full scale up to 500 V, decreasing linearly to crest factor  1.5 at 1000 V.                                     up and running.
3. In the 9999 F range for measurements to 1000 F, the measurement accuracy is 1.2 % for all models.
4. In mA and A ranges, frequency measurement is specified to 30 kHz.

5. Frequency < 10 kHz are not specified in 600 mV AC, 60 mA AC, and 6 A AC ranges.

Function Overload Protection 1 Input Impedance                 Common Mode Rejection Ratio Normal Mode Rejection              Fluke Corporation
                                            (Nominal)          (1 k Unbalanced)                                               PO Box 9090, Everett, WA USA 98206

Volts AC 1000 V RMS or DC > 10 M < 100 pF                      > 60 dB @ DC, 50 or 60 Hz                                      Fluke Europe B.V.
Volts DC 1000 V RMS or DC > 10 M <100 pF                                                                                      PO Box 1186, 5602 BD
                                                               >120 dB @ DC, 50 or 60 Hz   > 60 dB @ 50 Hz or 60 Hz           Eindhoven, The Netherlands
                                            Open Circuit Test
                                            Voltage            Full Scale Voltage To:      Short Circuit Current              For more information call:
                                                                                                                              U.S.A. (800) 443-5853 or
Ohms 1000V RMS or DC < 1.5 V DC                                6.0 M         50 M                                             Fax (425) 446-5116
                                                                                                                              Europe (31 40) 2 675 200 or
Diode test 1000V RMS or DC 2.4 to 3.0 V DC                     < 600 mV DC < 1.5 V DC      < 500 A                           Fax (31 40) 2 675 222
                                                                                                                              Canada (905) 890-7600 or
1. 10 7 V-Hz maximum.                                          2.4 V DC                    < 1.2 mA typical                   Fax (905) 890-6866
                                                                                                                              Other countries (425) 446-5500 or
                                                                                                                              Fax (425) 446-5116
                                                                                                                              Web access: http://www.fluke.com

                                                                                                                              2001 Fluke Corporation. All rights reserved.
                                                                                                                              Specifications subject to change without notice.
                                                                                                                              Printed in U.S.A. 3/2001

                                                                                                                              Printed on recycled paper.
Mouser Electronics

Authorized Distributor

Click to View Pricing, Inventory, Delivery & Lifecycle Information:

Fluke:

  179-NISTW/DATA
This datasheet has been downloaded from:
            datasheet.eeworld.com.cn

                 Free Download
           Daily Updated Database
      100% Free Datasheet Search Site
  100% Free IC Replacement Search Site
     Convenient Electronic Dictionary

               Fast Search System
             www.EEworld.com.cn

                                                 All Datasheets Cannot Be Modified Without Permission
                                                                Copyright Each Manufacturing Company

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2017 EEWORLD.com.cn, Inc. All rights reserved