                                                                 8845A/8846A
                                                                                                  Digital Multimeter

                                                                                                  Calibration Manual

January 2007, Rev. 1, 11/07

© 2007 Fluke Corporation, All rights reserved.  Specifications subject to change without notice.

All product names are trademarks of their respective companies.
                       LIMITED WARRANTY AND LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and

service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and

services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of

a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in

Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal

conditions of operation or handling. Fluke warrants that software will operate substantially in accordance

with its functional specifications for 90 days and that it has been properly recorded on non-defective media.

Fluke does not warrant that software will be error free or operate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers

only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is

available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the

applicable international price. Fluke reserves the right to invoice Buyer for importation costs of

repair/replacement parts when product purchased in one country is submitted for repair in another country.

Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair,

or replacement of a defective product which is returned to a Fluke authorized service center within the

warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center to obtain return

authorization information, then send the product to that service center, with a description of the difficulty,

postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following

warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke

determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal

condition of operation or handling, including overvoltage failures caused by use outside the product’s

specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair

costs and obtain authorization before commencing the work. Following repair, the product will be returned to

the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges

(FOB Shipping Point).

THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER

WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY

OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE

FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES,

INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or

limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not

apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other

decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other

provision.

                       Fluke Corporation                   Fluke Europe B.V.

                       P.O. Box 9090                       P.O. Box 1186

                       Everett, WA  98206-9090             5602 BD Eindhoven

                       U.S.A.                              The Netherlands

11/99

To register your product online, visit register.fluke.com
                Table of Contents

Chapter  Title                                                                                                                 Page

1        Introduction and Specifications.........................................................                              1-1

         Introduction........................................................................................................  1-3

         Safety Information .............................................................................................      1-3

         Symbols .........................................................................................................     1-4

         General Safety Summary...............................................................................                 1-4

         Description of IEC 61010 Measurement Categories .........................................                             1-6

         Organization of the Calibration Manual ............................................................                   1-7

         Chapter 1 – Introduction and Specifications .................................................                         1-7

         Chapter 2 – General Maintenance .................................................................                     1-7

         Chapter 3 – Performance Test and Calibration..............................................                            1-7

         Chapter 4 – List of Replaceable Parts ...........................................................                     1-7

         Operating Instructions........................................................................................        1-7

         Accessories ........................................................................................................  1-7

         General Specifications .......................................................................................        1-9

         Power .............................................................................................................   1-9

         Dimensions ....................................................................................................       1-9

         Display...........................................................................................................    1-9

         Environment ..................................................................................................        1-9

         Safety .............................................................................................................  1-9

         EMC ..............................................................................................................    1-9

         Triggering ......................................................................................................     1-9

         Memory .........................................................................................................      1-10

         Math Functions..............................................................................................          1-10

         Electrical........................................................................................................    1-10

         Remote Interfaces ..........................................................................................          1-10

         Warranty ........................................................................................................     1-10

         Electrical Specifications ....................................................................................        1-10

         DC Voltage Specifications ............................................................................                1-10

         Input Characteristics..................................................................................               1-11

         8846A Accuracy........................................................................................                1-11

         8845A Accuracy........................................................................................                1-11

         Additional Errors.......................................................................................              1-11

         AC Voltage Specifications ............................................................................                1-11

         Input Characteristics..................................................................................               1-12

         8846A Accuracy........................................................................................                1-12

         i
8845A/8846A

Calibration Manual

                       8845A Accuracy........................................................................................                  1-13

                       Additional Low Frequency Errors.............................................................                            1-13

                       Resistance ......................................................................................................       1-14

                       Input Characteristics..................................................................................                 1-14

                       8846A Accuracy........................................................................................                  1-14

                       8845A Accuracy........................................................................................                  1-15

                       Additional Ohms Errors ............................................................................                     1-15

                       DC Current ....................................................................................................         1-15

                       Input Characteristics..................................................................................                 1-15

                       Accuracy (8846A).....................................................................................                   1-16

                       Accuracy (8845A).....................................................................................                   1-16

                       Additional Current Errors .........................................................................                     1-16

                       AC Current ....................................................................................................         1-17

                       Input Characteristics..................................................................................                 1-17

                       8846A Accuracy........................................................................................                  1-17

                       8845A Accuracy........................................................................................                  1-18

                       Additional Low Frequency Errors.............................................................                            1-19

                       Frequency ......................................................................................................        1-19

                       8846A Accuracy........................................................................................                  1-19

                       8845A Accuracy........................................................................................                  1-20

                       Gate Time vs. Resolution..........................................................................                      1-20

                       Additional Low Frequency Errors.............................................................                            1-20

                       Capacitance (8846A Only) ............................................................................                   1-20

                       Temperature (8846A only) ............................................................................                   1-21

                       Continuity ......................................................................................................       1-21

                       Diode Test .....................................................................................................        1-21

                       Measurement Rates (IEEE488[4]).................................................................                         1-22

                       Measurement Uncertainty..................................................................................               1-22

                       Interpreting Accuracy Specifications.................................................................                   1-23

                       24-Hour Accuracy .........................................................................................              1-23

                       90-Day and 1-Year Accuracy ........................................................................                     1-23

                       Temperature Coefficients ..............................................................................                 1-23

                       Configuring for Highest Accuracy Measurements ............................................                              1-23

                       DC Voltage, DC Current, and Resistance Measurements .............................                                       1-23

                       AC Voltage and AC Current Measurements: ................................................                                1-23

                       Frequency and Period Measurements:...........................................................                           1-23

                    2  General Maintenance ..........................................................................                          2-1

                       Introduction........................................................................................................    2-3

                       Warranty Repairs and Shipping Information .....................................................                         2-3

                       General Maintenance Information .....................................................................                   2-3

                       Required Equipment......................................................................................                2-3

                       Power Requirements......................................................................................                2-3

                       Static Safe Handling ......................................................................................             2-3

                       Cleaning .............................................................................................................  2-4

                       Fuse Replacement ..............................................................................................         2-4

                       Line-Power Fuse............................................................................................             2-4

                       Current-Input Fuses .......................................................................................             2-5

                       If the Meter Does Not Turn On..........................................................................                 2-7

                       Display Tests......................................................................................................     2-7

                       Disassembly Procedures ....................................................................................             2-7

                       General Disassembly .....................................................................................               2-7

                       Main Chassis Disassembly ............................................................................                   2-8

                       Front Panel Disassembly ...............................................................................                 2-8

                       ii
                                                                                                                              Contents  (continued)

   Assembly Procedures.........................................................................................               2-8

3  Performance Test and Calibration .....................................................                                     3-1

   Introduction........................................................................................................       3-3

   Required Equipment ..........................................................................................              3-3

   Test Considerations............................................................................................            3-5

   Performance Tests..............................................................................................            3-5

      Volts DC Verification....................................................................................               3-5

      Volts DC Ratio Verification ..........................................................................                  3-8

      Volts AC and Frequency Verification ...........................................................                         3-9

      4-Wire Ohms Verification .............................................................................                  3-14

      2-Wire Ohms Verification .............................................................................                  3-16

      2X4 Test Lead Verification Steps .................................................................                      3-18

      Rear Panel Terminal Verification Steps ........................................................                         3-18

      Capacitance Verification Steps (8846A only) ...............................................                             3-21

      DC Current Verification Steps ......................................................................                    3-22

      AC Current Verification Steps ......................................................................                    3-25

   Adjustment (Calibration) ...................................................................................               3-27

      Unlocking the Meter for Adjustments (Calibration) .....................................                                 3-28

      Unlocking the Meter for Adjustments Over a Remote Interface...................                                          3-28

      Changing the Calibration Password ..............................................................                        3-28

      Resetting the Calibration Password...............................................................                       3-28

      Changing the Calibration Date Remotely......................................................                            3-29

      Equipment for Calibration .............................................................................                 3-29

      Adjustment Process .......................................................................................              3-30

      Aborting a Calibration Process......................................................................                    3-36

      Sample Adjustment Program.........................................................................                      3-36

4  List of Replaceable Parts....................................................................                              4-1

   Introduction........................................................................................................       4-3

   How to Obtain Parts...........................................................................................             4-3

   How to Contact Fluke ........................................................................................              4-3

   Parts ...................................................................................................................  4-4

Appendices

   A  Verification Forms ......................................................................................               A-1

   B  Example Adjustment Program ....................................................................                         B-1

            iii
8845A/8846A

Calibration Manual

                    iv
              List of Tables

Table  Title                                                                                                                     Page

1-1.   Safety and Electrical Symbols................................................................................             1-4

1-2.   Safety Information .................................................................................................      1-5

1-3.   Accessories.............................................................................................................  1-7

2-1.   Line Voltage to Fuse Rating...................................................................................            2-5

3-1.   Required Test Equipment.......................................................................................            3-3

3-2.   8846A DC Volts Verification Steps.......................................................................                  3-6

3-3.   8845A DC Volts Verification Steps.......................................................................                  3-7

3-4.   8846A DC Volts Ratio Verification Steps .............................................................                     3-9

3-5.   8845A DC Volts Ratio Verification Steps .............................................................                     3-9

3-6.   8846A AC Volts Verification Steps.......................................................................                  3-11

3-7.   8845A AC Volts Verification Steps.......................................................................                  3-12

3-8.   8845A/8846A AC Volts Frequency Verification Steps .........................................                               3-13

3-9.   8846A 4-Wire Ohms Verification Steps ................................................................                     3-15

3-10.  8845A 4-Wire Ohms Verification Steps ................................................................                     3-15

3-11.  8846A 2-Wire Ohms Verification Steps ................................................................                     3-17

3-12.  8845A 2-Wire Ohms Verification Steps ................................................................                     3-17

3-13.  8846A Rear-Panel Terminal Verification Steps (Optional Test) ...........................                                  3-20

3-14.  8845A Rear-Panel Terminal Verification Steps (Optional Test) ...........................                                  3-20

3-15.  8846A Capacitance Verification Steps ..................................................................                   3-21

3-16.  8846A DC Current Verification Steps ...................................................................                   3-23

3-17.  8845A DC Current Verifications Steps..................................................................                    3-24

3-19.  8845A AC Current Verification Steps ...................................................................                   3-27

3-20.  8845A/8846A Adjustment Steps............................................................................                  3-31

4-1.   Replaceable Parts ...................................................................................................     4-4

A-1.   Blank Verification Record for 90-Day Specifications ...........................................                           A-2

A-2.   Blank Verification Record for 1-Year Specifications ............................................                          A-10

       v
8845A/8846A

Calibration Manual

                    vi
               List of Figures

Figure  Title                                                                                                                  Page

1-1.    IEC 61010 Measurement Category (CAT) Levels.................................................                           1-6

2-1.    Line Fuse Replacement ..........................................................................................       2-5

2-2.    Current Input Fuse Replacement............................................................................             2-6

3-1.    DC Volts Test Equipment Setup with 5520A ........................................................                      3-6

3-2.    DC Volts Ratio Test Equipment Setup with 5520A and 5720A ............................                                  3-8

3-3.    AC Volts Test Equipment Setup with 5520A ........................................................                      3-10

3-4.    4-Wire Ohms Test Equipment Setup .....................................................................                 3-14

3-5.    2-Wire Ohms Test Equipment Setup .....................................................................                 3-16

3-6.    Rear-Panel Terminals Equipment Setup ................................................................                  3-19

3-7.    Capacitance Equipment Setup................................................................................            3-21

3-8.    100 mA DC Current Equipment Setup ..................................................................                   3-22

3-9.    AC Current Equipment Setup ................................................................................            3-25

3-10.   Calibration Jumper Location..................................................................................          3-29

4-1.    Final Assembly.......................................................................................................  4-7

        vii
8845A/8846A

Calibration Manual

                    viii
       Chapter 1

Introduction and Specifications

Title                                                                                                                 Page

Introduction........................................................................................................  1-3

Safety Information .............................................................................................      1-3

Symbols .........................................................................................................     1-4

General Safety Summary...............................................................................                 1-4

Description of IEC 61010 Measurement Categories .........................................                             1-6

Organization of the Calibration Manual ............................................................                   1-7

Chapter 1 – Introduction and Specifications .................................................                         1-7

Chapter 2 – General Maintenance .................................................................                     1-7

Chapter 3 – Performance Test and Calibration..............................................                            1-7

Chapter 4 – List of Replaceable Parts ...........................................................                     1-7

Operating Instructions........................................................................................        1-7

Accessories ........................................................................................................  1-7

General Specifications .......................................................................................        1-9

Power .............................................................................................................   1-9

Dimensions ....................................................................................................       1-9

Display...........................................................................................................    1-9

Environment ..................................................................................................        1-9

Safety .............................................................................................................  1-9

EMC ..............................................................................................................    1-9

Triggering ......................................................................................................     1-9

Memory .........................................................................................................      1-10

Math Functions..............................................................................................          1-10

Electrical........................................................................................................    1-10

Remote Interfaces ..........................................................................................          1-10

Warranty ........................................................................................................     1-10

Electrical Specifications ....................................................................................        1-10

DC Voltage Specifications ............................................................................                1-10

AC Voltage Specifications ............................................................................                1-11

Resistance ......................................................................................................     1-14

DC Current ....................................................................................................       1-15

AC Current ....................................................................................................       1-17

Frequency ......................................................................................................      1-19

Capacitance (8846A Only) ............................................................................                 1-20

Temperature (8846A only) ............................................................................                 1-21

Continuity ......................................................................................................     1-21

                                                                                                                            1-1
8845A/8846A

Calibration Manual

                    Diode Test .....................................................................................................  1-21

                    Measurement Rates (IEEE488[4]).................................................................                   1-22

                    Measurement Uncertainty..................................................................................         1-22

                    Interpreting Accuracy Specifications.................................................................             1-23

                    24-Hour Accuracy .........................................................................................        1-23

                    90-Day and 1-Year Accuracy ........................................................................               1-23

                    Temperature Coefficients ..............................................................................           1-23

                    Configuring for Highest Accuracy Measurements ............................................                        1-23

                    DC Voltage, DC Current, and Resistance Measurements .............................                                 1-23

                    AC Voltage and AC Current Measurements: ................................................                          1-23

                    Frequency and Period Measurements:...........................................................                     1-23

1-2
                                                                  Introduction and Specifications        1

                                                                                           Introduction

Introduction

The 8845A and 8846A are 6-1/2 digit, dual-display multimeters designed for bench-top,

field service, and system applications. Their full complement of measurement functions

plus its RS-232, IEEE 488, and Ethernet Remote Interfaces makes these multimeters

ideal candidates for precision manual measurements and use in automated systems. For

portability, these multimeters include a carrying handle that also serves as a bail for

bench top operation.

There are a few additional features in the 8846A that are not present in the 8845A. These

features will be identified with the annotation of “8846A Only” by each feature that is

found only in that model. Separate specification tables are also used to clarify the

differences between these two models.

The following is a list of some of the features and functions:

•  Bright, large-digit, wide-viewing-angle display

•  Dual display for displaying two properties of an input signal (e.g., ac voltage in one

   display and frequency in the other).

•  Remote operation via IEEE 488, RS-232, and Ethernet interface.

•  Trigger in and measurement-complete out

•  Front panel USB port for optional memory (8846A Only)

•  6-1/2 digit resolution

•  Half-rack width

•  True rms ac

•  2 and 4-wire resistance measurements

•  Extended 10 Ω and 1 GΩ ranges (8846A Only)

•  Frequency measurements to 300 kHz (8846A to 1 MHz)

•  Capacitance measurements (8846A Only)

•  Temperature measurement (8846A Only)

•  10 A current capability

•  Decibels (dB and dBm) with variable reference impedance and audio power

   measurement capability

•  Input terminals on both front and rear panels of the meter

•  Closed-case calibration (no internal calibration adjustments)

This calibration manual focuses on performance verification and calibration of the Fluke

8845A and 8846A Digital Multimeters (hereafter referred to as the Meter).

Safety Information

This section addresses safety considerations and describes symbols that may appear on

the Meter or in the manual.

A Warning statement identifies conditions or practices that could result in injury or

death.

A Caution statement identifies conditions or practices that could result in damage to the

Meter or equipment to which it is connected.

                                                                                                         1-3
8845A/8846A

Calibration Manual

                                                         XW Warning

                        To avoid electric shock, personal injury, or death, carefully read

                        the information under “Safety Information” before attempting to

                        install, use, or service the Meter.

     Symbols

                        Table 1-1 is a list of safety and electrical symbols that appear on the Meter or in this

                        manual.

                                             Table 1-1. Safety and Electrical Symbols

             Symbol               Description                    Symbol                    Description

             W          Risk of danger. Important                O       Display ON / OFF

                        information. See manual

             X          Hazardous voltage. Voltage > 30       V  J       Earth ground

                        dc or ac peak might be present

                    B   AC (Alternating Current)                 E       Capacitance

                    F   DC (Direct Current)                      G       Diode

                    D                                            I       Fuse

                    or  AC or DC (Alternating or Direct

                    C   Current)                                 Y       Digital signal

             R          Continuity test or continuity beeper     U       Maintenance or Service

                        tone

                    Y   Potentially hazardous voltage            CAT II  IEC 61010 Overvoltage (installation or

                                                                         measurement) Category 2.

                    T   Double insulated                         <       Recycle

             h          Static awareness. Static discharge       ~       Do not dispose of this product as unsorted

                        can damage part(s)                               municipal waste. Go to Fluke’s website for

                                                                         recycle information.

     General Safety Summary

                        This instrument has been designed and tested in accordance with the European standard

                        publication EN 61010-1:2001 and U.S. / Canadian standard publications UL 61010-1A1

                        and CAN/CSA-C22.2 No.61010.1. The Meter has been supplied in a safe condition.

                        This manual contains information and warnings that must be observed to keep the

                        instrument in a safe condition and ensure safe operation.

                        To use the Meter correctly and safely, read and follow the precautions in Table 1-2, and

                        follow all the safety instructions or warnings given throughout this manual that relate to

                        specific measurement functions. In addition, follow all generally accepted safety

                        practices and procedures required when working with and around electricity.

1-4
                                                                   Introduction and Specifications      1

                                                                                    Safety Information

                           Table 1-2. Safety Information

                                        XW Warning

To avoid possible electric shock, personal injury, or death, read the following before using

the Meter:

• Use the Meter only as specified in this manual, or the protection provided by the Meter

   might be impaired.

• Do not use the Meter in wet environments.

•  Inspect the Meter before using it.  Do not use the Meter if it appears damaged.

• Inspect the test leads before use. Do not use them if insulation is damaged or metal is

   exposed. Check the test leads for continuity. Replace damaged test leads before using

   the Meter.

• Verify the Meter's operation by measuring a known voltage before and after using it. Do

   not use the Meter if it operates abnormally. Protection may be impaired. If in doubt, have

   the Meter serviced.

• Whenever it is likely that safety protection has been impaired, make the Meter inoperative

   and secure it against any unintended operation.

• Servicing of the Meter should be performed by qualified service personnel.

• Do not apply more than the rated voltage, as marked on the Meter, between the terminals

   or between any terminal and earth ground.

• While in IEC Measurement Category II environments, do not apply voltages above 600 V

   ac to the input of the Meter. See “Description of IEC 61010 Measurement Categories”

   later in this chapter.

• Always use the power cord and connector appropriate for the voltage and outlet of the

   country or location in which you are working.

• Always use a power cord with a ground connection and ensure the ground is properly

   connected to the power distribution system.

• Remove test leads from the Meter before opening the case.

• Never remove the cover or open the case of the Meter without first removing it from the

   main power source.

• Use caution when working with voltages above 30 V ac rms, 42 V ac peak, or 42 V dc.

   These voltages pose a shock hazard.

• Use only the replacement fuse(s) specified by the manual.

• Use the proper terminals, function, and range for your measurements.

• Do not operate the Meter around explosive gas, vapor, or dust.

• When using probes, keep your fingers behind the finger guards.

• When making electrical connections, connect the common test lead before connecting

   the live test lead; when disconnecting, disconnect the live test lead before disconnecting

   the common test lead.

• Disconnect circuit power and discharge all high-voltage capacitors before testing

   resistance, continuity, diodes, or capacitance.

• Before measuring current, check the Meter's fuses and turn OFF power to the circuit

   before connecting the Meter to the circuit.

• When servicing the Meter, use only specified replacement parts.

                                                                                                        1-5
8845A/8846A

Calibration Manual

                                                          WCaution

                    To prevent damage to the Meter, do not change the position of

                    the Front/Rear switch while signals are applied to either the

                    front or rear input terminals.

     Description of IEC 61010 Measurement Categories

                    The IEC 61010 safety standard defines four Overvoltage (Installation) Categories (CAT         I

                    to CAT IV) based on the magnitude of danger from transient impulses as shown in

                    Figure 1-1.

                    CAT I        CAT II                   CAT III            CAT IV

                    Electronic   Appliances,     MC Panels, etc.

                                 PCs, TVs

                                                          Service

                                                          Entrance

                                                                    Meter

                                                          Service                                 Outbuilding

                                                          Entrance  Meter

                                                                             Underground Service

                                                          Service            Transformer          Outbuilding

                                                          Entrance                        Meter

                                                                             Underground Service

                                                                                                               cat_levels.eps

                                 Figure 1-1. IEC 61010 Measurement Category (CAT) Levels

                    The IEC 61010 Measurement CAT level indicates the level of protection the instrument

                    provides against impulse withstand voltage.

                    CAT I equipment is designed to protect against transients from high-voltage, low-energy

                    sources, such as electronic circuits or a copy machine.

                    CAT II equipment is designed to protect against transients from energy-consuming

                    equipment supplied from the fixed installation, such as TVs, PCs, portable tools, and

                    other household appliances.

                    CAT III equipment is designed to protect against transients in equipment in fixed

                    equipment installations, such as distribution panels, feeders and short branch circuits, and

                    lighting systems in large buildings.

                    CAT IV equipment is designed to protect against transients from the primary supply

                    level, such as an electricity meter or an overhead or underground utility service.

1-6
                                                                        Introduction and Specifications   1

                                                                  Organization of the Calibration Manual

Organization of the Calibration Manual

       This calibration manual is divided into the following chapters:

Chapter 1 – Introduction and Specifications

       This chapter introduces the Fluke 8845A and 8846A Digital Multimeters, describing their

       features, and accessories. This chapter also discusses use of the Calibration Manual and

       the various conventions used in describing the meter’s circuitry and presents a complete

       set of specifications.

Chapter 2 – General Maintenance

       Chapter 2 provides maintenance information covering handling, cleaning, and fuse

       replacement. Access and reassembly procedures are also explained in this chapter.

Chapter 3 – Performance Test and Calibration

       This chapter provides performance verification procedures related to the specifications

       presented in Chapter 1. To maintain these specifications, a full adjustment/calibration

       procedure is also presented.

Chapter 4 – List of Replaceable Parts

       Chapter 4 includes parts lists for all standard assemblies and information on how and

       where to order parts.

Operating Instructions

       Full operating instructions are provided in the Fluke 8845A/8846A Users Manual.

       Reference to these instructions may be necessary during some of the maintenance and

       repair procedures presented in this Calibration Manual.

Accessories

       Table 1-3 lists the available accessories for the 8845A and 8846A.

                                     Table 1-3. Accessories

Model/Fluke PN                                       Description

TL71            Test Lead Set, Premium DMM

TL910           Precision Electronic Probe Set

TL80A           Basic Electronic DMM Test Set

8845A-TPIT      Test Probe Ic Tip (Set Red & Black)

8845A-EFPT      Extended Fine Pt Tip Adap Set (Set   Red/Black)

884X-SHORT      4-Wire short

TL2X4W-PTII     2x4 Wire Ohms 1000V Test lead

884X-RTD        100 Ohm RTD Temperature Probe

884X-512M       USB Memory 512M

                                                                                                          1-7
8845A/8846A

Calibration Manual

                                         Table 1-3. Accessories (cont)

             Model/Fluke PN                                         Description

             884X-1G         USB Memory 1 GB

             884X-USB        USB to RS232 cable adapter

             884X-ETH        Ethernet Interface Cable

             RS43            RS232 Cable (RS43, 2m)

             Y8021           IEEE488 cable (1m)

             Y8022           IEEE488 cable (2m)

             884X-CASE       Black Case

             FVF-UG          FlukeView Forms Software Upgrade - NO cable

             FVF-SC4         Flukeview Forms W/Cable For 8845/8846

             FVF-SC5         Flukeview Forms -Basic For 8845/8846

             Y8846S          Rack Mount Kit 8845A & 8846A Single

             Y8846D          Rack Mount Kit 8845A & 8846A Dual

             FUSE            Fuse,11A,1000V,Fast.406inx1.5in,Bulk

             FUSE            Fuse,.440 mA,1000V,Fast,.406X1.375,Bulk

             2132558         Calibration, Traceable, w/ Data

             1259800         Calibration, Traceable, w/o Data

             1256480         Calibration, Z540 Traceable, w/ Data

             1258910         Calibration, Z540 Traceable, w/o Data

             1256990         Calibration, Accredited

             1024830         Agreement, Extended Warranty

             2426684         Agreement, Calibration, Traceable, w/ Data

             1028820         Agreement, Calibration, Traceable, w/o Data

             1259170         Agreement, Calibration, Z540 Traceable, w/ Data

             1258730         Agreement, Calibration, Z540 Traceable, w/o Data

             1259340         Agreement, Calibration, Accredited

             2441827         Agreement, Calibration, Primary Standards Lab

             1540600         Agreement, Calibration, Artifact

1-8
                                                                                                                      Introduction and Specifications       1

                                                                                                                                    General Specifications

General Specifications

Power

Voltage

100 V Setting ......................................................       90 V to 110 V

120 V Setting ......................................................       108 V to 132 V

220 V Setting ......................................................       198 V to 242 V

240 V Setting ......................................................       216 V to 264 V

Frequency ...............................................................  47 Hz to 440 Hz. Automatically  sensed     at  power-on.

Power Consumption................................................          28 VA peak (12 Watt average)

Dimensions

Height...................................................................... 88 mm (3.46 in.)

Width....................................................................... 217 mm (8.56 in.)

Depth ...................................................................... 297 mm (11.7 in.)

Weight..................................................................... 3.6 kg (8.0 lbs)

Shipping Weight...................................................... 5.0 kg (11.0 lbs)

Display

Vacuum Fluorescent Display, dot matrix

Environment

Temperature

Operating ............................................................     0 °C to 55 °C

Storage ...............................................................    -40 °C to 70 °C

Warm Up.............................................................       1 hour to full uncertainty specifications

Relative Humidity (non-condensing)

Operating ............................................................     0 °C to 28 °C <90 %

                                                                           28 °C to 40 °C <80 %

                                                                           40 °C to 55 °C <50 %

Storage ...............................................................    -40 °C to 70 °C <95 %

Altitude

Operating ............................................................     2,000 Meters

Storage ...............................................................    12,000 Meters

Vibration and Shock................................................        Complies with Mil-T-28800F Type III, Class     5  (Sine  only)

Safety

Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT                                       II  600V

EMC

Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has

shown susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz.

Triggering

Samples per Trigger ........................................... 1 to 50,000

Trigger Delay ...................................................... 0 s to 3600 s; in 10 µS increments

External Trigger Delay ........................................ <1 mS

External Trigger Jitter ......................................... <500 µS

Trigger Input ....................................................... TTL Levels

Trigger Output..................................................... 5 V maximum (open collector)

                                                                                                                                                            1-9
8845A/8846A

Calibration Manual

      Memory

             8845A ................................................................. 10,000 measurements, internal only

             8846A ................................................................. 10,000 measurements, internal, and up to 2 Gigabyte capacity with

                                 USB memory module (available separately, see “Accessories”) through

                                 front-panel USB port

      Math Functions

      Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics (min/max/average/standard deviation), and

      Limit Test

      Electrical

      Input Protection....................................................... 1000 V all ranges

      Overrange ............................................................... 20 % on all ranges except 1000 V dc, 1000 V ac (8846A),

                                 750 V ac (8845A), Diode, and 10 A ranges

      Remote Interfaces

             RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the Meter to a PC USB port.

             See Accessories)

             IEEE 488.2

             LAN and “Ethernet 10/100 base T with DHCP (for IP_ADDRess) option”

      Warranty

             One year

      Electrical Specifications

      Accuracy specifications are valid for 6½ digit resolution mode after at least a 1-hour warm-up with Auto Zero enabled.

      24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per

      EN 61326-1:2000-11

      DC Voltage Specifications

      Maximum Input ....................................................... 1000 V on any range

      Common Mode Rejection ....................................... 140 dB at 50 or 60 Hz ±0.1 % (1 kΩ unbalance)

      Normal Mode Rejection .......................................... 60 dB for NPLC of 1 or greater with analog filter off and power line

                                 frequency ±0.1 %

                                 100 dB for NPLC of 1 or greater with analog filter on and power line

                                 frequency ±0.1 %

      Measurement Method ............................................. Multi-ramp A/D

      A/D Linearity ........................................................... 0.0002 % of measurement +0.0001 % of range

      Input Bias Current ................................................... <30 pA at 25 °C

      Autozero Off Operation ........................................... Following instrument warm-up at calibration temperature ±1 °C and

                                 less than 10 minutes, add error: 0.0002 % range additional error + 5

                                 μV.

      Analog Filter............................................................ When using the analog filter, specifications are relative to within one

                                 hour of using the ZERO function for that range and NPLC setting.

      DC Ratio ................................................................. Accuracy is +/- (Input accuracy + Reference accuracy), where Input

                                 accuracy = DC Voltage accuracy for the HI to LO Input (in ppm of the

                                 Input voltage), and Reference accuracy = DC Voltage accuracy for the

                                 HI to LO (Sense) Reference (in ppm of the Reference voltage).

      Settling Considerations ........................................... Measurement settling times are affected by source impedance, cable

                                 dielectric characteristics, and input signal changes.

1-10
                                                                                                                  Introduction and Specifications       1

                                                                                                                             Electrical Specifications

Input Characteristics

Range   Resolution                                               Resolution                                                  Input Impedance

                                  4½ Digits                      5½ Digits                                  6½    Digits

100 mV  100.0000 mV               10 µV                   1 µV                                        100 nV                 10 MΩ or >10 GΩ[1]

1V      1.000000 V                100 µV                  10 µV                                       1 µV                   10 MΩ or >10 GΩ[1]

10 V    10.00000 V                1 mV                    100 µV                                      10 µV                  10 MΩ or >10 GΩ[1]

100 V   100.0000 V                10 mV                   1 mV                                        100 µV                 10 MΩ ±1%

1000 V  1,000.000 V               100 mV                  10 mV                                       1 mV                   10 MΩ ±1%

[1] Inputs beyond ±14 V are clamped through 200      kΩ  typical. 10 MΩ is default           input    impedance.

8846A Accuracy

Accuracy is given as ± (% measurement + % of range)

                      24 Hour             90 Days                                            1 Year            Temperature

Range                 (23 ±1 °C)          (23 ±5 °C)                                      (23 ±5 °C)        Coefficient/ °C

                                                                                                      Outside 18 to 28 °C

100 mV          0.0025 + 0.003            0.0025 + 0.0035        0.0037 + 0.0035                           0.0005 + 0.0005

1V              0.0018 + 0.0006           0.0018 + 0.0007        0.0025 + 0.0007                           0.0005 + 0.0001

10 V            0.0013 + 0.0004           0.0018 + 0.0005        0.0024 + 0.0005                           0.0005 + 0.0001

100 V           0.0018 + 0.0006           0.0027 + 0.0006        0.0038 + 0.0006                           0.0005 + 0.0001

1000 V          0.0018 + 0.0006           0.0031 + 0.001         0.0041 + 0.001                            0.0005 + 0.0001

8845A Accuracy

Accuracy is given as ± (% measurement + % of range)

                      24 Hour             90 Days                                            1 Year            Temperature

Range                 (23 ±1 °C)          (23 ±5 °C)                                      (23 ±5 °C)        Coefficient/ °C

                                                                                                      Outside 18 to 28 °C

100 mV          0.003 + 0.003             0.004 + 0.0035         0.005 + 0.0035                            0.0005 + 0.0005

1V              0.002 + 0.0006            0.003 + 0.0007         0.004 + 0.0007                            0.0005 + 0.0001

10 V            0.0015 + 0.0004           0.002 + 0.0005         0.0035 + 0.0005                           0.0005 + 0.0001

100 V           0.002 + 0.0006            0.0035 + 0.0006        0.0045 + 0.0006                           0.0005 + 0.0001

1000 V          0.002 + 0.0006            0.0035 + 0.0010        0.0045 + 0.0010                           0.0005 + 0.0001

Additional Errors

        Digits                            NPLC                   Additional NPLC Noise Error

        6½                                100                                                0 % of range

        6½                                       10                                          0 % of range

        5½                                       1                                           0.001 % of range

        5½                                       .2               0.0025 % of range +12 μV

        4½                                0.02                    0.017 % of range +17 μV

AC Voltage Specifications

AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,

add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.

Maximum Input ....................................................... 750 V rms or 1000 V peak (8845A), 1000 V rms or 1414 V peak

                                                         (8846A) or 8 x 107 volts-Hertz product (whichever is less) for any

                                                         range.

Measurement Method ............................................. AC-coupled true-rms. Measures the ac component of input with up to

                                                         1000 V dc bias on any range.

AC Filter Bandwidth:

Slow .................................................................... 3 Hz – 300 kHz

Medium ............................................................... 20 Hz – 300 kHz

Fast ..................................................................... 200 Hz – 300 kHz

Common Mode Rejection ....................................... 70 dB at 50 Hz or 60 Hz ±0.1 % (1 kΩ unbalance)

                                                                                                                                                        1-11
8845A/8846A

Calibration Manual

      Maximum Crest Factor............................................ 5:1 at Full Scale

      Additional Crest Factor Errors (<100 Hz)................ Crest factor 1-2, 0.05 % of full scale

                                                                    Crest factor 2-3, 0.2 % of full scale

                                                                    Crest factor 3-4, 0.4 % of full scale

                                                                    Crest factor 4-5, 0.5 % of full scale

      Input Characteristics

      Range         Resolution                                      Resolution                                              Input Impedance

                                         4½ Digits                        5½ Digits                    6½ Digits

      100 mV        100.0000 mV          10 µV                      1 µV                  100 nV

      1V            1.000000 V           100 µV                     10 µV                 1 µV                            1 MΩ ±2 % shunted

      10 V          10.00000 V           1 mV                       100 µV                10 µV                           by <100 pf

      100 V         100.0000 V           10 mV                      1 mV                  100 µV

      1000 V        1,000.000 V          100 mV                     10 mV                 1 mV

      8846A Accuracy

      Accuracy is given as ± (% measurement + % of range)

                                                        24 Hour     90 Days               1 Year           Temperature

      Range             Frequency                       (23 ±1 °C)  (23 ±5 °C)            (23 ±5 °C)       Coefficient/ °C

                                                                                                           Outside 18 to 28 °C

      100 mV        3 – 5 Hz                   1.0 + 0.03           1.0 + 0.04            1.0 + 0.04       0.1 + 0.004

                    5 – 10 Hz                  0.35 + 0.03          0.35 + 0.04           0.35 + 0.04      0.035 + 0.004

                    10  Hz – 20 kHz            0.04 + 0.03          0.05 + 0.04           0.06 + 0.04      0.005 + 0.004

                    20 – 50 kHz                0.1 + 0.05           0.11 + 0.05           0.12 + 0.05      0.011 + 0.005

                    50 – 100 kHz               0.55 + 0.08          0.6 + 0.08            0.6 + 0.08       0.06 + 0.008

                    100 – 300 kHz[1]           4.0 + 0.50           4.0 + 0.50            4.0 + 0.50       0.20 + 0.02

      1V            3 – 5 Hz                   1.0 + 0.02           1.0 + 0.03            1.0 + 0.03       0.1 + 0.003

                    5 – 10 Hz                  0.35 + 0.02          0.35 + 0.03           0.35 + 0.03      0.035 + 0.003

                    10  Hz – 20 kHz            0.04 + 0.02          0.05 + 0.03           0.06 + 0.03      0.005 + 0.003

                    20 – 50 kHz                0.1 + 0.04           0.11 + 0.05           0.12 + 0.05      0.011 + 0.005

                    50 – 100 kHz               0.55 + 0.08          0.6 + 0.08            0.6 + 0.08       0.06 + 0.008

                    100 – 300 kHz[1]           4.0 + 0.50           4.0 + 0.50            4.0 + 0.50       0.2 + 0.02

      10 V          3 – 5 Hz                   1.0 + 0.02           1.0 + 0.03            1.0 + 0.03       0.1 + 0.003

                    5 – 10 Hz                  0.35 + 0.02          0.35 + 0.03           0.35 + 0.03      0.035 + 0.003

                    10  Hz – 20 kHz            0.04 + 0.02          0.05 + 0.03           0.06 + 0.03      0.005 + 0.003

                    20 – 50 kHz                0.1 + 0.04           0.11 + 0.05           0.12 + 0.05      0.011 + 0.005

                    50 – 100 kHz               0.55 + 0.08          0.6 + 0.08            0.6 + 0.08       0.06 + 0.008

                    100 – 300 kHz[1]           4.0 + 0.50           4.0 + 0.50            4.0 + 0.50       0.2 + 0.02

      100 V         3 – 5 Hz                   1.0 + 0.02           1.0 + 0.03            1.0 + 0.03       0.1 + 0.003

                    5 – 10 Hz                  0.35 + 0.02          0.35 + 0.03           0.35 + 0.03      0.035 + 0.003

                    10  Hz – 20 kHz            0.04 + 0.02          0.05 + 0.03           0.06 + 0.03      0.005 + 0.003

                    20 – 50 kHz                0.1 + 0.04           0.11 + 0.05           0.12 + 0.05      0.011 + 0.005

                    50 – 100 kHz               0.55 + 0.08          0.6 + 0.08            0.6 + 0.08       0.06 + 0.008

                    100 –- 300 kHz[1]          4.0 + 0.50           4.0 + 0.50            4.0 + 0.50       0.2 + 0.02

      1000 V        3 – 5 Hz                   1.0 + 0.015          1.0 + 0.0225          1.0 + 0.0225     0.1 + 0.00225

                    5 – 10 Hz                  0.35 + 0.015         0.35 + 0.0225         0.35 + 0.0225    0.035 + 0.00225

                    10  Hz – 20 kHz            0.04 + 0.015         0.05 + 0.0225         0.06 + 0.0225    0.005 + 0.00225

                    20 – 50 kHz                0.1 + 0.03           0.11 + 0.0375         0.12 + 0.0375    0.011 + 0.00375

                    50 – 100 kHz[2]            0.55 + 0.06          0.6 + 0.06            0.6 + 0.06       0.06 + 0.006

                    100 – 300 kHz[1][2]        4.0 + 0.375          4.0 + 0.375           4.0 + 0.375      0.2 + 0.015

             [1] Typically 30 % reading error at 1 MHz

             [2] 1000 Volt range is limited to 8 X 107 volt-Hertz

1-12
                                                                                          Introduction and Specifications           1

                                                                                                         Electrical Specifications

8845A Accuracy

Accuracy is given as ± (% measurement + % of range)

                 Frequency                   24 Hour        90 Days          1 Year       Temperature

Range            (Hz)                 (23 ±1 °C)            (23 ±5 °C)       (23 ±5 °C)   Coefficient/ °C

                                                                                          Outside 18 to 28 °C

100 mV       3 – 5 Hz                 1.0 + 0.03            1.0 + 0.04       1.0 + 0.04   0.10 + 0.004

             5 – 10 Hz                0.35 + 0.03           0.35 + 0.04      0.35 + 0.04  0.035 + 0.004

             10  Hz – 20 kHz          0.04 + 0.03           0.05 + 0.04      0.06 + 0.04  0.005 + 0.004

             20 – 50 kHz              0.1 + 0.05            0.11 + 0.05      0.12 + 0.05  0.011 + 0.005

             50 – 100 kHz             0.55 + 0.08           0.6 + 0.08       0.6 + 0.08   0.06 + 0.008

             100 – 300 kHz[1]         4.0 + 0.50            4.0 + 0.50       4.0 + 0.50   0.2 + 0.02

1V           3 – 5 Hz                 1.0 + 0.02            1.0 + 0.03       1.0 + 0.03   0.1 + 0.003

             5 – 10 Hz                0.35 + 0.02           0.35 + 0.03      0.35 + 0.03  0.035 + 0.003

             10  Hz – 20 kHz          0.04 + 0.02           0.05 + 0.03      0.06 + 0.03  0.005 + 0.003

             20 – 50 kHz              0.1 + 0.04            0.11 + 0.05      0.12 + 0.05  0.011 + 0.005

             50 – 100 kHz             0.55 + 0.08           0.6 + 0.08       0.6 + 0.08   0.06 + 0.008

             100 – 300 kHz[1]         4.0 + 0.50            4.0 + 0.50       4.0 + 0.50   0.2 + 0.02

10 V         3 – 5 Hz                 1.0 + 0.02            1.0 + 0.03       1.0 + 0.03   0.1 + 0.003

             5 – 10 Hz                0.35 + 0.02           0.35 + 0.03      0.35 + 0.03  0.035 + 0.003

             10  Hz – 20 kHz          0.04 + 0.02           0.05 + 0.03      0.06 + 0.03  0.005 + 0.003

             20 – 50 kHz              0.1 + 0.04            0.11 + 0.05      0.12 + 0.05  0.011 + 0.005

             50 – 100 kHz             0.55 + 0.08           0.6 + 0.08       0.6 + 0.08   0.06 + 0.008

             100 – 300 kHz[1]         4.0 + 0.50            4.0 + 0.50       4.0 + 0.50   0.2 + 0.02

100 V        3 – 5 Hz                 1.0 + 0.02            1.0 + 0.03       1.0 + 0.03   0.1 + 0.003

             5 – 10 Hz                0.35 + 0.02           0.35 + 0.03      0.35 + 0.03  0.035 + 0.003

             10  Hz – 20 kHz          0.04 + 0.02           0.05 + 0.03      0.06 + 0.03  0.005 + 0.003

             20 – 50 kHz              0.1 + 0.04            0.11 + 0.05      0.12 + 0.05  0.011 + 0.005

             50 – 100 kHz             0.55 + 0.08           0.6 + 0.08       0.6 + 0.08   0.06 + 0.008

             100 – 300 kHz[1]         4.0 + 0.50            4.0 + 0.50       4.0 + 0.50   0.2 + 0.02

750 V        3 –- 5 Hz                1.0 + 0.02            1.0 + 0.03       1.0 + 0.03   0.1 + 0.003

             5 – 10 Hz                0.35 + 0.02           0.35 + 0.03      0.35 + 0.03  0.035 + 0.003

             10  Hz – 20 kHz          0.04 + 0.02           0.05 + 0.03      0.06 + 0.03  0.005 + 0.003

             20 – 50 kHz              0.1 + 0.04            0.11 + 0.05      0.12 + 0.05  0.011 + 0.005

             50 – 100 kHz [2]         0.55 + 0.08           0.6 + 0.08       0.6 + 0.08   0.06 + 0.008

             100 – 300 kHz[1] [2]     4.0 + 0.5             4.0 + 0.5        4.0 + 0.5    0.2 + 0.02

[1] Typically 30 % reading error at   1 MHz

[2] 750 Volt range is limited to 8 X  107 volt-Hertz

Additional Low Frequency Errors

Error is stated as % of reading.

      Frequency                                          AC Filter

                               3HZ (slow)    20HZ (medium)                200HZ (fast)

10 – 20 Hz                 0                          0.25             –

20 – 40 Hz                 0                          0.02             –

40 – 100 Hz                0                          0.01             0.55

100 – 200 Hz               0                          0                0.2

200 Hz – 1 kHz             0                          0                0.02

>1 kHz                     0                          0                0

                                                                                                                                    1-13
8845A/8846A

Calibration Manual

      Resistance

      Specifications are for 4-wire resistance function, 2 x 4-wire resistance, or 2-wire resistance with zero. If zero is not used,

      add 0.2 Ω for 2-wire resistance plus lead resistance, and add 20 mΩ for 2 x 4-wire resistance function.

      Measurement Method ............................................. Current source referenced to LO input

      Max. Lead Resistance (4-wire ohms) ..................... 10 % of range per lead for 10 Ω, 100 Ω, 1 kΩ ranges. 1 kΩ per lead on

                                                                    all other ranges

      Input Protection....................................................... 1000 V on all ranges

      Common Mode Rejection ....................................... 140 dB at 50 or 60 Hz ± 0.1 % (1 kΩ unbalance)

      Normal Mode Rejection .......................................... 60 dB for NPLC of 1 or greater with analog filter off and power line

                                                                    frequency ±0.1 %

                                                                    100 dB for NPLC of 1 or greater with analog filter on and power line

                                                                    frequency ±0.1 %

      Analog Filter............................................................ When using the analog filter, specifications are relative to within one

                                                                    hour of using the ZERO function for that range and NPLC setting.

      Input Characteristics

             Range      Resolution                                       Resolution                                      Source Current

                                                    4½ Digits              5½ Digits                      6½ Digits

      10 Ω[1]           10.00000  Ω           1 mΩ                   100 μΩ                         10 μΩ                5 mA/13 V

      100 Ω             100.0000  Ω           10 mΩ                  1 mΩ                           100 μΩ               1 mA/6 V

      1 kΩ              1.000000  kΩ          100 mΩ                 10 mΩ                          1 mΩ                 1 mA/6 V

      10 kΩ             10.00000  kΩ          1Ω                     100 mΩ                         10 mΩ                100 μA/6 V

      100 kΩ            100.0000  kΩ          10 Ω                   1Ω                             100 mΩ               100 μA/13 V

      1 MΩ              1.000000  MΩ          100 Ω                  10 Ω                           1Ω                   10 μA/13 V

      10 MΩ             10.00000  MΩ          1 kΩ                   100 Ω                          10 Ω                 1 μA/13 V

      100 MΩ            100.0000  MΩ          10 kΩ                  1 kΩ                           100 Ω                1 μA || 10 MΩ/10                V

      1.0 GΩ[1]         1.000000  GΩ          100 kΩ                 10 kΩ                          1 kΩ                 1 μA || 10 MΩ/10                V

             [1] 8846A  Only

      8846A Accuracy

      Accuracy is given as ± (% measurement + % of range)

                              24 Hour                 90 Days                1 Year                       Temperature

             Range            (23 ±1 °C)              (23 ±5 °C)             (23 ±5 °C)                 Coefficient/ °C

                                                                                                    Outside 18 to 28 °C

      10 Ω                    0.003 + 0.01            0.008 + 0.03         0.01+ 0.03               0.0006 + 0.0005

      100 Ω                   0.003 + 0.003           0.008 + 0.004        0.01 + 0.004             0.0006 + 0.0005

      1 kΩ                    0.002 + 0.0005          0.008 + 0.001        0.01 + 0.001             0.0006 + 0.0001

      10 kΩ                   0.002 + 0.0005          0.008 + 0.001        0.01 + 0.001             0.0006 + 0.0001

      100 kΩ                  0.002 + 0.0005          0.008 + 0.001        0.01 + 0.001             0.0006 + 0.0001

      1 MΩ                    0.002 + 0.001           0.008 + 0.001        0.01 + 0.001             0.001 + 0.0002

      10 MΩ                   0.015 + 0.001           0.02 + 0.001         0.04 + 0.001             0.003 + 0.0004

      100 MΩ                  0.3 + 0.01              0.8 + 0.01           0.8 + 0.01               0.15 + 0.0002

      1 GΩ                    1.0 + 0.01              1.5 + 0.01           2.0 + 0.01               0.6 + 0.0002

1-14
                                                                                                       Introduction and Specifications             1

                                                                                                                 Electrical Specifications

8845A Accuracy

Accuracy is given as ± (% measurement + % of range)

                    24 Hour                      90 Days              1 Year                    Temperature

Range               (23 ±1 °C)                   (23 ±5 °C)           (23 ±5 °C)                Coefficient/ °C

                                                                                     Outside 18 to 28 °C

100 Ω           0.003 + 0.003   0.008 + 0.004                         0.01 + 0.004   0.0006 + 0.0005

1 kΩ            0.002 + 0.0005  0.008 + 0.001                         0.01 + 0.001   0.0006 + 0.0001

10 kΩ           0.002 + 0.0005  0.008 + 0.001                         0.01 + 0.001   0.0006 + 0.0001

100 kΩ          0.002 + 0.0005  0.008 + 0.001                         0.01 + 0.001   0.0006 + 0.0001

1 MΩ            0.002 + 0.001   0.008 + 0.001                         0.01 + 0.001   0.0010 + 0.0002

10 MΩ           0.015 + 0.001   0.02 + 0.001                          0.04 + 0.001   0.0030 + 0.0004

100 MΩ          0.3 + 0.01      0.8 + 0.01                            0.8 + 0.01     0.1500 + 0.0002

Additional Ohms Errors

            Digits                               NPLC                 Additional NPLC Noise Error

            6½                                   100                              0 % of range

            6½                                   10                               0 % of range

            5½                                   1                    0.001 % of range

            5½                                   0.2                  0.003 % of range ±7 mΩ

            4½                                   0.02                 0.017 % of range ±15 mΩ

DC Current

Input Protection....................................................... Tool-accessible 11 A/1000 V and 440 mA/1000 V fuses, limits of 400

                                                          mA continuous 550 mA for 2 minutes on, 1 minute off.

Common Mode Rejection ....................................... 140 dB at 50 or 60 Hz ±0.1 % (1 kΩ unbalance)

Normal Mode Rejection .......................................... 60 dB for NPLC of 1 or greater with analog filter off and power line

                                                          frequency ±0.1 %

                                                          100 dB for NPLC of 1 or greater with analog filter on and power line

                                                          frequency ±0.1 %

Analog Filter............................................................ When using the analog filter, specifications are relative to within one

                                                          hour of using the ZERO function for that range and NPLC setting.

Input Characteristics

Range           Resolution                             Resolution                    Shunt Resistance        Burden Voltage

                                4½ Digits              5½ Digits      6½ Digits                 (Ohms)

100 µA          100.0000 µA     10 nA                  1 nA           100 pA                    100 Ω            <0.015 V

1 mA            1.000000 mA     100 nA                 10 nA          1 nA                      100 Ω            <0.15 V

10 mA           10.00000 mA     1 μA                   100 nA         10 nA                     1Ω               <0.025 V

100 mA          100.0000 mA     10 µA                  1 μA           100 nA                    1Ω               <0.25 V

400 mA [3]      400.000 mA      100 µA                 10 µA          1 μA                      1Ω               <0.50 V

1 A [2]         1.000000 A      100 µA                 10 µA          1 μA                      0.01 Ω           <0.05 V

3 A[1]          3.00000A        1 mA                   100 µA         10 µA                     0.01 Ω           <0.15 V

10 A            10.00000 A      1 mA                   100 µA         10 µA                     0.01 Ω           <0.5 V

[1] Part of 10 A range.

[2] Available on the front panel terminal only.

[3] 400 mA available in software version 2.0 or greater only. 400 mA  continuously;  550 mA for 2 minutes on, 1 minute off.

                                                                                                                                                   1-15
8845A/8846A

Calibration Manual

      Accuracy (8846A)

      Accuracy is given as ± (% measurement + % of range)

                                     24 Hour               90 Days               1 Year                  Temperature

             Range                   (23 ±1 °C)            (23 ±5 °C)            (23 ±5 °C)              Coefficient/ °C

                                                                                                        Outside 18 to 28 °C

      100 µA                     0.01 + 0.02               0.04 + 0.025          0.05 + 0.025            0.002 + 0.003

      1 mA                       0.007 + 0.005             0.030 + 0.005         0.05 + 0.005            0.002 + 0.0005

      10 mA                      0.007 + 0.02              0.03 + 0.02           0.05 + 0.02             0.002 + 0.002

      100 mA                     0.01 + 0.004              0.03 + 0.005          0.05 + 0.005            0.002 + 0.0005

      400 mA [3]                 0.03 + 0.004              0.04 + 0.005          0.05 + 0.005            0.005 + 0.0005

      1 A[2]                     0.03 + 0.02               0.04 + 0.02           0.05 + 0.02             0.005 + 0.001

      3 A[1][2]                  0.05 + 0.02               0.08 + 0.02           0.1 + 0.02              0.005 + 0.002

      10 A[2]                    0.1 + 0.008               0.12 + 0.008          0.15 + 0.008            0.005 + 0.0008

             [1] Part of 10 A range

             [2] Available at front panel connectors only

             [3] 400 mA available in software version 2.0 or greater only.  400  mA continuously;  550  mA for 2 minutes on, 1

                    minute off.

      Accuracy (8845A)

      Accuracy is given as ± (% measurement + % of range)

                                     24 Hour               90 Days               1 Year                  Temperature

             Range                   (23 ±1 °C)            (23 ±5 °C)            (23 ±5 °C)              Coefficient/ °C

                                                                                                        Outside 18 to 28 °C

      100 µA                     0.01 + 0.02               0.04 + 0.025          0.05 + 0.025            0.002 + 0.003

      1 mA                       0.007 + 0.005             0.030 + 0.005         0.05 + 0.005            0.002 + 0.0005

      10 mA                      0.007 + 0.02              0.03 + 0.02           0.05 + 0.02             0.002 + 0.002

      100 mA                     0.01 + 0.004              0.03 + 0.005          0.05 + 0.005            0.002 + 0.0005

      400 mA [3]                 0.03 + 0.004              0.04 + 0.005          0.05 + 0.005            0.005 + 0.0005

      1 A[2]                     0.03 + 0.02               0.04 + 0.02           0.05 + 0.02             0.005 + 0.001

      3 A[1][2]                  0.05 + 0.02               0.08 + 0.02           0.10 + 0.02             0.005 + 0.002

      10 A[2]                    0.10 + 0.008              0.12 + 0.008          0.15 + 0.008            0.005 + 0.0008

             [1] Part of 10 A range

             [2] Available at front panel connectors only

             [3] 400 mA available in software version 2.0 or greater only.  400  mA continuously;  550  mA for 2 minutes on, 1

                    minute off.

      Additional Current Errors

                    Digits                       NPLC                       Additional NPLC Noise Error for 1  Additional NPLC Noise Error

                                                                            mA, 100 mA, 400 mA, 3A and 10A              for 100 μA, 10 mA, 1A

                    6½                           100                             0 % of range                                   0 % of range

                    6½                           10                              0 % of range                                   0 % of range

                    5½                           1                               0.001 % of range                               0.01 % of range

                    5½                           0.2                             0.011 % of range ±4 μA                 0.11 % of range ±4 μA

                    4½                           0.02                            0.04 % of range ±4 μA                  0.28 % of range ±4 μA

1-16
                                                                                                               Introduction and Specifications             1

                                                                                                                                Electrical Specifications

AC Current

The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs

from 1 % to 5 % of range, add an additional error of 0.1 % of range.

Input Protection....................................................... Tool accessible 11 A/1000 V and 440 mA/1000 V fuses, limits of 400

                                                       mA continuous 550 mA for 2 minutes on, 1 minute off.

Measurement Method ............................................. ac-coupled true-rms, dc-coupled to the fuse and shunt (no blocking

                                                       capacitor)

AC Filter Bandwidth

Slow .................................................................... 3 Hz to 10 kHz

Medium ............................................................... 20 Hz to 10 kHz

Fast ..................................................................... 200 Hz to 10 kHz

Maximum Crest Factor............................................ 5:1 at full scale

Additional Crest Factor Errors (<100 Hz)................ Crest factor 1-2, 0.05 % of full scale

                                                       Crest factor 2-3, 0.2 % of full scale

                                                       Crest factor 3-4, 0.4 % of full scale

                                                       Crest factor 4-5, 0.5 % of full scale

Input Characteristics

Range       Resolution                        Resolution                                     Shunt Resistance          Burden Voltage

                                 4½ Digits     5½ Digits        6½ Digits                    (Ohms)

100 μA[1]   100.0000 μA          10 nA         1 nA             100 pA                       100 Ω                     <0.015 V

1 mA[1]     1.000000 mA          100 nA        10 nA            1 nA                         100 Ω                     <0.15 V

10 mA       10.00000 mA          1 μA          100 nA           10 nA                        1Ω                        <0.025 V

100 mA      100.0000 mA          10 µA         1 μA             100 nA                       1Ω                        <0.25 V

400 mA [4]  400.000 mA           100 µA        10 µA            1 μA                         1Ω                        <0.50 V

1 A[3]      1.000000 A           100 µA        10 µA            1 μA                         0.01 Ω                    <0.05 V

3 A[2][3]   3.00000 A            1 mA          100 µA           10 μA                        0.01 Ω                    <0.05 V

10 A[3]     10.00000 A           1 mA          100 µA           10 µA                        0.01 Ω                    <0.5 V

[1] 8846A Only

[2] Part of 10 A range

[3] Available at front panel connectors only

[4] 400 mA available in software version 1.0.700.18 or greater  only. 400 mA                 continuously; 550 mA for  2 minutes on, 1

minute off; maximum crest factor 3:1 at 400mA

8846A       Accuracy

Accuracy    is given as ± (% measurement + % of        range)

                Frequency                     24 Hour           90 Days                      1 Year                    Temperature

Range                   (Hz)             (23 ±1 °C)             (23 ±5 °C)                   (23 ±5 °C)                Coefficient/ °C

                                                                                                               Outside 18 to 28 °C

100 μA      3 – 5 Hz                   1.1 + 0.06               1.1 + 0.06                   1.1 + 0.06                0.2 + 0.006

            5 – 10 Hz                  0.35 + 0.06              0.35 + 0.06                  0.35 + 0.06               0.1 + 0.006

            10 Hz       – 5 kHz        0.15 + 0.06              0.15 + 0.06                  0.15 + 0.06               0.015 + 0.006

            5 – 10 kHz                 0.35 + 0.7               0.35 + 0.7                   0.35 + 0.7                0.03 + 0.006

1 mA        3 – 5 Hz                   1.0 + 0.04               1.0 + 0.04                   1.0 + 0.04                0.1 + 0.006

            5 – 10 Hz                  0.3 + 0.04               0.3 + 0.04                   0.3 + 0.04                0.035 + 0.006

            10 Hz – 5 kHz              0.1 + 0.04               0.1 + 0.04                   0.1 + 0.04                0.015 + 0.006

            5 – 10 kHz                 0.2 + 0.25               0.2 + 0.25                   0.2 + 0.25                0.03 + 0.006

10 mA       3 – 5 Hz                   1.1 + 0.06               1.1 + 0.06                   1.1 + 0.06                0.2 + 0.006

            5 – 10 Hz                  0.35 + 0.06              0.35 + 0.06                  0.35 + 0.06               0.1 + 0.006

            10 Hz – 5 kHz              0.15 + 0.06              0.15 + 0.06                  0.15+ 0.06                0.015 + 0.006

            5 – 10 kHz                 0.35 + 0.7               0.35 + 0.7                   0.35 + 0.7                0.03 + 0.006

100 mA      3 – 5 Hz                   1.0 + 0.04               1.0 + 0.04                   1.0 + 0.04                0.1 + 0.006

            5 – 10 Hz                  0.3 + 0.04               0.3 + 0.04                   0.3 + 0.04                0.035 + 0.006

            10 Hz – 5 kHz              0.1 + 0.04               0.1 + 0.04                   0.1 + 0.04                0.015 + 0.006

            5 – 10 kHz                 0.2 + 0.25               0.2 + 0.25                   0.2 + 0.25                0.03 + 0.006

                                                                                                                                                           1-17
8845A/8846A

Calibration Manual

      8846A Accuracy (cont)

                      Frequency            24 Hour         90 Days      1 Year                                         Temperature

             Range                   (Hz)  (23 ±1 °C)      (23 ±5 °C)   (23 ±5 °C)                                     Coefficient/ °C

                                                                                                                       Outside 18 to 28 °C

      400 mA [3]      3 – 5 Hz             1.0 + 0.1       1.0 + 0.1    1.0 + 0.1                                      0.1 + 0.006

                      5 – 10 Hz            0.3 + 0.1       0.3 + 0.1    0.3 + 0.1                                      0.035 + 0.006

                      10 Hz – 1 kHz        0.1 + 0.1       0.1 + 0.1    0.1 + 0.1                                      0.015 + 0.006

                      1kHz – 10 kHz        0.2 + 0.7       0.2 + 0.7    0.2 + 0.7                                      0.03 + 0.006

      1 A[2]          3 – 5 Hz             1.0 + 0.04      1.0 + 0.04   1.0 + 0.04                                     0.1 + 0.006

                      5 – 10 Hz            0.3 + 0.04      0.3 + 0.04   0.3 + 0.04                                     0.035 + 0.006

                      10 Hz – 5 kHz        0.1 + 0.04      0.1 + 0.04   0.1 + 0.04                                     0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

      3 A[1][2]       3 – 5 Hz             1.1 + 0.06      1.1 + 0.06   1.1 + 0.06                                     0.1 + 0.006

                      5 – 10 Hz            0.35 + 0.06     0.35 + 0.06  0.35 + 0.06                                    0.035 + 0.006

                      10 Hz – 5 kHz        0.15 + 0.06     0.15 + 0.06  0.15 + 0.06                                    0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

      10 A[2]         3 – 5 Hz             1.1 + 0.06      1.1 + 0.06   1.1 + 0.06                                     0.1 + 0.006

                      5 – 10 Hz            0.35 + 0.06     0.35 + 0.06  0.35 + 0.06                                    0.035 + 0.006

                      10 Hz – 5 kHz        0.15 + 0.06     0.15 + 0.06  0.15 + 0.06                                    0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

             [1] Part of 10 A range

             [2] Available only on front panel connectors

             [3] 400 m available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1

                    minute off; maximum crest factor 3:1 at 400mA; specification for current above 329 mA is typical.

      8845A Accuracy

      Accuracy is given as ± (% measurement + % of range)

                      Frequency            24 Hour         90 Days      1 Year                                         Temperature

             Range                   (Hz)  (23 ±1 °C)      (23 ±5 °C)   (23 ±5 °C)                                     Coefficient/ °C

                                                                                                                       Outside 18 to 28 °C

      10 mA           3 – 5 Hz             1.1 + 0.06      1.1 + 0.06   1.1 + 0.06                                     0.2 + 0.006

                      5 – 10 Hz            0.35 + 0.06     0.35 + 0.06  0.35 + 0.06                                    0.1 + 0.006

                      10 Hz – 5 kHz        0.15 + 0.06     0.15 + 0.06  0.15+ 0.06                                     0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

      100 mA          3 – 5 Hz             1.0 + 0.04      1.0 + 0.04   1.0 + 0.04                                     0.1 + 0.006

                      5 – 10 Hz            0.3 + 0.04      0.3 + 0.04   0.3 + 0.04                                     0.035 + 0.006

                      10 Hz – 5 kHz        0.1 + 0.04      0.1 + 0.04   0.1 + 0.04                                     0.015 + 0.006

                      5 – 10 kHz           0.2 + 0.25      0.2 + 0.25   0.2 + 0.25                                     0.03 + 0.006

      400 mA [3]      3 – 5 Hz             1.0 + 0.1       1.0 + 0.1    1.0 + 0.1                                      0.1 + 0.006

                      5 – 10 Hz            0.3 + 0.1       0.3 + 0.1    0.3 + 0.1                                      0.035 + 0.006

                      10 Hz – 1 kHz        0.1 + 0.1       0.1 + 0.1    0.1 + 0.1                                      0.015 + 0.006

                      1kHz – 10 kHz        0.2 + 0.7       0.2 + 0.7    0.2 + 0.7                                      0.03 + 0.006

      1      A[2]     3 – 5 Hz             1.0 + 0.04      1.0 + 0.04   1.0 + 0.04                                     0.1 + 0.006

                      5 – 10 Hz            0.3 + 0.04      0.3 + 0.04   0.3 + 0.04                                     0.035 + 0.006

                      10 Hz – 5 kHz        0.1 + 0.04      0.1 + 0.04   0.1 + 0.04                                     0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

      3      A[1][2]  3 – 5 Hz             1.1 + 0.06      1.1 + 0.06   1.1 + 0.06                                     0.1 + 0.006

                      5 – 10 Hz            0.35 + 0.06     0.35 + 0.06  0.35 + 0.06                                    0.035 + 0.006

                      10 Hz – 5 kHz        0.15 + 0.06     0.15 + 0.06  0.15 + 0.06                                    0.015 + 0.006

                      5 – 10 kHz           0.35 + 0.7      0.35 + 0.7   0.35 + 0.7                                     0.03 + 0.006

1-18
                                                                                                                 Introduction and Specifications  1

                                                                                                                     Electrical Specifications

8845A Accuracy (cont)

                Frequency                     24 Hour           90 Days            1 Year                        Temperature

Range           (Hz)               (23 ±1 °C)                (23 ±5 °C)            (23 ±5 °C)                    Coefficient/ °C

                                                                                                                 Outside 18 to 28 °C

10 A[2]       3 – 5 Hz             1.1 + 0.06                1.1 + 0.06        1.1 + 0.06                        0.2 + 0.006

              5 – 10 Hz            0.35 + 0.06               0.35 + 0.06       0.35 + 0.06                       0.035 + 0.006

              10 Hz – 5 kHz        0.15 + 0.06               0.15 + 0.06       0.15 + 0.06                       0.015 + 0.006

              5 – 10 kHz           0.35 + 0.7                0.35 + 0.7        0.35 + 0.7                        0.03 + 0.006

[1] Part of the 10 A range

[2] Available only at front panel connectors

[3] 400 m available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1

minute off; maximum crest factor 3:1 at 400mA; specification for current above 329 mA is typical.

Additional Low Frequency Errors

Error is stated as % of reading.

Frequency                                              AC Filter

                               3HZ (slow)     20HZ (medium)                200HZ (fast)

10 – 20 Hz                  0                 0.25                      –

20 – 40 Hz                  0                 0.02                      –

40 – 100 Hz                 0                  0.01                     0.55

100 – 200 Hz                0                 0                         0.2

200 Hz – 1 kHz              0                 0                         0.02

> 1 kHz                     0                 0                         0

Frequency

Gate Times ............................................................. Programmable to 1 s, 100 ms, and 10 ms

Measurement Method ............................................. Flexible counting technique. AC-coupled input using the ac voltage

                                                            measurement function.

Settling Considerations ........................................... When measuring frequency or period after a dc offset voltage change,

                                                            errors may occur. For the most accurate measurement, wait up to 1

                                                            second for the input blocking capacitor to settle.

Measurement Considerations ................................. To minimize measurement errors, shield inputs from external noise

                                                            when measuring low-voltage, low-frequency signals.

8846A Accuracy

Accuracy is given as ± % measurement

                                   24 Hour             90 Days     1 Year          Temperature

Range           Frequency          (23 ±1 °C)          (23 ±5 °C)  (23 ±5 °C)      Coefficient/ °C

                                                                               Outside 18 to 28 °C

100 mV to     3 – 5 Hz             0.1                 0.1         0.1         0.005

1000 V[1][2]  5 – 10 Hz            0.05                0.05        0.05        0.005

              10 – 40 Hz           0.03                0.03        0.03        0.001

              40 Hz – 300 kHz      0.006               0.01        0.01        0.001

              300 kHz – 1 MHz      0.006               0.01        0.01        0.001

[1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.

[2] Limited to 8 X 107 volt-Hertz

                                                                                                                                                  1-19
8845A/8846A

Calibration Manual

      8845A Accuracy

      Accuracy is given as ± % measurement

                                                24 Hour              90 Days         1 Year      Temperature

             Range         Frequency            (23 ±1 °C)           (23 ±5 °C)     (23 ±5 °C)   Compensation/ °C

                                                                                                 Outside 18 to 28 °C

      100 mV to          3 – 5 Hz               0.1                  0.1            0.1          0.005

      750 V[1][2]        5 – 10 Hz              0.05                 0.05           0.05         0.005

                         10 – 40 Hz             0.03                 0.03           0.03         0.001

                         40 Hz – 300 kHz        0.006                0.01           0.01         0.001

             [1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.

             [2] Limited to 8 X 107 volt-Hertz

      Gate Time vs. Resolution

                         Gate Time                                       Resolution

                           0.01                                                 5½

                           0.1                                                  6½

                           1.0                                                  6½

      Additional Low Frequency Errors

      Error stated as percent of measurement for inputs                   >100  mV.  For  10  –  100 mV, multiply  percent  by  10.

                    Frequency                                                                    Resolution

                                                           6½                                    5½                                  4½

      3 – 5 Hz                                              0                                    0.12                                0.12

      5 – 10 Hz                                             0                                    0.17                                0.17

      10 – 40 Hz                                            0                                    0.2                                 0.2

      40 – 100 Hz                                           0                                    0.06                                0.21

      100 – 300 Hz                                          0                                    0.03                                0.21

      300 Hz – 1 kHz                                        0                                    0.01                                0.07

      > 1 kHz                                               0                                    0                                   0.02

      Capacitance (8846A Only)

      Accuracy is stated as ±(% of measurement + % of range)

                    Range                       Resolution                                1 Year Accuracy[1]       Temperature Coefficient/      °C

                                                                                                 (23 ±5 °C)                 Outside 18 to 28 °C

      1 nF                             1 pF                                          2% ± 2.5 %                       0.05 + 0.05

      10 nF                            10 pF                                         1% ± 0.5 %                       0.05 + 0.01

      100 nF                           100 pF                                        1% ± 0.5 %                       0.01 + 0.01

      1 µF                             1 nF                                          1% ± 0.5 %                       0.01 + 0.01

      10 µF                            10 nF                                         1% ± 0.5 %                       0.01 + 0.01

      100 µF                           100 nF                                        1% ± 0.5 %                       0.01 + 0.01

      1 mF                             1 µF                                          1% ± 0.5 %                       0.01 + 0.01

      10 mF                            10 µF                                         1% ± 0.5 %                       0.01 + 0.01

      100 mF                           100 µF                                        4% ± 0.2 %                       0.05 + 0.05

             [1] Stated  accuracy  is  attained when Zero  function  is  used.

1-20
                                                                                                               Introduction and Specifications         1

                                                                                                                            Electrical Specifications

Temperature (8846A only)

Test Current ............................................................ 1 mA

Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 ohms lead

resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function.

Specifications do not include probe accuracy, which must be added.

                                                                                            Accuracy                         Temperature

Range             Resolution                                                    90 Days                 1 Year               Coefficient/ °C

                                                                                (23 ±5 °C)              (23 ±5 °C)          Outside 18 to 28 °C

-200 °C           0.001 °C                                                      0.06                    0.09                 0.0025

-100 °C           0.001 °C                                                      0.05                    0.08                 0.002

          0 °C    0.001 °C                                                      0.04                    0.06                 0.002

100 °C            0.001 °C                                                      0.05                    0.08                 0.002

300        °C     0.001 °C                                                      0.1                     0.12                 0.002

600 °C            0.001 °C                                                      0.18                    0.22                 0.002

          Digits                 NPLC                 Additional NPLC Noise Error

           6½                    100                                                        0 °C

           6½                    10                                                         0 °C

           5½                    1                                                        0.03 °C

           5½                    0.2                                                      0.12 °C

           4½                    0.02                                                       0.6 °C

Continuity

Continuity Threshold ............................................... Selectable between 1 Ω and 1000 Ω

Test Current ............................................................ 1 mA

Response Time....................................................... 300 samples/sec with audible tone

Accuracy is given as ± (% measurements + % of range)

                  24 Hour              90 Days                                              1 Year             Temperature

Range             (23 ±1 °C)           (23 ±5 °C)                                           (23 ±5 °C)  Coefficient/ °C

                                                                                                        Outside 18 to 28 °C

1000.0 Ω          0.002 + 0.01       0.008 + 0.02                               0.01 + 0.02             0.001 + 0.002

Diode Test

Test Current ............................................................ 100 μA or 1 mA

Response Time....................................................... 300 samples/sec with audible       tone.

Accuracy is given as ± (% measurements + % of range)

                  24 Hour              90 Days                                              1 Year             Temperature

Range             (23 ±1 °C)           (23 ±5 °C)                                           (23 ±5 °C)  Coefficient/ °C

                                                                                                        Outside 18 to 28 °C

5.0000 V          0.002 + 0.002      0.008 + 0.002                              0.01 + 0.002            0.001 + 0.002

10.0000 V         0.002 + 0.001      0.008 + 0.002                              0.01 + 0.002            0.001 + 0.002

                                                                                                                                                       1-21
8845A/8846A

Calibration Manual

      Measurement Rates (IEEE488[4])

                    Function        Digits        Setting                               Integration Time             Measurements/Second[1]

                                                                                        60 Hz (50 Hz)                8845A      8846A

      DC Volts, DC Current, and     6½            100 NPLC                              1.67 (2) s                   0.6 (0.5)  0.6 (0.5)

      Resistance                    6½            10 NPLC                               167 (200) ms                 6 (5)      6 (5)

                                    5½            1 NPLC                                16.7 (20) ms                 60 (50)    60 (50)

                                    5½            0.2 NPLC                              3.3 ms                       270        270

                                    4½            0.02 NPLC                             500 us                       995        995

      AC Voltage and AC Current[2]  6½            3 Hz                                                               0.47       0.47

                                    6½            20 Hz                                                              1.64       1.64

                                    6½            200 Hz[3]                                                          4.5        4.5

      Frequency and Period          6½                                    1s                                         1          1

                                    5½            100 ms                                                             9.8        9.8

                                    4½            10 ms                                                              80         80

      Capacitance                   6 1/2                                                                            NA         2

             [1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.

             [2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.

             [3] For remote operation or external trigger using default settling delay

             [4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the

                    baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus

                    has a maximum measurement rate of 963 measurement/s.

      Measurement Uncertainty

                    The Meter's measurement uncertainties are expressed in the form ( % of reading + % of

                    range ). In addition to the reading error and range error, you may need to add additional

                    errors for certain operating conditions. If the Meter is operated outside the temperature

                    range specified, an additional temperature coefficient error must be applied. For dc

                    voltage, dc current, and resistance measurements, apply an additional reading-speed

                    error. For ac voltage and ac current measurements, apply an additional low frequency

                    error or crest factor error.

                    The "% of reading" error varies according to the input level on the selected range. This

                    error is expressed in percent of input measurement. The “% of range” error represents the

                    floor noise of the range and represents the lowest meaningful resolution for that range.

                    The following example shows the reading error applied to the Meter's 24-hour 10 Vdc

                    specification: 0.0013% of input + 0.0004% of range.

                    Assuming the Meter is set to the 10V range with an input voltage of 1 V, the

                    measurement uncertainty would be: +/- [(0.0013% x 1V) + (.0004% x 10V)].

                    Permissible High Value = 1 + 0.000053V = 1.000053 V

                    Permissible Low Value = 1 - 0.000053V = 0.999947 V

1-22
                                                                Introduction and Specifications  1

                                                     Interpreting Accuracy Specifications

Interpreting Accuracy Specifications

The following sections provide a clearer understanding of specifications over time and

with temperature variations.

24-Hour Accuracy

The 24-hour accuracy specification indicates the Meter's relative accuracy over its full

measurement range for short time intervals and within a stable environment. Short-term

accuracy is usually specified for a 24-hour period and for a ±1 °C temperature range.

90-Day and 1-Year Accuracy

The longer duration accuracy specifications are valid at the calibration temperature (Tcal)

±5 °C temperature range. These specifications include the initial calibration errors plus

the Meter's long-term drift errors.

Temperature Coefficients

Accuracy is usually specified at the calibration temperature (Tcal) ±5 °C temperature

range. This is a common temperature range for many operating environments. Add

additional temperature coefficient errors to the accuracy specification if the Meter is

operated outside the ±5 °C temperature range (the specification is per °C).

Configuring for Highest Accuracy Measurements

The measurement configurations shown below assume that the Meter is in its power–on

or reset state. It is also assumed that auto-ranging is enabled to ensure proper full-scale

range selection.

DC Voltage, DC Current, and Resistance Measurements

Select NPLC and 100 (NPLCs) for highest instrument resolution and accuracy.

For the best dc voltage accuracy, set INPUT HIGH INPUT Z (impedance) to GOhm (for

the 100 mV, 1 V, and 10 V ranges).

For the best resistance measurement accuracy, use the 4-wire ohms function (4W).

For 2-wire ohms, dc voltage and dc current measurements, set AUTOZERO to ON to

remove thermal EMF and offset errors.

Zero the test lead resistance for 2-wire and 4-wire ohms measurements and zero to

remove any interconnection offset for dc voltage measurements.

AC Voltage and AC Current Measurements:

Set the AC FILTER to 3 Hz: SLOW.

Frequency and Period Measurements:

Set the GATE TIME to 1 sec.

                                                                                                 1-23
8845A/8846A

Calibration Manual

1-24
       Chapter 2

General Maintenance

Title                                                                                                                   Page

Introduction........................................................................................................    2-3

Warranty Repairs and Shipping Information .....................................................                         2-3

General Maintenance Information .....................................................................                   2-3

Required Equipment......................................................................................                2-3

Power Requirements......................................................................................                2-3

Static Safe Handling ......................................................................................             2-3

Cleaning .............................................................................................................  2-4

Fuse Replacement ..............................................................................................         2-4

Line-Power Fuse............................................................................................             2-4

Current-Input Fuses .......................................................................................             2-5

If the Meter Does Not Turn On..........................................................................                 2-7

Display Tests......................................................................................................     2-7

Disassembly Procedures ....................................................................................             2-7

General Disassembly .....................................................................................               2-7

Main Chassis Disassembly ............................................................................                   2-8

Front Panel Disassembly ...............................................................................                 2-8

Assembly Procedures.........................................................................................            2-8

                                                                                                                              2-1
8845A/8846A

Calibration Manual

2-2
                                                                                     General Maintenance            2

                                                                                                      Introduction

Introduction

        This chapter provides handling, cleaning, fuse replacement, disassembly, and assembly

        instructions for the Meter.

Warranty Repairs and Shipping Information

        If your meter is still under warranty, see the warranty information at the front of this

        manual for instructions on returning the unit. A list of Fluke telephone numbers and the

        website address can be found in the “How to Contact Fluke” section of Chapter 4.

General Maintenance Information

        The following sections describe how to maintain the Meter.

Required Equipment

        Equipment required for calibration, troubleshooting, and repair of the Meter is listed in

        Table 3-1.

Power   Requirements

                                     XW Warning

        To avoid electric shock, connect the Meter’s power cord to a

        power receptacle with earth ground.

        The Meter operates on power distribution standards found throughout the world, and

        must be set up to operate on the correct line voltage power it. The Meter is packed ready

        for use with a line voltage determined at the time of ordering. If the selected line voltage

        does not match the power the Meter will be plugged into, then the Meter’s line voltage

        setting must be changed and the line fuse possibly replaced. See the 8845A/8846A Users

        Manual for information on switching the Meter’s line voltage.

        If you have not already done so, plug the line cord into the connector on the rear of the

        Meter.

Static  Safe Handling

        All integrated circuits, including surface mounted ICs, are susceptible to damage from

        electrostatic discharge (ESD). Modern integrated circuit assemblies are more susceptible

        to damage from ESD than ever before.

        Integrated circuits today can be built with circuit lines less than one micron thick,

        allowing more than a million transistors on a 1/4-inch square chip. These submicron

        structures are sensitive to static voltages under 100 volts. This much voltage can be

        generated on a dry day by simply moving your arm. A person can develop a charge of

        2,000 volts by walking across a vinyl tile floor, and polyester clothing can easily generate

        5,000 to 15,000 volts during movement against the wearer. These low voltage static

        problems are often undetected, because a static charge must be in the 30,000 to 40,000

        volt range before a person will feel a shock.

        Most electronic components manufactured today can be degraded or destroyed by ESD.

        While protection networks are used in CMOS devices, they can only reduce, not

        eliminate, component susceptibility to ESD.

        ESD may not cause an immediate failure in a component; a delayed failure or

        "wounding" effect is caused when the semiconductor’s insulation layers or junctions are

                                                                                                                    2-3
8845A/8846A

Calibration Manual

                    punctured. The static problem is thus complicated in that failure may occur anywhere

                    from two hours to six months after the initial damage.

                    Two failure modes are associated with ESD. First, a person who has acquired a static

                    charge can touch a component or assembly and cause a transient discharge to pass

                    through the device. The resulting current ruptures the junctions of a semiconductor. The

                    second failure mode does not require contact with another object. Simply exposing a

                    device to the electric field surrounding a charged object can destroy or degrade a

                    component. MOS devices can fail when exposed to static fields as low as 30 volts.

                    Observe the following rules for handling static-sensitive devices:

                    1.  Handle all static-sensitive components in a static-safe work area.

                        Use grounded static-control table mats on all repair benches, and always wear a

                        grounded wrist strap. Handle boards by their nonconductive edges only. Store plastic,

                        vinyl, and Styrofoam objects outside the work area.

                    2.  Store and transport all static-sensitive components and assemblies in static shielding

                        bags or containers.

                    Static-shielding bags and containers protect components and assemblies from direct static

                    discharge and external static fields. Store components in their original packages until they

                    are ready for use.

     Cleaning

                                             XW Warning

                        To avoid electric shock or damage to the Meter, never get water

                        inside the meter.

                                                           W Caution

                        To avoid damaging the Meter’s housing, do not apply solvents

                        to the Meter.

                    If the Meter requires cleaning, wipe it down with a cloth lightly dampened with water or

                    a mild detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-

                    based fluids to wipe down the meter.

     Fuse Replacement

                    The Meter employs fuses to protect both the line-power and current measurement inputs.

     Line-Power Fuse

                    The Meter has a line-power fuse in series with the power supply. Table 2-1 indicates the

                    proper fuse for each of the four line voltage selections. This fuse is located on the rear

                    panel.

                    To replace this fuse:

                    1.  Unplug the power cord from the Meter and remove any test leads.

                    2.  Remove the fuse holder by inserting a small screwdriver blade in the narrow recess to

                        the left of the fuse holder and pry to the right until the holder pops out, as shown in

                        Figure 2-1. The Meter is shipped with a replacement fuse of the same rating as the

                        fuse installed in the fuse block.

2-4
                                                                               General Maintenance  2

                                                                               Fuse Replacement

3.  Remove the fuse and replace with one rated appropriately for the selected  line-power

    voltage. See Table 2-1 for fuse ratings with specific line voltage.

4.  Replace the selector block back into the fuse holder.

                               XW Warning

    To avoid electric shock or fire, do not use makeshift fuses or

    short-circuit the fuse holder. Use only Fluke fuses.

                     Table 2-1. Line Voltage to Fuse Rating

    Line Voltage Selection     Fuse Rating                     Fluke Part No.

       100                  0.25 A, 250 V (slow blow)               166306

       120                  0.25 A, 250 V (slow blow)               166306

       220                  0.125 A, 250 V (slow blow)              166488

       240                  0.125 A, 250 V (slow blow)              166488

                                                               120

                                                                               caw0201f.eps

                            Figure 2-1. Line Fuse Replacement

Current-Input Fuses

The 100 mA and 10 A inputs are protected by user-replaceable fuses.

    •  The 100 mA input is protected by a fuse (A1F2) rated at 440 mA, 1000 V (fast

       blow), 10,000 A minimum breaking capacity.

    •  The 10 A input is protected by a fuse (A1F1) rated at 11 A, 1000 V (fast blow),

       10,000 A minimum breaking capacity.

                               W Warning

    For protection against fire or arc flash, replace a blown fuse

    only with one from Fluke.

                                                                                                    2-5
8845A/8846A

Calibration Manual

                    To  test for a blown Current Input fuse:

                        1.  With the Meter powered up, plug a test lead into the VΩGER connector.

                            If the fuse is good, the Meter will read less than 200 Ω. If the fuse is blown, the

                            Meter will read overload.

                        2.  Remove the probe from the 100 mA connector and insert into the 10 A

                            connector.

                            If the fuse is good, the Meter will read less than 1 Ω. If the fuse is blown, the

                            Meter will read overload.

                    To  replace the Current Input fuses:

                        1.  Turn the Meter off, unplug the power cord from the Meter, and remove all test

                            leads.

                        2.  Turn the Meter on its back.

                        3.  Unscrew the retaining screw on the fuse access door, as depicted in Figure 2-2.

                        4.  Remove the protective cover from the fuse holders by slightly depressing the

                            back edge of the cover to unlatch it from the printed circuit board. Pull up on the

                            back edge of the cover and remove it from the fuse compartment.

                        5.  Remove the defective fuse and replace with one having the appropriate rating

                            (See table 4-1 for fuse ratings and Fluke part numbers).

                        6.  Replace the protective cover by pushing it over the fuses while aligning the

                            catches with the holes in the printed circuit board. Press the cover down until the

                            catches engage the printed circuit board.

                        7.  Replace the fuse door and secure it by tightening the retaining screw.

                                    Fuses  A1 F1

                                        A1 F2

                                                                       Bottom front

                                                                                    left corner

                                                                                                                 caw020.eps

                                        Figure 2-2. Current Input Fuse Replacement

2-6
                                                                             General Maintenance            2

                                                                             If the Meter Does Not Turn On

If  the Meter Does Not Turn On

    Use the following steps to help solve problems encountered when turning on the Meter.

    1.  Verify the Meter’s power switch is in the “On” position.

    2.  Make sure that the power cord is firmly plugged into the power module on the rear of

        the Meter.

    3.  Make sure the power source the Meter is plugged into is energized.

    4.  Ensure the Meter’s power line voltage is set to the proper value for your country. See

        the “Fuse Replacement” section earlier in this chapter for instructions on changing

        the Meter’s voltage setting.

    5.  Verify that the power-line fuse is good.

    If these steps don’t solve the problem, then contact Fluke for more help. See the

    “Contacting Fluke” section in Chapter 4 for contact information.

Display Tests

    To test the pixels on the front panel display, use the following steps.

    1.  Press I.

    2.  Press the SYSTEM softkey.

    3.  Press the DISPLAY softkey.

    All pixels of the display should be illuminated.

Disassembly Procedures

                                      XWWarning

        To avoid electric shock, disconnect the Meter from power

        before removing the cover.

        Only qualified service personnel should attempt servicing this

        Meter.

    To disassemble the Meter, a #2 Phillips screwdriver and small crescent wrench are

    required. There are three sets of disassembly instructions: general, main chassis, and front

    panel.

General Disassembly

    To disassemble the Meter:

    1.  Turn off the power by turning off the mains power at the rear of the Meter and

        removing the power cord. The front panel power key only puts the Meter in a power-

        save mode and does not remove mains power from the Meter.

    2.  Remove all cables from the Meter.

    3.  Remove the Meter bumpers by pulling from a corner and stretching the bumpers off

        the Meter.

    4.  Remove the bail by rotating the handle upright to a 90° angle from the top of the

        Meter and pull bail out from the sides of the Meter.

    5.  Remove the top cover by removing the four screws on the bottom of the chassis, and

        slide the cover towards the back of the Meter.

                                                                                                            2-7
8845A/8846A

Calibration Manual

                    6.  Remove the wedges from the front and rear input modules by rotating the top toward

                        the chassis middle and pulling up. See Figure 4-1.

                    7.  Remove both sets of screws holding the front panel to the chassis (8-32 pan head and

                        6-32 flat head undercut). See Figure 4-1.

                    8.  Gently remove the front panel by pulling it forward and set it aside.

                    9.  Remove the plastic handle caps from the chassis sides by rotating the front slightly

                        inward and pulling forward.

     Main           Chassis Disassembly

                    To disassemble the Meter’s main chassis:

                    1.  Carefully remove the transformer connectors from the main board.

                    2.  Carefully remove the screws that hold the transformer and its bracket to the chassis

                        while holding the transformer so that it cannot drop on the circuits.

                    3.  Carefully lift the transformer out of the chassis and place the transformer behind the

                        instrument with the power module leads still attached.

                    4.  Remove the jackscrews for the RS-232 and IEEE488 connectors to the chassis.

                    5.  Remove the screws holding the main circuit board to the chassis (three 6-32 pan-head

                        screws).

                    6.  Remove the main circuit board.

     Front Panel Disassembly

                    To disassemble the Meter’s front panel:

                    1.  Remove the one 6-32 pan head and two 6-32 flat head undercut screws holding the

                        display shield to the rest of the front panel assembly and remove the shield.

                    2.  Remove the thread-forming screw holding the front panel shield to the front panel

                        and remove the shield.

                    3.  Remove the three thread-forming screws from the keypad assembly and remove

                        display module.

     Assembly Procedures

                    To assemble the three parts of the Meter, follow the disassembly instructions in reverse

                    order.

2-8
       Chapter 3

Performance Test and Calibration

Title                                                                                                                 Page

Introduction........................................................................................................  3-3

Required Equipment ..........................................................................................         3-3

Test Considerations............................................................................................       3-5

Performance Tests..............................................................................................       3-5

Volts DC Verification....................................................................................             3-5

Volts DC Ratio Verification ..........................................................................                3-8

Volts AC and Frequency Verification ...........................................................                       3-9

4-Wire Ohms Verification .............................................................................                3-14

2-Wire Ohms Verification .............................................................................                3-16

2X4 Test Lead Verification Steps .................................................................                    3-18

Rear Panel Terminal Verification Steps ........................................................                       3-18

Capacitance Verification Steps (8846A only) ...............................................                           3-21

DC Current Verification Steps ......................................................................                  3-22

AC Current Verification Steps ......................................................................                  3-25

Adjustment (Calibration) ...................................................................................          3-27

Unlocking the Meter for Adjustments (Calibration) .....................................                               3-28

Unlocking the Meter for Adjustments Over a Remote Interface...................                                        3-28

Changing the Calibration Password ..............................................................                      3-28

Resetting the Calibration Password...............................................................                     3-28

Changing the Calibration Date Remotely......................................................                          3-29

Equipment for Calibration .............................................................................               3-29

Adjustment Process .......................................................................................            3-30

Aborting a Calibration Process......................................................................                  3-36

Sample Adjustment Program.........................................................................                    3-36

                                                                                                                            3-1
8845A/8846A

Calibration Manual

3-2
                                                                       Performance Test and Calibration           3

                                                                                                    Introduction

Introduction

          This chapter of the Calibration Manual provides performance tests to verify the Meter is

          operating within published specifications as well as a complete calibration procedure.

          The performance test and, if necessary, the calibration procedure can be performed both

          periodically and after service or repair.

          The performance tests can be used as an acceptance test upon receipt of the Meter. Use

          the 90-day specifications when performing an acceptance test after performing a

          calibration.

Required Equipment

          Table 3-1 lists the equipment required for performance testing and calibration of the

          Meter.

                                   Table 3-1. Required Test Equipment

Function          Instrument Type  Model                               Comments

Volts dc   Standard                Fluke 5520A               Must be characterized with 8508A

           8½ digit meter          Fluke 8508A               Used to characterize the 5520A

           4-wire short            Fluke low thermal 4-wire  Fluke PN 2653346

                                   short or equivalent

           Alternate standard[1]   Fluke 5720A

Volts ac   Standard                Fluke 5520A               Must be characterized with 8508A

           8½ digit meter          Fluke 8508A               Used to characterize the 5520A.

                                                             Note: TURs <4:1 at 1 V, 10 V, and

                                                             100 V at 20 kHz

           4-wire short            Fluke low thermal 4-wire  Fluke PN 2653346

                                   short or equivalent

           Alternate standard[1]   Fluke 5720A

Frequency  Standard                Fluke 5520A

           Alternate standard      Fluke 5520A with any

                                   scope option

           Alternate standard      Function generator        Specifications include0.075 %

                                                             frequency accuracy from 3 – 40 Hz

                                                             and 0.0025 % accuracy for

                                                             frequencies up to 1 MHz

                                                                                                                  3-3
8845A/8846A

Calibration Manual

                                       Table 3-1. Required Test Equipment (cont.)

             Function  Instrument Type                                  Model         Comments

     Ohms              Standard               Fluke 5520A                      Must be characterized with 8508A

                       8½ digit meter         Fluke 8508A                      Used to characterize the 5520A

                       4-wire short           Fluke low thermal 4-wire         Fluke PN 2653346

                                              short or equivalent

                       Alternate standard     Fluke 5720A or equivalent

                       Alternate standard     Fluke 8508A-7000K                Used for calibrating/verifying 1 GΩ

                       resistor[1]            1 Gohm resistor or               range in the 8846A when a 5520A

                                              equivalent (better than          is not available.

                                              ±0.35 % maximum

                                              uncertainty)

     Capacitance       Standard               Fluke 5520A                      TURS <4:1 at 1 nF, 10 nF, 100 μF,

     (8846A only)                                                              1 mF, and 10 mF.

                       Alternate standards    1 nF, 10 nF, 100 nF, 1 μF,       Standards must be ±0.25 % and

                                              10 μF, 100 μF, 1 mF,             rated for at least 5 V

                                              10 mF, and 100 mF

                                              standards

     Current dc        Standard               Fluke 5520A                      Must be characterized with 8508A

                       8½ digit meter         Fluke 8508A                      Only used to characterize the

                                                                               5520A

                       Alternate standard[1]  Fluke 5720A with Fluke           Note: TUR <4 at 10 A

                                              5725A

     Current ac        Standard               Fluke 5520A                      Must be characterized with 8508A

                       8½ digit meter         Fluke 8508A                      Only used to characterize the

                                                                               5520A or 5720A

                       Alternate standard[1]  Fluke 5720A with Fluke           TUR at 1 A at 5 kHz < 4:1. Must

                                              5725A                            characterize with 8508A at 100 μA

                                                                               at 5 kHz. 100 μA, 1 mA, 10 mA, 100

                                                                               mA, 1 A and 2 A at 10 kHz.

     Cables            To reduce the possibility of inducing errors with ac signals picked up by the test leads,

                       use short, shielded twisted-pair PTFE-insulated test cables between the test equipment

                       and the Meter. Fluke makes a 2 foot (PN 738716) and 4 foot (PN 738724) PTFE

                       insulated test cable for this purpose.

                       Fluke USB to RS232 cable (PN 2675479) or Fluke IEEE 488 cable (2 meter, PN 708297)

             [1] Other alternate standards beside those listed can be used as long as they provide sufficient traceable [Test uncertaintity

             Ratios(TURs)) at each calibration and verification point.

3-4
                                                                  Performance Test and Calibration             3

                                                                                          Test Considerations

Test   Considerations

       For optimum performance, all test procedures should comply with the following

       recommendations:

           •  Assure the calibration ambient temperature (Tcal) is stable and between 18 °C and

              28 °C. Ideally the calibration should be performed at 23 °C ±2 °C.

           •  Assure ambient relative humidity is less than 80%.

           •  Allow a 60-minute warm-up period.

           •  Use shielded twisted-pair PTFE-insulated cables to reduce settling and noise

              errors.

           •  Keep all input cables as short as possible.

           •  Ensure that the calibration standards and test procedures used do not introduce

              additional errors.

                                               Note

           Ideally, the standards used to verify and adjust the Meter should be four

           times more accurate than each full-scale error specification of the Meter.

           •  User Fluke’s low thermal 4-Wire short for all voltages and ohmic shorts. See

              Table 3-1 for Fluke part number..

Performance Tests

       The following performance tests are provided to ensure that the Meter is in proper

       operating condition. If the Meter fails any of the performance tests, calibration

       adjustment and/or repair is needed. The performance test works best if executed in the

       sequence shown in Table 3-2.

       Each of the measurements listed in the following tests assumes the Meter is being tested

       after a one-hour warm-up in an environment with an ambient temperature of 18 to 28ºC

       and a relative humidity of less than 80%.

                                               Note

           All instrument settings for verification use power up conditions except as

           noted by the verification step.

Volts  DC Verification

       Connect the Meter to the test equipment as shown in Figure 3-1 and, depending on which

       meter you are calibrating, apply the voltages listed in Table 3-2 or Table 3-3. Some

       verification test steps are firmware dependant. To determine the firmware revision of the

       UUT:

       1.  Press I,

       2.  Press the softkey labeled System.

       3.  Press the softkey labeled Version.

       Note the OutG SW value.

       Verification forms can be found in Appendix A which can be copied and used to record

       each meter reading.

                                                                                                               3-5
8845A/8846A

Calibration Manual

                                                                                                                                     Note

                          For the     zero              (0)    V tests, use the 4-wire short to short the Hi/Lo                                                                                                                   and  Sense

                          inputs.

                                                                                INPUT                SENSE                           8846A 6-1/2 DIGIT PRECISION MULTIMETER

                                                                            V                            4W

                                                                     2W/4W

                                                                     HI                                         HI

                                                                     1000   V   CCFAAUTTSIIEI D              300 V                   +1.000002                                          V
                                                                     600    V

                                                                     LO                                         LO

                                                                                                 1V

                                                                                FUSED

                                                                     100                                        10 A     BACK        F1        F2               F3                  F4                 F5            MEMORY

                                                                     mA

                                                                                                                                         DC V  AC V    DC I       AC I                  TRIG           INSTR

                                                                                REAR                     FRONT                                                                                      SETUP

                                                                                                                         RANGE

                                                                                                                                               FREQ             TEMP         ZERO       ANALYZE        MEAS

                                                                                                                                               PERIOD                                               SETUP

                                                                                                 5520A CALIBRATOR

                                                                            NORMAL                           AUX         SCOPE

                                                                            V,  ,  , RTD             A,  -SENSE, AUX V

                                                                                                                         OUT                                                                                              PREV

                                                                                                     HI                                        STBY          OPR             EARTH      SCOPE          EXGRD              MENU

                                                                                FUSED                                                                                                            µ                dBm        sec

                                                                                                 LO                                            7             8               9                      m                V          Hz

                                                                                                                         TRIG

                                                                                                                                               4             5               6                   n                W          F

                                                                                                                                                                                                    k                A          C

                                                                                                                                               1             2               3                   p

                                                                                                 GUARD              20A                                                                             M                           F

                                                                                                                                               +/            0               •                   SHIFT                  ENTER

                                                                            20V PK MAX                          TC    20V PK MAX

                                             UUT                                                             A.                   :  OFF               :     OFF                                                        5720A Calibrator

                                             INPUT      SENSE                                                            EX SNS                EX GRD                                                             OUTPUT            SENSE

                                          V             4W

                                   2W/4W                                                                                                                                                                               V  A         V

                                   HI                          HI

                                   1000 V CAT I             300 V                                                                                                                                             HI                       HI  WIDEBAND
                                   600V CAT II

                                   LO                          LO

                                                    1V

                                             FUSED                                                                                                                                                            LO                       LO

                                   100                         10 A

                                   mA

                                             REAR       FRONT                                                                                                                                                 HI

                                                                                                                                                                                                                     AUX            GUARD  GROUND

                                                                                                                                                                                                              CURRENT

                                                 NC

                                                                                                                                                                                                                                                          caw023.eps

                                                    Figure           3-1. DC Volts Test Equipment                                                                            Setup            with 5520A

                                                                   Table 3-2. 8846A                                               DC Volts Verification Steps

                Nominal                                                                                      90-day               Test Limits                                                                                1-year        Test Limits

                Input (V)                 Range

                                                                                                     High                                            Low                                                          High                               Low

             0                     0.100                             3.5 μV                                                          -3.5 μV                                            3.5 μV                                             -3.5 μV

             100.0 mV[1]           0.100                             100.006 mV                                                      99.994 mV                                          100.0072 mV                                        99.9928 mV

             -100.0 mV[1]          0.100                             -99.994 Mv                                                      -100.006 mV                                        -99.9928 mV                                        -100.0072 mV

             0V                    1                                 7.0 μV                                                          -7.0 μV                                            7.0 μV                                             -7.0 μV

3-6
                                                                                   Performance Test and Calibration  3

                                                                                              Performance Tests

                            Table 3-2. 8846A DC Volts Verification Steps (cont.)

Nominal                                  90-day Test Limits                           1-year  Test Limits

Input (V)               Range

                                High         Low                                   High                Low

1 V[1]               1          1.000025 V   0.999975 V                    1.000032 V         0.999968 V

-1 V[1]              1          -0.999975 V  -1.000025 V                   -0.999968 V        -1.000032 V

0V                   10         50.0 μV      -50.0 μV                      50.0 μV            -50.0 μV

5 V[1]               10         5.000140 V   4.999860 V                    5.000170 V         4.999830 V

-5 V[1]              10         -4.999860 V  -5.000140 V                   -4.999830 V        -5.000170 V

10 V[1]              10         10.000230 V  9.999770 V                    10.000290 V        9.999710 V

-10 V[1]             10         -9.999770 V  -10.000230 V                  -9.999710 V        -10.000290 V

0V                   100        600.0 μV     -600.0 μV                     600.0 μV           -600.0 μV

100 V[1]             100        100.0033 V   99.9967 V                     100.0044 V         99.9956 V

-100 V[1]            100        -99.9967 V   -100.0033 V                   -99.9956 V         -100.0044 V

0V                   1000       10.0 mV      -10.0 mV                      10.0 mV            -10.0 mV

1 kV[1]              1000       1000.0410 V  999.9590 V                    1000.0510 V        999.9490 V

-1 kV[1]             1000       -999.9590 V  -1000.0410 V                  -999.9490 V        -1000.0510 V

[1] 5520A  must  be  used with  8508A to obtain suitable test uncertainty  ratio.

                                Table 3-3. 8845A DC Volts Verification Steps

Nominal                                  90-day Test Limits                           1-year  Test Limits

Input (V)               Range

                                High         Low                                   High                Low

0V                   0.100      3.5 μV       -3.5 μV                       3.5 μV             -3.5 μV

100.0 mV[1]          0.100      100.0075 mV  99.9925 mV                    100.0085 mV        99.9915 mV

-100.0 mV[1]         0.100      -99.9925 mV  -100.0075 mV                  -99.9915 mV        -100.0085 mV

0V                   1          7.0 μV       -7.0 μV                       7.0 μV             -7.0 μV

1 V[1]               1          1.000037 V   999.963 mV                    1.000047 V         999.953 mV

-1 V[1]              1          -999.963 mV  -1.000037 V                   -999.953 mV        -1.000047 V

0V                   10         50.0 μV      -50.0 μV                      50.0 μV            -50.0 μV

5 V[1]               10         5.00015 V    4.99985 V                     5.000225 V         4.999775 V

-5 V[1]              10         -4.99985 V   -5.00015 V                    -4.999775 V        -5.000225 V

10 V[1]              10         10.00025 V   9.99975 V                     10.0004 V          9.9996 V

-10 V                10         -9.99975 V   -10.00025 V                   -9.9996 V          -10.0004 V

0V                   100        600.0 μV     -600.0 μV                     600.0 μV           -600.0 μV

100 V                100        100.0041 V   99.9959 V                     100.0051 V         99.9949 V

-100 V               100        -99.9959 V   -100.0041 V                   -99.9949 V         -100.0051 V

0V                   1000       10.0 mV      -10.0 mV                      10.0 mV            -10.0 mV

1 kV                 1000       1.000045 kV  999.955E+0                    1.000055 kV        999.945 V

-1 kV                1000       -999.955 V   -1.000045 kV                  -999.945 V         -1.000055 kV

[1] 5520A must   be  used with  8508A to obtain suitable test uncertainty  ratio.

                                                                                                                     3-7
8845A/8846A

Calibration Manual

     Volts DC Ratio Verification

                    Connect the Meter to the test equipment as shown in Figure 3-2 and depending on which

                    meter you are calibrating, apply the voltages listed in Table 3-4 or Table 3-5. Verification

                    forms can be found in Appendix A which can be copied and used to record each meter

                    reading.

                                                                                                                                Note

                    Note that the Ratio function is only available in Outguard SW version 2.0

                    and later.

                                                                          INPUT                SENSE                        8846A 6-1/2 DIGIT PRECISION MULTIMETER

                                                                      V                            4W

                                                              2W/4W

                                                                  HI                                      HI

                                                                  1000 V  CFCUAATTSIIEI D          FUSE3D00 V                   +0.938193
                                                                  600 V

                                                                  LO                                      LO

                                                                                           1V

                                                                          FUSED

                                                              100                                         10 A      BACK    F1        F2               F3                  F4                 F5          MEMORY

                                                              mA

                                                                                                                                DC V  AC V    DC I       AC I                  TRIG           INSTR

                                                                          REAR                     FRONT                                                                                   SETUP

                                                                                                                    RANGE

                                                                                                                                      FREQ             TEMP         ZERO       ANALYZE        MEAS

                                                                                                                                      PERIOD                                               SETUP

                                                                                           5520A CALIBRATOR

                                                                      NORMAL                           AUX          SCOPE

                                                                      V,  ,  , RTD             A,  -SENSE, AUX V

                                                                                                                    OUT                                                                                     PREV

                                                                                               HI                                     STBY          OPR             EARTH      SCOPE          EXGRD         MENU

                                                                                                                                      7             8               9                   µ            dBm          sec

                                                                                           LO                                                                                              m             V           Hz

                                                                                                                    TRIG

                                                                                                                                      4             5               6                   n            W            F

                                                                                                                                                                                           k             A           C

                                                                                                                                      1             2               3                   p

                                                                                           GUARD               20A                                                                         M                         F

                                                                                                                                      +/            0               •                   SHIFT               ENTER

                                                                      20V PK MAX                          TC    20V PK MAX

                                        UUT                                                                                                                                                                       5720A Calibrator

                                        INPUT      SENSE                                                                                                                                                 OUTPUT          SENSE

                                     V             4W

                              2W/4W                                                                                                                                                                         V     A      V

                              HI                              HI

                              1000 V CAT I             300 V                                                                                                                                         HI                     HI  WIDEBAND
                              600V CAT II

                              LO

                                               1V             LO

                                        FUSED                                                                                                                                                 LO                            LO

                              100                         10 A

                              mA

                                        REAR       FRONT                                                                                                                                             HI

                                                                                                                                                                                                            AUX          GUARD  GROUND

                                                                                                                                                                                                     CURRENT

                                                                                                                                                                                                                                          caw037.eps

                    Figure 3-2. DC Volts Ratio Test                                                                         Equipment Setup with 5520A and 5720A

3-8
                                                                    Performance Test and Calibration        3

                                                                                         Performance Tests

                             Table 3-4. 8846A DC Volts Ratio Verification Steps

Nominal                      90-day Test Limits                                  1-year  Test Limits[1]

Input (V)           Range

                             High        Low                        High                 Low

100 mV[2]           0.100 V  1.00012 V   0.99988 V                  1.000144 V           0.999856 V

1 V[2]              1V       1.00005 V   0.99995 V                  1.000064 V           0.999936 V

-10 V               10 V     1.000046 V  0.999954 V                 1.000058 V           0.999942 V

[1] Using 5520A     output to the Hi/Lo and Sense Hi/Lo terminals.

[2] Optional test.

                             Table 3-5. 8845A DC Volts Ratio Verification Steps

Nominal                      90-day Test Limits                                  1-year  Test Limits[1]

Input (V)           Range

                             High        Low                        High                 Low

100 mV[2]           0.100 V  1.00015 V   0.99985 V                  1.00017 V            0.99983 V

1 V[2]              1V       1.000074 V  0.999926 V                 1.000094 V           0.999906 V

-10 V               10 V     1.00005 V   0.99995 V                  1.00008 V            0.99992 V

[1] Using 5520A     output to the Hi/Lo and Sense Hi/Lo terminals.

[2] Optional test.

Volts AC and Frequency Verification

        Connect the Meter to the test equipment as shown in Figure 3-3 and, depending on which

        meter you are calibrating, apply the voltage listed in Table 3-6 or Table 3-7. Verification

        forms can be found in Appendix A which can be copied and used to record each meter

        reading.

                                                                                                            3-9
8845A/8846A

Calibration Manual

                                                                 INPUT               SENSE                        8846A 6-1/2 DIGIT PRECISION MULTIMETER

                                                             V                           4W

                                                      2W/4W

                                                      HI                                        HI

                                                      1000 V     CCFAAUTTSIIIED              300 V                    0.999944                                       V
                                                      600 V

                                                      LO                                        LO

                                                                                 1V

                                                                 FUSED

                                                      100                                       10 A     BACK     F1        F2               F3                  F4                 F5              MEMORY

                                                      mA

                                                                                                                      DC V  AC V    DC I       AC I                  TRIG           INSTR

                                                                 REAR                    FRONT                                                                                   SETUP

                                                                                                         RANGE

                                                                                                                            FREQ             TEMP         ZERO       ANALYZE        MEAS

                                                                                                                            PERIOD                                               SETUP

                                                                                 5520A CALIBRATOR

                                                             NORMAL                          AUX         SCOPE

                                                             V,  ,  , RTD            A,  -SENSE, AUX V

                                                                                                         OUT                                                                                           PREV

                                                                                     HI                                     STBY          OPR             EARTH      SCOPE          EXGRD              MENU

                                                                 FUSED                                                                                                        µ                dBm          sec

                                                                                 LO                                         7             8               9                      m                V            Hz

                                                                                                         TRIG

                                                                                                                            4             5               6                   n                W            F

                                                                                                                                                                                 k                A            C

                                                                                                                            1             2               3                   p

                                                                                 GUARD              20A                                                                          M                             F

                                                                                                                            +/            0               •                   SHIFT                  ENTER

                                                             20V PK MAX                         TC    20V PK MAX

                              UUT                                                            A.                   : OFF             : OFF                                                            5720A Calibrator

                              INPUT      SENSE                                                           EX SNS             EX GRD                                                             OUTPUT              SENSE

                           V             4W

                    2W/4W                                                                                                                                                                           V  A           V

                    HI                          HI

                    1000 V CAT I             300 V                                                                                                                                         HI                         HI  WIDEBAND
                    600V CAT II

                    LO                          LO

                                     1V

                              FUSED                                                                                                                                                        LO                         LO

                    100                         10 A

                    mA

                              REAR       FRONT                                                                                                                                             HI

                                                                                                                                                                                                  AUX              GUARD  GROUND

                                                                                                                                                                                           CURRENT

                                  NC

                                                                                                                                                                                                                                    caw024.eps

                                         Figure                  3-3.                AC Volts                     Test      Equipment Setup with 5520A

3-10
                                                                                 Performance Test and Calibration  3

                                                                                            Performance Tests

                            Table 3-6. 8846A AC Volts Verification Steps

    Nominal Input                       90-day Test Limits                       1-year     Test Limits

                            Range

Ampl.        Freq.                      High       Low                           High       Low

100.0 mV[1]  10 Hz      0.100           100.09 mV  99.91 mV                      100.1 mV   99.9 mV

100.0 mV     20 kHz     0.100           100.09 mV  99.91 mV                      100.1 mV   99.9 mV

100.0 mV     50 kHz     0.100           100.16 mV  99.84 mV                      100.17 mV  99.83 mV

100.0 mV     100 kHz    0.100           100.68 mV  99.32 mV                      100.68 mV  99.32 mV

100.0 mV     300 kHz    0.100           104.5 mV   95.5 mV                       104.5 mV   95.5 mV

1 V[1]       10 Hz      1               1.0008 V   999.2 mV                      1.0009 V   999.1 mV

1V           20 kHz     1               1.0008 V   999.2 mV                      1.0009 V   999.1 mV

1V           50 kHz     1               1.0016 V   998.4 mV                      1.0017 V   998.3 mV

1V           100 kHz    1               1.0068 V   993.2 mV                      1.0068 V   993.2 mV

1V           300 kHz    1               1.045 V    955.0 mV                      1.045 V    955.0 mV

10 V[1]      10 Hz      10              10.008 V   9.992 V                       10.009 V   9.991 V

10 V[1]      20 kHz     10              10.008 V   9.992 V                       10.009 V   9.991 V

10 V         50 kHz     10              10.016 V   9.984 V                       10.017 V   9.983 V

10 V         100 kHz    10              10.068 V   9.932 V                       10.068 V   9.932 V

3V           300 kHz    10              3.17 V     2.83 V                        3.17 V     2.83 V

100 V        45 Hz      100             100.08 V   99.92 V                       100.09 V   99.91 V

100 V[1]     20 kHz     100             100.08 V   99.92 V                       100.09 V   99.91 V

100 V        50 kHz     100             100.16 V   99.84 V                       100.17 V   99.83 V

100 V[1]     100 kHz    100             100.68 V   99.32 V                       100.68 V   99.32 V

1000 V       45 Hz      1000            1000.800   999.200                       1.0009 kV  999.1 V

1000 V       1 kHz      1000            1000.800   999.200                       1.0009 kV  999.1 V

1000 V[1]    10 kHz     1000            1000.800   999.200                       1.0009 kV  999.1 V

320 V        20 kHz     1000            320.460    319.540                       320.492 V  319.508 V

320 V        50 kHz     1000            320.852 V  319.148 V                     320.884 V  319.116 V

320 V        100 kHz    1000            322.72 V   317.28 V                      322.72 V   317.28 V

[1] 5520A must be used  with 8508A  to  obtain suitable test uncertainty ratio.

                                                                                                                   3-11
8845A/8846A

Calibration Manual

                                           Table 3-7. 8845A AC Volts Verification Steps

                     Nominal Input                90-day Test Limits                         1-year     Test Limits

                                           Range

             Ampl.        Freq.                   High       Low                             High       Low

             100.0 mV[1]  10 Hz            0.100  100.09 mV  99.91 mV                        100.1 mV   99.9 mV

             100.0 mV     20 kHz           0.100  100.09 mV  99.91 mV                        100.1 mV   99.9 mV

             100.0 mV     50 kHz           0.100  100.16 mV  99.84 mV                        100.17 mV  99.83 mV

             100.0 mV     100 kHz          0.100  100.68 mV  99.32 mV                        100.68 mV  99.32 mV

             100.0 mV     300 kHz          0.100  104.5 mV   95.5 mV                         104.5 mV   95.5 mV

             1 V[1]       10 Hz            1      1.0008 V   999.2 mV                        1.0009 V   999.1 mV

             1V           20 kHz           1      1.0008 V   999.2 mV                        1.0009 V   999.1 mV

             1V           50 kHz           1      1.0016 V   998.4 mV                        1.0017 V   998.3 mV

             1V           100 kHz          1      1.0068 V   993.2 mV                        1.0068 V   993.2 mV

             1V           300 kHz          1      1.045 V    955.0 mV                        1.045 V    955.0 mV

             10 V[1]      10 Hz            10     10.008 V   9.992 V                         10.009 V   9.991 V

             10 V[1]      20 kHz           10     10.008 V   9.992 V                         10.009 V   9.991 V

             10 V         50 kHz           10     10.016 V   9.984 V                         10.017 V   9.983 V

             10 V         100 kHz          10     10.068 V   9.932 V                         10.068 V   9.932 V

             3V           300 kHz          10     3.17 V     2.83 V                          3.17 V     2.83 V

             100 V        45 Hz            100    100.08 V   99.92 V                         100.09 V   99.91 V

             100 V[1]     20 kHz           100    100.08 V   99.92 V                         100.09 V   99.91 V

             100 V        50 kHz           100    100.16 V   99.84 V                         100.17 V   99.83 V

             100 V[1]     100 kHz          100    100.68 V   99.32 V                         100.68 V   99.32 V

             750 V        45 Hz            750    750.600    749.400                         750.675 V  749.325 V

             750 V        1 kHz            750    750.600    749.400                         750.675 V  749.325 V

             750 V        10 kHz           750    750.600    749.400                         750.675 V  749.325 V

             320 V        20 kHz           750    320.385    319.615                         320.417 V  319.583 V

             320 V        50 kHz           750    320.727 V  319.273 V                       320.759 V  319.241 V

             320 V        100 kHz          750    322.52 V   317.48 V                        322.52 V   317.48 V

             [1] 5520A must be used  with  8508A to obtain suitable test uncertainty ratio.

3-12
                                                      Performance Test and Calibration               3

                                                                                  Performance Tests

                   Table  3-8. 8845A/8846A AC Volts Frequency Verification Steps

    Nominal Input         90-day Test Limits          1-year Test Limits

Ampl.            Freq.    High          Low           High                        Low

1V      10 Hz             10.00300 Hz   9.99700 Hz    10.00300 Hz                 9.99700 Hz

1V      40 Hz             40.0040 Hz    39.9960 Hz    40.00400 Hz                 39.9960 Hz

100 mV  300 kHz           300.030 kHz   29.99700 kHz  300.0300 kHz                29.99700 kHz

100 mV  1 MHz[1]          1.000100 MHz  999.9000 kHz  1.0001000 MHz               999.9000 kHz

[1] 8846A Only.

                                                                                                     3-13
8845A/8846A

Calibration Manual

      4-Wire        Ohms Verification

                    Connect the Meter to the test equipment as shown in Figure 3-4 and, depending on which

                    meter you are calibrating, apply the resistance listed in Table 3-9 or Table 3-10.

                    Verification forms can be found in Appendix A which can be copied and used to record

                    each meter reading.

                                                                                                                              Note

                    For zero (0) ohms tests, use the 4-wire short to short the Hi/Lo and Sense

                    inputs.

                                                                        INPUT                SENSE                        8846A     6-1/2 DIGIT  PRECISION  MULTIMETER

                                                                    V                            4W

                                                             2W/4W

                                                               HI                                       HI

                                                               1000 V   CFCUAATTSIIEI D          FUSE3D00 V                                                                       w4
                                                               600 V

                                                               LO                                       LO

                                                                                         1V

                                                                        FUSED

                                                             100                                        10 A      BACK    F1        F2                         F3             F4                 F5          MEMORY

                                                             mA

                                                                                                                              DC V  AC V         DC I            AC I             TRIG           INSTR

                                                                        REAR                     FRONT                                                                                        SETUP

                                                                                                                  RANGE

                                                                                                                                    FREQ                       TEMP     ZERO      ANALYZE        MEAS

                                                                                                                                    PERIOD                                                    SETUP

                                                                                         5520A CALIBRATOR

                                                                    NORMAL                           AUX          SCOPE

                                                                    V,  ,  , RTD             A,  -SENSE, AUX V

                                                                                                                  OUT                                                                                          PREV

                                                                                             HI                                     STBY                    OPR        EARTH      SCOPE          EXGRD         MENU

                                                                        FUSED                        FUSED                                                                                 µ            dBm          sec

                                                                                         LO                                         7                       8           9                     m             V           Hz

                                                                                                                  TRIG

                                                                                                                                    4                       5           6                  n            W            F

                                                                                                                                                                                              k             A           C

                                                                                                                                    1                       2           3                  p

                                                                                         GUARD               20A                                                                              M                         F

                                                                                                                                    +/                      0           •                  SHIFT               ENTER

                                                                    20V PK MAX                          TC    20V PK MAX

                                       UUT                                                                                                                                                                           5720A Calibrator

                                       INPUT      SENSE                                                                                                                                                     OUTPUT          SENSE

                                    V             4W

                             2W/4W                                                                                                                                                                             V     A      V

                             HI                          HI

                             1000 V CAT I             300 V                                                                                                                                             HI                     HI  WIDEBAND
                             600V CAT II

                             LO                          LO                                                                                                                       NC

                                              1V                        NC

                                       FUSED                                                                                                                                                            LO                     LO

                             100                         10 A

                             mA

                                       REAR       FRONT                                                                                                                                                 HI

                                                                                                                                                                                                               AUX          GUARD  GROUND

                                                                                                                                                                                                        CURRENT

                                                  NC                                                                                                                                                              NC

                                                                                                                                                                                                                                             caw025.eps

                                                      Figure 3-4. 4-Wire Ohms                                                       Test                    Equipment                            Setup

3-14
                                                                                    Performance    Test and Calibration  3

                                                                                                   Performance Tests

                              Table 3-9. 8846A 4-Wire Ohms Verification Steps

Nominal                          90-day Test Limits                                  1-year  Test  Limits

     Input             Range

                                 High             Low                               High               Low

0 Ω[1]             10            3.0 mΩ       0Ω                            3.0 mΩ                 0Ω

10 Ω[1]            10            10.00380 Ω   9.99620 Ω                     10.00400 Ω             9.99600 Ω

0 Ω[2]             100           4.0 mΩ       0Ω                            4.0 mΩ                 0Ω

100 Ω[1]           100           100.0120 Ω   99.9880 Ω                     100.01400 Ω            99.98600 Ω

0Ω                 1000          10.0 mΩ      0Ω                            10.0 mΩ                0Ω

1 kΩ               1000          1000.090 Ω   999.910                       1.00011000 kΩ          999.89000

0Ω                 10000         100.0 mΩ     0Ω                            100.0 mΩ               0Ω

10 kΩ              10000         10.00090 kΩ  9999.10                       1.000110 kΩ            9998.90

0Ω                 100000        1.000000     0Ω                            1.000000               0Ω

100 kΩ             100000        100.0090 kΩ  99991.0                       10.00110 kΩ            99989.0

[1]   5520A  must  be used with  8508A to obtain suitable test uncertainty  ratio.

                           Table 3-10. 8845A 4-Wire Ohms Verification Steps

Nominal                          90-day Test Limits                                  1-year  Test  Limits

     Input             Range

                                 High             Low                               High               Low

0 Ω[1]             100           4.0 mΩ       0Ω                            4.0 mΩ                 0Ω

100 Ω[1]           100           100.0120 Ω   99.9880 Ω                     100.01400 Ω            99.98600 Ω

0Ω                 1000          10.0 mΩ      0Ω                            10.0 mΩ                0Ω

1 kΩ               1000          1.00009 kΩ   999.910 Ω                     1.000110 kΩ            999.89000 Ω

0Ω                 10000         100.0 mΩ     0Ω                            100.0 mΩ               0Ω

10 kΩ              10000         10.00090 kΩ  9999.10 Ω                     10.00110 kΩ            9998.90 Ω

0Ω                 100000        1.000000 Ω   0Ω                            1.000000 Ω             0Ω

100 kΩ             100000        100.0090 kΩ  99.9910 kΩ                    100.0110 kΩ            99.9890 kΩ

[1]   5520A  must  be used with  8508A to obtain suitable test uncertainty  ratio.

                                                                                                                         3-15
8845A/8846A

Calibration Manual

      2-Wire        Ohms Verification

                    Connect the Meter to the test equipment as shown in Figure 3-5 and, depending on which

                    meter you are calibrating, apply the resistance listed in Table 3-11 or Table 3-12.

                    Verification forms can be found in Appendix A which can be copied and used to record

                    each meter reading.

                                                                                                                             Note

                    For zero (0) ohms tests, use the 4-wire short to short the Hi/Lo and Sense

                    inputs.

                                                                        INPUT               SENSE                        8846A     6-1/2 DIGIT  PRECISION  MULTIMETER

                                                                    V                           4W

                                                             2W/4W

                                                             HI                                        HI

                                                             1000 V     CCFAAUTTSIIIED              300 V
                                                             600 V

                                                             LO                                        LO

                                                                                        1V

                                                                        FUSED

                                                             100                                       10 A     BACK     F1        F2                         F3             F4                 F5         MEMORY

                                                             mA

                                                                                                                             DC V  AC V         DC I            AC I             TRIG           INSTR

                                                                        REAR                    FRONT                                                                                        SETUP

                                                                                                                RANGE

                                                                                                                                   FREQ                       TEMP     ZERO      ANALYZE        MEAS

                                                                                                                                   PERIOD                                                    SETUP

                                                                                        5520A CALIBRATOR

                                                                    NORMAL                          AUX         SCOPE

                                                                    V,  ,  , RTD            A,  -SENSE, AUX V

                                                                                                                OUT                                                                                          PREV

                                                                                            HI                                     STBY                    OPR        EARTH      SCOPE          EXGRD        MENU

                                                                        FUSED                                                                                                             µ            dBm         sec

                                                                                        LO                                         7                       8           9                     m            V           Hz

                                                                                                                TRIG

                                                                                                                                   4                       5           6                  n            W           F

                                                                                                                                                                                             k            A           C

                                                                                                                                   1                       2           3                  p

                                                                                        GUARD              20A                                                                               M                        F

                                                                                                                                   +/                      0           •                  SHIFT              ENTER

                                                                    20V PK MAX                         TC    20V PK MAX

                                       UUT                                                                                                                                                                   5720A Calibrator

                                       INPUT      SENSE                                                                                                                                                      OUTPUT       SENSE

                                    V             4W

                             2W/4W                                                                                                                                                                           V     A      V

                             HI                          HI

                             1000 V CAT I             300 V                                                                                                                                            HI                    HI  WIDEBAND
                             600V CAT II

                             LO                          LO

                                              1V

                                       FUSED                                                                                                                                                           LO                    LO

                             100                         10 A

                             mA

                                       REAR       FRONT                                                                                                                                                HI

                                                                                        NC                                                                                       NC                          AUX

                                                                                                                                                                                                          CURRENT         GUARD  GROUND

                                           NC                                                                                                                                                                      NC

                                                                                                                                                                                                                                           caw026.eps

                                                      Figure 3-5. 2-Wire Ohms Test Equipment Setup

3-16
                                                                                    Performance Test and Calibration     3

                                                                                                     Performance Tests

                                  Table 3-11. 8846A 2-Wire Ohms Verification        Steps

     Nominal                                     90-day Test Limits[1]                       1-year  Test Limits[1]

          Input           Range

                                         High             Low                              High          Low

0 Ω[2] [3]                100            4.0 mΩ       0Ω                            4.0 mΩ           0Ω

100 Ω[2] [3]              100            100.01200 Ω  99.988 Ω                      100.0140 Ω       99.9860 Ω

0Ω                        1000           10.0 mΩ      0Ω                            10.0 mΩ          0Ω

1.000 kΩ                  1000           1.00009 kΩ   999.910 Ω                     1.000110 kΩ      999.890 Ω

0Ω                        10000          0.100000 Ω   0Ω                            0.100000 Ω       0Ω

10.000 kΩ                 10000          10.00090 kΩ  9.99910 kΩ                    10.00110 kΩ      9.99890 kΩ

0Ω                        100000         1.000000 Ω   0Ω                            1.000000 Ω       0Ω

100.000 kΩ                100000         100.0090 kΩ  99.9910 kΩ                    100.0110 kΩ      99.9890 kΩ

0Ω                        1000000        10.00000 Ω   0Ω                            10.00000 Ω       0Ω

1.00000 MΩ[2]             1000000        1.000090 MΩ  0.999910 MΩ                   1.000110 MΩ      0.999890        MΩ

0 Ω[2]                    10000000       100.0000 Ω   0Ω                            100.0000 Ω       0Ω

10.0000 MΩ[2]             10000000       10.00210 MΩ  9.99790 MΩ                    10.00410 MΩ      9.995900 MΩ

0                         100000000      10.00000 kΩ  0Ω                            10.00000 kΩ      0Ω

100.000 MΩ                100000000      100.8100 MΩ  99.1900 MΩ                    100.8100 MΩ      99.1900 MΩ

0                         1000000000     100.0000 kΩ  0Ω                            1.000000 MΩ      0Ω

1.00000 GΩ[2]             1000000000     1.015100 GΩ  0.984900 GΩ                   1.020100 GΩ      0.979900 GΩ

[1]       Zero Meter before each measurement.

[2]       5520A must be used with 8508A to obtain suitable test uncertainty ratio.

[3]       Optional test.

                                  Table  3-12. 8845A 2-Wire Ohms Verification       Steps

     Nominal              Range                  90-day Test Limits[1]                       1-year  Test Limits[1]

          Input                          High             Low                              High          Low

0[2] [3]                  100            4.0 mΩ       0Ω                            4.0 mΩ           0Ω

100 Ω[2] [3]              100            100.01200 Ω  99.9880 kΩ                    100.0140 Ω       99.9860 Ω

0                         1000           10.0 mΩ      0Ω                            10.0 mΩ          0Ω

1.000 kΩ                  1000           1.000090 kΩ  0.99991 kΩ                    1.000110 kΩ      0.999890 kΩ

0Ω                        10000          0.100000 Ω   0Ω                            0.100000 Ω       0Ω

10.000 kΩ                 10000          10.00090 kΩ  9.99910 kΩ                    10.00110 kΩ      9.99890 kΩ

0Ω                        100000         1.000000 Ω   0Ω                            1.000000 Ω       0Ω

                                                                                                                         3-17
8845A/8846A

Calibration Manual

                                   Table 3-10. 8845A 2-Wire Ohms Verification Steps (cont.)

                   Nominal         Range                90-day Test Limits[1]                 1-year        Test Limits[1]

                   Input                         High                      Low                High              Low

             100.000 kΩ            100000     100.0090 kΩ            99.9910 kΩ               100.0110 kΩ   99.9890 kΩ

             0Ω                    1000000    10.00000 Ω             0Ω                       10.00000 Ω    0Ω

             1.00000 MΩ[2]         1000000    1.000090 MΩ            0.999910 MΩ              1.000110 MΩ   0.999890 MΩ

             0[1]                  10000000   100.0000 Ω             0Ω                       100.0000 Ω    0Ω

             10.0000 MΩ[2]         10000000   10.002100 MΩ           9.997900 MΩ              10.004100 MΩ  9.995900 MΩ

             0                     100000000  10000.00 Ω             0Ω                       10000.00 Ω    0Ω

             100.0000 MΩ           100000000  100.8100 MΩ            99.19000 MΩ              100.8100 MΩ   99.19000 MΩ

             [1]   Zero Meter before each measurement.

             [2]   5520A must be used with 8508A to obtain suitable  test uncertainty ratio.

             [3]   Optional test.

      2X4          Test Lead Verification Steps

                    This optional test verifies Meter operation with the TL2X4W-PT 2X4-Wire Test Leads.

                    To verify 2X4 Test Lead performance:

                    1.      Plug the TL2X4W-PT 2X4-Wire Test Leads into the HI and LO terminals of the

                            Meter (correct orientation is with the bump down)

                    2.      Press N.

                    3.      If not already highlighted, press the 2X4WIRE soft key.

                    4.      Connect the probe tips together to get the lowest reading.

                    The Meter should read under 3 milliohms. If the Meter reads above 3 milliohms, clean

                    off the probes with a damp cloth and repeat step 4 above.

                                                                     Note

                          If the leads are damaged, replace them.

      Rear Panel Terminal Verification Steps

                    This optional test verifies Meter operation through the rear-panel input terminals.

                    After connecting the Meter to the test equipment as shown in Figure 3-6, apply the

                    nominal values listed in Table 3-13 or Table 3-14, depending on which meter you are

                    calibrating.

3-18
                                                                                                                                                                          Performance             Test and Calibration         3

                                                                                                                                                                                                            Performance Tests

                                                    1/4 A SLOW FUSE,100/120 VAC   47 - 440 Hz                                                    INPUT            SENSE

                                                    1/8 A SLOW FUSE, 220/240 VAC  25 VA MAX                                                   V                   4W

                                                                                                     CAUTION:

                                                                                                     FOR FIRE PROTECTION REPLACE          HI                                 HI

                                                                                                     ONLY WITH A 250V FUSE

                                                                                                     AS STATED IN MANUAL.

                                                                                                                                          1000 V CAT I                    300 V
                                                                                                                                          600V CAT II

                                                                                                         CAL STICKER

                                                                                                                                          LO                                 LO

                                                                                                                                                              1V

                                                    IEEE488                                                                               FUSED

                                                                                        RS-232       LAN

                                                                                                               TRIG I/O              100  mA

                                             WARNING:                                                                             SERIAL TAG                      N10140

                                             TO AVOID ELECTRIC SHOCK

                                             GROUNDING CONNECTOR                                                                                              FLUKE CORPORATION

                                             IN POWER CORD                              NO INTERNAL  USER SERVICEABLE PARTS                                       MADE IN USA

                                             MUST BE CONNECTED                    REFER SERVICE TO   QUALIFIED SERVICE PERSONNEL                                  www.fluke.com

                                                    5520A CALIBRATOR

                                      NORMAL                AUX        SCOPE

                                      V,  ,  , RTD  A,  -SENSE, AUX V

                                                                                  OUT                                                                                            PREV

                                                    HI                                               STBY      OPR                EARTH   SCOPE               EXGRD              MENU

                                                                                                     7         8                  9                     µ         dBm             sec

                                                    LO                                                                                                     m              V            Hz

                                                                                  TRIG

                                                                                                     4         5                  6                     n         W               F

                                                                                                                                                           k              A            C

                                                                                                     1         2                  3                     p

                                             GUARD               20A                                                                                       M                         F

                                                                                                     +/        0                  •                     SHIFT                ENTER

                                      20V PK MAX             TC   20V PK MAX

            UUT                                                                                                                                                                  5720A Calibrator

            INPUT      SENSE                                                                                                                                              OUTPUT           SENSE

         V             4W

     HI                        HI                                                                                                                                            V    A        V

     1000 V CAT I              300 V                                                                                                                              HI                          HI  WIDEBAND
     600V CAT II

     LO                        LO                                                                                                                NC

                   1V                                        NC                                                                                                   LO

     FUSED                                                                                                                                                                                    LO

100  mA

                       FRONT                                                                                                                                      HI

                       N10140

                   FLUKE CORPORATION

                       MADE IN USA                                                                                                                                           AUX

                       www.fluke.com                                                                                                                                                       GUARD  GROUND

                                                                                                                                                                          CURRENT

                               NC                                                                                                                                                NC

                                                                                                                                                                                                            caw027.eps

                       Figure         3-6.          Rear-Panel                                 Terminals                     Equipment Setup

                                                                                                                                                                                                                               3-19
8845A/8846A

Calibration Manual

                             Table 3-13. 8846A Rear-Panel Terminal Verification Steps (Optional Test)

             Nominal                       90-day Test Limits                                 1-year  Test Limits

             Input           Range

                                           High        Low                                    High     Low

             10 V dc[1]      10 V          10.00023 V  9.99977 V                      10.00029 V       9.99971 V

             1000 Ω          4-W 1 kΩ      1000.09 Ω   999.91 Ω                       1000.11 Ω        999.89 Ω

             100 mA          100 mA        0.100035 A  0.099965 A                     0.100055 A       0.099945 A

             [1] 5520A must  be used with  8508A to obtain suitable test uncertainty  ratio.

                             Table 3-14. 8845A Rear-Panel Terminal Verification Steps (Optional Test)

             Nominal                       90-day Test Limits                                 1-year  Test Limits

             Input           Range

                                           High        Low                                    High     Low

             10 V dc[1]      10 V          10.00025 V  9.99975 V                      10.0004 V        9.9996 V

             1000 Ω          4-W 1 kΩ      1000.09 Ω   999.91 Ω                       1000.11 Ω        999.89 Ω

             100 mA          100 mA        0.100035 A  0.099965 A                     0.100055 A       0.099945 A

             [1] 5520A must  be used with  8508A to obtain suitable test uncertainty  ratio.

3-20
                                                                                                                                              Performance Test and Calibration                    3

                                                                                                                                                                               Performance Tests

Capacitance Verification Steps (8846A only)

         Connect the Meter to the test equipment as shown in Figure 3-7, and apply the

         capacitance values listed in Table 3-15. Verification forms can be found in Appendix A

         which can be copied and used to record each meter reading.

                             INPUT         SENSE                        8846A 6-1/2 DIGIT PRECISION MULTIMETER

                         V                     4W

                  2W/4W

                  HI                                  HI

                  1000 V CAT I                     300 V                100.8                                              nf
                  600V CAT II

                  LO                                  LO

                                       1V

                             FUSED

                  100                                 10 A     BACK     F1        F2               F3                  F4                 F5          MEMORY

                  mA

                                                                            DC V  AC V    DC I       AC I                  TRIG           INSTR

                             REAR              FRONT                                                                                   SETUP

                                                               RANGE

                                                                                  FREQ             TEMP         ZERO       ANALYZE        MEAS

                                                                                  PERIOD                                               SETUP

                                       5520A CALIBRATOR

                         NORMAL                    AUX         SCOPE

                         V,  ,  , RTD      A,  -SENSE, AUX V

                                                               OUT                STBY                                                                 PREV

                                           HI                                                   OPR             EARTH      SCOPE          EXGRD        MENU

                                                                                  7             8               9                   µ            dBm          sec

                                       LO                                                                                              m            V            Hz

                                                               TRIG

                                                                                  4             5               6                   n            W            F

                                                                                                                                       k            A            C

                                                                                  1             2               3                   p

                                    GUARD                 20A                                                                          M                         F

                                                                                  +/            0               •                   SHIFT              ENTER

                         20V PK MAX                   TC    20V PK MAX

                                                                                                                                                                                     caw029.eps

                  Figure 3-7. Capacitance Equipment Setup

                  Table 3-15. 8846A Capacitance Verification Steps

                                                                                                                           1-year                      Test Limits

Nominal   Output                       Range

                                                                                                       High                (F)                                             Low  (F)

0 ηF              1.0 ηF                                                          25.0 pF                                                                            0 ηF

1.0 ηF            1.0 ηF                                                          1.045 ηF                                                                           955.0 pF

10.0 ηF           10.0 ηF                                                         10.15 ηF                                                                           9.85 ηF

100.0 ηF          100.0 ηF                                                        101.5 ηF                                                                           98.5 ηF

1.0 μF            1.0 μF                                                          1.015 μF                                                                           985.0 ηF

10.0 μF           10.0 μF                                                         10.15 μF                                                                           9.85 μF

100.0 μF          100.0 μF                                                        101.5 μF                                                                           98.5 μF

1.0 mF            1.0 mF                                                          1.015 mF                                                                           985.0 μF

10.0 mF           10.0 mF                                                         10.15 mF                                                                           9.85 mF

100.0 mF          100.0 mF                                                        104.2 mF                                                                           95.8 mF

                                                                                                                                                                                                  3-21
8845A/8846A

Calibration Manual

      DC Current Verification Steps

                    Connect the Meter to the test equipment as shown in Figure 3-8 and, depending on which

                    meter you are calibrating, apply the nominal values listed in Table 3-16 or Table 3-17.

                    Verification forms can be found in Appendix A which can be copied and used to record

                    each meter reading.

                                                                                                                     Note

                    The 400 mA range is available only in with Outguard SW version 2.0 and

                    later.

                                                                         INPUT         SENSE                         8846A 6-1/2 DIGIT PRECISION MULTIMETER

                                                                     V                     4W

                                                              2W/4W

                                                              HI                                  HI

                                                              1000 V CAT I                     300 V                 +100.0031mA
                                                              600V CAT II

                                                              LO                                  LO

                                                                                   1V

                                                                         FUSED

                                                              100                                 10 A        BACK   F1          F2                  F3             F4                 F5          MEMORY

                                                              mA

                                                                                                                           DC V  AC V       DC I       AC I             TRIG           INSTR

                                                                         REAR              FRONT                                                                                    SETUP

                                                                                                              RANGE

                                                                                                                                 FREQ                TEMP    ZERO       ANALYZE        MEAS

                                                                                                                                 PERIOD                                             SETUP

                                                                                   5520A CALIBRATOR

                                                                     NORMAL                    AUX         SCOPE

                                                                     V,  ,  , RTD      A,  -SENSE, AUX V

                                                                                                              OUT                STBY                                                                  PREV

                                                                                       HI                                                         OPR        EARTH      SCOPE          EXGRD           MENU

                                                                                                                                 7                8          9                   µ            dBm          sec

                                                                                   LO                                                                                               m            V            Hz

                                                                                                              TRIG

                                                                                                                                 4                5          6                   n            W            F

                                                                                                                                                                                    k            A            C

                                                                                                                                 1                2          3                   p

                                                                                GUARD                 20A                                                                           M                         F

                                                                                                                                 +/               0          •                   SHIFT              ENTER

                                                                     20V PK MAX                   TC    20V PK MAX

                                      UUT                                                  A.                       : OFF                :  OFF                                               5720A Calibrator

                                      INPUT      SENSE                                                EX SNS               EX    GRD                                                          OUTPUT          SENSE

                                   V             4W

                            2W/4W                                                                                                                                                             V     A             V

                            HI                          HI

                            1000 V CAT I             300 V                                                                                                                             HI                            HI  WIDEBAND
                            600V CAT II

                            LO                          LO

                                             1V

                                      FUSED                                                                                                                                            LO                            LO

                            100                         10 A

                            mA

                                      REAR       FRONT                                                                                                                                 HI

                                                                                                                                                                                              AUX             GUARD      GROUND

                                                                                                                                                                                        CURRENT

                                                 NC

                                                                                                                                                                                                                                   caw022.eps

                                                     Figure 3-8. 100 mA DC Current Equipment Setup

3-22
                                                                              Performance Test and Calibration  3

                                                                                          Performance Tests

                         Table 3-16. 8846A DC Current Verification Steps

Nominal                           90-day Test Limits                                   1-year Test Limits

        Input      Range

                          High        Low                                     High        Low

0 A[1]         100.0 μA   25.0 ηA     -25.0 ηA                                25.0 ηA     -25.0 ηA

100.0 μA[1]    100.0 μA   100.065 μA  99.935 μA                               100.075 μA  99.935 μA

-100.0 μA[1]   100.0 μA   -99.935 μA  -100.065 μA                             -99.925 μA  -100.065 μA

0 A[1]         1.0 mA     50.0 ηA     -50.0 ηA                                50.0 ηA     -50.0 ηA

1.0 mA         1.0 mA     1.00035 mA  999.65 μA                               1.00055 mA  999.45 μA

-1.0 mA        1.0 mA     -999.65 μA  -1.00035 mA                             -999.45 μA  -1.00055 mA

0A             10.0 mA    2.0 μA      -2.0 μA                                 2.0 μA      -2.0 μA

10.0 mA        10.0 mA    10.005 mA   9.995 mA                                10.007 mA   9.993 mA

-10.0 mA       10.0 mA    -9.995 mA   -10.005 mA                              -9.993 mA   -10.007 mA

0A             100.0 mA   5.0 μA      -5.0 μA                                 5.0 μA      -5.0 μA

100.0 mA       100.0 mA   100.035 mA  99.965 mA                               100.055 mA  99.945 mA

-100.0 mA      100.0 mA   -99.965 mA  -100.035 mA                             -99.945 mA  -100.055 mA

0A             400 mA     20.0 μA     -20.0 μA                                20.0 μA     -20.0 μA

400 mA         400 mA     400.18 mA   399.82 mA                               400.22 mA   399.78 mA

-400 mA        400 mA     -399.82 mA  -400.18 mA                              -399.78 mA  -400.22 mA

Move the connector from the 400 mA jack to the 10 A jack on the UUT for the following steps.

0A             1A         200.0 μA    -200.0 μA                               200.0 μA    -200.0 μA

1 A[1]         1A         1.0006 A    999.4 mA                                1.0007 A    999.3 mA

-1 A[1]        1A         -999.4 mA   -1.0006 A                               -999.3 mA   -1.0007 A

0A             3A         600.0 μA    -600.0 μA                               600.0 μA    -600.0 μA

1.9 A          3A         1.90212 A   1.89788 A                               1.9025 A    1.8975 A

-1.9 A         3A         -1.9978 A   -2.0022 A                               -1.9974 A   -2.0026 A

0A             10 A       800.0 μA    -800.0 μA                               800.0 μA    -800.0 μA

10 A[1]        10 A       10.0128 A   9.9872 A                                10.0158 A   9.9842 A

-10 A[1]       10 A       -9.9872 A   -10.0128 A                              -9.9842 A   -10.0158 A

[1] 5520A must be used with 8508A to obtain suitable test uncertainty ratio.

                                                                                                                3-23
8845A/8846A

Calibration Manual

                                     Table 3-17. 8845A DC Current Verifications Steps

             Nominal                              90-day Test Limits                     1-year Test Limits

                     Input      Range

                                          High        Low                       High        Low

             0 A[1]         100.0 μA      25.0 ηA     -25.0 ηA                  25.0 ηA     -25.0 ηA

             100.0 μA [1]   100.0 μA      100.065 μA  99.935 μA                 100.075 μA  99.935 μA

             -100.0 μA[1]   100.0 μA      -99.935 μA  -100.065 μA               -99.925 μA  -100.065 μA

             0 A[1]         1.0 mA        50.0 ηA     -50.0 ηA                  50.0 ηA     -50.0 ηA

             1.0 mA         1.0 mA        1.00035 mA  999.65 μA                 1.00055 mA  999.45 μA

             -1.0 mA        1.0 mA        -999.65 μA  -1.00035 mA               -999.45 μA  -1.00055 mA

             0A             10.0 mA       2.0 μA      -2.0 μA                   2.0 μA      -2.0 μA

             10.0 mA        10.0 mA       10.005 mA   9.995 mA                  10.007 mA   9.993 mA

             -10.0 mA       10.0 mA       -9.995 mA   -10.005 mA                -9.993 mA   -10.007 mA

             0A             100.0 mA      5.0 μA      -5.0 μA                   5.0 μA      -5.0 μA

             100.0 mA       100.0 mA      100.035 mA  99.965 mA                 100.055 mA  99.945 mA

             -100.0 mA      100.0 mA      -99.965 mA  -100.035 mA               -99.945 mA  -100.055 mA

             0A             400 mA        20.0 μA     -20.0 μA                  20.0 μA     -20.0 μA

             400 mA         400 mA        400.18 mA   399.82 mA                 400.22 mA   399.78 mA

             -400 mA        400 mA        -399.82 mA  -400.18 mA                -399.78 mA  -400.22 mA

             Move the connector from the  400 mA jack to the 10 A jack on  the  UUT for the following steps.

             0A             1A            200.0 μA    -200.0 μA                 200.0 μA    -200.0 μA

             1 A[1]         1A            1.001 A     999.0 mA                  1.0007 A    998.75 mA

             -1 A[1]        1A            -999.0 mA   -1.001 A                  -999.3 mA   -1.0007 A

             0A             3A            600.0 μA    -600.0 μA                 600.0 μA    -600.0 μA

             1.9 A[1]       3A            1.90288 A   1.89712 A                 1.9025 A    1.8975 A

             -1.9 A[1]      3A            -1.89788 A  -1.90212 A                -1.8975 A   -1.9025 A

             0A             10 A          800.0 μA    -800.0 μA                 800.0 μA    -800.0 μA

             10 A[1]        10 A          10.0128 A   9.9872 A                  10.017 A    9.983 A

             -10 A[1]       10 A          -9.9872 A   -10.0128 A                -9.983 A    -10.017 A

3-24
                                                                                                                                                                   Performance Test and Calibration                           3

                                                                                                                                                                                                           Performance Tests

AC  Current Verification Steps

    Connect the Meter to the test equipment as shown in Figure 3-9 and, depending on which

    meter you are calibrating, apply the nominal values listed in Table 3-18 or Table 3-19.

    Verification forms can be found in Appendix A which can be copied and used to record

    each meter reading.

                                                                                             Note

    The 400 mA range is only available in Outguard version 2.0 and later.

                                                 INPUT         SENSE                         8846A 6-1/2 DIGIT PRECISION MULTIMETER

                                             V                     4W

                                      2W/4W

                                      HI                                  HI

                                      1000 V CAT I                     300 V                 +100.0031mA
                                      600V CAT II

                                      LO                                  LO

                                                           1V

                                                 FUSED

                                      100                                 10 A        BACK   F1          F2                  F3             F4                 F5          MEMORY

                                      mA

                                                                                                   DC V  AC V       DC I       AC I             TRIG           INSTR

                                                 REAR              FRONT                                                                                    SETUP

                                                                                      RANGE

                                                                                                         FREQ                TEMP    ZERO       ANALYZE        MEAS

                                                                                                         PERIOD                                             SETUP

                                                           5520A CALIBRATOR

                                             NORMAL                    AUX         SCOPE

                                             V,  ,  , RTD      A,  -SENSE, AUX V

                                                                                      OUT                STBY                                                                  PREV

                                                               HI                                                         OPR        EARTH      SCOPE          EXGRD           MENU

                                                                                                         7                8          9                   µ            dBm          sec

                                                           LO                                                                                               m            V            Hz

                                                                                      TRIG

                                                                                                         4                5          6                   n            W            F

                                                                                                                                                            k            A            C

                                                                                                         1                2          3                   p

                                                        GUARD                 20A                                                                           M                         F

                                                                                                         +/               0          •                   SHIFT              ENTER

                                             20V PK MAX                   TC    20V PK MAX

              UUT                                                  A.                       : OFF                :  OFF                                               5720A Calibrator

              INPUT      SENSE                                                EX SNS               EX    GRD                                                          OUTPUT          SENSE

           V             4W

    2W/4W                                                                                                                                                             V     A             V

    HI                          HI

    1000 V CAT I             300 V                                                                                                                             HI                            HI  WIDEBAND
    600V CAT II

    LO                          LO

                     1V

              FUSED                                                                                                                                            LO                            LO

    100                         10 A

    mA

              REAR       FRONT                                                                                                                                 HI

                                                                                                                                                                      AUX             GUARD      GROUND

                                                                                                                                                                CURRENT

                         NC

                                                                                                                                                                                                           caw022.eps

                                      Figure 3-9. AC Current Equipment                                                                      Setup

                                                                                                                                                                                                                              3-25
8845A/8846A

Calibration Manual

                                        Table 3-18. 8846A AC Current Verification Steps

                      Nominal   Output            90-day      Test Limits                1-year Test Limits

                                        Range

             Amplitude          Freq.             High        Low                        High        Low

             100.0 μA[1]        10 Hz   100.0 μA  100.14 μA   99.86 μA                   100.14 μA   99.86 μA

             100.0 μA           1 kHz   100.0 μA  100.14 μA   99.86 μA                   100.14 μA   99.86 μA

             100.0 μA[1]        5 kHz   100.0 μA  100.14 μA   99.86 μA                   100.14 μA   99.86 μA

             100.0 μA[1]        10 kHz  100.0 μA  100.45 μA   99.55 μA                   100.45 μA   99.55 μA

             1.0 mA[1]          10 Hz   1.0 mA    1.0014 mA   998.6 μA                   1.0014 mA   998.6 μA

             1.0 mA             1 kHz   1.0 mA    1.0014 mA   998.6 μA                   1.0014 mA   998.6 μA

             1.0 mA[1]          5 kHz   1.0 mA    1.0014 mA   998.6 μA                   1.0014 mA   998.6 μA

             1.0 mA[1]          10 kHz  1.0 mA    1.0045 mA   995.5 μA                   1.0045 mA   995.5 μA

             10.0 mA[1]         10 Hz   10.0 mA   10.014 mA   9.986 mA                   10.014 mA   9.986 mA

             10.0 mA            1 kHz   10.0 mA   10.014 mA   9.986 mA                   10.014 mA   9.986 mA

             10.0 mA[1]         5 kHz   10.0 mA   10.014 mA   9.986 mA                   10.014 mA   9.986 mA

             10.0 mA[1]         10 kHz  10.0 mA   10.045 mA   9.955 mA                   10.045 mA   9.955 mA

             100.0 mA[1]        10 Hz   100.0 mA  100.14 mA   99.86 mA                   100.14 mA   99.86 mA

             100.0     mA       1 kHz   100.0 mA  100.14 mA   99.86 mA                   100.14 mA   99.86 mA

             100.0     mA[1]    5 kHz   100.0 mA  100.14 mA   99.86 mA                   100.14 mA   99.86 mA

             100.0     mA[1]    10 kHz  100.0 mA  100.45 mA   99.55 mA                   100.45 mA   99.55 mA

             329.0     mA[1]    10 Hz   400 mA    330.387 mA  327.613 mA                 330.387 mA  327.613 mA

             329.0     mA[1]    1 kHz   400 mA    329.729 mA  328.271 mA                 329.729 mA  328.271 mA

             329.0     mA[1]    5 kHz   400 mA    332.458 mA  325.542 mA                 332.458 mA  325.542 mA

             329.0     mA[1]    10 kHz  400 mA    332.458 mA  325.542 mA                 332.458 mA  325.542 mA

             1 A[1]             45 Hz   1A        1.0014 A    998.6 mA                   1.00140 A   998.6 mA

             1A                 1 kHz   1A        1.0014 A    998.6 mA                   1.00140 A   998.6 mA

             1 A[1]             5 kHz   1A        1.0014 A    998.6 mA                   1.00140 A   998.6 mA

             1 A[1]             10 kHz  1A        1.0105 A    989.5 mA                   1.01050 A   989.5 mA

             1.9 A[1]           45 Hz   3A        1.90465 A   1.89535 A                  1.90465 A   1.89535 A

             1.9 A              1 kHz   3A        1.90465 A   1.89630 A                  1.90465 A   1.89535 A

             1.9 A[1]           10 kHz  3A        1.92765 A   1.87235 A                  1.92765 A   1.87235 A

             10 A[1]            45 Hz   10 A      10.02100 A  9.97900 A                  10.02100 A  9.97900 A

             10 A               1 kHz   10 A      10.02100 A  9.97900 A                  10.02100 A  9.97900 A

             Notes:

             [1] Optional test

3-26
                                                              Performance Test and Calibration                 3

                                                                                     Adjustment (Calibration)

                           Table 3-19. 8845A AC Current Verification Steps

         Nominal Output              90-day      Test Limits                   1-year Test Limits

                           Range

    Ampl           Freq              High        Low                           High               Low

10.0 mA[1]         10 Hz   10.0 mA   10.014 mA   9.986 mA                      10.014 mA      9.986 mA

10.0 mA            1 kHz   10.0 mA   10.014 mA   9.986 mA                      10.014 mA      9.986 mA

10.0 mA[1]         5 kHz   10.0 mA   10.014 mA   9.986 mA                      10.014 mA      9.986 mA

10.0 mA[1]         10 kHz  10.0 mA   10.045 mA   9.955 mA                      10.045 mA      9.955 mA

100.0 mA[1]        10 Hz   100.0 mA  100.14 mA   99.86 mA                      100.14 mA      99.86 mA

100.0 mA           1 kHz   100.0 mA  100.14 mA   99.86 mA                      100.14 mA      99.86 mA

100.0 mA[1]        5 kHz   100.0 mA  100.14 mA   99.86 mA                      100.14 mA      99.86 mA

100.0 mA[1]        10 kHz  100.0 mA  100.45 mA   99.55 mA                      100.45 mA      99.55 mA

329.0 mA[1]        10 Hz   400 mA    330.387 mA  327.613 mA                    330.387 mA     327.613 mA

329.0 mA[1]        1 kHz   400 mA    329.729 mA  328.271 mA                    329.729 mA     328.271 mA

329.0 mA[1]        5 kHz   400 mA    332.458 mA  325.542 mA                    332.458 mA     325.542 mA

329.0 mA[1]        10 kHz  400 mA    332.458 mA  325.542 mA                    332.458 mA     325.542 mA

1 A[1]             45 Hz   1A        1.0014 A    998.6 mA                      1.00140 A      998.6 mA

1A                 1 kHz   1A        1.0014 A    998.6 mA                      1.00140 A      998.6 mA

1 A[1]             5 kHz   1A        1.0014 A    998.6 mA                      1.00140 A      998.6 mA

1 A[1]             10 kHz  1A        1.0105 A    989.5 mA                      1.01050 A      989.5 mA

1.9 A[1]           45 Hz   3A        1.90465 A   1.89535                       1.90465 A      1.89535 A

1.9 A              1 kHz   3A        1.90465 A   1.89630                       1.90465 A      1.89535 A

1.9 A[1]           10 kHz  3A        1.92765 A   1.87235                       1.92765 A      1.87235 A

10 A[1]            45 Hz   10 A      10.01900 A  9.98100 A                     10.02100 A     9.97900 A

10 A               1 kHz   10 A      10.01900 A  9.98100 A                     10.02100 A     9.97900 A

Notes:

[1] Optional test

Adjustment (Calibration)

          Meter adjustments, or calibration, should be performed at the desired interval, or

          whenever a verification test indicates a Meter function is out of tolerance. The Meter

          accuracy will stay within specifications only if the adjustment procedure is performed at

          regular intervals. A one-year interval is adequate for most applications. The Meter’s

          accuracy specifications are not valid beyond the one-year interval.

          Meter adjustments are accessed only through the remote interface with a series of

          adjustment steps. The remote program directs the test equipment to apply a series of

          shorts, opens, voltages, currents, and capacitance (8846A only) to the Meter. At each

          step, the Meter internally makes the necessary adjustment to bring the Meter into

          specification. No internal mechanical adjustments are necessary.

                                                                                                               3-27
8845A/8846A

Calibration Manual

                    Using an automated, computer-controlled procedure, the calibration and verification

                    procedures can be performed on the Meter in under 60 minutes. A sample adjustment

                    program is listed on the “Sample Adjustment Program” section later in this manual. A

                    MetCal program is available at www.fluke.com to adjust the Meter.

                    The Meter’s adjustments are password protected to prevent accidental or unauthorized

                    adjustments. The security password must be entered through the front panel or remote

                    interface before adjustments can be made to the Meter.

      Unlocking the Meter for Adjustments (Calibration)

                    To unlock the Meter for adjustments from the front panel:

                    1.  Press I.

                    2.  Press CAL.

                    3.  Press UNLOCK       CAL.

                        Press the soft key labeled -- to decrement the character or ++ to increment the

                        character. The character can be set to 0 through 9, A through Z, period (.), and

                        dash (-).

                        To move to the next character, press -->.

                    4.  Press ENTER to enter the password and unlock the Meter for adjustments.

                    Find the CALIBRATION:SECURE:STATE command in the “Supported SCPI

                    Commands” section of the 8846A/8845A Programmers Manual for information on

                    unlocking the Meter for calibration.

                    The Meter is shipped from the factory with the password set to FLUKE884X.

      Unlocking the Meter for Adjustments Over a Remote Interface

                    To unlock the Meter, send the following command:

                        “CAL:SEC:STAT      OFF,  FLUKE884X”

                    To relock the Meter, send the following command:

                        “CAL:SEC:STAT      ON,   FLUKE884X”

      Changing the Calibration Password

                    The calibration password can be changed only through the remote interface. Find the

                    CALIBRATION:SECURE:CODE command in the “Supported SCPI Commands”

                    section of the 8846A/8845A Programmers Manual for information on changing the

                    calibration password.

      Resetting the Calibration Password

                    If the calibration password has is lost or forgotten, the password can be reset to

                    FLUKE884X by performing the following actions.

                                                             Note

                        Before taking the following steps, try to use the factory default password

                        FLUKE884X.

                    1.  Perform the general disassembly steps in the “Disassembly Instructions” section.

                    2.  Connect a jumper across W2, as shown in Figure 3-10.

3-28
                                                                     Performance Test and Calibration  3

                                                                       Adjustment (Calibration)

3.  Reconnect the power cord between the Meter and a power outlet.

4.  Turn the Meter on.

    On power-up, the password will automatically be reset to FLUKE884X.

5.  Turn the Meter off and disconnect the power cord.

6.  Remove the jumper connected above in step 2.

7.  Reassemble the Meter.

                                 Rear 8845A/8846A

                             Trig       LAN Port

                             IQ                             Main Input Module

Rear Terminals

                                         W2       W1

                Location of Calibration

                           Jumper

                                                                     Transformer

                                                                                        caw028.eps

                           Figure 3-10. Calibration Jumper Location

Changing the Calibration Date Remotely

The 8845A’s calibration date can only be changed remotely.

                                         Note

    Calibration must be enabled before sending the following command.

    “CAL:DATE   mm/dd/yyyy”

The 8846A calibration date is automatically updated when the “CAL:REC” command          is

sent. See the 8845A 8846A Programmers Manual for more details.

Equipment for Calibration

The required equipment for calibration is that same the equipment listed in Table 3-1.

                                                                                                       3-29
8845A/8846A

Calibration Manual

      Adjustment Process

                    The adjustment steps differ slightly between the 8845A and 8846A. In both cases, they

                    are divided into four areas: open adjust, zero adjust, rear panel zero adjust, and gain

                    adjust.

                    Table 3-20 lists the step numbers, the description of the adjustment, the measurement

                    adjustment type (open, zero, or gain adjust), the Meter value/range being adjusted, the

                    amplitude of the adjustment signal, and if required the frequency of the adjustment

                    signal.

3-30
                                                               Performance Test and Calibration        3

                                                                             Adjustment (Calibration)

                     Table 3-20.  8845A/8846A Adjustment       Steps

   Step  Modes  Value Range               Input Signal                Description      Series

Open

0        ORES   100000000         open                         OHM 100M open           Y

                                                               terminals

1        ORES   1000000000        open                         OHM 1G open terminals   Y

                                                               (8846A only)

2        ZCAP   1.00E-09          open                         CAP 1 nF open           N

                                                               terminals (8846A only)

ACV

Zero

3        ZVAC   100.0E-3          4-wire  low-thermal  short   AC 100 mV               Y

4        ZVACS  100.0E-3          4-wire  low-thermal  short   AC 100 mV               Y

5        ZVAC   1                 4-wire  low-thermal  short   AC 1V                   Y

6        ZVACS  1                 4-wire  low-thermal  short   AC 1V                   Y

7        ZVAC   10                4-wire  low-thermal  short   AC 10V                  Y

8        ZVACS  10                4-wire  low-thermal  short   AC 10V                  Y

9        ZVAC   100               4-wire  low-thermal  short   AC 100V                 Y

10       ZVACS  100               4-wire  low-thermal  short   AC 100V                 Y

11       ZVAC   1000              4-wire  low-thermal  short   AC 1000V                Y

12       ZVACS  1000              4-wire  low-thermal  short   AC 1000V                N

DCV

Zero

13       ZVDC   1000              4-wire  low-thermal  short   DC 1000V                Y

14       ZVDC   100               4-wire  low-thermal  short   DC 100V                 Y

15       ZVDC   10                4-wire  low-thermal  short   DC 10V                  Y

16       ZVDC   1                 4-wire  low-thermal  short   DC 1V                   Y

17       ZVDC   0.1               4-wire  low-thermal  short   DC 100mV                N

18       DFVDC  0.1               4-wire  low-thermal  short   DC 100mV                N

Ohm

Zero

19       ZRES   10000000          4-wire  low-thermal   short  4W Ohm 10 MOHM          Y

20       ZRES   1000000           4-wire  low-thermal   short  4W 1 MOHM               Y

21       ZRES   100000            4-wire  low-thermal   short  4W 100 kOHM             Y

22       ZRES   10000             4-wire  low-thermal  short   4W 10 kOHM              Y

23       ZRES   1000              4-wire  low-thermal  short   4W 1 kOHM               Y

24       ZRES   100               4-wire  low-thermal  short   4W 100 OHM              Y

                                                                                                       3-31
8845A/8846A

Calibration Manual

                               Table 3-18. 8845A/8846A Adjustment Steps (cont)

             Step       Modes  Value Range           Input Signal               Description     Series

             25         ZRES   10           4-wire low-thermal short  4W 10 OHM     (8846A      N

                                                                      only)

             Rear Ω

             Zero

             26         ZRES   100000       4-wire low-thermal short  4W 100 kOHM rear input    Y

             27         ZRES   10000        4-wire low-thermal short  4W 10 kOHM    rear input  Y

             28         ZRES   1000         4-wire low-thermal short  4W 1 kOHM     rear input  Y

             29         ZRES   100          4-wire low-thermal short  4W 100 OHM    rear input  Y

                        ZRES   10           4-wire low-thermal short  4W 10 OHM     rear input  N

             30                                                       (8846A only)

             Rear

             DCV

             Zero

             31         ZVDC   1            4-wire low-thermal short  DC 1V     rear input      Y

             32         ZVDC   0.1          4-wire low-thermal short  DC 100 mV     rear input  N

             Low I

             Zero

             33         ZIDC   100.0E-3     100mA    to  Lo  short    DC 100 mA                 Y

             34         ZIDC   1.0E-3       100mA    to  Lo  short    DC 1 mA                   Y

             35         ZIDC   10.0E-3      100mA    to  Lo  short    DC 10 mA                  Y

             36         ZIDC   100.0E-6     100mA    to  Lo  short    DC 100 uA                 Y

             37         ZIAC   0.0          100mA    to  Lo  short    AC 100 uA                 Y

             38         ZIACS  0.0          100mA    to  Lo  short    AC 100 uA                 Y

             39         ZIAC   1.0E-3       100mA    to  Lo  short    AC 1 mA                   Y

             40         ZIACS  1.0E-3       100mA    to  Lo  short    AC 1 mA                   Y

             41         ZIAC   10.0E-3      100mA    to  Lo  short    AC 10 A                   Y

             42         ZIACS  10.0E-3      100mA    to  Lo  short    AC 10 mA                  Y

             43         ZIAC   100.0E-3     100mA    to  Lo  short    AC 100 mA                 Y

             44         ZIACS  100.0E-3     100mA    to  Lo  short    AC 100 mA                 N

             Hi I Zero

             45         ZIDC   10           10 A to  Lo short         DC 10 A                   Y

             46         ZIDC   1            10 A to  Lo short         DC 1 A                    Y

             47         ZIAC   1            10 A to  Lo short         AC 1 A                    Y

3-32
                                                             Performance Test and Calibration       3

                                                                          Adjustment (Calibration)

                 Table  3-18. 8845A/8846A Adjustment  Steps  (cont)

Step  Modes      Value  Range         Input Signal           Description           Series

48    ZIACS      1             10 A to Lo short              AC 1 A                Y

49    ZIAC       10            10 A to Lo short              AC 10 A               Y

AC    ZIACS      10            10 A to Lo short              AC 10 A               N

50    Linearity

51    ACLIN      1.19          1.19 @1200 Hz                 AC 1 V                N

52    ACLIN      0.8           0.8 @1200 Hz                  AC 1 V                N

53    ACLIN      0.4           0.4 @1200 Hz                  AC 1 V                N

54    ACLIN      0.005         0.005 @1200 Hz                AC 1 V                N

ACV

Gain

55    GVAC       0.1           0.1 @1200 Hz                  AC 100 mV             Y

56    GVACS      0.1           0.1 @1200 Hz                  AC 100 mV             N

57    ACPOLE     0.1           0.1 @50000 Hz                 A 100 mV              N

58    GVAC       1             1 @1200 Hz                    AC 1 V                Y

59    GVACS      1             1 @1000 Hz                    AC 1 V                N

60    FVAC       1             1 @10 Hz                      AC 1 V                N

61    ACPOLE     1             1 @50000 Hz                   AC 1 V                N

62    GVAC       10            10 @1200 Hz                   AC 10 V               Y

63    GVACS      10            10 @1200 Hz                   AC 10 V               N

64    ACPOLE     10            10 @50000 Hz                  AC 10 V               N

65    GVAC       100           100 @1200 Hz                  AC 100 V              Y

66    GVACS      100           100 @1200 Hz                  AC 100 V              N

67    ACPOLE     100           100 @50000 Hz                 AC 100 V              N

68    GVAC       1000          1000 @1200 Hz                 AC 1000 V (8845A use  Y

                                                             750V on 750V range)

69    GVACS      1000          1000 @1200 Hz                 AC 1000 V (8845A use  N

                                                             750V on 750V range)

70    ACPOLE     1000          329 @50000 Hz                 AC 1000 V (use 8845A  N

                                                             750V range)

VDC

Gain

71    GVDC       1000          1000                          DC 1000 V             N

72    GVDC       -1000         -1000                         DC 1000 V             N

73    GVDC       100           100                           DC 100 V              N

74    GVDC       -100          -100                          DC 100 V              N

                                                                                                    3-33
8845A/8846A

Calibration Manual

                             Table 3-18. 8845A/8846A Adjustment Steps (cont)

             Step     Modes  Value Range          Input Signal                Description  Series

             75       GVDC   10           10                    DC 10 V                    N

             76       GVDC   -10          -10                   DC 10 V                    N

             77       GVDC   1            1                     DC 1 V                     N

             78       GVDC   -1           -1                    DC 1 V                     N

             79       GVDC   0.1          0.1                   DC 100 mV                  N

             80       GVDC   -0.1         -0.1                  DC 100 mV                  N

             Hi IDC

             Gain

             81       GIDC   1            1                     DC 1 A                     N

             82       GIDC   -1           -1                    DC 1 A                     N

             83       GIDC   10           10                    DC 10 A                    N

             84       GIDC   -10          -10                   DC 10 A                    N

             Hi IAC

             Gain

             85       GIAC   10           10                    AC 10 A                    Y

             86       GIACS  10           10                    AC 10 A                    N

             87       GIAC   1            1                     AC 1 A                     Y

             88       GIACS  1            1                     AC 1 A                     N

             Low IAC

             Gain

             89       GIAC   100.0E-3     100.0E-3              AC 100 mA                  Y

             90       GIACS  100.0E-3     100.0E-3              AC 100 mA                  N

             91       GIAC   10.0E-3      10.0E-3               AC 10 mA                   Y

             92       GIACS  10.0E-3      10.0E-3               AC 10 mA                   N

             93       GIAC   1.0E-3       1.0E-3                AC 1 mA                    Y

             94       GIACS  1.0E-3       1.0E-3                AC 1 mA                    N

             95       GIAC   100.0E-6     100.0E-6              AC 100 uA                  N

             96       GIACS  100.0E-6     100.0E-6              AC 100 uA                  N

             Lo IDC

             Gain

             97       GIDC   100.0E-6     100.0E-6              DC 100 uA                  N

             98       GIDC   -100.0E-6    -100.0E-6             DC 100 uA                  N

             99       GIDC   1.0E-3       1.0E-3                DC 1 mA                    N

             100      GIDC   -1.0E-3      -1.0E-3               DC 1 mA                    N

3-34
                                                                 Performance Test and Calibration          3

                                                                              Adjustment (Calibration)

                          Table 3-18. 8845A/8846A Adjustment Steps (cont)

Step    Modes             Value Range             Input  Signal            Description             Series

101     GIDC              10.0E-3          10.0E-3               DC 10 mA                            N

102     GIDC              -10.0E-3         -10.0E-3              DC 10 mA                            N

103     GIDC              100.0E-3         100.0E-3              DC 100 mA                           N

104     GIDC              -100.0E-3        -100.0E-3             DC 100 mA                           N

Ω Gain

105     GRES              100000000        100000000             R 100M Ω                            N

106     GRES              10000000         10000000              4W 10M Ω                            N

107     GRES              1000000          1000000               4W 1M Ω                             N

108     GRES              100000           100000                4W 100 kΩ                           N

109     GRES              10000            10000                 4W 10 kΩ                            N

110     GRES              1000             1000                  4W 1 kΩ                             N

111     GRES              100              100                   4W 100 Ω                            N

112     GRES              10               10                    4W 10 Ω   (8846A       only)        N

Misc

Gain

113     GRES              1000000000       1000000000            R1G Ω (8846A only)                  N

114     GCAP1             10.0E-9          10.0E-9               C10NF (8846A only)                  Y

115     GCAP2             10.0E-9          10.0E-9               C10NF (8846A only)                  N

        Once familiar with the calibration series of setups, the calibration time may be sped up by

        using the command “CAL OFF”. This command allows the instrument to automatically

        go to the next logical step in the series. The last column in Table 3-20 identifies which

        steps in the series can be run automatically. For example, all of the Open steps can be run

        by entering:

        CAL:CAL        ORES,    100000000

        CAL?      OFF

        Another example is automatically running the ACV zeros from step 3 through step 12 in

        Table 3-20 with:

        CAL:CAL        ZVAC,    100.0E-3

        CAL?      OFF

                                                  Notes

        Using the CAL? command without an argument turns off the single step

        feature.

        AC linearity steps must be run with AC Gain steps to complete the AC

        calibration.

                                                                                                           3-35
8845A/8846A

Calibration Manual

      Aborting a Calibration Process

                                                        WCaution

                    Aborting a calibration process while the Meter is attempting to

                    write new calibration constants to memory may corrupt the

                    calibration constants memory.

                    To abort a running calibration process, stop the program or issue a device clear command

                    over the remote interface. No constants are saved until the Meter receives a record

                    command.

      Sample Adjustment Program

                    The example below shows an IEEE 488 program that uses some of the commands to

                    adjust the 1V ACV portion of the Meter. Note that the ACV linearity adjustment must be

                    followed with a gain adjustment. Print is an output command to the Meter. Input Line

                    reads a response from the Meter.

                    INIT      PORT  0

                    CLEAR     PORT  0

                    #  Enable       Calibration

                    PRINT     @,  "CAL:SEC:STAT   OFF,       FLLUKE448X"

                    #  Zeros

                    ###   Calibrate  AC    Linearity    Set    input  value    to  1.19V@1200Hz

                    #  V@1200Hz

                    PRINT     @,  "CAL:VAL      ACLIN,1.19"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$

                    ###   Calibrate  AC    Linearity    Set    input  value    to  0.8V@1200Hz

                    #  0.8V@1200Hz

                    PRINT     @,  "CAL:VAL      ACLIN,0.8"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$

                    ###   Calibrate  AC    Linearity    Set    input  value    to  0.4V@1200Hz

                    #  0.4V@1200Hz

                    PRINT     @,  "CAL:VAL      ACLIN,0.4"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$

                    ###   Calibrate  AC    Linearity    Set    input  value    to  0.05V@1200Hz

                    #  0.005V@1200Hz

                    PRINT     @,  "CAL:VAL      ACLIN,0.005"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$PRINT  @,  "CAL:VAL"

                    ###   ACV  Gain  for   1V

                    PRINT     @,  "CAL:VAL      ZVAC,1"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$

                    PRINT     @,  "CAL:VAL      ZVACS,1"

                    PRINT     @,  "CAL?  ON"

                    INPUT     LINE  @,  A$

                    #  Disable      Calibration

                    PRINT     @,  "CAL:SEC:STAT   ON,        FLUKE884X”

                    For more information about writing a program to remotely control the Meter, refer       to  the

                    8845A/8846A Programmer’s Manual.

3-36
       Chapter 4

List of Replaceable Parts

Title                                                                                                                      Page

Introduction........................................................................................................       4-3

How to Obtain Parts...........................................................................................             4-3

How to Contact Fluke ........................................................................................              4-3

Parts ...................................................................................................................  4-4

                                                                                                                                 4-1
8845A/8846A

Calibration Manual

4-2
                                                                         List of Replaceable Parts          4

                                                                                              Introduction

Introduction

     This chapter contains an illustrated list of replaceable parts for the Fluke 8845A and

     8846A Digital Multimeters. Parts are listed by assembly and alphabetized by reference

     designator. Each assembly is accompanied by an illustration showing the location of each

     part and its reference designator. The parts lists provide the following information:

     •  Reference designator

     •  An indication if the part is subject to static discharge damage

     •  Description

     •  Fluke stock number

     •  Total quantity

     •  Any special notes (i.e., factory-selected part)

                                 WCaution

        An asterisk (*) indicates a part that may be damaged by static

        discharge.

How  to Obtain Parts

     Electrical components may be ordered directly from the manufacturer by using the

     manufacturer’s part number, or from the Fluke Corporation and its authorized

     representatives by using the part number under the heading FLUKE STOCK NO.

     Parts price information is available from the Fluke Corporation or its representatives.

     Prices are also available in a Fluke Replacement Parts Catalog, which is available on

     request.

     In the event that the part ordered has been replaced by a new or improved part, the

     replacement will be accompanied by an explanatory note and if necessary, installation

     instructions.

     To ensure prompt delivery of the correct part, include the following information when

     you place an order:

     •  Instrument model and serial number

     •  Part number and revision level of the pca containing the part.

     •  Reference designator

     •  Fluke stock number

     •  Description (as given under the DESCRIPTION heading)

     •  Quantity

How to Contact Fluke

     To contact Fluke, visit Fluke’s web site at www.fluke.com or call one of the following

     telephone numbers:

        USA and Canada: 1-888-99-FLUKE (1-888-993-5853)

        Europe: +31 402-675-200

        Japan: +81-3-3434-0181

                                                                                                            4-3
8845A/8846A

Calibration Manual

                     Singapore: +65-738-5655

                     Anywhere in the world: +1-425-446-5500

                     Visit Fluke's web site at www.fluke.com.

     Parts

                     Table 4-1 lists the replaceable parts for the 8845A and 8846A Digital Multimeters.  Figure

                     4-1 identifies the parts within the Meter.

                                            Table 4-1. Replaceable Parts

             Ref.                               Description                                 Part         Qty

             Desig.                                                                         Number

             A1      8846A-4001,MAIN PCA*                                                   2456914      1

                     8845A-4001,MAIN PCA*                                                   2165053      1

             A1F1    FUSE 11A. 1000v, FAST .406” X 1.5”                                     803293       1

             A1F2    FUSE 440A, 1000V, FAST .406” X 1.5”                                    943121       1

             A2      8846A-4002,PCA, DISPLAY/KEYBOARD                                       2454039      1

                     8845A-4002,PCA, DISPLAY/KEYBOARD                                       2430557      1

             F1      FUSE, 0.25A, 250V,SLOW .25” X 1.25” (for 100 to 120 VAC mains power)   166306       1

                     FUSE, 0.125, 250V, SLOW .25” X 1.25” (for 200 to 240 VAC mains power)  166488

             H1      SCREW,SCREW,FHU,P,LOCK,MAG SS,6-32,.250                                320093       9

             H2      SCREW,8-32,.250,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK                     228890       2

             H3      SCREW,SCREW,WH,P,THD FORM,STL,5-20,.312                                494641       5

             H4      SCREW,6-32,.250,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK                     152140       4

             H5      SCREW,SCREW,FH,P,LOCK,STL,8-32,.312                                    281725       2

             H6      NUT,NUT,HEX,STL,8-32                                                   110544       1

                     CONNECTOR,TERMINAL BLOCK,PLUGGABLE,3.5MM CTR,RT ANG,3                               1

             H7      POS,BULK                                                               2434588

                     CONNECTOR ACCESSORY,MICRO-RIBBON,SCREW LOCK,M3.5,6-                                 2

             H8      32,STEEL,BLACK ZINC                                                    854737

                     CONNECTOR ACCESSORY,D-SUB JACK SCREW,4-40,.250 L,W/FLAT                             2

             H9      WASHER                                                                 1777348

             H10     NUT,NUT,EXT LOCK,STL,8-32                                              195263       1

                     SCREW,4-24,.250,PAN,PHILLIPS,STEEL,ZINC-CLEAR,TYPE B THD                            1

             H11     FORM                                                                   1626602

             H12     SCREW,6-32,1.000,PAN,PHILLIPS,STEEL,ZINC-CLEAR,LOCK                    114215       1

                     SCREW,6-32,.250,PAN,SLOT/PHILLIPS COMBO,STEEL,ZINC-                                 1

             H13     CHROMATE                                                               855192

             MP1     8845A-2020, FUSE DOOR                                                  2439239      1

             MP2     8845A-2001,TOP COVER                                                   2439097      1

4-4
                                                                        List of Replaceable  Parts  4

                                                                                             Parts

                            Table 4-1. Replaceable Parts (cont)

Ref.                               Description                          Part     Qty

Desig.                                                                  Number

MP3     8846A-2001,FRONT PANEL                                          2454063  1

        8845A-2005,FRONT PANEL                                          2439123  1

MP4     8845A-2006,REAR PANEL TERMINAL BLOCK                            2439138  1

MP5     8845A-2008,HANDLE                                               2439150  1

MP6     8845A-2009,REAR BOOT                                            2439161  1

MP7     8845A-2010,FRONT BOOT                                           2439177  1

        POWER ENTRY MODULE,KG10.6101.105,AC INLET, ON/OFF LINE                   1

MP8     SWITCH,FUSE DRAWER,BULK                                         2437071

MP9     FILTER PART,FILTER,LINE,PART,FUSE DRWR W/SHRT BAR               944277   1

MP10    FILTER PART,FILTER,LINE,PART,VOLTAGE SELECTOR                   944272   1

MP11    8845A-2004,FRONT PANEL SHIELD                                   2439200  1

MP12    8845A-8003,FRONT PANEL DECAL                                    2454088  1

MP13    8846A-2002,LENS                                                 2454074  1

        8845a-2014,LENS                                                 2454056  1

MP14    8845A-2015,AC TOP SHIELD                                        2454111  1

MP15    8845A-2003,CHASSIS                                              2439189  1

MP16    8845A-2002,WEDGE                                                2439114  2

MP17    FOAM PAD,URETHANE,.250 IN X.375 IN X .062 IN THK,ADHESIVE       2567386  4

MP18    8845A-2025,TRANSFORMER BRACKET                                  2546476  1

MP19    8845A-2026,DISPLAY SHIELD                                       2559056  1

MP20    8845A-8004,REAR PANEL DECAL                                     2454095  1

MP21    8845A-2023,PUSH ROD                                             2526849  1

MP22    FLUKE 27-2002,COVER, FUSE                                       665031   1

MP23    8845A-2027,HANDLE CAP, PLASTIC PART                             2663315  2

S1      8846A-8001,KEYPAD                                               2454042  1

        8845a-8001,KEYPAD                                               2439221  1

S2      8845A-2013,SWITCHBODY                                           2440417  1

S3      8845A-4201,FRONT / REAR SWITCH ASSEMBLY                         2571214  1

        TRANSFORMER,POWER,100/120/220/240V,50/60HZ,31:7:1:1:7,14W,8845           1

T1      A-6501,EI100,BULK                                               2455793

W1      2620A-4403,WIRE ASSY,GROUND                                     874099   1

W2      8845A-4402,WIRE ASSY                                            2584378  1

TM1     CD-ROM, PROGRAMMER AND USERS MANUAL, 8845A/8846A                2453193  1

                                                                                                    4-5
8845A/8846A

Calibration Manual

                     Table 4-1. Replaceable Parts (cont)

             Ref.    Description                                          Part    Qty

             Desig.                                                       Number

             LC1     LC1 LINE CORD,NORTH AM,10A,5-15/IEC,18/3,SVT,6.0 FT  284174  1

                     TEST LEAD SET,600V/1KV,PROBE-R/A PLUG,BLACK/RED      802980  1

4-6
                                                            List of Replaceable               Parts  4

                                                                                              Parts

           MP12

     MP13

                     S1

                         A2

MP3

           MP17                              H3

                                                 MP11

                                                                MP19

                                                       H11

                 H1

                                                            H4

                                                            884XA Final Assembly

                                                                      (1 of 3)

                                                                                caw0301f.emf

                 Figure 4-1. Final Assembly

                                                                                                     4-7
8845A/8846A

Calibration Manual

                    H6

                        H10                                                                        MP14

                    W1

                                                                                                                       H12

                                                                 W2

                                                                                      A1

                                                                                                         MP22      S3

                                                                                  H3

                                                                      MP23

                                                             H5                                                              H3

                                                         H4                                                            MP21

                                                                                                               S2

                             T1

                                 MP18           H13

                                       MP10 F1                                               MP15

                                       MP8 MP9       H8                               MP4

                                                         H9

                                                                            MP20                   884XA Final Assembly

                                                                                                               (2 of 3)

                                                                                                                             caw0302f.emf

                                                         Figure 4-1.  Final Assembly (cont)

4-8
                                                        List of Replaceable     Parts  4

                                                                                Parts

                                 MP7

                        H2

           H1

      MP5

                                                   H1

MP16                                               MP1

                                      H1  H2

                                               H1

                                 MP2

                            MP6

                                                        884XA Final Assembly

                                                        (3 of 3)

                                                                  caw0303f.emf

           Figure 4-1.  Final Assembly (cont)

                                                                                       4-9
8845A/8846A

Calibration Manual

4-10
                 Appendices

Appendix  Title                                                                                                       Page

A  Verification Forms ..............................................................................................  A-1

B  Example Adjustment Program ............................................................................            B-1
8845A/8846A

Calibration Manual
                           Appendix A

                           Verification Forms

Introduction

The following tables are forms used to collect Meter readings while performing the

verification procedures contained in Chapter 3. Appendix pages may be copied as needed

to record meter readings.

                                                                                        A-1
8845A/8846A

Calibration Manual

                     Table   A-1. Blank  Verification Record for 90-Day Specifications

             Nominal Output                         Test Limits

                                Range                                 Results           Pass/Fail

             Ampl.   Freq.               High                    Low

     8846A DC Volts

     0                       0.100       3.5E-6       -3.5E-6

     100.0E-3                0.100       100.006E-3   99.994E-3

     -100.0E-3               0.100       -99.994E-3   -100.006E-3

     0                       1           7.0E-6       -7.0E-6

     1                       1           1.000025     0.999975

     -1                      1           -0.999975    -1.000025

     0                       10          50.0E-6      -50.0E-6

     5                       10          5.000140     4.999860

     -5                      10          -4.999860    -5.000140

     10                      10          10.000230    9.999770

     -10                     10          -9.999770    -10.000230

     0                       100         600.0E-6     -600.0E-6

     100                     100         100.0033     99.9967

     -100                    100         -99.9967     -100.0033

     0                       1000        10.0E-3      -10.0E-3

     1000                    1000        1000.0410    999.9590

     -1000                   1000        -999.9590    -1000.0410

     8846A DC Ratio

     100.0E-3                0.100       1.00012      0.99988

     1                       1           1.00005      0.99995

     -10                     10          1.000046     0.999954

     8845A DC Volts

     0                       0.100       3.5E-6       -3.5E-6

     100.0E-3                0.100       100.0075E-3  99.9925E-3

     -100.0E-3               0.100       -99.9925E-3  -100.0075E-3

     0                       1           7.0E-6       -7.0E-6

     1                       1           1.000037E+0  999.963E-3

     -1                      1           -999.963E-3  -1.000037E+0

     0                       10          50.0E-6      -50.0E-6

     5                       10          5.00015E+0   4.99985E+0

     -5                      10          -4.99985E+0  -5.00015E+0

A-2
                                                                                        Appendices          A

                                                                                        Verification Forms

                Table A-1.  Blank Verification Record for 90-Day Specifications (cont)

     Nominal Output                             Test Limits

                               Range                              Results               Pass/Fail

    Ampl.       Freq.                 High                   Low

8845A DC Volts (cont)

10                          10        10.00025E+0  9.99975E+0

-10                         10        -9.99975E+0  -10.00025E+0

0                           100       600.0E-6     -600.0E-6

100                         100       100.0041E+0  99.9959E+0

-100                        100       -99.9959E+0  -100.0041E+0

0                           1000      10.0E-3      -10.0E-3

1000                        1000      1.000045E+3  999.955E+0

-1000                       1000      -999.955E+0  -1.000045E+3

8845A DC Ratio

100.0E-3                    0.100     1.00015      0.99985

1                           1         1.000074     0.999926

-10                         10        1.00005      0.99995

8845A/8846A AC Volts

100.0E-3   10               0.100     100.09E-3    99.91E-3

100.0E-3   20000            0.100     100.09E-3    99.91E-3

100.0E-3   50000            0.100     100.16E-3    99.84E-3

100.0E-3   100000           0.100     100.68E-3    99.32E-3

100.0E-3   300000           0.100     104.5E-3     95.5E-3

1          10               1         1.0008E+0    999.2E-3

1          20000            1         1.0008E+0    999.2E-3

1          50000            1         1.0016E+0    998.4E-3

1          100000           1         1.0068E+0    993.2E-3

1          300000           1         1.045E+0     955.0E-3

10         10               10        10.008E+0    9.992E+0

10         20000            10        10.008E+0    9.992E+0

10         50000            10        10.016E+0    9.984E+0

10         100000           10        10.068E+0    9.932E+0

3          300000           10        3.17E+0      2.83E+0

                                                                                                            A-3
8845A/8846A

Calibration Manual

                    Table A-1. Blank Verification Record for 90-Day Specifications (cont)

             Nominal Output                        Test Limits

                                  Range                              Results               Pass/Fail

             Ampl.  Freq.                High                   Low

     8845A/8846A AC Volts (cont)

     100            45       100         100.08E+0   99.92E+0

     100            20000    100         100.08E+0   99.92E+0

     100            50000    100         100.16E+0   99.84E+0

     100            100000   100         100.68E+0   99.32E+0

     8846A Only AC Volts

     1000           45       1000        1000.800    999.200

     1000           1000     1000        1000.800    999.200

     1000           10000    1000        1000.800    999.200

     320            20000    1000        320.460     319.540

     320            50000    1000        320.852E+0  319.148E+0

     320            100000   1000        322.72E+0   317.28E+0

     8845A Only AC Volts

     750            45       750         750.600     749.400

     750            1000     750         750.600     749.400

     750            10000    750         750.600     749.400

     320            20000    750         320.385     319.615

     320            50000    750         320.727E+0  319.273E+0

     320            100000   750         322.52E+0   317.48E+0

     8845A/8846A Frequency

     1              10                   10.00300    9.99700

     1              40                   40.00400    39.99600

     0.1            300000               300030.0    299970.0

     8846A Only Frequency

     0.1            1000000              1000100.0   999900.0

     8846A Only 4-Wire Resistance

     0                       10          3.0E-3      0

     10                      10          10.00380    9.99620

     8845A/8846A 4-Wire Resistance

     0                       100         4.0E-3      0

     100                     100         100.0120    99.9880

     0                       1000        10.0E-3     0

A-4
                                                                                   Appendices          A

                                                                                   Verification Forms

            Table A-1. Blank Verification Record for 90-Day Specifications (cont)

     Nominal Output                             Test  Limits

                     Range                                         Results         Pass/Fail

   Ampl.    Freq.                     High                    Low

8845A/8846A 4-Wire Resistance (cont)

1000                 1000             1000.090        999.910

0                    10000            100.0E-3        0

10000                10000            10000.90        9999.10

0                    100000           1.000000        0

100000               100000           100009.0        99991.0

0                    100              0.0040          0

100                  100              100.0120        99.9880

0                    1000             10.0E-3         0

1000                 1000             1000.090        999.910

0                    10000            100.0E-3        0

10000                10000            10000.90        9999.10

0                    100000           1.000000        0

100000               100000           100009.0        99991.0

0                    1000000          10.00000        0

1000000              1000000          1000090         999910

0                    10000000         100.0000        0

10000000             10000000         10002100        9997900

0                    100000000        10000.00        0

100000000            100000000        100810000       99190000

8846A Only 2-Wire Resistance

0                    1000000000       100000.0        0

1000000000           1000000000       1015100000      984900000

8845A/8846A Optional 2X4 Test Lead (Split Lead) test

0                    100              4.0E-3          0

100                  100              100.0120        99.9880

8846A Optional Rear Panel Test

10 V                 DCV (10V)        10.00023 V      9.99977 V

1000 Ω               4-W Ω (1 kΩ)     1000.09 Ω       999.91 Ω

100 mA               DCI (100 mA)     0.100035 A      0.099965 A

                                                                                                       A-5
8845A/8846A

Calibration Manual

                    Table A-1. Blank Verification Record for 90-Day Specifications (cont)

             Nominal Output                           Test Limits

                                Range                                   Results            Pass/Fail

             Ampl.  Freq.                  High                    Low

     8845A Optional Rear Panel Test

     10 V                    DCV (10V)     10.00025 V  9.99975 V

     1000 Ω                  4-W Ω (1 kΩ)  1000.09 Ω   999.91 Ω

     100 mA                  DCI (100 mA)  0.100035 A  0.099965 A

     8846A Capacitance

     0                       1.0E-9        25.0E-12    0

     1.0E-9                  1.0E-9        1.045E-9    955.0E-12

     10.0E-9                 10.0E-9       10.15E-9    9.85E-9

     100.0E-9                100.0E-9      101.5E-9    98.5E-9

     1.0E-6                  1.0E-6        1.015E-6    985.0E-9

     10.0E-6                 10.0E-6       10.15E-6    9.85E-6

     100.0E-6                100.0E-6      101.5E-6    98.5E-6

     1.0E-3                  1.0E-3        1.015E-3    985.0E-6

     10.0E-3                 10.0E-3       10.15E-3    9.85E-3

     100.0E-3                100.0E-3      104.2E-3    95.8E-3

     8846A DC Current

     0                       100.0E-6      25.0E-9     -25.0E-9

     100.0E-6                100.0E-6      100.065E-6  99.935E-6

     -100.0E-6               100.0E-6      -99.935E-6  -100.065E-6

     0                       1.0E-3        50.0E-9     -50.0E-9

     1.0E-3                  1.0E-3        1.00035E-3  999.65E-6

     -1.0E-3                 1.0E-3        -999.65E-6  -1.00035E-3

     0                       10.0E-3       2.0E-6      -2.0E-6

     10.0E-3                 10.0E-3       10.005E-3   9.995E-3

     -10.0E-3                10.0E-3       -9.995E-3   -10.005E-3

     0                       100.0E-3      5.0E-6      -5.0E-6

     100.0E-3                100.0E-3      100.035E-3  99.965E-3

     -100.0E-3               100.0E-3      -99.965E-3  -100.035E-3

     0                       400.0E-3      20.0E-6     20.0E-6

     400.0E-3                400.0E-3      400.18E-3   399.82E-3

     -400.0E-3               400.0E-3      399.82E-3   400.18E-3

     0                       1             200.0E-6    -200.0E-6

A-6
                                                                                     Appendices          A

                                                                                     Verification Forms

           Table A-1.    Blank Verification Record for 90-Day Specifications (cont)

      Nominal Output                          Test Limits

                            Range                               Results              Pass/Fail

    Ampl.  Freq.                   High                    Low

8846A DC Current (cont)

1                        1         1.0006E+0    999.4E-3

-1                       1         -999.4E-3    -1.0006E+0

0                        3         600.0E-6     -600.0E-6

1.9                      3         1.90212E+0   1.89788E+0

-1.9                     3         -1.89788E+0  -1.90212E+0

0                        10        800.0E-6     -800.0E-6

10                       10        10.0128E+0   9.9872E+0

-10                      10        -9.9872E+0   -10.0128E+0

8845A Only DC Current

0                        100.0E-6  25.0E-9      -25.0E-9

100.0E-6                 100.0E-6  100.065E-6   99.935E-6

-100.0E-6                100.0E-6  -99.935E-6   -100.065E-6

0                        1.0E-3    50.0E-9      -50.0E-9

1.0E-3                   1.0E-3    1.00035E-3   999.65E-6

-1.0E-3                  1.0E-3    -999.65E-6   -1.00035E-3

0                        10.0E-3   2.0E-6       -2.0E-6

10.0E-3                  10.0E-3   10.005E-3    9.995E-3

-10.0E-3                 10.0E-3   -9.995E-3    -10.005E-3

0                        100.0E-3  5.0E-6       -5.0E-6

100.0E-3                 100.0E-3  100.035E-3   99.965E-3

-100.0E-3                100.0E-3  -99.965E-3   -100.035E-3

0                        400.0E-3  20.0E-6      20.0E-6

400.0E-3                 400.0E-3  400.18E-3    399.82E-3

-400.0E-3                400.0E-3  399.82E-3    400.18E-3

0                        1         200.0E-6     -200.0E-6

1                        1         1.001E+0     999.0E-3

-1                       1         -999.0E-3    -1.001E+0

0                        3         600.0E-6     -600.0E-6

1.9                      3         1.90288E+0   1.89712E+0

-1.9                     3         -1.89712E+0  -1.90288E+0

0                        10        800.0E-6     -800.0E-6

                                                                                                         A-7
8845A/8846A

Calibration Manual

                    Table A-1. Blank Verification Record for 90-Day Specifications (cont)

             Nominal Output                     Test Limits

                                Range                             Results                  Pass/Fail

             Ampl.  Freq.              High                  Low

     8845A Only DC Current   (cont)

     10                      10        10.0128E+0  9.9872E+0

     -10                     10        -9.9872E+0  -10.0128E+0

     8846A Only AC Current

     100.0E-6       10       100.0E-6  100.14E-6   99.86E-6

     100.0E-6       1000     100.0E-6  100.14E-6   99.86E-6

     100.0E-6       5000     100.0E-6  100.14E-6   99.86E-6

     100.0E-6       10000    100.0E-6  100.45E-6   99.55E-6

     1.0E-3         10       1.0E-3    1.0014E-3   998.6E-6

     1.0E-3         1000     1.0E-3    1.0014E-3   998.6E-6

     1.0E-3         5000     1.0E-3    1.0014E-3   998.6E-6

     1.0E-3         10000    1.0E-3    1.0045E-3   995.5E-6

     8845A/8846A Current

     10.0E-3        10       10.0E-3   10.014E-3   9.986E-3

     10.0E-3        1000     10.0E-3   10.014E-3   9.986E-3

     10.0E-3        5000     10.0E-3   10.014E-3   9.986E-3

     10.0E-3        10000    10.0E-3   10.045E-3   9.955E-3

     100.0E-3       10       100.0E-3  100.14E-3   99.86E-3

     100.0E-3       1000     100.0E-3  100.14E-3   99.86E-3

     100.0E-3       5000     100.0E-3  100.14E-3   99.86E-3

     100.0E-3       10000    100.0E-3  100.45E-3   99.55E-3

     329.0E-3       10       400.0E-3  330.387E-3  327.613E-3

     329.0E-3       1000     400.0E-3  329.729E-3  328.271E-3

     329.0E-3       5000     400.0E-3  332.458E-3  325.542E-3

     329.0E-3       10000    400.0E-3  332.458E-3  325.542E-3

     1              45       1         1.0014E+0   998.6E-3

     1              1000     1         1.0014E+0   998.6E-3

     1              5000     1         1.0014E+0   998.6E-3

     1              10000    1         1.0105E+0   989.5E-3

     1.9            45       3         1.90465     1.89535

     1.9            1000     3         1.90465     1.89630

     1.9            10000    3         1.92765     1.87235

A-8
                                                                                  Appendices          A

                                                                                  Verification Forms

           Table A-1. Blank Verification Record for 90-Day Specifications (cont)

    Nominal Output                           Test Limits

                            Range                              Results            Pass/Fail

    Ampl.  Freq.                   High                   Low

8845A/8846A Current (cont)

10         45               10     10.02100  9.97900

10         1000             10     10.02100  9.97900

                                                                                                      A-9
8845A/8846A

Calibration Manual

                           Table    A-2. Blank Verification Record for 1-Year Specifications

                    Nominal Output                       Test Limits

                                       Range                               Results            Pass/Fail

                    Ampl.  Freq.              High                    Low

      8846A DC Volts

      0                             0.100     3.5E-6        -3.5E-6

      100.0E-3                      0.100     100.0072E-3   99.9928E-3

      -100.0E-3                     0.100     -99.9928E-3   -100.0072E-3

      0                             1         7.0E-6        -7.0E-6

      1                             1         1.000032      0.999968

      -1                            1         -0.999968     -1.000032

      0                             10        50.0E-6       -50.0E-6

      5                             10        5.000170      4.999830

      -5                            10        -4.999830     -5.000170

      10                            10        10.000290     9.999710

      -10                           10        -9.999710     -10.000290

      0                             100       600.0E-6      -600.0E-6

      100                           100       100.0044      99.9956

      -100                          100       -99.9956      -100.0044

      0                             1000      10.0E-3       -10.0E-3

      1000                          1000      1000.0510     999.9490

      -1000                         1000      -999.9490     -1000.0510

      8846A DC Ratio

      100.0E-3                      0.100     1.000144E+0   0.999856E+0

      1                             1         1.000064E+0   0.999936E+0

      -10                           10        1.000058E+0   0.999942E+0

      8845A DC Volts

      0                             0.100     3.5E-6        -3.5E-6

      100.0E-3                      0.100     100.0085E-3   99.9915E-3

      -100.0E-3                     0.100     -99.9915E-3   -100.0085E-3

      0                             1         7.0E-6        -7.0E-6

      1                             1         1.000047E+0   999.953E-3

      -1                            1         -999.953E-3   -1.000047E+0

      0                             10        50.0E-6       -50.0E-6

      5                             10        5.000225E+0   4.999775E+0

      -5                            10        -4.999775E+0  -5.000225E+0

A-10
                                                                                   Appendices          A

                                                                                   Verification Forms

           Table A-2.  Blank Verification Record for 1-Year Specifications (cont)

     Nominal Output                        Test Limits

                          Range                              Results               Pass/Fail

    Ampl.       Freq.            High                   Low

8845A DC Volts (cont)

10                     10        10.0004E+0   9.9996E+0

-10                    10        -9.9996E+0   -10.0004E+0

0                      100       600.0E-6     -600.0E-6

100                    100       100.0051E+0  99.9949E+0

-100                   100       -99.9949E+0  -100.0051E+0

0                      1000      10.0E-3      -10.0E-3

1000                   1000      1.000055E+3  999.945E+0

-1000                  1000      -999.945E+0  -1.000055E+3

8845A DC Ratio

100.0E-3               0.100     1.00017E+0   0.99983E+0

1                      1         1.000094E+0  0.999906E+0

-10                    10        1.000080E+0  0.99992E+0

8845A/8846A AC Volts

100.0E-3   10          0.100     100.1E-3     99.9E-3

100.0E-3   20000       0.100     100.1E-3     99.9E-3

100.0E-3   50000       0.100     100.17E-3    99.83E-3

100.0E-3   100000      0.100     100.68E-3    99.32E-3

100.0E-3   300000      0.100     104.5E-3     95.5E-3

1          10          1         1.0009E+0    999.1E-3

1          20000       1         1.0009E+0    999.1E-3

1          50000       1         1.0017E+0    998.3E-3

1          100000      1         1.0068E+0    993.2E-3

1          300000      1         1.045E+0     955.0E-3

10         10          10        10.009E+0    9.991E+0

10         20000       10        10.009E+0    9.991E+0

10         50000       10        10.017E+0    9.983E+0

10         100000      10        10.068E+0    9.932E+0

3          300000      10        3.17E+0      2.83E+0

                                                                                                       A-11
8845A/8846A

Calibration Manual

                           Table A-2. Blank Verification Record for 1-Year Specifications (cont)

                    Nominal Output                   Test  Limits

                                    Range                               Results                   Pass/Fail

                    Ampl.  Freq.           High                    Low

      8845A/8846A AC Volts (cont)

      100                  45       100    100.09E+0       99.91E+0

      100                  20000    100    100.09E+0       99.91E+0

      100                  50000    100    100.17E+0       99.83E+0

      100                  100000   100    100.68E+0       99.32E+0

      8846A Only AC Volts

      1000                 45       1000   1.0009E+3       999.1E+0

      1000                 1000     1000   1.0009E+3       999.1E+0

      1000                 10000    1000   1.0009E+3       999.1E+0

      320                  20000    1000   320.492E+0      319.508E+0

      320                  50000    1000   320.884E+0      319.116E+0

      320                  100000   1000   322.72E+0       317.28E+0

      8845A Only AC Volts

      750                  45       750    750.675E+0      749.325E+0

      750                  1000     750    750.675E+0      749.325E+0

      750                  10000    750    750.675E+0      749.325E+0

      320                  20000    750    320.417E+0      319.583E+0

      320                  50000    750    320.759E+0      319.241E+0

      320                  100000   750    322.52E+0       317.48E+0

      8845A/8846A Frequency

      1                    10              10.00300        9.99700

      1                    40              40.00400        39.99600

      0.1                  300000          300030.0        299970.0

      8846A Only Frequency

      0.1                  1000000         1000100.0       999900.0

      8846A Only 4-Wire Resistance

      0                             10     3.0E-3          0

      10                            10     10.00400        9.99600

      8845A/8846A 4-Wire Resistance

      0                             100    4.0E-3          0

      100                           100    100.01400       99.98600

      0                             1000   10.0E-3         0

A-12
                                                                                   Appendices          A

                                                                                   Verification Forms

            Table A-2. Blank Verification Record for 1-Year Specifications (cont)

     Nominal Output                             Test  Limits

                     Range                                         Results         Pass/Fail

   Ampl.    Freq.                     High                    Low

8845A/8846A 4-Wire Resistance (cont)

1000                 1000             1000.11000      999.89000

0                    10000            100.0E-3        0

10000                10000            10001.10        9998.90

0                    100000           1.000000        0

100000               100000           100011.0        99989.0

8845A/8846A 2-Wire Resistance

0                    100              0.0040          0

100                  100              100.0140        99.99

0                    1000             10.0E-3         0

1000                 1000             1000.11000      999.89000

0                    10000            100.0E-3        0

10000                10000            10001.10        9998.90

0                    100000           1.000000        0

100000               100000           100011.0        99989.0

0                    1000000          10.00000        0

1000000              1000000          1000110         999890

0                    10000000         100.0000        0

10000000             10000000         10004100        9995900

0                    100000000        10000.00        0

100000000            100000000        100810000       99190000

8846A Only 2-Wire Resistance

0                    1000000000       100000.0        0

1000000000           1000000000       1020100000      979900000

8845A/8846A Optional 2X4 Test Lead (Split Lead) test

0                    100              4.0E-3          0

100                  100              100.01400       99.98600

8846A Optional Rear Panel Test

10 V                 DCV (10V)        10.00029 V      9.99971 V

1000 Ω               4-W Ω (1 kΩ)     1000.11 Ω       999.89 Ω

100 mA               DCI (100 mA)     0.100055 A      0.099945 A

                                                                                                       A-13
8845A/8846A

Calibration Manual

                           Table A-2. Blank Verification Record for 1-Year Specifications (cont)

                    Nominal Output                           Test  Limits

                                       Range                                    Results           Pass/Fail

                    Ampl.  Freq.                  High                     Low

      8845A Optional Rear Panel Test

      100 mV                        DCV (10V)     0.1000085 V      0.0999915 V

      1000 Ω                        4-W Ω (1 kΩ)  1000.11 Ω        999.89 Ω

      100 mA                        DCI (100 mA)  0.100055 A       0.099945 A

      8846A Capacitance

      0                             1.0E-9        25.0E-12         0

      1.0E-9                        1.0E-9        1.045E-9         955.0E-12

      10.0E-9                       10.0E-9       10.15E-9         9.85E-9

      100.0E-9                      100.0E-9      101.5E-9         98.5E-9

      1.0E-6                        1.0E-6        1.015E-6         985.0E-9

      10.0E-6                       10.0E-6       10.15E-6         9.85E-6

      100.0E-6                      100.0E-6      101.5E-6         98.5E-6

      1.0E-3                        1.0E-3        1.015E-3         985.0E-6

      10.0E-3                       10.0E-3       10.15E-3         9.85E-3

      100.0E-3                      100.0E-3      104.2E-3         95.8E-3

      8846A DC Current

      0                             100.0E-6      25.0E-9          -25.0E-9

      100.0E-6                      100.0E-6      100.075E-6       99.935E-6

      -100.0E-6                     100.0E-6      -99.925E-6       -100.065E-6

      0                             1.0E-3        50.0E-9          -50.0E-9

      1.0E-3                        1.0E-3        1.00055E-3       999.45E-6

      -1.0E-3                       1.0E-3        -999.45E-6       -1.00055E-3

      0                             10.0E-3       2.0E-6           -2.0E-6

      10.0E-3                       10.0E-3       10.007E-3        9.993E-3

      -10.0E-3                      10.0E-3       -9.993E-3        -10.007E-3

      0                             100.0E-3      5.0E-6           -5.0E-6

      100.0E-3                      100.0E-3      100.055E-3       99.945E-3

      -100.0E-3                     100.0E-3      -99.945E-3       -100.055E-3

      0                             400.0E-3      20.0E-6          20.0E-6

      400.0E-3                      400.0E-3      400.22E-3        399.78E-3

      -400.0E-3                     400.0E-3      399.78E-3        400.22E-3

      0                             1             200.0E-6         -200.0E-6

A-14
                                                                                  Appendices          A

                                                                                  Verification Forms

           Table A-2. Blank Verification Record for 1-Year Specifications (cont)

      Nominal Output                          Test  Limits

                            Range                                Results          Pass/Fail

    Ampl.  Freq.                   High                     Low

8846A DC Current (cont)

1                        1         1.0007E+0        999.3E-3

-1                       1         -999.3E-3        -1.0007E+0

0                        3         600.0E-6         -600.0E-6

1.9                      3         1.9025           1.8975

-1.9                     3         -1.8975E+0       -1.9025E+0

0                        10        800.0E-6         -800.0E-6

10                       10        10.0158E+0       9.9842E+0

-10                      10        -9.9842E+0       -10.0158E+0

8845A Only DC Current

0                        100.0E-6  25.0E-9          -25.0E-9

100.0E-6                 100.0E-6  100.075E-6       99.935E-6

-100.0E-6                100.0E-6  -99.925E-6       -100.065E-6

0                        1.0E-3    50.0E-9          50.0E-9

1.0E-3                   1.0E-3    1.00055E-3       999.45E-6

-1.0E-3                  1.0E-3    -999.45E-6       -1.00055E-3

0                        10.0E-3   2.0E-6           -2.0E-6

10.0E-3                  10.0E-3   10.007E-3        9.993E-3

-10.0E-3                 10.0E-3   -9.993E-3        -10.007E-3

0                        100.0E-3  5.0E-6           -5.0E-6

100.0E-3                 100.0E-3  100.055E-3       99.945E-3

-100.0E-3                100.0E-3  -99.945E-3       -100.055E-3

0                        400.0E-3  20.0E-6          20.0E-6

400.0E-3                 400.0E-3  400.22E-3        399.78E-3

-400.0E-3                400.0E-3  399.78E-3        400.22E-3

0                        1         200.0E-6         -200.0E-6

1                        1         1.0012E+0        998.8E-3

-1                       1         -998.8E-3        -1.0012E+0

0                        3         600.0E-6         -600.0E-6

1.9                      3         1.90288E+0       1.89712E+0

-1.9                     3         -1.89712E+0      -1.90288E+0

0                        10        800.0E-6         -800.0E-6

                                                                                                      A-15
8845A/8846A

Calibration Manual

                           Table A-2. Blank Verification Record for 1-Year Specifications (cont)

                    Nominal Output                     Test  Limits

                                       Range                              Results                 Pass/Fail

                    Ampl.  Freq.              High                   Low

      8845A Only DC Current (cont)

      10                            10        10.0158E+0     9.9842E+0

      -10                           10        -9.9842E+0     -10.0158E+0

      8846A Only AC Current

      100.0E-6             10       100.0E-6  100.14E-6      99.86E-6

      100.0E-6             1000     100.0E-6  100.14E-6      99.86E-6

      100.0E-6             5000     100.0E-6  100.14E-6      99.86E-6

      100.0E-6             10000    100.0E-6  100.45E-6      99.55E-6

      1.0E-3               10       1.0E-3    1.0014E-3      998.6E-6

      1.0E-3               1000     1.0E-3    1.0014E-3      998.6E-6

      1.0E-3               5000     1.0E-3    1.0014E-3      998.6E-6

      1.0E-3               10000    1.0E-3    1.0045E-3      995.5E-6

      8845A/8846A AC Current

      10.0E-3              10       10.0E-3   10.014E-3      9.986E-3

      10.0E-3              1000     10.0E-3   10.014E-3      9.986E-3

      10.0E-3              5000     10.0E-3   10.014E-3      9.986E-3

      10.0E-3              10000    10.0E-3   10.045E-3      9.955E-3

      100.0E-3             10       100.0E-3  100.14E-3      99.86E-3

      100.0E-3             1000     100.0E-3  100.14E-3      99.86E-3

      100.0E-3             5000     100.0E-3  100.14E-3      99.86E-3

      100.0E-3             10000    100.0E-3  100.45E-3      99.55E-3

      329.0E-3             10       400.0E-3  330.387E-3     327.613E-3

      329.0E-3             1000     400.0E-3  329.729 E-3    328.271E-3

      329.0E-3             5000     400.0E-3  332.458E-3     325.542E-3

      329.0E-3             10000    400.0E-3  332.458E-3     325.542E-3

      1                    45       1         1.00140        998.6E-3

      1                    1000     1         1.00140        998.6E-3

      1                    5000     1         1.00140        998.6E-3

      1                    10000    1         1.01050        989.5E-3

      1.9                  45       3         1.90465        1.89535E+0

      1.9                  1000     3         1.90465        1.89535E+0

      1.9                  10000    3         1.92765        1.87235E+0

A-16
                                                                                  Appendices          A

                                                                                  Verification Forms

           Table A-2. Blank Verification Record for 1-Year Specifications (cont)

    Nominal Output                       Test Limits

                    Range                                  Results                Pass/Fail

    Ampl.  Freq.               High                   Low

8845A/8846A AC Current (cont)

10         45       10         10.02100  9.97900

10         1000     10         10.02100  9.97900

                                                                                                      A-17
8845A/8846A

Calibration Manual

A-18
                                                                               Appendix B

                        Example Adjustment Program

Introduction

Shown below is an adjustment program example for the 8845A and 8846A.

###   Eastwood            8845A/8846A     Calibration    Adjustment  Meta  Program

*RST;SYST:REM

CAL:SEC:STAT              OFF,   FLUKE884X

#   Open          Adjust

##   Remove          all     connections     to  front  terminals

CAL:VAL           ORES,100000000

CAL?          ON

#

#   8846A         Only    -  comment    out  if  calibrating   a   8845A

#

CAL:VAL           ORES,1000000000

CAL?          ON

CAL:VAL           ZCAP,1.00E-09

CAL?          ON

#   end       comment

CAL:REC

###   CALIBRATION            OF  ZEROs

#   Zero          adjusts

##   Place        a  4-wire      short  on   front  and  rear  of  instrument

CAL:VAL           ZVAC,100.0E-3

CAL?          ON

CAL:VAL           ZVACS,100.0E-3

CAL?          ON

CAL:VAL           ZVAC,1

CAL?          ON

CAL:VAL           ZVACS,1

                                                                                           B-1
8845A/8846A

Calibration Manual

                    CAL?    ON

                    CAL:VAL     ZVAC,10

                    CAL?    ON

                    CAL:VAL     ZVACS,10

                    CAL?    ON

                    CAL:VAL     ZVAC,100

                    CAL?    ON

                    CAL:VAL     ZVACS,100

                    CAL?    ON

                    CAL:VAL     ZVAC,1000

                    CAL?    ON

                    CAL:VAL     ZVACS,1000

                    CAL?    ON

                    CAL:VAL     ZVDC,1000

                    CAL?    ON

                    CAL:VAL     ZVDC,100

                    CAL?    ON

                    CAL:VAL     ZVDC,10

                    CAL?    ON

                    CAL:VAL     ZVDC,1

                    CAL?    ON

                    CAL:VAL     ZVDC,0.1

                    CAL?    ON

                    #  Four     50/60  Hz  zero  adjust  steps   -  OutGuard  version  2.0  and  above

                    CAL:VAL     DFVDC60,   100E-3

                    CAL?    ON

                    CAL:VAL     DFVDC60_1,     100E-3

                    CAL?    ON

                    CAL:VAL     DFVDC50,   100E-3

                    CAL?    ON

                    CAL:VAL     DFVDC50_1,     100E-3

                    CAL?    ON

                    #  end   comment

                    CAL:VAL     ZRES,10000000

                    CAL?    ON

                    CAL:VAL     ZRES,1000000

                    CAL?    ON

                    CAL:VAL     ZRES,100000

                    CAL?    ON

                    CAL:VAL     ZRES,10000

                    CAL?    ON

                    CAL:VAL     ZRES,1000

                    CAL?    ON

                    CAL:VAL     ZRES,100

                    CAL?    ON

                    CAL:VAL     ZRES,10

                    CAL?    ON

                    #  Three    Ratio   zero   adjust  steps  -  OutGuard  version  2.0  and  above

                    CAL:VAL     ZVDCREF,   10

                    CAL?    ON

                    CAL:VAL     ZVDCREF,   1

                    CAL?    ON

                    CAL:VAL     ZVDCREF,   100E-3

                    CAL?    ON

                    CAL:REC

B-2
                                                                                                         Appendices  B

                                                                                        Example Adjustment Program

#   end  comment

##   Rear    short     adjust     -  press   the  F/R    switch  to     REAR  -  OutGuard  version       2.0

and   above

CAL:VAL      ZRESR,100000

CAL?     ON

CAL:VAL      ZRESR,10000

CAL?     ON

CAL:VAL      ZRESR,1000

CAL?     ON

CAL:VAL      ZRESR,100

CAL?     ON

CAL:VAL      ZRESR,10

CAL?     ON

#   end  comment

#   Two  dcV     rear  zero    adjust     steps   -  OutGuard    version      2.0  and     above

CAL:VAL      ZVDCR,1

CAL?     ON

CAL:VAL      ZVDCR,100E-3

CAL?     ON

#   end  comment

#   Two  rear     Ratio     zero     adjust  steps   -   OutGuard    version     2.0  and  above

CAL:VAL      ZVDCRREF,1

CAL?     ON

CAL:VAL      ZVDCRREF,100E-3

CAL?     ON

#   end  comment

##   Front    adjust     -  press    the     F/R  switch     to  FRONT

#   400  mA   dc  Current      zero  adjust       steps   -  OutGuard   version       2.0  and    above

CAL:VAL      ZIDC,400.0E-3

CAL?     ON

#   end  comment

#   Front    dc   Current    zeros   adjust

CAL:VAL      ZIDC,100.0E-3

CAL?     ON

CAL:VAL      ZIDC,1.0E-3

CAL?     ON

CAL:VAL      ZIDC,10.0E-3

CAL?     ON

CAL:VAL      ZIDC,100.0E-6

CAL?     ON

CAL:REC

#   Front    ac   Current    zeros   adjust

CAL:VAL      ZIAC,0.0

CAL?     ON

CAL:VAL      ZIACS,0.0

CAL?     ON

CAL:VAL      ZIAC,1.0E-3

CAL?     ON

CAL:VAL      ZIACS,1.0E-3

CAL?     ON

CAL:VAL      ZIAC,10.0E-3

CAL?     ON

CAL:VAL      ZIACS,10.0E-3

CAL?     ON

CAL:VAL      ZIAC,100.0E-3

                                                                                                                     B-3
8845A/8846A

Calibration Manual

                    CAL?     ON

                    CAL:VAL       ZIACS,100.0E-3

                    CAL?     ON

                    #   Two   400  mA  ac     zero   adjust    steps    -  OutGuard  version  2.0  and  above

                    CAL:VAL       ZIAC,400.0E-3

                    CAL?     ON

                    CAL:VAL       ZIACS,400.0E-3

                    CAL?     ON

                    #

                    CAL:VAL       ZIDC,10

                    CAL?     ON

                    CAL:VAL       ZIDC,1

                    CAL?     ON

                    CAL:VAL       ZIAC,1

                    CAL?     ON

                    CAL:VAL       ZIACS,1

                    CAL?     ON

                    CAL:VAL       ZIAC,10

                    CAL?     ON

                    CAL:VAL       ZIACS,10

                    CAL?     ON

                    CAL:REC

                    ###   Calibrate       AC     Linearity

                    #   1.19V@1200Hz

                    ##   set     calibrator      to  output    1.19V@1200Hz

                    CAL:VAL       ACLIN,1.19

                    CAL?     ON

                    ##   set     calibrator      to  output    0.8V@1200Hz

                    #   0.8V@1200Hz

                    CAL:VAL       ACLIN,0.8

                    CAL?     ON

                    ##   set     calibrator      to  output    0.4  V@1200Hz

                    #   0.4V@1200Hz

                    CAL:VAL       ACLIN,0.4

                    CAL?     ON

                    ##   set     calibrator      to  output    0.005V@1200Hz

                    #   0.005V@1200Hz

                    CAL:VAL       ACLIN,0.005

                    CAL?     ON

                    CAL:REC

                    ###   Calibrate       100    mV  AC  Gain  @1200       Hz

                    ##   set     calibrator      to  output    0.1  V@1200Hz

                    #   100   mV   AC  range

                    CAL:VAL       GVAC,0.1

                    CAL?     ON

                    ###   Calibrate       100    mV  AC  Gain  @1200       Hz

                    #   100   mV   AC  Gain

                    CAL:VAL       GVACS,0.1

                    CAL?     ON

                    ###   Calibrate       100    mV  AC  Pole  @50000      Hz

                    ##   set     calibrator      to  output    0.1  V@50000Hz

                    #   100   mV   AC  Pole

                    CAL:VAL       ACPOLE,0.1

                    CAL?     ON

                    ###   Calibrate       1   V  AC  Gain   @1200   Hz

B-4
                                                                            Appendices                  B

                                                                            Example Adjustment Program

##     set       calibrator           to     output  1.0     V@1200Hz

#   1   V     AC     Gain

CAL:VAL           GVAC,1

CAL?       ON

###     Calibrate           1   V     AC     Gain   @1000    Hz

##     set       calibrator           to     output  1.0     V@1000Hz

#   1   V     AC     Gain

CAL:VAL           GVACS,1

CAL?       ON

###     Calibrate           1   V     AC     Slow   @10  Hz

##     set       calibrator           to     output  1.0     V@10Hz

#   AC     Slow

CAL:VAL           FVAC,1

CAL?       ON

###     Calibrate           1   V     AC     Pole   @50000   Hz

##     set       calibrator           to     output  1.0     V@50000Hz

#   1   V     AC     Fast

CAL:VAL           ACPOLE,1

CAL?       ON

###     Calibrate           10     V     AC   Gain   @1200   Hz

##     set       calibrator           to     output  10.0    V@1200Hz

#   10     V     AC   Gain

CAL:VAL           GVAC,10

CAL?       ON

###     Calibrate           10     V     AC   Gain   @1200   Hz

#   10     V     AC   Gain

CAL:VAL           GVACS,10

CAL?       ON

###     Calibrate           10     V     AC   Pole   @50000      Hz

##     set       calibrator           to     output  10.0    V@50000Hz

#   10     V     AC   Pole

CAL:VAL           ACPOLE,10

CAL?       ON

###     Calibrate           100       V   AC  Gain   @1200       Hz

##     set       calibrator           to     output  100.0       V@1200Hz

#   100       V   AC  Gain

CAL:VAL           GVAC,100

CAL?       ON

###     Calibrate           100       V   AC  Gain   @1200       Hz

#   100       V   AC  Gain

CAL:VAL           GVACS,100

CAL?       ON

###     Calibrate           100       V   AC  Pole   @50000      Hz

##     set       calibrator           to     output  100.0       V@50000Hz

#   100       V   AC  Pole

CAL:VAL           ACPOLE,100

CAL?       ON

#

#   8845A         -   uncomment           if  adjusting      8845A

#

###     Calibrate           750       V   AC  Gain   @1200       Hz

#   set       calibrator           to     output     750.0   V@1200Hz

#   750       V   AC  Gain

#CAL:VAL             GVAC,750

#CAL?         ON

                                                                                                        B-5
8845A/8846A

Calibration Manual

                    #   Calibrate             750   V     AC     Gain   @1200       Hz

                    #   750   V     AC     Gain

                    #CAL:VAL           GVACS,750

                    #CAL?     ON

                    #   end   comment

                    #

                    #   8846A  &nb