datasheet

电子工程世界电子工程世界电子工程世界

型号

产品描述

搜索

117-NIST-W-DATA

器件型号:117-NIST-W-DATA
器件类别:测试与测量    万用表与电压表   
厂商名称:Fluke
下载文档

器件描述

Digital Multimeters ELECTRICIANS TRUE RMS MM NIST W/ DATA

参数

产品属性属性值
产品种类:
Product Category:
Digital Multimeters
制造商:
Manufacturer:
Fluke
RoHS:YES
产品:
Product:
Multimeters
类型:
Type:
Hand Held
Accuracy:5 %
Voltage Range:600 V
Resistance Range:0.001 kOhms to 40 MOhms
Capacitance Range:1 nF to 1000 uF
Display Count:6000
True RMS:True RMS
AC Supply Voltage:600 V
商标:
Brand:
Fluke
Number of Digits:4
封装:
Packaging:
Bulk
系列:
Series:
117
工厂包装数量:
Factory Pack Quantity:
1
零件号别名:
Part # Aliases:
1256480

117-NIST-W-DATA器件文档内容

Fluke 117

Electrician’s Multimeter

with Non-Contact Voltage

                                                        Technical Data

                          Compact true-rms meter for

                          commercial applications

                          The Fluke 117 is the ideal meter for

                          demanding settings like commercial

                          buildings, hospitals and schools. The 117

                          includes integrated non-contact voltage

                          detection to help get the job done faster.

                          Features include:

                          •  VoltAlert™ technology for non-contact voltage detection

                          •  AutoVolt automatic ac/dc voltage selection

                          •  LoZ: helps prevent false readings due to ghost voltage

                          •  Large white LED backlight to work in poorly lit areas

                          •  True-rms for accurate measurements on non-linear loads

                          •  Measures 10 A (20 A overload for 30 seconds)

                          •  Resistance, continuity, frequency and capacitance

                          •  Min/Max/Average to record signal fluctuations

                          •  Compatible with optional magnetic hanger (ToolPak™) for

                             hands free operation

                          • CAT III 600 V safety rated

                          General specifications

                          Accuracy is specified for 1 year after calibration, at operating

                          temperatures of 18 °C to 28 °C, with relative humidity at

                          0 % to 90 %.

                          The accuracy specifications take the form of:

                          ± ( [ % of reading ] + [ counts ] )

                          Maximum voltage between    any       600 V

                          terminal and earth ground

Actual size               Surge protection                     6 kV peak per IEC 61010-1 600 V

                                                               CAT III, Pollution Degree 2

                          Fuse for A input                     11 A, 1000 V FAST Fuse

                                                               (Fluke PN 803293)

                          Display                              Digital: 6,000 counts, updates 4/sec

                          Bar graph                            33 segments, updates 32/sec

                          Operating temperature                -10 °C to + 50 °C

                          Storage temperature                  -40 °C to + 60 °C

                          Battery                              9 volt Alkaline, NEDA 1604A/

R                                                              IEC 6LR61

             N10140       Battery life                         400 hours typical, without backlight
Accuracy specifications

Measurement         Range              Resolution      Accuracy ± ([% of reading] + [counts])                              Ordering

DC millivolts       600.0 mV           0.1 mV          0.5 % + 2                                                           information

DC volts            6.000 V            0.001 V

                    60.00 V            0.01 V          0.5 % + 2                                                           Fluke-117              Electrician’s

                    600.0 V            0.1 V                                                                                                      Multimeter with

Auto volts          600.0 V            0.1 V           2.0 % + 3 (dc, 45 Hz to 500 Hz)                                                            Non-Contact Voltage

                                                       4.0 % + 3 (500 Hz to 1 kHz)                                         Included

AC millivolts1      600.0 mV           0.1 mV          1.0 % + 3 (dc, 45 Hz to 500 Hz)                                     TL75 Test leads, holster, User’s

true-rms                                               2.0 % + 3 (500 Hz to 1 kHz)                                         manual and 9 V battery (installed).

AC volts1 true-rms  6.000 V            0.001 V

                    60.00 V            0.01 V          1.0 % + 3 (45 Hz to 500 Hz)

                                                       2.0 % + 3 (500 Hz to 1 kHz)

                    600.0 V            0.1 V

Continuity          600 Ω              1Ω              Beeper on < 20 Ω off > 250 Ω; detects

                                                       opens or shorts of 500 µs or longer.

Ohms                600.0 Ω            0.1Ω            0.9 % + 2

                    6.000 kΩ           0.001 kΩ

                    60.00 kΩ           0.01 kΩ         0.9 % + 1

                    600.0 kΩ           0.1 kΩ

                    6.000 MΩ           0.001 MΩ

                    40.00 MΩ           0.01 MΩ         1.5 % + 2

Diode test          2.000 V            0.001 V         0.9 % + 2

Capacitance         1000 nF            1 nF

                    10.00 µF           0.01 µF         1.9 % + 2

                    100.0 µF           0.1 µF

                    9999 µF            1 µF

                    100 µF to 1000 µF                  1.9 % + 2

                    > 1000 µF                          5 % + 20

Lo-Z capacitance    1 nF to 500 µF                     10 % + 2  typical

AC amps true-rms    6.000 A            0.001 A

(45 Hz to 500 Hz)   10.00 A            0.01 A          1.5 % + 3

                    20 A overload for 30 seconds max.

DC amps             6.000 A            0.001 A         1.0 % + 3

                    10.00 A            0.01 A

                    20 A overload for 30 seconds max.                                                                      Fluke. Keeping your world

Hz (V or A input)2  99.99 Hz           0.01 Hz                                                                             up and running.™

                    999.9 Hz           0.1 Hz          0.1 % + 2

                    9.999 kHz          0.001 kHz                                                                           Fluke Corporation

                                                                                                                           PO Box 9090, Everett, WA USA 98206

                    50.00 kHz          0.01 kHz                                                                            Fluke Europe B.V.

Notes:                                                                                                                     PO Box 1186, 5602 BD

1 All ac voltage ranges except Auto-V/LoZ are specified from 1 % to 100 % of range. Auto-V/LoZ is specified from 0.0 V.    Eindhoven, The Netherlands

Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-       For more information call:

zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of ≤ 3  In the U.S.A. (800) 443-5853 or

at 4000 counts, decreasing linearly to 1.5 at full scale. AC volts is ac-coupled. Auto-V LoZ, and ac mV are dc-coupled.    Fax (425) 446-5116

2 Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.          In Europe/M-East/Africa (31 40) 2 675 200 or

                                                                                                                           Fax (31 40) 2 675 222

                                                                                                                           In Canada (800)-36-FLUKE or

                                                                                                                           Fax (905) 890-6866

                                                                                                                           From other countries +1 (425) 446-5500 or

                                                                                                                           Fax +1 (425) 446-5116

                                                                                                                           Web access: http://www.fluke.com

                                                                                                                           ©2006 Fluke Corporation. All rights reserved.

                                                                                                                           Specifications subject to change without notice.

                                                                                                                           Printed in U.S.A. 10/2006  2634028 D-EN-N Rev B

  Fluke Corporation     117 Electrician’s Multimeter with Non-Contact Voltage
Mouser Electronics

Authorized Distributor

Click to View Pricing, Inventory, Delivery  &  Lifecycle  Information:

Fluke:

FLUKE-117  FLUKE-117-NIST  117-NIST-W/DATA

小广播

117-NIST-W-DATA器件购买:

About Us 关于我们 客户服务 联系方式 器件索引 网站地图 最新更新 手机版

站点相关: 大学堂 TI培训 Datasheet 电子工程

北京市海淀区知春路23号集成电路设计园量子银座1305 电话:(010)82350740 邮编:100191

电子工程世界版权所有 京ICP证060456号 京ICP备10001474号 电信业务审批[2006]字第258号函 京公海网安备110108001534 Copyright © 2005-2017 EEWORLD.com.cn, Inc. All rights reserved